Chapter 6 Parameter Setup Quick SYS Measurement Duration ・・・・・・・・・・・・・ 32 End of Measurement Tone ・・・・・・・・・・・・・・・・・・・・ 33 Quick SYS List ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 33 PR Display ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 34 Mean BP Display ・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 34 1-Minute Measurement Duration ・・・・・・・・・・・・・・・ 35 NIBP Speed ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 35 − Temperature − ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 36 Temperature Alarm ・・・・・・・・・・・・・・・・・・・・・・・・・・ 36 ON/OFF of Parameter Display ・・・・・・・・・・・・・・・・ 37 (D
− Parameter Setup − Setting the Monitoring Condition This menu allows setup of measurement condition, waveform size, scale, etc. of ECG, BP, NIBP, SpO2, RESP, TEMP, and CO2. To Display the Parameter Setup Menu Press the Menu → Parameter keys to display the parameter setup menu, and select the parameter. On the parameter setup menu, BP zero balance can be performed. The parameter setup menu for each parameter can be also accessed by pressing the parameter key where numeric data is displayed.
Zero Balance of All Pressure Lines (BP1, BP2) 1. Open the three-way cock of all pressure transducers to air. 2. Press the Zero All BP key. Verify the BP waveform is positioned at zero, and “0” is displayed for the BP value. A message, “BP zero complete” will be displayed when the procedure is complete. A message, “BP zero failed” will be displayed when the process fails. The three-way cock may not be opened to air, artifact may be present, or the transducer may be defective.
− ECG − The measurement condition for ECG can be set on this menu. Lead, Size HR Alarm Arrhy. Alarm Configuration Arrhy. Learn : : : : : Sets the waveform size and lead for ECG display and recording. Sets ON/OFF of HR alarm, and sets upper and lower alarm limit. Sets ON/OFF and detection threshold for each arrhythmia alarm. Sets the condition for measuring ECG and HR. The monitor learns the normal QRS at ECG electrode replacement or at misdetection of the arrhythmia analysis.
Pressing the Auto key will automatically adjust the ECG amplitude to 10mm. The automatic adjustment will function only when the key is pressed. The automatic adjustment will not function when the monitor is learning arrhythmia. 4. Select the lead. The leads can be selected from 3 leads, 6 leads, 7 leads depending on the connected ECG relay cable. ECG Relay Cable Lead 3-electrode 4-electrode 5-electrode 6 5. Set the baseline position.
HR Alarm 1. Press the HR Alarm key to display the alarm setup menu. Select ON/OFF of HR/PR alarm, and set the upper and lower alarm limit. The common alarm value for HR measured from ECG, PR measured from SpO2, PR measured from BP can be set. The upper and lower limit can be set in 5 bpm increments. Key Item Description Selecting ON will generate the HR/PR alarm. Individual Alarm Selecting OFF will not generate the HR/PR alarm. Sets the lower alarm limit (20〜295bpm).
Filter Mode Selection The waveform frequency characteristic can be selected from Monitor Mode, ESIS Mode, or ST Display Mode according to the monitoring purpose. Each mode has different frequency characteristic. The AC filter is always set to ON. 1. Press the Config. key to display the setup menu for selecting the filter. 2. Select a frequency characteristic to monitor ECG from Monitor , ESIS , or ST Display . Monitor Mode 6 ECG Patient Type Frequency Characteristic Adult / Child 0.
HR Average Selection The averaging method of HR measured from ECG can be selected. 1. Press the Config. key to display the setup menu for HR Average selection. 2. Select Instant or Average . Selecting Instant will display the HR measured from RR interval of each heartbeat. Selecting Average will display the HR measured from 6 seconds of heartbeat for adult and child, and 3 seconds of heartbeat for neonate.
ECG Source The ECG source to display on the home display can be selected. The alarm will be generated based on this selection. The tabular trend and graphic trend will be also based on this selection. 1. Press the Config. key to display the setup menu for selecting the ECG source. 6 2. Select a parameter. Selecting ECG will measure the HR from ECG. “HR” will be displayed inside the parameter key. ECG Selecting SpO2 will measure the PR from SpO2. “PR_SpO2” will be displayed inside the parameter key.
1. Press the Config. lead switching. → Page Down keys to display the setup menu for setting the auto 2. Select ON or OFF . OFF will not switch the lead when an electrode comes off. ON will automatically switch to another lead when an electrode comes off. Pacemaker Pulse The artificial pace pulse can be displayed by superimposing on the ECG waveform. The artificial pace pulse will be displayed in yellow. Pacemaker Pulse Detection Algorithm ③ ①ECG Signal Input Inputs ECG signal.
1. Press the selection. Config. → Page Down key to display the setup menu for pacemaker pulse 2. Select ON or OFF . OFF will not display the pacemaker artificial pulse. ON will display the pacemaker artificial pulse in a different color from the ECG waveform. This will automatically set to ON when “Used” is selected for pacemaker use on the patient admit / discharge menu.
1. Press the Config. configuration menu. → Page Down keys to display the second page of the 2. Select the pace pulse mask time. Select from 10ms , 20ms , 40ms depending on the pace spike amplitude or presence of fusion beat. Selecting OFF will set the mask time to 0ms. Selecting Auto will switch between 20ms and 40ms depending on the pace spike amplitude.
ON/OFF of Parameter Display 1. Press the Display ON/OFF will appear. 2. Select Display ON or key. The confirmation display for ON/OFF of ECG display Display OFF . Pressing the Display ON key will display the waveform and numeric data. Pressing the Display OFF key will not display the waveform and numeric data. The Display OFF message will be displayed inside the parameter key.
− Respiration − This menu allows setup for the impedance respiration measurement and CO2 respiration measurement. Size RR Alarm APNEA Alarm Configuration : : : : Selects the waveform size to display impedance respiration. Selects ON/OFF of respiration rate alarm, and sets upper and lower alarm limits. Selects ON/OFF of apnea alarm and sets upper alarm limit. Sets the respiration monitoring configuration. ? When the following relay cables are used, respiration can not be measured.
RR Alarm 1. Press the RR Alarm key to display alarm setup menu. Set ON/OFF of RR alarm and upper and lower alarm limit. The alarm will be set common to RR measured from impedance respiration waveform or RR measured from CO2 waveform. The adjustable increment for upper and lower limit depends on the patient type.
Key Item Individual Setup Upper Alarm Limit Automatic Setup Description Selecting ON will generate the apnea alarm. Selecting OFF will not generate the apnea alarm. Sets the upper alarm limit (5〜20sec.). Setting a value equal to or above 20sec. will turn OFF the alarm. Sets the apnea alarm value set for the currently selected alarm mode.
Respiration Source The parameter to measure respiration rate and apnea time can be selected from impedance, CO2, or automatic. RR and apnea alarm will be generated according to the selected parameter. These will be also stored as graphic trend or tabular trend. 1. Press the Configuration key to display configuration menu for RR source selection. 2. Select the parameter. Auto will automatically select the parameter to measure the respiration rate with the priority order of CO2>impedance.
RR Synchronization Mark The RR mark synchronized to impedance respiration or CO2 waveform will be displayed inside the parameter key. RR Mark 1. Press the Mark. Configuration key to display the setup menu for setting the RR synchronized 2. Select ON or OFF . OFF will not display the synchronization mark. ON will display the synchronization mark.
ON/OFF of Parameter Display 1. Press the Display ON/OFF will appear. 2. Select Display ON or key. The confirmation display for ON/OFF of RESP display Display OFF . The Display OFF message will be displayed inside the parameter key. When ECG electrodes are attached to the patient with the respiration display set to OFF, the respiration waveform and numeric data will be automatically displayed after 30 seconds.
− Invasive Blood Pressure (BP1, BP2)− This menu allows setup of the measurement condition for BP1, BP2. Scale BP Alarm : Selects the scale for BP waveform display. : Sets the upper and lower alarm limit of systolic, diastolic, mean blood pressure and ON/OFF of the alarm. BP Zero : Performs zero balance. Configuration : Sets the BP monitoring condition. When the main power is turned ON, the BP value will not be displayed until C A U T I O N zero balance is performed. Make sure to perform the zero balance.
BP Alarm (BP1, BP2) 1. Press the BP Alarm key to display the alarm setup menu. Select ON/OFF of BP alarm and set the upper and lower alarm limit for systolic (SYS), diastolic (DIA), and mean (MEAN) BP. The alarm value is to be set for each measurement unit. (mmHg / kPa) The adjustable increment for upper and lower limit changes from 50mmHg / 7kPa. mmHg : 0〜50mmHg / 2mmHg increment 55〜300mmHg / 5mmHg increment kPa : 0〜7kPa / 0.2kPa increment 7.5〜40.0kPa / 0.
Verify the BP waveform is positioned at zero, and “0” is displayed for the BP value. A message, “BP zero complete” will be displayed when the procedure is complete. A message, “BP zero failed” will be displayed when the process fails. The three-way cock may not be opened to air, artifact is present, or the transducer may be defective. Check the cause and try the zero balance procedure again. A message, “BP zero drift” will be displayed when the interface cable is not connected.
If the mean BP display is set to OFF, the mean BP alarm will not be C A U T I O N generated. Also, the mean BP will not be displayed on the tabular trend. Be cautious when setting the mean BP display OFF. ECG Source (BP1) The HR/PR source to display on the home display can be selected. The alarm will be generated based on this selection. The graphic trend and tabular trend will be also stored based on this selection. BP2 can not be set as ECG source. 1.
ON/OFF of Parameter Display 1. Press the appear. 2. Select Display ON/OFF Display ON or key. The confirmation display for ON/OFF of BP display will Display OFF . Pressing the Display ON key will display the waveform and numeric data. Pressing the Display OFF key will not display the waveform and numeric data. The Display OFF message will be displayed inside the parameter key.
− SpO2 − This menu allows setup of the measurement condition for the SpO2. Size : Sets the waveform size for SpO2 waveform display. SpO2 Alarm : Sets ON/OFF of alarm, upper and lower alarm limit, and SEC alarm. Configuration : Sets the SpO2 monitoring configuration. 6 SpO2 Take the following precautions when monitoring over long periods of time. CAUTION ? To avoid skin rash or low-temperature burn, it is recommended to change the measurement position several times a day.
SpO2 Waveform Size 1. Press the Size key to display the SpO2 waveform size setup menu. Select the waveform size for displaying and recording. 2. Select the waveform size. Select the size from × 1/4 , × 1/2 , × 1, × 2, × 4. SpO2 Alarm 1. Press the SpO2 Alarm key to display the alarm setup menu. Select ON/OFF of SpO2 alarm, and set the upper and lower alarm limit.
ECG Source The HR/PR source to display on the home display can be selected. The alarm will be generated based on this selection. The graphic trend and tabular trend will be also stored based on this selection. 1. Press the Configuration key to display the setup menu to set the ECG source. 2. Select a parameter. Selecting ECG will measure the HR from ECG. “HR” will be displayed inside the parameter key. 6 Selecting SpO2 will measure the PR from SpO2. “PR_SpO2” will be displayed inside the parameter key.
ON/OFF of Parameter Display 1. Press the Display ON/OFF will appear. 2. Select Display ON or key. The confirmation display for ON/OFF of SpO2 display Display OFF . Pressing the Display ON key will display the waveform and numeric data. Pressing the Display OFF key will not display the waveform and numeric data. The Display OFF message will be displayed inside the parameter key.
− Non-Invasive Blood Pressure − This menu allows the setup of NIBP monitoring condition. Auto Mode : Sets the automatic interval measurement and starts the 1-minute interval measurement and Quick SYS measurement. NIBP Alarm : Sets the ON/OFF of NIBP alarm and upper / lower limit of systolic, diastolic, and mean BP. Configuration : Sets the NIBP monitoring configuration. NIBP For the following situation, measurements will be terminated.
2. Select an interval time. Select from 2 min / 2.5 min / 3 min / 5 min / 10 min 30 min / 60 min / 120 min . Select OFF if not performing the interval measurement. / 15 min / 20 min / The measurement time will be the integral multiple of the selected interval time beginning with 0 minute. Ex.)If the present time is 13:14, the measurement time will be as follows for each interval time. 2 min. : 13:16, 13:18, 13:20, ... 2.5 min. : 13:15, 13:17:30, 13:20, ... 3 min. : 13:15, 13:18, 13:21, ... 5 min.
Quick SYS Start The NIBP measurement can be continuously performed for 3 min. / 5 min. / 10 min. If any abnormality on the cuff hose, etc. is found during the Quick SYS, the continuous measurement will be ceased. Only the systolic blood pressure will be measured and displayed. 1. Press the Auto Mode Quick SYS. key to display the measurement interval setup menu to start the Measurement Duration 6 NIBP 2. Start the Quick SYS. Pressing the Quick SYS Start key will start the continuous measurement.
Key Item Individual Alarm Description Selecting Selecting ON will generate the NIBP alarm. OFF will not generate the NIBP alarm. Select from SYS (systolic BP), DIA (diastolic BP), or MEAN (mean BP) Lower Alarm Limit Upper Alarm Limit Automatic Setup Sets the lower alarm limit (10〜295mmHg/1.5〜 39.5kPa). Setting a value 10mmHg/1.5kPa or below will turn OFF the alarm. Sets the upper limit (15〜300mmHg /2.0〜40.0kPa). Setting a value 300bpm/40.0kPa or above will turn OFF the alarm.
End of Measurement Tone By selecting ON for the “End Tone”, a tone will be generated when the NIBP measurement completes. 1. Press the Configuration the End Tone. key to display the NIBP configuration menu to set ON/OFF for 2. Select ON or OFF . ON will generate a tone when the measurement completes. OFF will not generate a tone when the measurement completes. 6 NIBP Quick SYS List The systolic blood pressure measured by Quick SYS can be included in the NIBP list. 1.
PR Display The measured pulse rate can be displayed. This selection is for display only, and alarm function and tabular trend function will be ineffective. 1. Press the display. 2. Select Configuration ON or key to display the NIBP configuration menu for setting the PR OFF . Selecting Selecting Pulse Rate ON will display the pulse rate. OFF will not display the pulse rate. Mean BP Display The ON/OFF of mean BP display can be selected. 1. Press the Configuration mean BP display. 2.
1-Minute Measurement Duration The duration for 1-minute measurement can be selected from 10 minutes or 20 minutes. When the previous measurement is prolonged due to patient motion, etc, the cuff pressure release time until the next measurement will be shortened, and the measured location may congest. Be cautious when performing long duration of continuous measurement. 1. Press the Configuration → Page Down keys to display the NIBP configuration menu for setting the 1-minute measurement duration. 2.
− Temperature − This menu allows the setup of the temperature monitoring condition. TEMP Alarm : Sets ON/OFF of temperature alarm, and upper and lower alarm limits. Temperature Alarm 1. Press the TEMP Alarm key to display the alarm setup menu. Select ON/OFF of temperature alarm, and set the upper and lower alarm limit. The alarm limit can be set for each measurement unit (?C / ?F). The upper and lower limit can be set in increments of 0.5?C / 0.5?F.
ON/OFF of Parameter Display 1. Press the Display ON/OFF will appear. 2. Select Display ON or key. The confirmation display for ON/OFF of TEMP display Display OFF . The Display OFF message will be displayed inside the parameter key. When the waveform and numeric data display is set to OFF, the alarm C A U T I O N generation and tabular/graphic trend will be also set to OFF. 6− 37 6 Temperature Pressing the Display ON key will display the numeric data.
− CO2 Concentration − (DS-7141) This menu allows setup of CO2 concentration measurement. Scale : EtCO2 Alarm : InspCO2 Alarm : Configuration : Sets the CO2 waveform scale. Sets ON/OFF of EtCO2 alarm, and upper and lower alarm limits. Sets ON/OFF of InspCO2 alarm and upper alarm limit. Sets CO2 monitoring conditions. CO2 Scale 1. Press the Scale key to display the scale setup menu. <Scale setup menu for the unit in mmHg> 2. Select the CO2 waveform scale for displaying and recording.
Key Item Individual Alarm Lower Alarm Limit Upper Alarm Limit Automatic Setup Description Selecting ON will generate the EtCO2 alarm. Selecting OFF will not generate the EtCO2 alarm. Sets the lower alarm limit (1〜98mmHg, 0.1〜13.1kPa, 0.1〜13.1%). Setting a value equal to or below 1mmHg, 0.1kPa, 0.1% will turn the alarm OFF. Sets the upper alarm limit (3〜100mmHg, 0.4〜 13.3kPa, 0.3〜13.3%). Setting a value equal to or above 100mmHg, 13.3kPa, 13.3% will turn the alarm OFF.
EtCO2 Average Duration The duration to average the EtCO2 value can be selected from 10 sec., 20 sec., 30 sec., or OFF. 1. Press the selection. Configuration key to display configuration menu for EtCO2 average duration 2. Select the average duration. Select the duration to average the EtCO2 value from 10sec , 20sec , 30sec . If OFF is selected, EtCO2 value for each respiration will be displayed.
CO2 Calibration CO2 calibration can be performed using calibration gas. Calibration should be conducted every 6 months or when any measurement error is found. 1. Press the display. Configuration key to display the configuration menu for CO2 Cal. key <Preparing for calibration> 3. Press the Start Cal <Start calibration> key and conduct calibration according to the displayed messages. 4. The message, “Feed CAL. GAS” will be displayed. Press the injection button to inject the calibration gas. 5.
Restarting the CO2 Unit The sampling tube will cease functioning when erroneous condition such as blocking of exhaust tube, sampling tube or nasal prong is detected. When the pump ceases functioning, “Check CO2 unit” message will be displayed. After resolving the problem, press the Restart CO2 key and restart the measurement. 1. Press the Configuration key to display the configuration menu for Restart CO2 key display. 2. Press the Restart CO2 key for 2 seconds.
2. Select Display ON or Display OFF . Pressing the Display ON key will display the waveform and numeric data. Pressing the Display OFF key will not display the waveform and numeric data. The Display OFF message will be displayed inside the parameter key. When filter line is attached to the patient with the CO2 display set to OFF, and 2 or more respiration is detected within 30 seconds, the CO2 waveform and numeric data will be automatically displayed.
Blank Page 6− 44
Chapter 7 Function This chapter describes the functions such as arrhythmia analysis, trend, and recall. Definition, etc. ・・・・・ 2 − Arrhythmia Analysis − Arrhythmia Definition ・・・・・・・・・・・・・・・・・・・・・・・・・・ 2 − − − ST Display, Alarm Setup, etc.
− Arrhythmia Analysis − Definition, etc. This section explains the arrhythmia analysis and alarm setup procedure. Arrhythmia Definition The arrhythmia detection is performed by learning the normal waveform of the patient and determining VPC by comparing the waveform (QRS pattern) and R-R interval of each heart beat. A pattern matching is performed with the VPC detected from R-R interval, QRS amplitude, QRS area, QRS polarity, etc., and determines as VPC after discriminating the noise and VPC.
WARNING Objective and constant arrhythmia detection is possible through the fixed algorithm incorporated in this monitor. However, excessive waveform morphology change, motion artifact, or the inability to determine the waveform pattern may cause an error, or fail to make adequate detection. Therefore, physicians should make final decisions using manual recording, alarm recording and recall waveform for evaluation.
<Arrhythmia Reference Range> Arrhythmia Reference Range Default ASYSTOLE 3 sec.〜10 sec. 5 sec. RUN 2 beats〜8beats 3 beats PAUSE 1.5 sec. 〜5 sec. 3 sec. FREQUENT 1 beat〜50 beats/min. 10 beats/min. 3. Select ON or OFF for the alarm. Alarm will generate. Alarm will not generate 4. Select ON or OFF for recall factor. ON/OFF of recall factor can be set on the alarm setup menu. Will be selected as recall factor. Will not be selected as recall factor.
Pressing the key while learning arrhythmia will not stop the learning. 3. During arrhythmia learning, a message will be displayed.
− Graphic Trend Data− Display / Record This section explains the graphic trend function and recording procedure. To Display the Graphic Trend The graphic trend menu can be accessed from the menu, or from the preprogrammed user key. To display from the user key. To display from the menu The 24 hours of graphic trend data in 1-minute interval will be automatically stored and displayed if the data is displayed on the home display. 1.
2. Select the parameter to display. Pressing the parameter selection key will display the selection for display. Select a parameter and press the Parameter EVENT1 EVENT2 key.
Parameter Scale HR ST VPC BP1, BP2 NIBP SpO2 PR_ SpO2 TEMP RR APNEA CO2 EVENT Unit 100, 200, 300 ± 0.2, ± 0.5, ± 1.0, ± 2.0 bpm mV ± 2, ± 5, ± 10, ± 20 20, 50, 100 20, 50, 100, 150, 200, 300 4, 8, 16, 20, 24, 40 100, 150, 200, 300 16, 20, 24, 40 0〜100, 50〜100, 80〜100 100, 200, 300 20〜45, 30〜40 mm beat mmHg kPa mmHg kPa % bpm 68〜113, 86〜104 50, 100, 150 15, 30 4.0, 8.0, 10.0 50, 100 none ?F Bpm Sec %, kPa mmHg ?C 4. Select the display time range.
6. Move the cursor. Scrolls to past Scrolls to present The data of selected time can be displayed by moving the cursor. Pressing the ? key will scroll to the past data with the selected time range. Pressing the ? key will scroll to the present data. 7. Enlarge the display. Pressing the Zoom key will display the 1-hour data with the cursor time in center. Directly pressing the graphic trend display area will also display the 1-hour data with the pressed time in center. 8. Store the graphic trend data.
The Description of the Display Cursor Time Cursor Scale Measurement Data at Cursor Time Real time scale The measured data will be compressed for the 12-hour / 24-hour display.
− Tabular Trend − Display / Record This section explains the tabular trend function and recording procedure. To Display the Tabular Trend The tabular trend menu can be accessed from the menu, or from the preprogrammed user key. The 24 hours of data in 1-minute interval will be automatically stored and displayed if the data is displayed on the home display. 7 1. Press the Menu Tabular Trend Display from the user key. Display from the menu → Tabular Trend keys to display the tabular trend. 2.
3. Shift the page. Shift to past Shift to present The page can be shifted past or present by page with the selected time range. Pressing the ? key will shift one page to the past data. Pressing the ? key will shift one page to the present data. The data will be listed in 8 columns. If 5-minute time range is selected and the starting time on the list is 10:00, 35 minutes from 10:00 to 9:25 will be listed in 1 page. Pressing the ? key will display the list from 9:20 to 8:45. 4. Shift the displayed column.
Parameter Setup for Tabular Trend The parameters for tabular trend can be selected. 1. Press the menu. List Setup key on the tabular trend menu to display the tabular trend setup 2. Select the position on the list. 7 3. Select the parameter for display. Select the parameter to display for the previously selected position. The position will automatically shift downward so that consecutive parameter selection is possible. 7− 13 Tabular Trend Select the position.
− Recall Data− Display / Record This section explains the recall menu function and recording procedure. To Display the Recall Menu The recall menu can be accessed from the menu, or from the preprogrammed user key. Display from the menu Display from the user key When the alarm factor assigned on the recall setup occurs, the assigned waveform and the numeric data at alarm occurence can be stored for up to 100 data. The recall data to be displayed can be selected on the display selection menu.
2. Select the recall factor to display on the recall list. Press the Display Selection key and select the recall factor. Select the numeric data, arrhythmia to display as recall factor. If the key LED is lighted, recall data will be displayed. If the key LED extinguished, recall data will not be displayed. 3. Shift the recall list display. The newest 5 data will be displayed from the recall list. Shift the recall list to newer data by 1 page (5 data). 7 4. Press the Recall Setup the recall setup menu.
1. Pressing one of the recall factors will display the enlarged recall waveform. QRS Classification Recall Waveform 1 Recall Waveform 2 Scale in seconds Measured Data 2. Shift the waveform left or right. The recall waveform display can be shifted to left or right. ? key will shift to the older data. ? key will shift to the newer data. 3. The alarm factor occurred at the same time will be displayed. Pressing the recall factor key will display the recall factor occurred at the same time. 4.
5. Erase the recall waveform. Erase the unnecessary recall waveform. Pressing the Erase key will display the confirmation message. YES will erase the waveform and displays the recall list display. NO will return to the previous display. To Set the Recall Condition On the recall menu, the storing condition at alarm occurrence can be set. The recall waveform and recall factor (numeric data, arrhythmia) can be selected. 7 Recall Data 1. Select the recall waveform.
Select the recall factor by pressing the keys. The key with the LED lighted will be the recall factor. 3. Select the recall factor (arrhythmia). Pressing the Arrhy. key will display the menu to select the arrhythmia alarm factor. Select an arrhythmia for recall factor. The key with LED lighted will be the recall factor. The recall waveform will start with the following delay time tracing back from the alarm occurrence. NOTE Delay Time 7− 18 Adult Child 12sec. 12sec. Neonate Meas. Data Alarm Arrhy.
− OCRG − Display This section describes the procedure for OCRG display. The OCRG display can be accessed from the menu, or from the preprogrammed user key. Display from the menu Display from the user key On the OCRG display, compressed respiration waveform, HR trend and SpO2 trend are displayed simultaneously. Menu → OCRG OCRG 1. Press the keys to display the OCRG menu. The trend scale is a fixed scale. SpO2 trend HR : 0〜300bpm SpO2 : 70〜100% HR trend Compressed respiration waveform 2.
− ST Display − ST Display, Alarm Setup, etc. This section describes the operation procedure for the ST display and alarm setup. To Display the ST Measurement Menu The ST display can be accessed from the menu, or from the preprogrammed user key. Display from the menu Display from the user key On the ST display, the averaged ECG waveform of 16 beats will be superimposed for 5 minutes. 3 frames of superimposed waveform will be displayed.
4. Select the trend scale. Select the displayed scale for the trend. Trend HR ST Scale Unit 100, 200, 300 ± 0.2, ± 0.5, ± 1.0, ± 2.0 bpm mV ± 2, ± 5, ± 10, ± 20 mm ● To Set the Reference Waveform Set the reference waveform for the ST display and set the reference point and measurement point on the reference waveform. 1. Press the Menu → waveform setup menu. ST Display → Reference Size / Lead keys to display the reference Measurement Point ST Measurement Reference Point 2.
● ST Alarm Setup The ST upper value and lower value compared with the reference waveform will be set. The alarm value is to be set for each measurement unit (mm / mV). The upper and lower limit can be set in 1mm / 0.1mV increment. 1. Press the Menu → ST Display → Alarm keys to display the alarm setup menu. Upper limit cursor Lower limit cursor 2. Set the upper and lower alarm limit. Use the Item Lower Alarm Limit Upper Alarm Limit ?, ? keys to adjust the alarm limit.
− NIBP List − Display / Record This section explains the NIBP list function and recording procedure. To Display the NIBP List The NIBP list display can be accessed from the menu, or from the preprogrammed user key. 7 Display from the menu On the NIBP list, NIBP data and HR, SpO2 pulse rate, SpO2 value at the commencement of NIBP measurement will be stored and displayed for 120 NIBP measurements. 1. Press the Menu → NIBP List keys to display the NIBP list display. 2. Shift the displayed column.
The Description of the Display The mean BP will be displayed on the NIBP list only if it is displayed on the home display. If HR or SpO2 is not measured, or not correctly measured at the commencement of NIBP measurement, the measured data will be displayed as “? ? ? ”. For Quick SYS measurement, only the SYS (highest BP value) will be displayed. NOTE 7− 24 If the NIBP measurement was not completed, the data will not be displayed on the NIBP list.
− Other Bed ― Display / Alarm This section explains about the function to display the waveform and numeric data and to set alarms for other bedside monitors. To use this function, wired network connection is required. Other Bed Display The other bed display can be accessed from the menu or from the preprogrammed user key. Also, by setting the other bed alarm ON, Other Bed Alarm key will be displayed when other bedside monitor generates an alarm.
Other Bed Alarm Setup From the 48 bedside monitors, the bed to generate the other bed alarm and ON/OFF of other bed alarm display can be performed. 1. Select the bed to generate the other bed alarm. Press the Set other alarm key to display the other bed alarm setup menu. Select the bed to generate the other bed alarm. The key LED for the bed selected as the other bed alarm generating bed will be lighted. 2. Select ON for the other bed alarm. Press the Prev. Disp.