Press the ID key. Enter the ID number using the numeric keypad. 2 Press the Age key. Enter the birth date using the numeric keypad. 4. Select the patient type, and Used / Not used of pacemaker use. Select the patient type from Adult , Child , or Neonate . Select the pacemaker use from Used , Not used . The patient type and pacemaker use must be selected. The patient type W A R N I N G selection influences the precision of the QRS detection and NIBP measurement.
− Basic Operation − Scale, Lead, Baseline Position Setup (Parameter Key) 1. Select the parameter to set the scale / lead / baseline position. (Ex.: ECG) Press the parameter key where heart rate is displayed. 2. Press the Lead・Size displayed. key. The menu to adjust size / lead / baseline position will be 3. Adjust the waveform size. Select an appropriate waveform size for the monitoring waveform. The arrhythmia detection level corresponds with the displayed waveform C A U T I O N size.
4. Select the lead. Select an appropriate lead according to the monitoring purpose. 2 Basic Operation 5. Adjust the baseline position. Use the ? , ? keys to adjust the waveform baseline position. Reference Use the same procedure for each parameter. Refer to section for each parameter in “6. Parameter Setup”.
Scale, Lead, Baseline Position Setup (User Key) Pressing the Lead・Size key will display the arrow keys on the home display to adjust waveform size, scale, lead, baseline position. 1. Select the waveform size, lead, baseline position for ECG waveform. Adjust the waveform suitable for monitoring. Pressing the Lead key will sequentially change the lead. 3-electrode:Ⅰ→Ⅱ→Ⅲ→Ⅰ 4-electrode:Ⅰ→Ⅱ→Ⅲ→aVR→aVL→aVF→Ⅰ 5-electrode:Ⅰ→Ⅱ→Ⅲ→aVR→aVL→aVF→V→Ⅰ Pressing the Size key will sequentially change the size.
Alarm Setup for Each Parameter The alarm can be set for each parameter. By pressing the selected parameter key, upper and lower alarm limit and ON/OFF of alarm can be set. 1. Select the parameter to set the alarm. (Ex.: HR alarm) 2 Basic Operation 2. Press the HR Alarm key. The menu to adjust the alarm limit will be displayed. 3. Set the upper and lower alarm limit. Use the ? , ? keys to adjust the lower limit. ? , ? keys to adjust the upper limit.
ON/OFF of Parameter Display Waveform/Numeric Data Display The waveform and numeric data display for each parameter can be turned ON or OFF without changing the display configuration. If not performing the ECG or SpO2 measurement while the ECG cable or SpO2 sensor is connected to the monitor, the equipment status alarm such as “Lead Off” will generate.
Recording ● Start / Stop of Waveform Recording Pressing the Rec. START/STOP key on the home display will start the waveform recording. Up to 3 waveforms can be recorded. 2 Basic Operation 【Example of Recording】 Recorded Time Sex Age Patient ID Patient Type Measured Data Bed ID Patient Name Waveform Type, Lead, Size Recording Speed Rec. Mode Delay Time Telemetry Channel Filter Mode For the manual recording, number of recording waveforms and recording duration can be set. Refer to “4.
To Install the Paper 1. Press the button located at the right side of the recorder magazine. 2. The magazine will come out. 3. Pull out the magazine from the monitor. 4. Set the recording paper so that the mark printed on each paper comes to the right. Mark 5. Place the magazine back into the monitor. Push in until it locks into place with a click sound.
To Suspend Monitoring When not monitoring for a while, turning OFF the power will erase the recall data, ST measurement, OCRG data. However, using the cease monitoring function allows suspension of data measurement, alarm generation, automatic measurement, automatic recording without erasing the data or setup details. 1. Press the Cease key on the admit / discharge menu. Press the Menu → Admit / Discharge → Cease key. 2 Basic Operation 2. Suspend monitoring.
Discharging Procedure This procedure will erase the past data, such as patient name, ID, age, tabular / graphic trend, and recall. 1. Press the Discharge key on the admit / discharge menu. Press the Menu → Discharge key. Admit / Discharge 2. Perform discharge procedure. Pressing the OK key on the confirmation display will discharge the patient. Pressing the Cancel key will return to the previous display.
− Operation Flow − The operation flow of this system is as follows.
Blank Page 2− 22