SHARP PPRPPLPEPELPLP PIPPIN PGP % 3 Telemetry Transmitter Module : HEX-561 $ < @ Operation Manual 3 2 8660060006006 260 00600600066 @ Read this manual prior to use of the HEX-561. @ Store this operation manual in an easily accessible place near the unit for future reference.
WARNING FCC Radiation Exposure Statement: This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance 20cm between the radiator and your body (excluding extremities : hands, wrists, and feet) and must not be co-located or operated with any other antenna or transmitter. Federal law restricts this device to sale by or on the order of a physician.
Telemetry Precautions For proper management of the telemetry installation, consult your Fukuoka Dens hi representative concerning the following: ® Plan the installation of your telemetry system taking into account your entire medical facility needs and plant requirements. * Be sure the antenna system installed meets the facility and plant requirements. This Radio Frequency device is susceptible to interference from outside sources.
ii Users are advised to periodically contact the FCC or specified frequency coordinator and determine if other or your transmitter frequencies that may cause interference. The manufacturers, installers and users of Wireless Medical Telemetry System equipment are cautioned that the operation of this equipment could result in harmful interference to other nearby medical devices.
Thank you for purchasing the HEX-561 telemetry transmission module from Fukuoka Dens hi. Before use, read this operation manual thoroughly for correct handling and operation. If you are familiar with a similar type of product, it is recommended this operation manual be read and understood since the HEX-561 has unique operation and handling methods. Safety Information and Messages @ Be sure to observe the warning and cautionary messages related to important safety information.
Safety Considerations in Using Medical Electrical Equipment Contents described here are general safety considerations to be taken for proper operation of medical electrical equipment and for the safety of both the patient and operator. Safety considerations specific to the use of the HEX-561 telemetry transmission module are described on the next page. 1. Do not use the equipment unless familiar with its operation. 2.
4. Precautions during Operation * Do not to operate the equipment beyond the time period necessary that is required for diagnosis, monitoring or treatment. « Observe the equipment and patient constantly for any abnormality. « If the equipment or patient indicates any abnormality, stop the equipment from operating or take proper measures (e.g. disconnect transducers and electrodes) to ensure patient safety. « Instruct the patient not to contact the equipment. 5.
vi 9. When using an neurosurgical unit or defibrillator in conjunction with this equipment, take care of the following. » The high-frequency energy produced by electroscope equipment may result in burns to the patient or possible damage to the equipment. Be sure to observe the precautionary instructions described in the operation manual of the neurosurgical unit. + Some types of equipment may be damaged by the energy discharged by a defibrillator.
Contents Telemetry Precautions sksusescecodremrraraan i Safety Information and Messages PHOEBE eas sanction TRESS TRO HERD EAE iii Safety Considerations in Using Medical Electrical Equipment General Description HEH UP EG TE ERAS HAA BHIIADLAOBGI NOB ARIA EG 1 2. Controls and Indicators PEE EEE NRE ES SS SEA SAY RE AGED ESE EY 2 3. Mounting the HEX-561 to the Bedside Monitor [Mounting into the DD-5100] -vvverersrrmernsans 4 [Mounting into the DD-5300] -v-rxverrorrerenes 7 4.
1. General Description The HEX-561 is a telemetry transmission module to transmit physiological parameters monitored by a bedside monitor in the DD-5000 series via wireless technology. Thus, the HEX-561 allows the user to configure an integrated telemetry patient monitoring system in combination with a central telemetry receiver and central monitor such as the DS-5800N. For compatibility with monitors other than DD-5000 series, contact your Fukuoka Dens hi representative.
2.
(D Antenna Transmitting antenna for the physiological parameters. @ Channel Label Contains the channel identification number of the transmitter. The receiver must be set to the same channel number. If the channel number is changed, the channel label must be replaced. (3 Connector Connects to the bedside monitor, (@Wire Adapter Mounting Hole Connects to the optional wire adapter PA-287. (®Wire Adapter PA-287 (option) Allows for mounting the HEX-561 onto a bedside monitor.
3. Mounting the HEX-561 to the Bedside Monitor Described here are the procedures for mounting the HEX-561 to the DD-5100 and DD-5300 as typical bedside monitors. If the radio channel is changed, it is recommended the channel label be changed in advance. [Mounting to the DD-5100 ] NOTE: Before starting the mounting work, turn the main power switch on the DD-5100 OFF and disconnect the power cable from the wall outlet.
Switch The F-LINK / HLX switch is now accessible. Switch to the HLX position to effect internal communication (Setting this switch to the PHLOX position disables the external F-LINK connector.) HEX-561 Connect the cable from the bottom of the module housing to the connector at the bottom of the HEX-561. Remove the cap which covers the antenna hole in the top of HEX-561 transmission module housing cover. Place the HEX-561 into the housing, taking care not to jam the cable in the inside.
Place the cover back in its original position while at the same time inserting the antenna into the antenna hole. Replace the cover onto the DD-5100 by replacing the screw at the bottom. Reconnect the power cord and turn the DD-5100 ON. Set the telemetry channel from the telemetry set-up display on the DD-5100. For details in selecting the telemetry channel, refer to the operation manual for the DD-5100.
[Mounting into the DD-5300 ] NOTE: Before starting, be sure to turn the main power switch of the DD-5300 OFF and disconnect the power cable from the wall outlet. Connect the optional PA-287 adapter to the HILT-561, connector to connector. Attach the adapter to the HEX-561 with the accessory screw (M2.6x8). Attach the optional mounting fixture onto the HEX-561 with two accessory screws (M2.6 x 6).
Connect the cable from the PA-287 to the F-LINK connector of DD-5300. Reconnect the power cord and turn the DD-5300 ON. Select the telemetry channel from the telemetry set-up display on the DD-5300. For details on the telemetry channel set-up procedure, refer to the operation manual for to the DD-5300. Operation of this equipment requires the prior coordination with a frequency coordinator designated by the FCC for the Wireless Medical Telemetry Service.
To check the system, and verify proper transmission from the HEX-561 to the telemetry receiver, check the central receiver and monitor to ensure the waveform and alphanumeric data are properly displayed. The wireless network system is now functioning properly.
10 4. Setting the Receiving Channel of the Monitor Select the telemetry receiver channel to match that of the telemetry transmission module. The monitor will display the channel number. If the transmission module has a different group code than that of the receiver, set the telemetry receiver group code to correspond to the telemetry transmission module. (For details on selecting the channel and group code, refer to the operation manual for the appropriate monitor.
5. Cleaning, Sterilization and Disinfection Clean, sterilize and disinfect the HEX-561 and accessories as follows. MECHANIZE = v rss srr ssa tiara Use a clean, soft cloth dampened with alcohol or a weak acid/alkalescent/neutral cleanser. Clean the HEX-561 and accessories with the dampened cloth, taking care that the cleaning solution does not enter into any connectors or get inside the HEX-561. Do not use cleaning agents containing organic solutions, thinner, toluene and benzene.
12 6. Maintenance and Inspection To maintain the safety and reliable functionality and performance of the HEX-561, it is preferable to perform routine and periodic inspections before use. A recommendation is described in this section. Failure to perform routine maintenance and inspection may void the responsibility of Fukuoka Dens hi. @ Persons other than your authorized Fukuoka Dens hi service representative must not remove the external easing of the HEX-561.
Routine Inspection Record Inspection Date: Inspected by: Mendel: HEX-561 Serial No.: Location: No. Purchase Date: Items Inspection Judgment criteria Judgment Check the external No abnormalities appearance for scratches, breaks, cracks, loose External screws, or labels. Pass/Fail Appearance | check the antenna for bends or breaks. Under normal operating Wave forms and data conditions, check the received in normal functions at the central manner. Function monitor.
7. Specifications Transmitting Information (will depend on bedside monitor used) Wave forms Numeric Parameters Status Information Interface Communications Method : : ECG x 1, 5 additional waveform (e.g. Respiration, pulse, 2 blood pressure waveform) : NIB, Spot, Temp 1, Temp 2, BP1, BP2, Heart Rate, Respiration Rate, etc. : Low Battery, Electrode fail, Group code (64 types), Alarm, Alarm suspend, Remote record, Model ID number, Transmitter channel No.
Dimensions and Weight Dimensions : x x mum Weigh : Approximately 100 grams Environmental Conditions Operating Temperature Operating Humidity %R.H.
16 8. Accessories To ensure the HEX-561 will perform to its published specifications, use only those accessories specified by Fukuoka Dens hi. When ordering spare parts or accessories, specify the part name and model number. HM Standard Accessories No. Item Model Qty Remarks Operation Manual 1 WM Optional Accessories The following item is not included in standard accessories. Purchase it separately. No. Stem Model Remarks Comes complete with the mounting .