FUJI FILM DIGITAL RADIOGRAPHY For Safe Operation System FUJI FILM DR overate Configuration FDR D-EVE (Product Overview) | (DR-ID 600) . Basic Operation Operation Manual 48th Edition : January 2014 Troubleshooting Daily Inspection and Maintenance Appendix | i Maintenance and | Inspection | “This Operation Manual describes details on how to operate the FDR D-EVE and cautions to be observed when operating it.
| Introduction This Operation Manual includes descriptions of matters necessary when using the FDR D-EVE, such as the equipment overview, operation procedures and precautions to observe, as well as daily inspections and maintenance. Accompanying documents were originally drafted in the English language. Installation may only be conducted by authorized service personal. indications for use {for U.S.
| FDR D-EVE System Operation Manuals For the U.S. DIGITAL RADIOGRAPHY FDR D-EVE (DR-ID 800) Operation Manual DR-ID 600PU Operation Manual DR-ID 300CL Operation Manual Access Point Operation Manual For other countries DIGITAL RADIOGRAPHY [ae ia Boa {eS ve Re TIE BUR LETTER) DR-ID 660PL Operation Manual DR-D 300CL Operation Manual *1 See the operation manual for the designated access point. © See “DR-ID 300CL Operation Manual” along with the manuals for the related products.
| Contents at a Glance For Safe Operation “This chapter presents Warnings and Cautions we wish you to observe for safe operation of the FDR D-EVE. System Configuration (Product Overview) This chapter gives the various unit names and describes their functions and features of the FOR D-EVE. Basic Operation This chapter describes start-up, shut-down and other basic operations of the FDR D-EVE.
| How to Read This Manual Basic page layout Please have a good grasp of the basic page configuration of this Operation Manual, as illustrated below, for you to use it more efficiently. Shaw the tilts of an aeration procedure Describes Information we wish you to know described in the section. in advance of your operating the system or information that may help you to operate &.
information items to be observed when you are operating this system and the supplementary remarks are described in this manual with the respective marks. For the safe system operation, be sure to observe Contamination.
[ Contents Introduction FDR D-EVE System Operation MANUALS iii iv Contents at a Glance How to Read This Manual . Chapter 1 For Safe Operation Precautions Before Operating This Equipment Precautions to be Observed When Using the Electric Medical Equipment. Electromagnetic Compatibility (EMC) 1.4.1 DR-ID 14.2 DR-ID 600MC and DR-ID 300CL. Precautions in Using the FDR D-EVE. 1.5.4 Handling. 1.5.2 Before Exposure 1.6.3 During Exposure 1.54 During Cleaning 155 Storage 1.5.
3.2 33 3.1.7 Changing the Direction of the Flat Panel Sensor Connector 3.1.8 Charging the Battery Pack (Optional) for the Flat Panel Sensor {DR-ID 801ST, DR-ID 602SE, DR-ID 811SE, DR-ID 6125E and DR-D 3.1.9 Installing/Removing the Battery Pack {Optional} for the Flat Panel Sensor (DR-ID 601ST, DR-ID 602SE, DR-ID 811SE, DR-ID 612SE and DR-ID 6138E) Starting Up and Shutting Down the System 3.21 Starting Up the System... 3.22 Shutting Down the System .
Appendix Z Precautions for Exposure 2.1 Precautions for Exposure in AUTO MODE Z11 Radiation Field.. 2.1.2 Depiction of the Cervical Region Z.1.3 Depiction of the HIP JOINT AXL — 2 Menu 2.1.4 EDR image Data Analysis Z.2 Precautions for Exposure in SEMI-AUTO MODE 2.3 Precautions for Exposure in SEMI-X MODE ZA Precautions for Exposure in FIX MODE... 2.5 Precautions for the Automatic X-ray Detection Function Z.5.1 Precautions for Making an Exposure. 2.52 Precautions Related to the X-ray Exposure Z.
chapter 1 For Safe Operation 1.1 Precautions Before Operating This Equipment Before using this equipment, please read “Precautions Before Operating This Equipment’ carefully so that you can operate it correctly. Whenever you operate this equipment, be sure to observe those precautions. Failure to do so may cause you to subject to injuries or property damage to occur. The institution where the equipment is installed is responsible for its use and maintenance.
rr endogenous seq sod 1.2 Precautions to be Observed When Using the Electric Medical Equipment We ask that you observe these usage precautions and use the equipment correctly, 1, This equipment should be used only by people who have the proper skills. 2. Observe the following precautions when installing the equipment. 2-1. Install the equipment where water will not splash it. 2-2.
| 1.3 Safety Before using the FDR D-EVE, read this section thoroughly to ensure that you use the product property. Electric Shock Warnings A WARNING The power supply to the FDR D-EVE Is AC100 to 240V. To avoid electric shocks, users should always take the following precautions: « Never open any covers of the equipment, Install the equipment in a location where it will not be exposed to water. Check that the equipment is securely earthed.
BISHOPRIC AES 104 mw A WARNING — em Do not disassemble or remodel the equipment. Otherwise, fire or electric shock may result. Keep away from the parts inside the product, which may cause electric shock. if you touch them accidentally, death or severe injury may result. A WARNING Do not hit or drop the equipment or subject it to severe shock. Otherwise, the equipment may be damaged. If the damaged equipment is used, fire or electric shock may result.
A WARNING Flammable gasses may stay in the room after disinfection. Ventilate the room well before powering on the system following disinfection. Warnings for Abnormalities A WARNING a If any of the following occurs, immediately turn off the power of each unit, unplug the power cable from the outlet, and then contact a FUJI FILM dealer. « When smoke, strange odor, or abnormal sound is present. * When a foreign object {such as a metal object) or liquid enters the product.
Oneida SRS 104 Evenness ) When a setting of the network to which the equipment is connected has been changed, check that the change does not affect the system operation and take measures if necessary. The setting change may include the following: Change of connection destination * Addition of devices + Removal of devices « Update of devices « Upgrade of devices Cautions on Network AN Cautions Connect to the Ethernet Network of 1000BASE-T, 100BASE-TX, or ABASE-T prescribed in the IEEE standard 802.3.
Software Precaution (AN cautions Do not install additional software to the system. Do not uninstall any of the software preinstalled in the system. The system is preinstalled with the appropriate software. If other software Is installed or if the existing software is uninstalled, various operational errors may result. Disinfection Instructions dialogue feign 104 ume AN WARNING Confirm that the respiratory density of disinfectant including solvent Is under legal regulation.
Dolores 28g 104 + some » While charging the battery, do not allow the battery pack or battery charger get wet or dusty. +» Do not step on the AC adapter of the battery charger. Also, be careful not to trip over the power cable, « Do not subject the battery and battery charger to severe shock (by dropping them, etc.) « Do not place the battery charger within the reach of patients. « Do not charge the battery pack near fire or under strong sunshine.
Do not use or leave the battery pack in a place where it is exposed to high temperature (80°C or higher), such as fire or a heater. If the resin separator is damaged due to heat, the battery pack may short-circuit, causing it to overheat, emit smoke, explode or ignite. Do not immerse the battery pack in water or seawater, and do not allow it fo become wet. If the built-in protection mechanisms are damaged, the battery pack may overheat, emit smoke, explode or ignite.
fe uoeiad) Jeg Jog Warnings for Pediatric Use A WARNING « If the exposure conditions for average-size adults are applied to children, it may cause excessive radiation exposure. « Studies show that children are more radio sensitive than adults (i.e. children are at higher risk of developing cancer compared to adults exposed to the same dose of ionizing radiation).
I 1.4 Electromagnetic Compatibility (EMC) & Essential performance {1) DR-IL} 8028E/DR-D 6138E obtains images. (2) DR-ID 800MC stores images. (3) DR-ID 800MC corrects images. (4) Image transfer in order from DR-D 6028E/DR-D 61288/ DR-ID 6138E, DR-ID G0GMP, DR-ID 800MC, to DR-ID 360CL (5) DR-ID 300CL stores and displays images after correction. (6) 1t shall fill and maintain the safety requirement of the standards.
headphone JPEG 10d meeting Further information for IEC 60601 (ENG0601~ ) 1. Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the accompanying documents. 2. Portable and mobile RF communications equipment can affect medical electrical equipment. 3. Information regarding the cable affecting EMC is as follows.
Table 2 Guidance and manufacturer's declaration electromagnetic immunity The DR-ID B00PU is intended for use in the electromagnetic environment specified below. The customer or the user of the DR-ID 600PU should assure that iris used In such an environment. Immunity test EC 60601-1-2 Compliance level Electromagnetic environment « test level guidance Electrostatic Floors should be woad, concrete or discharge 6KV contact #6KV contact ceramic tie.
ral seasoned jests Jog Table 3 Guidance and manufacturer's declaration electromagnetic immunity The DR-ID BOOHOO is intended for use in the electromagnetic environment specified below. The customer or the user of the DR-ID 800PU should assure that it is used in such an environment.
Table 4 Recommended separation distances between Portable and mobile RF communications equipment and the DR-D 800PU The DR-ID 600PU is intended for use In the electromagnetic environment in which radiated RF disturbances are controlled, ‘The customer or the user of the DR-ID 600PU can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the DR-1 800PU as recommended below, according to the maximum output power
labeled ares 04 TETRAS 1.4.2 DR-ID 600MC and DR-ID 300CL This equipment has been tested and found to comply with the international standard for medical devices below, according to the requirement of the 1EC 60601-1-2. EMC Standard : CRISP 22/EN 56022 CRISP 24/EN 55024 This does not guarantee that there is no harmful electromagnetic interference under any installation environment. This equipment can generate, use and radiate radio frequency energy.
Table 1 Guidance and manufacturer's declaration electromagnetic emissions environment. The DR-ID 800MC and the DR-ID 300CL are intended for use In the electromagnetic environment specified below.
unworried) RS I04 Menkent Table 3 Guidance and manufacturer's declaration « electromagnetic immunity The DR-ID 800MC and the DR-ID 300CL are intended for use in the electromagnetic environment specified below, ‘The customer or the user of the DR-ID B00MG and the DR-ID 300CL should assure that they are used in such an environment.
I 1.5 Precautions in Using the FDR D-EVE This section describes the precautions In using the FDR D-EVE. 1.5.1 Handling Handle the flat panel sensor carefully since it is manufactured with precision. if the flat panel sensor or the SE cable connector is hit or dropped or is subject to severe shock, it may be damaged. ¥f the front and rear of the flat panel sensor are subject to impact by a projection, it may be damaged.
Sr nonplussed ekes 104 If any of the screw caps on the flat panel sensor comes off, attach a spare cap. Otherwise, artifacts may appear in the image due to static electricity. To ensure optimal image quality, # is recommended that you do not use the flat panel sensor near devices {motor, transformer, switching supply, etc.) that generate electromagnetic noise.
if an exposure is made with the front and rear of the fiat panel sensor facing the other way round, re-exposure and electric parts may be damaged. Da not use the flat panel sensor, which is communicating with the power supply unit in wired communication mods, for the radio graphic examination stand equipped with an automatic goading function. 1.5.3 During Exposure Before making an exposure, make sure that exposure conditions most appropriate for this system are set.
Cassiopeia jeS 404 TREE 1.5.6 Precautions Related to the Load Applied to the Flat Panel Sensor If excessive load is applied to the fiat panel sensor, use it on a flat floor or platform. When making an exposure for the patient in a wheelchair or adjustable bed or on a stretcher, the flat panel sensor may be deformed (deflection by some millimeters). Fiat panel sensor if DR-ID B00SE is used in such an exposure, insert the flat panel sensor inch the cassette holder (optional).