User’s manual Flir CM83 600 A power clamp
User’s manual Flir CM83 #T559825; r.
Table of contents 1 Disclaimers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 Copyright. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.2 Quality assurance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.3 Documentation updates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.
1 Disclaimers 1.1 Copyright 1.3 Documentation updates © 2013, Flir Systems, Inc. All rights reserved worldwide. No parts of the software including source code may be reproduced, transmitted, transcribed or translated into any language or computer language in any form or by any means, electronic, magnetic, optical, manual or otherwise, without the prior written permission of Flir Systems.
2 Safety information Note Before operating the device, you must read, understand, and follow all instructions, dangers, warnings, cautions, and notes. Note Flir Systems reserves the right to discontinue models, parts or accessories, and other items, or to change specifications at any time without prior notice. Note Remove the batteries if the device is not used for an extended period of time. WARNING Do not operate the device if you do not have the correct knowledge.
2 Safety information WARNING Do not measure the current on a circuit when the voltage increases to more than 600 V. This can cause damage to the instrument and can cause injury to persons. WARNING You must disconnect the test leads from the circuit that you did a test on before you change the range. If you do not do this, damage to the instrument and injury to persons can occur. WARNING Do not look directly into the laser beam. The laser beam can cause eye irritation.
2 Safety information WARNING Be careful when you do the measurements if the voltages are more than 25 VAC rms or 35 VDC. There is a risk of shock from these voltages. Injury to persons can occur. WARNING Do not do diode, resistance or continuity tests before you have removed the power from the capacitors and from a device during a test. Injury to persons can occur. WARNING Do not use the device as a tool to identify live terminals. You must use the correct tools.
2 Safety information WARNING Do not cause a short-circuit of the batteries. This can cause damage to the instrument and can cause injury to persons. WARNING Do not put the batteries into a fire. Injury to persons can occur. CAUTION Do not use the device for a procedure that it is not made for. This can cause damage to the protection. This symbol, adjacent to another symbol or terminal, indicates that the user must refer to the manual for further information.
3 Introduction Thank you for choosing a Flir CM83 clamp meter. This device is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. 3.1 Key features • • • • • • • • • • • • • • • • • • • • • • • • • • • • 10 000-count digital display. Active backlit, large-scale display. Analog bar graph. True RMS reading in AC and AC+DC mode. Torch lightening when clamping. Auto AC/DC 1000 A capability and selection. Auto AC/DC 1000 V capability and selection.
4 Description 4.1 Meter parts Figure 4.1 Front view 1. 2. 3. 4. 5. 6. 7. 8. Clamp jaw. Jaw opening trigger. Function buttons, see section 4.3 Function buttons, page 9. Navigation buttons. Work light. Function switch, see section 4.2 Function switch, page 8. LCD display. Probe/thermocouple terminals. #T559825; r.
4 Description Figure 4.2 Rear view 1. Battery compartment. 4.2 Function switch The meter can measure capacitance through the probe inputs. The meter can measure resistance, continuity, or diode polarity through the probe inputs. The type of measurement is selected by the button. The meter can measure power through the probe inputs and the clamp jaws. The meter can measure current through the clamp jaws. #T559825; r.
4 Description The meter can measure voltage through the probe inputs. The meter is in full power-saving mode. 4.3 Function buttons • • • • Use the button to select Auto select or Manual select mode, see section 5.2 Auto/Manual select mode, page 12. In Manual select mode, press the button to select the operating mode. Use the button to select Auto range or Manual range mode, see section 5.3 Auto/Manual range mode, page 13. In Manual range mode, press the button to change the range (scale).
4 Description 4.4 Display icons and indicators Figure 4.3 Display Indicates that MeterLink (Bluetooth) communication is active, see section 5.2 Auto/Manual select mode, page 12. Indicates that the meter is in Auto select mode. Indicates that the meter is displaying maximum reading values. Indicates that the meter is displaying minimum reading values. Indicates that the meter is displaying the average reading. Indicates that the meter is displaying peak maximum values.
4 Description Indicates that the meter is in Hold mode. Indicates the battery voltage status. Indicates that the auto power off function is enabled. Indicates that the measured voltage is greater than 30 V DC or AC RMS. Indicates that the meter is measuring AC current or voltage. Indicates that the meter is measuring DC current or voltage. Indicates that the meter is measuring AC+DC current or voltage. Indicates that the continuity function is active. Indicates that the diode test function is active.
5 Operation Note Before operating the device, you must read, understand, and follow all instructions, dangers, warnings, cautions, and notes. Note When the meter is not in use, the function switch should be set to the position. Note When connecting the probe leads to the device under test, connect the negative lead before connecting the positive lead. When removing the probe leads, remove the positive lead before removing the negative lead. 5.1 Powering the meter 1. 2.
5 Operation • To enter Manual select mode, press the operating mode, press the • button. To manually select the button repeatedly. To enter Auto select mode, press and hold the cator is displayed. button until the indi- 5.3 Auto/Manual range mode In Auto range mode, the meter automatically selects the most appropriate measurement scale. In Manual range mode, the desired range (scale) is set manually. • To enter Manual range mode, press the press the • button.
5 Operation 5.5.2 Basic current measurements WARNING Do not measure the current on a circuit when the voltage increases to more than 600 V. This can cause damage to the instrument and can cause injury to persons. Note If the measured voltage is greater than 30 V DC or AC RMS, the is displayed. indicator When measuring current using the clamp jaws, only one conductor should be enclosed by the jaws—refer to Figure 5.1. Figure 5.1 Correct and incorrect setup 1. 2. 3. 4.
5 Operation 5.5.3 Extended functionality In addition to the basic voltage and current measurements, the meter can be set to different modes for extended functionality. 5.5.3.1 Selecting the mode The mode icons applicable for the selected measurement type are displayed in the lower part of the display. When a mode is enabled, the icon is framed. Figure 5.
5 Operation 5.5.3.3 In-rush current mode In In-rush current mode, the meter displays the highest current reading in the first 100 ms after the trigger point. In-rush current mode is available when measuring AC current in Manual select mode. 1. Use the navigation buttons to select and enable In-rush current mode. Note If the in-rush current under testing may be more than 100 A AC, manually set the range to 600 A/1000 A before activating the in-rush current, see section 5.
5 Operation 3. Press and hold the button to disable Min/Max/Avg mode. 5.5.3.7 Harmonic Distortion mode In Harmonic Distortion mode, the meter displays the distortion percentage value for the first 25 harmonics as well as the total harmonic distortion. Harmonic Distortion mode is available when measuring AC current or voltage in Manual select mode. The harmonic distortion is expressed as Hn = (RMS of an individual harmonic n )/(RMS of the fundamentals) × 100%. 1. 2.
5 Operation 3. Press the trigger to open the clamp jaws. Fully enclose one conductor—refer to Figure 5.1. For optimum results, center the conductor in the jaws. Note The + symbol on the jaw should be directed toward the power source. 4. 5. Connect the probe leads in parallel to the part under test. Read the active power value on the display. • • If the value is displayed without a sign, the power is flowing from the power source to the load.
5 Operation Figure 5.3 Three-phase three-wire measurement 1. Set the function switch to the 2. Ensure that the meter is set to active power measurement. If the 3. indicator is displayed, press the button repeatedly until none of these indicators are displayed. Take two measurements of the active power, in accordance with Figure 5.3. position. or the 5.6.2.
5 Operation Figure 5.4 Three-phase four-wire measurement 1. Set the function switch to the 2. Ensure that the meter is set to active power measurement. If the 3. position. or the indicator is displayed, press the button repeatedly until none of these indicators are displayed. Take three measurements of the active power, in accordance with Figure 5.4. #T559825; r.
5 Operation 5.6.3 Phase rotation With the meter set to Phase rotation mode, it is possible to determine the phase rotation for a three-wire system. Note The system frequency must be stable. Figure 5.5 Phase rotation 1. Set the function switch to the 2. Enter Phase rotation mode by pressing the 3. indicator is displayed. Connect the red test lead to the presumed phase line 1 and the black test lead to the presumed phase line 3. #T559825; r.8008/8011; en-US position.
5 Operation 4. One of the following results is displayed: • • • • 5. 6. OLU flashes if the voltage is >1000 V. LoU flashes if the voltage is <30 V. outF flashes if the frequency is >65 Hz or <45 Hz. If normal, L1 flashes for about 3 seconds. Then L2 is displayed. Move the red test lead to the presumed phase line 2. One of the following results is displayed: • • • 123 indicates clockwise or forward rotation, which means that the presumed phase line 1 is ahead of the presumed phase line 2.
5 Operation 1. 2. 3. 4. Set the function switch to the position. Insert the black probe lead into the negative COM terminal and the red probe lead into the positive terminal. Touch the tips of the probe across the part under test. Read the capacitance value on the display. Note For very large capacitance values, it may take several minutes for the measurement to settle and the final reading to stabilize. 5.
5 Operation 3. 4. 5. 6. 7. Use the button to select the diode test function. The indicator will be displayed. Touch the tips of the probe across the diode or semiconductor junction under test. Make a note of the value on the display. Reverse the polarity of the probe, by interchanging the probe test locations. Touch the tips of the probe across the diode or semiconductor junction under test. Make a note of the new value on the display.
6 Maintenance 6.1 Cleaning and storage Clean the meter with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for an extended period, remove the batteries and store them separately. 6.2 Battery replacement 1. 2. 3. 4. To avoid electrical shock, disconnect the meter if connected to a circuit, remove the probe/thermocouple leads from the terminals, and set the function switch to the position before attempting to replace the batteries.
7 Technical specifications 7.1 Model specifications AC A Ranges 100.00 A, 600.0 A Resolution 0.01 A Basic accuracy ±(1.5% + 5 digits) at 50–500 Hz Conversion type: average sensing RMS indicating • • AC+DC True RMS DC A Ranges 100.00 A, 600.0 A Resolution 0.01 A Basic accuracy ±(1.5% + 5 digits) AC+DC A Ranges 100.00 A, 600.0 A Resolution 0.01 A Basic accuracy Same as AC A + (1.5% + 5 digits) AC V Ranges 100.00–1000 V Resolution 0.01 V Basic accuracy ±(1.
7 Technical specifications Ranges 100.00–1000 V Resolution 0.01 V Basic accuracy ±(0.7% + 2 digits) Input impedance 3.5 MΩ Overload protection 1000 V RMS AC+DC V Ranges 100.00–1000 V Resolution 0.01 V Basic accuracy ±(1.0% + 5 digits) Input impedance 3.5 MΩ Overload protection 1000 V RMS Watts Ranges 10–600 kW Resolution 1W Basic accuracy ±(2.5% + 5 digits) Power factor Ranges –1.00 to 1.00 Resolution 0.01 Basic accuracy ±3° Total harmonic distortion Ranges 0.1%~100.
7 Technical specifications Resolution Basic accuracy 0.1% • • ±(5.0% + 10 digits) for order 1–12 ±(10.0% + 10 digits) for order 13– 25 Ohms Ranges 1.0000 kΩ, 10.00 kΩ, 100.00 kΩ Resolution 0.01 Ω Accuracy ±( 1.0% + 3 digits) Overload protection 1000 V RMS Continuity beeper <30 Ω, 2 kHz tone buzzer Diode test Open circuit voltage ±1.8 V max Capacitance Ranges 400.0 μF, 4.000 mF Resolution 0.1 μF Basic accuracy ±(1.
7 Technical specifications 7.2 System specifications Dimensions 49 mm × 100 mm × 262 mm (1.9″ × 3.9″ × 10.3″) Weight 0.59 kg (1.29 lb.), including batteries Battery life 200 hours Battery type 6 × AAA (LR03) Agency approvals FCC Class B, CE, UL/CSA, GSA, SJQA 7.
8 Flir Global Limited Lifetime Warranty A qualifying FLIR Test and Measurement product (the “Product”) purchased either directly from FLIR Commercial Systems Inc and affiliates (FLIR) or from an authorized FLIR distributor or reseller that Purchaser registers on-line with FLIR is eligible for coverage under FLIR’s Limited Lifetime Warranty, subject to the terms and conditions in this document.
8 Flir Global Limited Lifetime Warranty to Purchaser any Product that FLIR repairs or replaces under warranty. FLIR reserves the right to determine, in its sole discretion, whether a returned Product is covered under Warranty.
A note on the technical production of this publication This publication was produced using XML — the eXtensible Markup Language. For more information about XML, please visit http://www.w3.org/XML/ A note on the typeface used in this publication This publication was typeset using Linotype Helvetica™ World. Helvetica™ was designed by Max Miedinger (1910–1980) LOEF (List Of Effective Files) T501025.xml; en-US; 8008; 2013-06-05 T505544.xml; en-US; 7912; 2013-05-31 #T559825; r.
Corporate last page Headquarters Flir Systems, Inc. 27700 SW Parkway Ave. Wilsonville, OR 97070 USA Telephone: +1-503-498-3547 Website http://www.flir.com Customer support http://support.flir.com Publ. No.