Chapter 4 Monitoring Setup Describes the procedures to set the monitor according to the monitoring purpose.
− Display Configuration − For Easier View The waveform and numeric data display can be configured according to the monitoring purpose. To Configure the Display The display can be configured by selecting the waveforms and numeric data to be displayed. Also, the numeric data display can be enlarged, or graphic trend data can be displayed with the waveform and numeric data. <Display Configuration Menu> ● To Enlarge the Numeric Data The numeric data can be enlarged for easier view.
● To Configure the Display The waveforms and numeric data to be displayed can be selected as desired. When performing telemetry transmission, the numeric data corresponding C A U T I O N to the waveform should be selected for display. Otherwise, the displayed waveform or numeric data may not be transmitted. 1. Press the Menu → Display Config. keys. The display configuration menu will be displayed. 4 2. Select the parameters for numeric data display. Up to 7 parameters can be selected.
<Waveform Selection Menu> First, select the position to display the waveform. By selecting the same waveform successively, the waveform display area can be enlarged. Next, select the waveform to be displayed. The selected parameter will be displayed inside the selected position key ECG1 Cascade displays the ECG in long duration. BP Overlap overlaps the BP1 and BP2 waveforms display. By selecting BP Overlap successively, the BP waveform display area can be enlarged.
● To Display the Short Trend The short trend data can be displayed with the waveform and numeric data. 1. Press the Menu → Display Config. → Wave keys. The waveform selection for the display configuration will be displayed. <Display Configuration / Waveform Selection> The home display with the short trend is shown on the left. The short trend can be displayed 5 ~ 30 min. in 5-minute increments.
Description of the Display This section explains the displayed item on the home display. ● ECG1, ECG2, BP, SpO2, Impedance RESP, CO2 Waveform Display Bed / Room ID No. Patient Name Event Key Patient Type Battery Mark Date / Time Telemetry Channel ECG Lead ECG Size BP Scale SpO2 Size RESP Size Respiratory Sweep Speed CO2 Scale CO2 Wave RESP Wave ECG Wave BP Wave SpO2 Wave Bed / Room ID No. Displays the 4-digit Bed ID and 3-digit (000~999) Room ID.
Date / Time Displays the current date (month, day) and time (hour, minute). Patient Name / Patient Type Displays the patient name and patient type (adult / child / neonate) selected on the admit menu. Respiratory Sweep Speed Indicates the displaying sweep speed for impedance respiration waveform and CO2 waveform. Event Key This touch key will be displayed at alarm occurrence. Even when the alarm is resolved, this key will be remained to be displayed until it is pressed.
● HR, BP, VPC/ST, BP1, BP2, NIBP, SpO2, TEMP, RR, CO2 Numeric Display HR/PR Value HR / PR Parameter BP 1 Value (SYS / DIA (Mean)) HR / PR Sync. Mark BP 2 Value (SYS/ DIA (Mean)) NIBP Meas. Interval Time of NIBP Measurement NIBP Cuff Pressure NIBP (SYS / DIA) SpO2 Value Alarm OFF Symbol Temperature RR Parameter RR Value RR Synchronization Mark EtCO2 Value InspCO2 Value HR / PR Parameter Displays the parameter measuring the HR/ PR.
Temperature Value Displays the temperature measurement value. The YSI-400 temperature sensor can be used. When the measurable range is exceeded, “× × × ” will be displayed. When the YSI-700 is used, “ー ー ー” will be displayed for the measurement value. RR Value Displays the impedance RR / CO2 RR measurement value corresponded to the respiration synchronization source. When the measurable range is exceeded, “× × × ” will be displayed.
● HR, BP, NIBP, SpO2, TEMP, RR, CO2 Enlarged Numeric Data Display Waveform Display Area Enlarged Numeric Data Display Area Waveform Display Area On the enlarged numeric data display, the top 3 waveforms selected on the display configuration will be displayed. Enlarged Numeric Data Display Area On the enlarged numeric data display, the numeric data will be located automatically.
● HR, BP, NIBP, SpO2, TEMP, RR, CO2 Short Trend Display 4 Trend Scale Short Trend Display Displays the short trend beside each numeric data. To change the time of displayed short trend, press a point on the waveform. The time of displayed short trend will change to the pressed point. The displayed time can be changed from 0 min. to 30 min. in 5-min. increments. Trend Scale The short trend scale will be displayed between the short trend and numeric data.
Description of Alarm Message and Alarm Sound This section explains about the message displayed on the home display. There are vital alarm message and equipment status alarm message which will be displayed on the top of the home display. The alarms are classified in level 1, level 2, level 3, level 4, and the alarm message will be displayed according to the priority of level 1>level 2>level 3>level 4.
Life Threatening Alarm (Alarm Level 1) Parameter HR PR (SpO2, BP) BP1 SpO2 Respiration (Impedance, CO2) NIBP Arrhythmia 4 Display Configuration CO2 Message “Lower HR alarm” “Upper HR alarm” “Lower PR alarm” “Upper PR alarm” “Lower BP1 alarm” “Upper BP1 alarm” “Lower SpO2 alarm” “Upper SpO2 alarm” “Apnea alarm” “Lower RR alarm” “Upper RR alarm” “Lower NIBP alarm” “Upper NIBP alarm” “Upper EtCO2 alarm” “Lower EtCO2 alarm” “ASYSTOLE” “VF” “VT” “SLOW VT” “TACHY” “BRADY” “RUN” Cautionary Alarm (Alarm Level
● Equipment Status Alarm Message The equipment status alarm message will be displayed when proper monitoring can not be performed. The alarm message will be displayed according to the priority of the alarm level. If more than one alarm with the same level is generated, the alarm message for the newer alarm will be displayed.
“Wrong temp probe” Temperature “TEMP auto check” “TEMP unit check” “Motion artifact” SpO2 “SpO2 unit error” CO2 “CO2 unit error” ECG “ECG unit error” NIBP “NIBP unit error” “Recorder error” “Paper out” Recorder “Magazine open” “Paper jammed” “Recorder busy” All Alarm “Alarm Mute” ECG, Impedance Respiration “Check electrode” Telemetry “Telemetry unit error” 4 Display Configuration ● Lead-Off Message If the ECG electrodes are detached, HR alarm and arrhythmia alarm will not be generated.
− Key Setup − For Easier Use The touch keys on the screen are the only operational keys for the DS-7100 system. The touch keys consist of fixed keys (menu, Home / Enlarge, NIBP Start/Stop) and 4 user keys which can be programmed according to the monitoring purpose. Fixed Key <Home Display> User Key 1 User Key 2 User Key 3 User Key 4 The unnecessary keys on the display can be erased.
To Set the User Keys Program the key function to each user key. 1. Press the Menu → System Config. → User Key keys. The user key setup menu will be displayed. Pre-Set → Ward Setup → 4 Key Setup 2. Select a position to set the user key. Press the key for the position to set the user key. User Key 1 User Key 2 User Key 3 User Key 4 3. Select a parameter for the user key. Press the key for the parameter to be set as user key. User Key Alarm Silence Alarm Admit / Discharge Rec.
Freeze NIBP Auto Mode Lead・Size HR Source Tone/Volume Other Bed Record Display Config. ST Display Night Mode BP Zero OFF Temporarily stops the waveform trace. By pressing the Rec. START/STOP key during freeze mode, the waveform in freeze mode can be recorded. Displays the NIBP measurement interval setup menu. Displays the keys to adjust the size, scale, baseline position of the displayed waveform. Sequentially selects the HR source in the order of ECG → SpO2 → BP1 → Auto → ECG.
To Set the Menu Keys The key display can be erased from the menu display, configuration menu display and preset menu display. 1. Press the Menu → System Config. → Key Mask keys. Display the key mask setup menu. Pre-Set → Ward Setup → 4 Key Setup <Key Mask Display> 2. Select the key to erase. Pressing the key will extinguish the LED. The key with the LED extinguished will not be displayed on the menu display. Display ON 3. Press the Display OFF System Config. key.
− Recording Setup − To Record the Waveform and Numeric Data On the DS-7100 system, the waveform recording by manual recording, periodic recording, alarm recording, or freeze recording, and graphic recording of graphic trend, tabular trend can be performed. This section describes the procedure for the following recording.
2. Select the waveform for recording. Up to 3 waveforms can be selected. The waveforms will be automatically located at recording. 3. Select the duration for recording. Select the recording duration from 24sec or Cont. . 24sec will automatically stop the recording after 24 seconds. 4. Select the delay time for recording. None will start the recording from the time the Rec. START/STOP key is pressed. 8sec , 16sec will start the recording 8 sec. / 16 sec. prior to the time the key is pressed.
5. Start the periodic recording. ON will activate the periodic recording with the selected interval. If periodic recording is not necessary, select OFF . Alarm Recording The recording will automatically start at occurrence of numeric alarm or arrhythmia alarm. If the alarm with the higher priority occurs during alarm recording, the C A U T I O N recording in process will be ceased and starts the alarm recording with the higher priority. 1. Press the Menu → System Config.
The delay time differs depending on the recording time. NOTE Delay Time Recording Time Adult Child 12 sec. 24 sec. 12 sec. 16 sec. 12 sec. 16 sec. Neonate Numeric Alarm Arrhythmia Alarm 8 sec. 12 sec. 16 sec. 16 sec. 5. Select the arrhythmia type. If arrhythmia is selected for the recording factor, select the arrhythmia type. Pressing the Arrhythmia Record key will display the arrhythmia selection window for alarm recording.
Freeze Recording The waveform display can be frozen and recorded from 6 seconds prior to the frozen point. The waveform to be recorded is the one selected for manual recording. The recording duration is fixed as 6 seconds. To freeze the waveform display, the Freeze key needs to be assigned as user key. 1. Freeze the waveform display. Press the Freeze key on the user key. 2. Start freeze recording. Press the Rec. START/STOP 4− 24 key to record the displayed waveform.
Graphic Recording (Graphic/Tabular Trend, etc.) Graphic trend recording, tabular trend recording, NIBP list recording, and recall recording can be performed. Reference Refer to “7. Function procedure. Graphic Trend, Tabular Trend, Recall Data, NIBP List” for recording Graphic Trend One parameter from the following can be selected for graphic recording.
Recorder Operation (QRS Symbol, etc.) This menu allows the setup of paper feed operation and ON/OFF of QRS classification symbol recording 1. Press the Menu → System Config. → Record The recorder operation setup menu will be displayed. → Setup keys. 2. Select ON/OFF of paper feed to top. ON will feed the paper to start recording from the next perforation. OFF will not feed the paper and starts recording from the position where the previous recording ended. 3. Select ON/OFF of paper feed to end.
− Volume Setup − Pulse Tone, Alarm Sound, etc. This menu allows volume setup of the pulse tone, alarm sound, key sound and other bed alarm sound. 1. Press the Menu → Tone / Volume keys. The volume setup menu will be displayed. 4 Volume Setup <Volume Setup Menu> Decreases the volume Tests the set volume. Current volume Increases the volume. 2. Set the volume for the pulse tone. The volume of the HR synchronized tone, SpO2 synchronous tone, BP synchronized tone can be adjusted. 3.
− Color / Brightness Setup − This menu allows the setup of the colors of numeric data / waveform and brightness of the display. Color Setup (Numeric Data, Waveform) The displayed colors can be set for each parameter. Assign the color from the 32-color palette to each parameter. 1. Press the Menu → System Config. → Color keys. <Color Setup Menu> 2. Select the color from the 32-color palette. Press the key of the desired color. 3. Assign the selected color to the parameter.
Brightness Setup The brightness of the display can be adjusted. 1. Press the Menu → System Config. → Brightness Setup keys. 4 Color / Brightness Setup <Brightness Setup Menu> 2. Adjust the brightness. Use the ? , ? keys to adjust the brightness. Darkens the display. NOTE Current brightness Brightens the display. The display panel utilizes exclusive fluorescent light for the backlight.
− Telemetry Setup − (DS-7141, DS-7101LT) This section explains the setup of telemetry transmitting channel. The DS-7141 and DS-7101LT incorporates a telemetry transmitter. Once the transmitting channel ID and group ID are programmed, these will remain stored even after the main power is turned OFF. The telemetry function is not supported for the DS-7101L. 1. Press the Menu → System Config. → Telemeter Setup keys. Pre-Set → Hospital Setup → <Telemetry Setup Menu> 2. Select ON/OFF of telemetry.
5. Enter the group ID. Press the Group key, and enter the group ID. Use the numeric keypad to enter the group ID in the range from 00 ~ 63. 6. Save the telemetry channel ID and group ID. Press the Save key to store the channel ID and group ID. Verify that the “Complete” message is displayed. 4 If an error is found on the password, channel ID, or group ID, the following message will be displayed. “Invalid data” “Telemetry unit error” “Install Program Card” : The entered password is incorrect.
Transmitting Waveform Setup This function allows to select the transmitting waveform. 1. Press the Menu → System Config. → Telemetry Wave Setup keys. 2. Select the waveform to transmit. Select the waveform to transmit from ECG1 , ECG1+2 . Selecting ECG1 will transmit ECG1, CO2, RESP, BP1, BP2, SpO2. RESP waveform will be transmitted when “Impedance” is selected for RR source. Selecting ECG1+2 will transmit ECG1, ECG2, CO2/RESP, BP1, SpO2.
− Wired Network Connection − (LAN Communication) This section describes the procedure to set the Room / Bed ID. The DS-7100 system incorporates Ethernet LAN unit. The set Room / Bed ID will be remain stored even when the power is turned off. Room / Bed ID Setup To connect to a wired network, it is necessary to set the Room / Bed ID. 1. Press the Menu → Admit / Discharge → Bed ID keys. 4 Use the numeric keypad to enter the password The entered number will be displayed as “****”. 3. Set the Room ID.
? The bed ID is factory set to “000”. If connected to the wired network with the bed ID unchanged, monitoring on the central monitor will not be possible. ? When connecting to a wired network, make sure that there are no other bedside monitors with the same ID. If there are more than one bedside monitor with the same bed ID, the duplicated bedside monitors can not be monitored on the central monitor. ? When connecting to a wired network, set the bed ID in the range from “001” to “048”.
− Alarm Pole Setup − Notifying the Alarm by Light This section describes the use of alarm pole function. This equipment incorporates an alarm pole. The alarm generation can be indicated by alarm pole flash so that it can be notified from distance. Alarm Level Setup On the alarm pole setup function, alarm pole function and alarm level can be linked. 1. Press the Menu → System Config. → Alarm Pole keys. The alarm pole setup menu will be displayed. Pre-Set → Ward Setup → 4 Alarm Pole Setup 2.
Alarm Level 1 (Life Threatening Alarm) Parameter HR PR (SpO2, BP) BP1 SpO2 Respiration (Impedance, CO2) NIBP CO2 Arrhythmia Message “Lower HR alarm” “Upper HR alarm” “Lower PR alarm” “Upper PR alarm” “Lower BP1 alarm” “Upper BP1 alarm” “Lower SpO2 alarm” “Upper SpO2 alarm” “Apnea alarm” “Lower RR alarm” “Upper RR alarm” “Lower NIBP alarm” “Upper NIBP alarm” “Upper EtCO2 alarm” “Lower EtCO2 alarm” “ASYSTOLE” “VF” “VT” “SLOW VT” “TACHY” “BRADY” “RUN” Alarm Level 2 (Cautionary Alarm) Parameter BP2 ST Temper
Alarm Level 3 (Treatment Needed Alarm) Parameter Equipment Status Message “Check NIBP hose” “CVA detected” “No pulse detect” “Pacemaker error” 4 Alarm Pole Setup 4− 37
Alarm Pole Flash Pattern Setup This function sets the alarm pole flash pattern for each alarm. 1. Press the Pattern Setup key on the alarm pole setup menu. The alarm pole flash pattern setup menu will be displayed. 2. Select the flash pattern for each item. First, select the flash pattern from Pattern 10 . Pattern 1 to The alarm pole consists of 3 blocks. (Left: Red, Center: Red/Orange/Green, Right: Red) The ways these 3 blocks flash are different for each pattern.
− Alarm Setup − To Set the Alarm Condition This section explains the setup of alarm suspension and upper / lower alarm limit. On the alarm setup menu, ON/suspend of system alarm, ON/OFF and upper / lower alarm limit of each parameter can be set. To Set the System Alarm The system alarm can be set to ON or suspend, but it can not be turned OFF. ? When the system alarm is suspended, all the alarm will be suspended even if the parameter alarm is set to ON. Also, the alarm event will not be stored as recall.
To Turn ON the System Alarm Turning the system alarm ON will activate the alarm setups for each parameter. 1. Press the Menu → Alarm keys. 2. Set the alarm ON . Press the Suspend key when the key LED is lighted. The key LED will extinguish. The alarm is ON when the key LED is extinguished. To Suspend the System Alarm The alarm can be temporarily suspended. During the alarm suspension, “Alarm Suspended** *sec.” message will be displayed. “***sec.” indicates the remaining suspended time.
Alarm Setup for Each Parameter The alarm for each parameter can be turned ON or OFF, and upper and lower alarm limit can be set. 1. Press the Menu → Alarm keys. The alarm setup menu will be displayed. 4 Alarm Setup 2. Select the parameter to display the alarm setup menu. Display Description Displays the upper and lower alarm limit and the current measurement (▲ ). The limits can be adjusted by directly pressing the bar display or using the arrow keys. Displays lower limit←current value→upper limit.
3. Select ON/OFF and set upper and lower alarm limit for the parameter alarm. Numeric Data Key Item Description ON, OFF ST ON, OFF ST1 ± 2.0mV ST2 ± 2.0mV BP1 ON, OFF 0− 300mmHg BP2 ON, OFF 0− 300mmHg SpO2 ON, OFF 50− 100% ON, OFF 5− 150bpm (Adult) 2− 150bpm (Child, Neonate) RR APNEA (Upper Limit) ON, OFF 5− 20 sec. TEMP ON, OFF 30− 50 NIBP ON, OFF 10− 300mmHg ON, OFF 1− 100mmHg 0.1− 13.3kPa 0.1− 13.3% ON, OFF 1− 4mmHg 0.1− 0.4kPa 0.1− 0.
Arrhythmia Alarm Setup The arrhythmia alarm can be turned ON or OFF, and arrhythmia detection level can be set. 1. Press the Menu → Alarm keys. The alarm setup menu will be displayed. 4 Alarm Setup 2. Press the Page Arrhy. key to display the arrhythmia alarm setup menu.
● To Set ON/OFF of Arrhythmia Alarm Selecting ON will generate the arrhythmia alarm. Selecting OFF will not generate the arrhythmia alarm. However, Asystole, VF, and VT alarm can not be turned OFF at the initial factory setting. Select ON/OFF for “Asystole, VF, VT” on the Hospital Setup menu to turn OFF the alarm. Reference Page Page 1/3 Page 2/3 Page 3/3 NOTE Item Asystole VF VT Slow_VT Run Bigeminy Trigeminy Pause Couplet Tachy Brady Frequent Refer to “8.
● Alarm Limit for TACHY, BRADY The arrhythmia detection level for tachycardia (TACHY) and bradycardia (BRADY) alarm links with the upper and lower alarm limit for HR / PR. The tachycardia (TACHY) alarm generates when the measurement exceeds the HR / PR upper alarm limit. When the upper alarm limit is OFF, alarm will not generate. The bradycardia (BRADY) alarm generates when the measurement is below the HR / PR lower alarm limit. When the lower alarm limit is OFF, alarm will not generate.
1. Press the Menu → Alarm keys. The alarm setup menu will be displayed. 2. Press the SpO2 key to display the alarm setup menu. 3. Select the SEC alarm value according to the alarm frequency. If 10 / 25 / 50 / 100 is selected, a circular SEC alarm indicator will be displayed inside the parameter key. As the integral value increases, the indicator will begin to fill, and when it is Indicator completely filled, an alarm will be generated.
Alarm Suspend / Alarm Silence Time The time for suspending the system alarm and suspending the alarm sound can be selected. 1. Press the Menu → Alarm → Page Down keys. The second page of the alarm setup menu will be displayed. 4 Alarm Setup 2. Select the time for “Alarm Suspend Time. Select the appropriate time for alarm suspend time. 3. Select the time for “Alarm Silence Time. Select the appropriate time for alarm silence time.
ON / OFF of Alarm Limit Display The alarm limit can be selected to display or not display on the home display according to the preference. <Display OFF> <Display ON> 1. Press the Menu → Alarm → Page Down keys. The second page of the alarm setup menu will be displayed. 2. Select ON or OFF for alarm limit display. NOTE 4− 48 ? To display the alarm limit, short trend display should be set to OFF on the display configuration menu.
Chapter 5 Admit / Discharge of a Patient This chapter describes the procedure to admit or discharge a patient. − − − − ● Pacemaker Use Selection ・・・・・・・・・・・・・・・・・・・ 6 ● Impedance Respiration Measurement ・・・・・・・・・ 6 Patient Age ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 7 Discharging a Patient − Erasing Name, Data, etc.
− Admit / Discharge of a Patient− This menu allows setup of admitting, discharging, suspend monitoring of a patient, and selection of the display configuration mode and alarm mode according to the monitoring purpose. 1. Press the menu. Menu → Admit / Discharge keys to display patient admit / discharge If you start monitoring a new patient without performing a discharge C A U T I O N procedure for the previous patient, new data will be added to the previous data which will result in inaccuracy.
− Admitting a Patient − Entering Name, Sex, and Age This menu allows entering of patient’s name, ID, age, and selection of patient type (adult, child, neonate) and pacemaker use (used, not used) which affects the monitoring accuracy. Patient Name For entering the patient’s name, up to 16 letters can be used. Patient Name 5 Enter the name using the alphanumeric keypad. The keypad can be selected from ABC or QWERTY arrangement. ABC Arrangement QWERTY Arrangement 2.
Patient ID Up to 10 characters of alphabets, numbers, or symbols can be used for the patient ID. Enter the ID according to the monitoring purpose. The entered ID will be output on the recording paper. Patient ID 1. Press the ID key. Enter the ID using the alphanumeric keypad. Patient Type The selection of patient type affects the accuracy of NIBP measurement, HR measurement, and RR measurement. Also the delay time to generate the measurement data alarm will change according to the patient type.
1. Select Adult , Child , or Neonate . If “ST Display” is selected for the filter mode (Monitor / ST Display / ESIS), Neonate can not be selected for the patient type. To select Neonate , set the filter mode to Monitor or ESIS mode. 2. The selected patient type will be displayed on the home display. Patient Type 5 Select the patient’s sex from male or female. The default is set as undetermined. The selected sex will be output on the recording paper. 1. Select Male or Female .
Pacemaker Use ● Pacemaker Use Selection If the patient is wearing a pacemaker, the monitor will identify the pacemaker pulse and insert an artificial pulse onto the ECG waveform for easy identification. By detecting the pacemaker pulse, it prevents to erroneously detect QRS as pacemaker pulse when pacing waveform does not appear (pacing failure). The arrhythmia analysis analyzes pacing beat as P (pacemaker beat) or F (fusion beat) to prevent erroneous judgement of VPC.
Patient Age There are two ways to set patient’s age. One is to enter the birth date which will automatically calculate the age, and the other is to directly enter the age using the numeric keypad. If Neonate is selected as patient type, the age in days will be displayed. 1. Press the Age key, and enter patient’s age. 5 Numeric Keypad 2. Enter the patient’s birth date using the numeric keypad. The age will be automatically calculated.
− Discharging a Patient − Erasing Name, Data, etc. This menu allows to clear the patient name, ID, age, and past measurement data such as tabular trend, graphic trend, and recall data. Discharging Procedure 1. Press the Discharge key. The confirmation display will appear. If the Discharge Cancel key to return to the previous display. key is pressed by mistake, press the 2. Erase the patient data.
− Monitoring Mode Selection − Alarm and Display Mode This menu allows to select the alarm mode and display mode. The alarm setting and display configuration can be each selected from 5 modes depending on the monitoring purpose. Select the appropriate mode when admitting a patient. The setup for the alarm mode and display mode remains stored even when the power is turned off or when discharging procedure is performed.
3. Select a display mode from the selection. Press one key from the Disp mode 1 ~ Disp mode 5 keys to select a display configuration mode which meets the monitoring purpose. The setup for the display mode remains stored even when the power is turned off or when discharging procedure is performed. The previously selected display configuration mode will be effective if the selection is not made.
● Display Modes Item Mode Selection Mode 1 Mode 2 Mode 4 Mode 5 NOTE 1 No. of Waveforms 3 Waveform No. of Numeric Data 4 Numeric Data Displayed Waveforms ECG1, SpO2, RESP Displayed Numeric Data HR, NIBP, SpO2, RR Enlarged Display OFF Short Trend OFF Comment CONFIG. 1 No. of Waveforms 3 Waveforms No. of Numeric Data 4 Numeric Data Displayed Waveforms ECG1, SpO2, RESP Displayed Numeric Data HR, NIBP, SpO2, RR Enlarged Display ON Short Trend OFF Comment CONFIG. 2 No.
● Alarm Modes Item Alarm Mode HR ASYSTOLE VF VT SLOW_VT RUN COUPLET PAUSE BIGEMINY TRIGEMINY FREQUENT TACHY BRADY ST BP1 Alarm Mode 1〜5 BP2 RR APNEA SpO2 NIBP TEMP Default Setting 1 ON, 40− 120 ON, 5 sec. ON ON ON ON, 3 beats OFF OFF, 2 sec. OFF OFF OFF, 10 beats ON ON OFF ON SYS 80− 180 DIA OFF− OFF MEAN OFF− OFF OFF SYS OFF− OFF DIA OFF− OFF MEAN OFF− OFF ON 5− 30 ON, 15 sec. ON, 90− OFF SEC Alarm OFF ON SYS 80− 180 DIA OFF− OFF MEAN OFF− OFF OFF OFF− OFF ON 30− 45mmHg 4.0− 6.0kPa 4.0− 6.0% 3mmHg 0.
− Suspend Monitoring − Suspend and Resume Monitoring This menu allows to suspend and resume monitoring when a patient temporarily leaves the bed. Turning the power OFF will erase the recall and ST measurement data, but with this suspend monitoring function, data measurement, alarm generation, automatic measurement, and automatic recording can be suspended without erasing any data and setup condition. To Suspend Monitoring 5 Cease key is pressed by mistake, press the 2.
To Resume Monitoring 1. Press the Resume key. The monitor suspend display will be cleared and monitoring will resume. C A U T I O N Resuming monitoring will also resume the alarm in suspension.
− Room / Bed ID Setup − for LAN Communication This section describes the procedure to set the Room / Bed ID. The DS-7100 system incorporates Ethernet LAN unit. The set Room / Bed ID will be remain stored even when the power is turned off. Room / Bed ID Setup To connect to a wired network, it is necessary to set the Room / Bed ID. 1. Press the Menu → Admit / Discharge → Bed ID keys. 5 Use the numeric keypad to enter the password The entered number will be displayed as “****”. 3. Set the Room ID.
? The bed ID is factory set to “000”. If connected to the wired network with the bed ID unchanged, monitoring on the central monitor will not be possible. ? When connecting to a wired network, make sure that there are no other bedside monitors with the same ID. If there are more than one bedside monitor with the same bed ID, the duplicated bedside monitors can not be monitored on the central monitor. ? When connecting to a wired network, set the bed ID in the range from “001” to “048”.