Coating Thickness Tester User Manual Group4 Group4 FE FE (44) (44) DEL DEL
Coating Thickness Tester User Manual Contents Page 1. General Information ............................................................ 4 1-1-1. Application ................................................................... 5 1-1-2. Description of the Gauge............................................... 5 1-1-3. Supply Schedule............................................................. 5 1-1-4. Probe............................................................................. 5 1-1-5. Specifications......
Coating Thickness Tester User Manual 1. General Information The coating thickness gauges work either on the magnetic induction principle or on the eddy current principle, depending on the type of probe used. You can select the type of probe via MENU system, or it will work automatically. Features • Measured Coatings: Non-magnetic coatings (e.g. paint, zinc)on steel; • Insulating coatings(e.g.
Coating Thickness Tester User Manual 1-1. Application • This compact and handy gauge is designed for non-destructive, fast and precise coating thickness measurements. The principal applications lie in the field of corrosion protection. It is ideal for manufacturers and their customers, for offices and specialist advisers, for paint shops and electroplaters, for the chemical, automobile, shipbuilding and aircraft industries and for light and heavy engineering.
Coating Thickness Tester User Manual 1-5. Specifications Sensor probe Working principle Measuring range Guaranteed tolerance (of reading) Precision Minimum curvature radius Diameter of Minimum area Minimum measurable thickness Overload display Working temperature Working relative humidity 6 F N Magnetic induction Eddy current principle 0~2000µm 0~2000µm 0~78.7mils 0~78.7mils 0 ~ 1000 um 0 ~ 1000 um (±2%±2um) (±2%±2um) 1000~ 2000 um 1000~ 2000 um (±3.5%) (±3.5%) 0~39.3mils 0~39.3mils (±2%±0.
Coating Thickness Tester User Manual 1-6. Front View 1 8 2 3 4 5 6 9 1-LCD display 2- Power ON/OFF key 3-Left Key: For choose and OK 4-Up key: For upward selection and working mode switch 5-Right Key: For return or delete the last measured data 6-Down key: For downward selection or working mode switch 7-Probe 8- Lanyard hole 9-Battery cover 1-7. LCD Display 1-Working mode 2-Auto measure mode 3-Bluetooth indication 4-Ferrous or No-ferrous 9-metals being measured.
Coating Thickness Tester User Manual 2. Ready to start 2-1. Power Supply For power on, please press and hold button. If no LCD display: please check if the battery is correctly installed or it is in low battery. Display: Please replace battery immediately. Note that the gauge will make faulty measurements if the voltage is very low. Note: When power on, please away from any metal at least 10 cm above Before each measurement, zero calibration is recommended. 2-2. Replacing the Battery • Power off the meter.
Coating Thickness Tester User Manual 2-3-1-2. select measure mode (Probe) The probe can work in three modes: AUTO: The probe can automatically select the working mode. When placed on steel (magnetic substrates), it will work in magnetic induction principle. When placed on no-ferrous metals, it will work in eddy current principle. Fe: The probe works in magnetic induction principle. No-Fe: The probe works in eddy current principle.
Coating Thickness Tester User Manual 3-3. Auto Power Off You can disable Auto-power-off via MENU system or the gauge will power off automatically in about 10 minutes if no measurement or operation. • Press left button to enter the menu • Press up/down button to select the setting • Press left button to enter the setting • Press up/down button to select Auto-power-off • Press left button to enter • Press up/down button to select • Press left button for ok 3-4.
Coating Thickness Tester User Manual • Press left button to enter the menu • Press up/down button to select the setting • Press left button to enter the setting • Press up/down button to select Bluetooth • Press left button to enter • Press up/down button to select • Press left button for ok 3-5. Adjust gray scale (contrast) The grayscale adjustment: dot matrix LCD display may not be good in different temperature & humidity, than you can adjust gray scale to get good LCD display.
Coating Thickness Tester User Manual • Press up/down button to view each data • Press right button to return or press left button to delete this group • data. 5. Calibration and Measurement The gauge can calibrate Zero-point of Fe or Non-Fe, or delete wrong calibrated zero-point of Fe or Non-Fe. It suggested that zero-point be calibrated before use. 6.
Coating Thickness Tester User Manual 8. Delete Functions • • • • • • In MENU system, you can find following function: Delete current data: in measurement interface, you can delete the last data by pressing right button. Delete all data: Enter menu-view; you can delete all data and statistics of the Current Delete Group data: Enter menu-view-group data; you can delete this group data by pressing left button Transfer data to PC or phone or other Bluetooth receiving device.
Coating Thickness Tester User Manual Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. This device complies with Part 15 of the FCC Rules.