Reference Guide Agilent Technologies 8719ET/20ET/22ET 8719ES/20ES/22ES Network Analyzers Part Number 08720-90393 Printed in USA February 2001 Supersedes July 2000 © Copyright 1999–2001 Agilent Technologies, Inc.
Notice The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
How to Use This Guide This guide uses the following conventions: Front-Panel Key This represents a key physically located on the instrument. SOFTKEY This represents a “softkey,” a key whose label is determined by the instrument’s firmware. Screen Text This represents text displayed on the instrument’s screen.
Documentation Map The Installation and Quick Start Guide provides procedures for installing, configuring, and verifying the operation of the analyzer. It also will help you familiarize yourself with the basic operation of the analyzer. The User’s Guide shows how to make measurements, explains commonly-used features, and tells you how to get the most performance from your analyzer. The Reference Guide provides reference information, such as specifications, menu maps, and key definitions.
Contents 1. 8719/20/22ES Specifications and Characteristics Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Specifications for Instruments with Multiple Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2 Corrected System Performance (8719/20ES) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3 Corrected System Performance (8722ES) . . . . . . . . . . . . . . . .
Contents Option 089, Frequency Offset Mode (ES Only). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4 Option 400, Four-Sampler Test Set (ES Only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4 Option 1CM, Rack Mount Flange Kit Without Handles . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4 Option 1CP, Rack Mount Flange Kit With Handles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4 Service and Support Options . . . . . . . . . . .
Contents 10. Determining System Measurement Uncertainties Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-2 Sources of Measurement Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-3 Sources of Systematic Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-3 Sources of Random Errors . . . . . . . . . . . . . .
1 8719/20/22ES Specifications and Characteristics 1-1
8719/20/22ES Specifications and Characteristics Definitions Definitions All specifications and characteristics apply over a 23 °C ±3 °C range (unless otherwise stated) and 1/2 hour after the instrument has been turned on. Specification (spec.): Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions. Characteristic (char.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Corrected System Performance (8719/20ES) The specifications in this section apply for measurements made using 10 Hz IF bandwidth, no averaging, and at an environmental temperature of 23 ±3 °C, with less than 1 °C deviation from the calibration temperature. Assumes that an isolation calibration was performed with an averaging factor of 8.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-2 System Dynamic Range, All Device Connector Types 8722ES, All Cal Kits, All Cables, 10 Hz IF BW Description Specification Supplemental Information System Transmission Dynamic Range 8722ES (Standard)a 50 MHz to 840 MHz 67 dB 840 MHz to 8 GHz 93 dB 8 GHz to 20 GHz 91 dB 20 GHz to 40 GHz 80 dBb 8722ES (Option 007)a 50 MHz to 840 MHz 72 dB 840 MHz to 8 GHz 98 dB 8 GHz to 20 GHz 96 dB 20 GHz to 40
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-3 3.5-mm Device Connector Type Network Analyzer: 8719ES/20ES, Standard Calibration Kit: 85052D (3.5-mm, 50 Ω) Cables: 85131F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 50 to 500 MHz 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-4 3.5-mm Device Connector Type Network Analyzer: 8719ES/20ES, Standard Calibration Kit: 85052B (3.5-mm, 50 Ω) Cables: 85131F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 50 to 500 MHz 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-5 3.5-mm Device Connector Type Network Analyzer: 8719ES/20ES, Option 400 Calibration Kit: 85052C (3.5-mm, 50 Ω) Cables: 85131F Calibration: TRL IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 50 to 500 MHz 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-6 3.5-mm Device Connector Type Network Analyzer: 8719ES/20ES, Option 007 Calibration Kit: 85052B (3.5-mm with Sliding Loads, 50 Ω) Cables: 85132F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.05 to 0.5 GHz 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-7 3.5-mm Device Connector Type Network Analyzer: 8719ES/20ES, Option 007 Calibration Kit: 85052D (3.5-mm 50 Ω) Cables: 85132F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.05 to 0.5 GHz 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-8 7-mm Device Connector Type Network Analyzer: 8719ES/20ES, Standard Calibration Kit: 85050B (7-mm, 50 Ω) Cables: 85132F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 50 to 500 MHz 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-9 7-mm Device Connector Type Network Analyzer: 8719ES/20ES, Option 400 Calibration Kit: 85050C (7-mm, 50 Ω) Cables: 85132F Calibration: TRL IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 50 to 500 MHz 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-10 Type-N Device Connector Type Network Analyzer: 8719ES/20ES, Standard Calibration Kit: 85054D (Type-N, 50 Ω) Cables: 85132F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 50 to 500 MHz 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8719/20ES) Table 1-11 Type-N Device Connector Type Network Analyzer: 8719ES/20ES, Standard Calibration Kit: 85054B (Type-N, 50 Ω) Cables: 85132F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 50 to 500 MHz 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Corrected System Performance (8722ES) The specifications in this section apply for measurements made using 10 Hz IF bandwidth, no averaging, and at an environmental temperature of 23 ±3 °C, with less than 1 °C deviation from the calibration temperature. Assumes that an isolation calibration was performed with an averaging factor of 8.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Table 1-12 2.4-mm Device Connector Type Network Analyzer: 8722ES, Standard Calibration Kit: 85056A (2.4-mm, 50 Ω) Cables: 85133F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Table 1-13 2.4-mm Device Connector Type Network Analyzer: 8722ES, Standard Calibration Kit: 85056D (2.4-mm, 50 Ω) Cables: 85133F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Table 1-14 3.5-mm Device Connector Type Network Analyzer: 8722ES, Standard Calibration Kit: 85052D (3.5-mm, 50 Ω) Cables: 85131F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.05 to 2 GHz 2 to 8 GHz 8 to 20 GHz 20 to 26.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Table 1-15 3.5-mm Device Connector Type Network Analyzer: 8722ES, Standard Calibration Kit: 85052B (3.5-mm, 50 Ω) Cables: 85131F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.05 to 2 GHz 2 to 8 GHz 8 to 20 GHz 20 to 26.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Table 1-16 3.5-mm Device Connector Type Network Analyzer: 8722ES, Option 400 Calibration Kit: 85052C (3.5-mm, 50 Ω) Cables: 85131F Calibration: TRL IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.05 to 2 GHz 2 to 8 GHz 8 to 20 GHz 20 to 26.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Table 1-17 Type-N Device Connector Type Network Analyzer: 8722ES, Standard Calibration Kit: 85054D (Type-N, 50 Ω) Cables: 85132F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Table 1-18 2.4-mm Device Connector Type Network Analyzer: 8722ES, Option 007 Calibration Kit: 85056A (2.4-mm with sliding loads, 50 Ω) Cables: 85133F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Table 1-19 2.4-mm Device Connector Type Network Analyzer: 8722ES, Option 007 Calibration Kit: 85056D (2.4-mm 50 Ω) Cables: 85133F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.
8719/20/22ES Specifications and Characteristics Corrected System Performance (8722ES) Table 1-20 Type-N Device Connector Type Network Analyzer: 8722ES, Standard Calibration Kit: 85054B (Type-N, 50 Ω) Cables: 85132F Calibration: Full 2-Port IF BW = 10 Hz, Avg off, Temp = 23 ± 3 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8 Description Specification 0.
8719/20/22ES Specifications and Characteristics Instrument Specifications Instrument Specifications Uncorrected Port Performance Table 1-21 3.5-mm Device Connector Type 8719ES/20ES (3.5-mm, 50 Ω) Characteristic Description 50 to 500 MHz 0.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-22 2.4-mm Device Connector Type 8722ES (2.4-mm, 50 Ω) Description Characteristic 50 to 500 MHz 0.
8719/20/22ES Specifications and Characteristics Instrument Specifications Test Port Output Table 1-23 Test Port Output 8719ES/20ES/22ES Test Port Output Description Specification Supplemental Information Frequency Range 8719ES 0.05 to 13.51 GHz 8720ES 0.05 to 20.05 GHz 8722ES 0.05 to 40.05 GHz Resolution 1 Hz Stability Standard ±7.5 ppm, 0˚ to 55˚C, typ. ±3 ppm/year, typ. Option 1D5 ±0.05 ppm, 0˚ to 55˚C, typ. ±0.5 ppm/year, typ.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-24 Test Port Output 8719ES/20ES/22ES Test Port Output Description Specification Supplemental Information Output Power Maximum Leveled Powera 8719/20ES (Standard) +5 dBm, char. 8719ES/20ES (Option 007) +10 dBm, char. 8722ES (Standard) 0.05 to 20 GHz −5 dBm, char. 20 to 40 GHz −10 dBm, char. 8722ES (Option 007) 0.05 to 20 GHz 0 dBm, char. 20 to 40 GHz −5 dBm, char.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-25 Test Port Output 8719ES/20ES Test Port Output Description Specification Supplemental Information Output Power Power Resolution 0.01 dB Attenuator Switch Points: Note: For Option 400, the switch point between Range 0 and Range 1 is −10 dBm.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-26 Test Port Output 8722ES Test Port Output Description Specification Supplemental Information Output Power Power Resolution 0.01 dB Attenuator Switch Points: Note: For Option 400, the switch point between Range 0 and Range 1 is 0 dBm.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-27 Test Port Output 8719ES/20ES Test Port Output Description Specification Supplemental Information −5 dB from reference ±0.35 dB +5 dB from reference ±0.35 dB Test Reference Powers: Standard 8719/20ES: −5 dBm 8719/20ES Option 007: 0 dBm −10 dB from reference ±0.6 dB +10 dB from reference ±1.0 dB Linearitya Impedance Standard 50 Ω, nominal Attenuator Accuracyb 0 dB reference; at 50 MHz 5 dB ±0.6 dB, char.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-28 Test Port Output 8722ES Test Port Output Description Specification Linearitya −5 dB from reference 50 MHz to 20 GHz ±0.35 dB 20 GHz to 40 GHz ±0.6 dB Supplemental Information Test Reference Powers: Standard 8722ES: −10 dBm 8722ES Option 007: −5 dBm +5 dB from reference 50 MHz to 20 GHz −10 dB from reference ±0.35 dB ±0.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-29 Test Port Output 8719ES/20ES/22ES Test Port Output Description Specification Supplemental Information Signal Purity 2nd Harmonic 0.05 GHz to one half the maximum source frequency at the maximum output level < −15 dBc, typ. Non-harmonic Spurious Mixer Related at 100 kHz offset < −40 dBc, typ. at 200 kHz offset < −45 dBc, typ. at > 200 kHz offset < −65 dBc, typ.
8719/20/22ES Specifications and Characteristics Instrument Specifications Test Port Input Table 1-30 Test Port Input 8719ES/20ES/22ES Test Port Input Description Specification Supplemental Information Frequency Range 8719ES 0.05 to 13.51 GHz 8720ES 0.05 to 20.05 GHz 8722ES 0.05 to 40.05 GHz Frequency Response (A, B, R) Channel R 0.05 to 20.05 GHz ±1.5 dB, char. 20.05 to 32 GHz ±2.5 dB, char. 32 to 40 GHz +2.5 dB to −6 dB, char. Channels A and B 50 to 500 MHz +2.5 dB to −28 dB, char. 0.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-31 Test Port Input 8719ES/20ES/22ES Test Port Input Description Specification Supplemental Information 0.1 dB Compression (Option 012), direct receiver input 0.05 to 0.5 GHz −5 dBm, typ. 0.5 to 2 GHz −5 dBm, typ. 2 to 8 GHz −5 dBm, typ. 8 to 20 GHz −5 dBm, typ. 20 to 40 GHz −5 dBm, typ. Average Noise Floor (Option 012) - 10 Hz IF BWa, direct receiver input 0.05 to 0.5 GHz −120 dBm, typ. 0.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-32 Test Port Input 8719/20/22ES Test Port Input Description Supplemental Information System Bandwidths 3000 Hz 10 Hz Trace Noisea Magnitude 0.05 GHz to 13.5 GHz < 0.03 dB rms, typ. < 0.003 dB rms, typ. 13.5 GHz to 20 GHz < 0.04 dB rms, typ. < 0.004 dB rms, typ. 20 GHz to 40 GHz < 0.15 dB rms, typ. < 0.015 dB rms, typ. Phase 0.05 GHz to 13.5 GHz < 0.3° rms, typ. < 0.03° rms, typ. 13.5 GHz to 20 GHz < 0.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-34 Test Port Input 8719/20ES Test Port Input Dynamic Accuracy (Characteristic) For input ports 1 and 2, accuracy of the test port input power reading relative to the reference input power level. • Inputs: testport 1 and 2 • For test port powers > −50 dBm and < 0 dBm, magnitude dynamic accuracy is 0.02 dB + 0.0015 dB/dB from the reference power, phase dynamic accuracy is 0.132 deg + 0.0066 deg/dB from the reference power.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-34 Test Port Input 8719/20ES Test Port Input 1-37
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-35 Test Port Input 8722ES Test Port Input Dynamic Accuracy (Characteristic) For input ports 1 and 2, accuracy of the test port input power reading relative to the reference input power level. • Inputs: testport 1 and 2 • For test port powers > −50 dBm and < 0 dBm, magnitude dynamic accuracy is 0.020 dB + 0.0015 dB/dB from the reference power, phase dynamic accuracy is 0.132 deg + 0.0066 deg/dB from the reference power.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-35 Test Port Input 8722ES Test Port Input 1-39
8719/20/22ES Specifications and Characteristics Instrument Specifications General Information Table 1-36 General Information 8719/20/22ES General Information Description Specification Display Range Magnitude ±200 dB (at 20 dB/div), max Phase ±180°, max Polar 10 pico units, min 1000 units, max Display Resolution Magnitude 0.001 dB/div, min Phase 0.01°/div, min Reference Value Range Magnitude ±500 dB, max Phase ±360 °, max Reference Level Resolution Magnitude 0.001 dB, min Phase 0.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-37 General Information 8719/20/22ES General Information Description Specification Supplemental Information Group Delaya Aperture (selectable) (frequency span)/(number of points −1) Maximum Aperture 20% of frequency span Range 1/2 × (1/minimum aperture) with smoothing enabled Maximum Delay Limited to measuring no more than 180° of phase change within the minimum aperture.) Accuracy See graph. Char.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-38 General Information 8719/20/22ES General Information Description Supplemental Information System Bandwidths IF bandwidth settings 6000 Hz, nom. 3700 Hz, nom. 3000 Hz, nom. 1000 Hz nom. 300 Hz, nom. 100 Hz, nom. 30 Hz, nom. 10 Hz, nom. Rear Panel External Auxiliary Input Connector Female BNC Range ±10 V, typ. External Trigger Triggers on a positive or negative TTL transition or contact closure to ground.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-39 General Information 8719/20/22ES General Information Description Specification Supplemental Information Rear Panel Test Port Bias Input Maximum voltage ±40 Vdc Maximum current ±500 mA External Reference In Input Frequency 1, 2, 5, and 10 MHz ±200 Hz at 10 MHz Input Power −10 dBm to +20 dBm, typ. Input Impedance 50 Ω, nom. VGA Video Output 15-pin mini D-Sub; female. Drives VGA compatible monitors.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-40 General Information 8719/20/22ES General Information Description Specification Front Panel Display Pixel Integrity Red, Green, or Blue Pixels Red, green, or blue “stuck on” pixels may appear against a black background.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-41 General Information 8719/20/22ES General Information Description Specification Supplemental Information General Environmental RFI/EMI Susceptibility Defined by CISPR Pub. 11 and FCC Class B standards. ESD Minimize using static-safe work procedures and an antistatic bench mat (part number 9300-0797). Dust Minimize for optimum reliability.
8719/20/22ES Specifications and Characteristics Instrument Specifications Speed Parameters Table 1-42 8719/20/22ES Measurement and Data Transfer Speed Performance Typical Time for Completion (ms) Number of Points Description 51 201 1601 401 Swept Stepped Swept Stepped Swept Stepped Swept Stepped Typical Time for Completion (in ms), Center 1 GHz, Span 10 MHz, IFBW=6000 Uncorrected 27 134 65 492 116 970 419 3836 1-port and Enh. Resp.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-43 8719/20/22ES Recall and Sweep Speed Performance Total Time, Typical (secs) Operations Channel Points Recall-Only Time, Typical (secs) Raw Offset Blank Off Blank On Blank Off Blank On Error Correction ON Recall and Sweep Single Chan. 201 On 0.654 0.546 0.497 0.394 Recall and Sweep Single Chan. 201 Off 0.605 0.496 0.448 0.344 Sweep only (no Recall) Single Chan. 201 N/A 0.158 0.
8719/20/22ES Specifications and Characteristics Instrument Specifications Table 1-44 Sweep Time vs. IF Bandwidth IF Bandwidth Typical Sweep Time (seconds)a 6000 0.066 3700 0.091 3000 0.116 1000 0.243 300 0.700 100 2.018 30 7.058 10 21.475 a. Preset condition, CF = 1 GHz, Span = 100 MHz; includes retrace time, 201 points. Table 1-45 Sweep Time vs. Number of Points Number of Points Typical Sweep Time (seconds) a 51 0.034 101 0.053 201 0.091 401 0.167 801 0.318 1601 0.621 a.
8719/20/22ES Specifications and Characteristics Instrument Specifications Power Meter Calibration Accuracy Table 1-46 Power Meter Calibration Sweep Speed and Accuracy Power Desired at Test Port +5 dBm −15 dBm −30 dBm Number of Readings Sweep Time Setting (seconds) a Characteristic Accuracy (dB) b 1 33 ±0.7 2 64 ±0.2 3 95 ±0.1 1 48 ±0.7 2 92 ±0.2 3 123 ±0.1 1 194 ±0.7 2 360 ±0.2 3 447 ±0.1 a.
2 8719/20/22ET Specifications and Characteristics 2-1
8719/20/22ET Specifications and Characteristics Definitions Definitions All specifications and characteristics apply over a 23 °C ±3 °C range (unless otherwise stated) and 1/2 hour after the instrument has been turned on. Specification (spec.): Warranted performance. Specifications include guardbands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions. Characteristic (char.
8719/20/22ET Specifications and Characteristics Corrected System Performance (8719/20ET) Corrected System Performance (8719/20ET) The specifications in this section apply for measurements made using 10 Hz IF bandwidth, no averaging, and at an environmental temperature of 23 ±3 °C, with less than 1 °C deviation from the calibration temperature. Assumes that an isolation calibration was performed with an averaging factor of 8.
8719/20/22ET Specifications and Characteristics Corrected System Performance (8719/20ET) Table 2-3 3.5-mm (50 Ω) Device Connector Type Network Analyzer: 8719/20ET Standard or Option 004 Attenuator Calibration Kit: 85052B (3.5-mm, 50 Ω) Cal Kit Cables: 85131E Calibration: One-Port, Response, or Enhanced Response IF BW = 10 Hz, Avg = off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8 Specification Description 50 to 500 MHz 0.
8719/20/22ET Specifications and Characteristics Corrected System Performance (8719/20ET) Transmissiona Uncertainty: Enhanced Response Calibration (Specification) Transmission Uncertainty: Response Calibration (Specification) Reflection Uncertainty: One-Port Calibration (Specification) a. Option 004 may degrade transmission source match as much as 2 dB, resulting in up to 0.05 dB additional uncertainty in transmission uncertainty.
8719/20/22ET Specifications and Characteristics Corrected System Performance (8719/20ET) This page is intentionally left blank.
8719/20/22ET Specifications and Characteristics Corrected System Performance (8722ET) Corrected System Performance (8722ET) The specifications in this section apply for measurements made using 10 Hz IF bandwidth, no averaging, and at an environmental temperature of 23 ±3 °C, with less than 1 °C deviation from the calibration temperature. Assumes that an isolation calibration was performed with an averaging factor of 8.
8719/20/22ET Specifications and Characteristics Corrected System Performance (8722ET) Table 2-4 2.4-mm (50 Ω) Device Connector Type Network Analyzer: 8722ET Standard or Option 004 Attenuator Calibration Kit: 85056A (2.4-mm, 50 Ω) Cables: 85133E Calibration: One-Port, Response, or Enhanced Response IF BW = 10 Hz, Avg = off, Temp = 23 ± 3°C with < 1°C deviation from cal temp, Isol cal with avg = 8 Specification Description 0.
8719/20/22ET Specifications and Characteristics Corrected System Performance (8722ET) Transmissiona Uncertainty: Enhanced Response Calibration (Specification) Transmission Uncertainty: Response Calibration (Specification) Reflection Uncertainty: One-Port Calibration (Specification) a. Option 004 may degrade transmission source match as much as 2 dB, resulting in up to 0.05 dB additional uncertainty in transmission uncertainty.
8719/20/22ET Specifications and Characteristics Instrument Specifications Instrument Specifications Uncorrected Port Performance Table 2-5 3.5-mm (50 Ω) Device Connector Type 8719/20ET (3.5-mm, 50 Ω) Specification Description 50 to 500 MHz 0.5 to 2 GHz 2 to 8 GHz 8 to 20 GHz Directivitya (dB) 24 27 21 16 Source Matcha (dB) 16 20 14 11 Load Matcha (dB) 22 22 22 17 Reflection Trackingb (dB) ±3 ±3 ±3 ±3 Transmissiona Tracking (dB) ±3 ±3 ±3 ±3 Magnitude (dB) 0.02/˚C, typ. 0.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-6 2.4-mm (50 Ω) Device Connector Type 8722ET (2.4-mm, 50 Ω) Specification Description 50 to 500 MHz 0.5 to 2 GHz 2 to 8 GHz 8 to 20 GHz 20 to 40 GHz Directivitya (dB) 23 23 21 16 15 Source Matcha (dB) 16 20 14 11 10 Load Matcha (dB) 22 20 20 20 15 Reflectionb Tracking (dB) ±3 ±3 ±3 ±3 ±4 Transmissiona Tracking (dB) ±3 ±3 ±3 ±3 ±3 Magnitude (dB) 0.02/˚C, typ. 0.03/˚C, typ. 0.03/˚C, typ.
8719/20/22ET Specifications and Characteristics Instrument Specifications Test Port Output Table 2-7 Test Port Output 8719ET/20ET/22ET Test Port Output Description Specification Supplemental Information Frequency Range 8719ET 0.05 to 13.51 GHz 8720ET 0.05 to 20.05 GHz 8722ET 0.05 to 40.05 GHz Resolution 1 Hz Stability Standard ±7.5 ppm, 0° to 55 °C, typ. ±3 ppm/year, typ. Option 1D5 ±0.05 ppm, 0° to 55°C, typ. ±0.5 ppm/year, typ.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-8 Test Port Output 8719ET/20ET/22ET Test Port Output Description Specification Supplemental Information 8719ET/20ET ± 2 dB at 5 dBm 8722ET ± 3 dB at −5 dBm Output Power Level Accuracya Maximum Leveled Powerb 8719/20ET +10 dBm, char. 8722ET 0.05 to 20 GHz 0 dBm, char. 20 to 40 GHz −5 dBm, char. Power Rangec 8719/20ET (Standard) −10 to +10 dBm 8719/20ET (Option 004) −65 to +10 dBm −18 to +12 dBm, typ.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-9 Test Port Output 8719ES/20ET Test Port Output Description Specification Output Power (Option 004 Only) Power Resolution Attenuator Switch Points: 2-14 0.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-10 Test Port Output 8722ET Test Port Output Description Specification Supplemental Information Output Power (Option 004 Only) Power Resolution 0.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-11 Test Port Output 8719ET/20ET Test Port Output Description Specification Supplemental Information Output Power Test Reference Power: 0 dBm Linearitya −5 dB from reference ± 0.35 dB +5 dB from reference ± 0.35 dB −10 dB from reference ± 0.6 dB +10 dB from reference ± 1.0 dB Impedance Standard 50 Ω, nominal Attenuator Accuracyb (Option 004 Only) 0 dB reference; at 50 MHz 5 dB ±0.6 dB, char. 10 dB ±0.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-12 Test Port Output 8722ET Test Port Output Description Specification Supplemental Information Output Power Test Reference Power: −5 dBm Linearitya −5 dB from reference 50 MHz to 20 GHz ± 0.35 dB 20 GHz to 40 GHz ± 0.6 dB +5 dB from reference 50 MHz to 20 GHz ± 0.35 dB −10 dB from reference ± 0.6 dB Impedance Standard 50 Ω, nominal Attenuator Accuracyb (Option 004 Only) 0 dB reference; at 50 MHz 5 dB ±0.
8719/20/22ET Specifications and Characteristics Instrument Specifications Test Port Input Table 2-13 Test Port Input 8719ET/20ET/22ET Test Port Input Description Specification Supplemental Information Frequency Range 8719ET 0.05 to 13.51 GHz 8720ET 0.05 to 20.05 GHz 8722ET 0.05 to 40.05 GHz Frequency Response (B, R) Channel R 0.05 to 20.05 GHz ± 1.5 dB, char. 20.05 to 32 GHz ± 2.5 dB, char. 32 to 40 GHz +2.5 dB to −6 dB, char. Channel B 0.05 to 20.05GHz ± 2.5 dB, char. 20 to 32 GHz ± 3.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-14 Test Port Input 8719/20/22ET Test Port Input Description Supplemental Information System Bandwidths 3000 Hz 10 Hz Trace Noisea Magnitude 0.05 GHz to 13.5 GHz < 0.03 dB rms, typ. < 0.003 dB rms, typ. 13.5 GHz to 20 GHz < 0.04 dB rms, typ. < 0.004 dB rms, typ. 20 GHz to 40 GHz < 0.15 dB rms, typ. < 0.015 dB rms, typ. Phase 0.05 GHz to 13.5 GHz < 0.3° rms, typ. < 0.03° rms, typ. 13.5 GHz to 20 GHz < 0.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-16 Test Port Input 8719/20ET Test Port Input (Transmission) Dynamic Accuracy (Characteristic) For transmission accuracy of the test port input power reading relative to the reference input power level. • Inputs: transmission port • For test port powers up to 0 dBm.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-16 Test Port Input 8719/20ET Test Port Input (Transmission) 2-21
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-17 Test Port Input 8719/20ET Test Port Input (Reflection) Dynamic Accuracy (Characteristic) For reflection accuracy of the test port input power reading relative to the reference input power level. • Inputs: reflection port • For test port powers up to 0 dBm.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-17 Test Port Input 8719/20ET Test Port Input (Reflection) 2-23
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-18 Test Port Input 8722ET Test Port Input (Transmission) Dynamic Accuracy (Characteristic) For transmission accuracy of the test port input power reading relative to the reference input power level. • Inputs: transmission port • For test port powers up to 0 dBm.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-18 Test Port Input 8722ET Test Port Input (Transmission) 2-25
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-19 Test Port Input 8722ET Test Port Input (Reflection) Dynamic Accuracy (Characteristic) For reflection accuracy of the test port input power reading relative to the reference input power level. • Inputs: reflection port • For test port powers up to 0 dBm.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-19 Test Port Input 8722ET Test Port Input (Reflection) 2-27
8719/20/22ET Specifications and Characteristics Instrument Specifications General Information Table 2-20 General Information 8719/20/22ET General Information Description Specification Display Range Magnitude ± 200 dB (at 20 dB/div), max Phase ± 180°, max Polar 10 pico units, min 1000 units, max Display Resolution Magnitude 0.001 dB/div, min Phase 0.1°/div, min Reference Level Range Magnitude ± 500 dB, max Phase ± 360 °, max Reference Level Resolution Magnitude 0.001 dB, min Phase 0.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-21 General Information 8719/20/22ET General Information Description Specification Supplemental Information Group Delaya Aperture (selectable) (frequency span)/(number of points − 1) Maximum Aperture 20% of frequency span Range 1/2 × (1/minimum aperture) Maximum Delay Limited to measuring no more than 180° of phase change within the minimum aperture.) Accuracy See graph. Char.
19/20/22ET Specifications and Characteristics Instrument Specifications Table 2-22 General Information 8719/20/22ET General Information Description Supplemental Information System Bandwidths IF bandwidth settings 6000 Hz, nom. 3700 Hz, nom. 3000 Hz, nom. 1000 Hz nom. 300 Hz, nom. 100 Hz, nom. 30 Hz, nom. 10 Hz, nom. Rear Panel External Auxiliary Input Connector Female BNC Range 10 V, typ. External Trigger Triggers on a positive or negative TTL transition or contact closure to ground.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-23 General Information 8719/20/22ET General Information Description Specification Supplemental Information Rear Panel External Reference In Input Frequency 1, 2, 5, and 10 MHz ± 200 Hz at 10 MHz Input Power −10 dBm to +20 dBm, typ. Input Impedance 50 Ω, nom. VGA Video Output 15-pin mini D-Sub; female. Drives VGA compatible monitors.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-24 General Information 8719/20/22ET General Information Description Specification Front Panel Display Pixel Integrity Red, Green, or Blue Pixels Red, green, or blue "stuck on" pixels may appear against a black background.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-25 General Information 8719/20/22ET General Information Description Specification Supplemental Information General Environmental RFI/EMI Susceptibility Defined by CISPR Pub. 11 and FCC Class B standards. ETD Minimize using static-safe work procedures and an antistatic bench mat (part number 9300-0797). Dust Minimize for optimum reliability.
8719/20/22ET Specifications and Characteristics Instrument Specifications Speed Parameters Table 2-26 8719/20/22ET Measurement and Data Transfer Speed Performance Typical Time for Completion (ms) Number of Points Description 51 201 1601 401 Swept Stepped Swept Stepped Swept Stepped Swept Stepped Typical Time for Completion (in ms), Center 1 GHz, Span 10 MHz, IFBW=6000 Uncorrected 27 134 65 492 116 970 419 3836 1-port and Enh. Resp.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-27 8719/20/22ET Recall and Sweep Speed Performance Total Time, Typical (secs) Operations Channel Points Recall-Only Time, Typical (secs) Raw Offset Blank Off Blank On Blank Off Blank On Error Correction ON Recall and Sweep Single Chan. 201 On 0.654 0.546 0.497 0.394 Recall and Sweep Single Chan. 201 Off 0.605 0.496 0.448 0.344 Sweep only (no Recall) Single Chan. 201 N/A 0.158 0.
8719/20/22ET Specifications and Characteristics Instrument Specifications Table 2-28 Sweep Time vs. IF Bandwidth IF Bandwidth Typical Sweep Time (seconds)a 6000 0.066 3700 0.091 3000 0.116 1000 0.243 300 0.700 100 2.018 30 7.058 10 21.475 a. Preset condition, CF = 1 GHz, Span = 100 MHz; includes retrace time, 201 points. Table 2-29 Sweep Time vs. Number of Points Number of Points Typical Sweep Time (seconds) a 51 0.034 101 0.053 201 0.091 401 0.167 801 0.318 1601 0.621 a.
8719/20/22ET Specifications and Characteristics Instrument Specifications Power Meter Calibration Accuracy Table 2-30 Power Meter Calibration Sweep Speed and Accuracy Power Desired at Test Port Number of Readings Sweep Time Setting (seconds) a Characteristic Accuracy (dB) b +5 dBm 1 33 ±0.7 2 64 ±0.2 3 95 ±0.1 1 48 ±0.7 2 92 ±0.2 3 123 ±0.1 1 194 ±0.7 2 360 ±0.2 3 447 ±0.1 −15 dBm −30 dBm a.
3 Front/Rear Panel 3-1
Front/Rear Panel Front Panel Features Front Panel Features CAUTION Do not mistake the line switch for the disk eject button. See the following illustrations. If the line switch is mistakenly pushed, the instrument will be turned off, losing all settings and data that have not been saved.
Front/Rear Panel Front Panel Features Figure 3-2 8719/20/22ET Front Panel The location of the following front panel features and key function blocks is shown in Figure 3-1 and Figure 3-2. These features are described in more detail later in this chapter, and in Chapter 5 , “Hardkey/Softkey Reference.” 1. LINE switch. This switch controls ac power to the analyzer. 1 is on, 0 is off. 2. Display. This shows the measurement data traces, measurement annotation, and softkey labels.
Front/Rear Panel Front Panel Features 9. The ENTRY block. This block includes the knob, the step keys, the number pad, and the backspace key. These allow you to enter numerical data and control the markers. You can use the numeric keypad to select digits, decimal points, and a minus sign for numerical entries. You must also select a units terminator to complete value inputs.
Front/Rear Panel Analyzer Display Analyzer Display Figure 3-3 Analyzer Display (Single Channel, Cartesian Format) The analyzer display shows various measurement information: • The grid where the analyzer plots the measurement data. • The currently selected measurement parameters. • The measurement data traces. Figure 3-3 illustrates the locations of the different information labels described below.
Front/Rear Panel Analyzer Display 2. Stimulus Stop Value. This value could be any one of the following: • The stop frequency of the source in frequency domain measurements. • The stop time in time domain measurements or CW sweeps. • The upper limit of a power sweep. When the stimulus is in center/span mode, the span is shown in this space. The stimulus values can be blanked, as described under the FREQUENCY BLANK softkey in Chapter 5 , “Hardkey/Softkey Reference.
Front/Rear Panel Analyzer Display ext Waiting for an external trigger. Ofs (ES) Frequency offset mode is on. (See the “Making Mixer Measurements” chapter in the user’s guide.) Of? (ES) Frequency offset mode error, the IF frequency is not within 10 MHz of expected frequency. LO inaccuracy is the most likely cause. (See the “Making Mixer Measurements” chapter in the user’s guide.) Gat Gating is on (time domain Option 010 only).
Front/Rear Panel Analyzer Display 4. Active Entry Area. This displays the active function and its current value. 5. Message Area. This displays prompts or error messages. 6. Title. This is a descriptive alphanumeric string title that you define and enter through an attached keyboard or as described the “Printing, Plotting, and Saving Measurement Results” chapter of the user’s guide. 7. Active Channel.
Front/Rear Panel Rear Panel Features and Connectors Rear Panel Features and Connectors Figure 3-4 8719/20/22ET/ES Rear Panel Figure 3-4 illustrates the features and connectors of the rear panel, described below. Requirements for input signals to the rear panel connectors are provided in the specifications and characteristics chapter. 1. GPIB connector. This allows you to connect the analyzer to an external controller, compatible peripherals, and other instruments for an automated system.
Front/Rear Panel Rear Panel Features and Connectors 7. Fan. This fan provides forced-air cooling for the analyzer. 8. 10 MHZ PRECISION REFERENCE OUTPUT. (Option 1D5) 9. 10 MHZ REFERENCE ADJUST. (Option 1D5) 10. EXTERNAL REFERENCE INPUT connector. This allows for a frequency reference signal input that can phase lock the analyzer to an external frequency standard for increased frequency accuracy.
4 Menu Maps 4-1
Menu Maps Menu Maps Menu Maps This chapter contains menus maps for the hardkeys listed below. The figure number of these menu maps is listed next to the name of the hardkey. Fold Outs are located at the end of this chapter.
Menu Maps Menu Maps Figure 4-2 Menu Map for Copy 4-3
Menu Maps Menu Maps Figure 4-3 Menu Map for Display 4-4
Menu Maps Menu Maps Figure 4-4 Menu Map for Format 4-5
Menu Maps Menu Maps Figure 4-5 Menu Map for Local 4-6
Menu Maps Menu Maps Figure 4-6 Menu Map for Marker, Marker Fctn, and Marker Search 4-7
Menu Maps Menu Maps Figure 4-7 Menu Map for Meas (ET Models Only) Meas S-Parameter Menu Aux Input Menu REFLECTION RESOLUTION [LOW] TRANSMISSN AUX OUT on OFF ANALOG IN Aux Input Input Ports Menu OFF A/R Z:Ref l B/R Z:Trans A/B COUNTER: OFF Y: Ref I A ANALOG BUS Y:Trans FRAC N * Conversion Menu 1/S DIV FRAC N B R CONVERSION [ ] INPUT PORTS RETURN RETURN RETURN * Aux input menu appears only when the analog bus on OFF (service menu key under the system hardkey) is turned to ON.
Menu Maps Menu Maps Figure 4-8 Menu Map for Meas (ES Models Only) 4-9
Menu Maps Menu Maps Figure 4-9 Menu Map for Power and Sweep Setup (ET Only) 4-10
Menu Maps Menu Maps Figure 4-10 Menu Map for Power and Sweep Setup (ES Only) 4-11
Menu Maps Menu Maps Figure 4-11 Menu Map for Preset 4-12
Menu Maps Menu Maps Figure 4-12 Menu Map for Save/Recall 4-13
Menu Maps Menu Maps Figure 4-13 Menu Map for Scale Ref 4-14
Correction Menu Cal ECal Menu MODULE A b CORRECTION on OFF REFLECTION 1-PORT INTERPOL ON off ECal Pause Menu ECal Config Menu CONTINUE ECal MODULE A b ECal Confidence ECal Service Menu Power Meter Cal. Main Menu Pwr. Loss/Sens.
Cal Correction Menu CORRECTION on OFF ECal Menu MODULE A b INTERPOL ON off S11 1-PORT 1 CALIBRATE MENU S22 1-PORT 1 RESUME CAL SEQUENCE FULL 1, 2 2-PORT ECal MENU S11/S21 1, 2 ENH. RESP. CAL KIT [ ] ECal Pause Menu ECal Config Menu CONTINUE ECal MODULE A b CONFIGURE [MODULE A] MORE RETURN SET CONF STANDARD CONFIDENCE CHECK ECal Service Menu Power Meter Cal. Main Menu Pwr. Loss/Sens.
Select Seq Menu A TTL I/O Menu New / Modify Seq Menu SEQUENCE X SEQ X TTL OUT SEQUENCE 1 SEQ 1 TTL OUT HIGH PARALLEL OUT ALL SEQUENCE 2 SEQ 2 Seq TTL Out Menu TTL OUT LOW SET BIT SEQUENCE 3 SEQ 3 CLEAR BIT END SWEEP HIGH PULSE SEQUENCE 4 SEQ 4 PARALL IN BIT NUMBER END SWEEP LOW PULSE SEQUENCE 5 SEQ 5 PARALL IN IF BIT H SEQUENCE 6 SEQ 6 IF BIT L Seq Spec Func Menu Seq Dec Making Menu New / Modify Seq Menu Spec Func More Menu DECISION MAKING IF LIMIT TEST PASS SEQUENCE 1 SEQ 1 E
Set Clock Menu TIME STAMP ON off User Settings Menu PRESET SETTINGS ROUND SECONDS SET MINUTES K36 MODE on OFF SET HOUR K39 MODE on OFF Preset Settings Menu Select Limits Menu Ripple Test Menu Edit Ripple Limits Menu LIMIT LINE RIPL LIMIT on OFF FREQUENCY BAND STEP SWP on OFF RIPPLE LIMIT RIPL TEST on OFF MINIMUM FREQUENCY BANDWIDTH LIMIT RIPL VALUE [ OFF ] MAXIMUM FREQUENCY BW DISPLAY on OFF MAXIMUM RIPPLE BW MARKER on OFF DELETE N DB POINTS ADD MINIMUM BANDWIDTH CLEAR LIST MAXIM
Set Clock Menu User Settings Menu TIME STAMP ON off Preset Settings Menu Select Limits Menu RIPL LIMIT on OFF FREQUENCY BAND RIPPLE LIMIT RIPL TEST on OFF MINIMUM FREQUENCY BANDWIDTH LIMIT RIPL VALUE [ OFF ] ROUND SECONDS SET MINUTES K36 MODE on OFF SET HOUR K39 MODE on OFF Edit Ripple Limits Menu LIMIT LINE CAL INTERP ON off PRESET SETTINGS Ripple Test Menu RETURN Configure Menu Offset Limits Menu RETRCE PWR on STD STIMULUS OFFSET TRANSFORM on OFF STEP SWP ON off AMPLITUDE OFFSE
5 Hardkey/Softkey Reference 5-1
Hardkey/Softkey Reference Key Reference Key Reference This chapter contains information on the following topics: • softkey and front-panel functions in alphabetical order (includes a brief description of each function) • cross reference of programming commands to key functions • cross reference of softkeys to front-panel access keys NOTE 5-2 The SERVICE MENU keys are not included in this chapter. Information on the SERVICE MENU keys can be found in the service guide.
Hardkey/Softkey Reference Where to Look for More Information Where to Look for More Information Additional information about many of the topics discussed in this chapter is located in the following areas: • "Making Measurements" in the user’s guide contains step-by-step procedures for making measurements or using particular functions.
Hardkey/Softkey Reference Analyzer Functions Analyzer Functions This section contains an alphabetical listing of softkey and front-panel functions, and a brief description of each function. . is used to add a decimal point to the number you are entering. − is used to add a minus sign to the number you are entering. is used to step up the current value of the active function. The analyzer defines the step size for different functions. No units terminator is required.
Hardkey/Softkey Reference Analyzer Functions ∆ MODE MENU goes to the delta marker menu, which is used to read the difference in values between the active marker and a reference marker. ∆ MODE OFF turns off the delta marker mode, so that the values displayed for the active marker are absolute values. ∆ REF = 1 establishes marker 1 as a reference. The active marker stimulus and response values are then shown relative to this delta reference.
Hardkey/Softkey Reference Analyzer Functions 2.92* 85056K selects the 85056K cal kit. 2.92mm other kits selects the 2.92 mm cal kit model. 3.5mmC 85033C selects the 85033C cal kit. 3.5mmD 85052 selects the 85052B or the 85052D cal kit. 3.5mmE 85033D/E selects the 85033D or the 85033E cal kit. 4X: [1] [2]/[3] [4] sets up a four-graticule display with channel 2 in the upper right quadrant and channel 3 in the lower left quadrant.
Hardkey/Softkey Reference Analyzer Functions ADD 1) displays the edit segment menu and adds a new segment to the end of the list. The new segment is initially a duplicate of the segment indicated by the pointer > and selected with the SEGMENT softkey. 2) adds a new frequency band to the Ripple Limit list which is indicated by the pointer >. The new frequency band is a duplicate of the most recently selected frequency band. ADDRESS: 8720 sets the GPIB address of the analyzer, using the entry controls.
Hardkey/Softkey Reference Analyzer Functions ARBITRARY IMPEDANCE defines the standard type to be a load, but with an arbitrary impedance (different from system Z0). ASSERT SRQ sets the sequence bit in the Event Status Register, which can be used to generate a SRQ (service request) to the system controller. AUTO FEED ON off turns the plotter auto feed function on or off when in the define plot menu. It turns the printer auto feed on or off when in the define print menu.
Hardkey/Softkey Reference Analyzer Functions AVERAGING RESTART averaging starts at 1 and averages each new sweep into the trace until it reaches the specified averaging factor. The sweep count is displayed in the status notations area below “Avg” and updated every sweep as it increments. Avg is used to access three different noise reduction techniques: sweep-to-sweep averaging, display smoothing, and variable IF bandwidth. Any or all of these can be used simultaneously.
Hardkey/Softkey Reference Analyzer Functions BRIGHTNESS adjusts the brightness of the color being modified. Refer to the section on adjusting the display color in the “Using Measurement Functions” chapter of the user’s guide for an explanation of using this softkey for color modification of display attributes. BW DISPLAY on OFF displays the measured bandwidth value to the right of the pass/fail message. BW MARKER on OFF displays the cutoff frequencies of the bandwidth using markers on the data trace.
Hardkey/Softkey Reference Analyzer Functions Cal key leads to a series of menus to perform measurement calibrations for vector error correction (accuracy enhancement), and for specifying the calibration standards used. The CAL key also leads to softkeys which activate interpolated error correction and power meter calibration. CAL FACTOR accepts a power sensor calibration factor % for the segment.
Hardkey/Softkey Reference Analyzer Functions — N 50Ω 85054 — N 75Ω 85036 — TRL 3.5mm 85052C CAL Z0: LINE Z0 this default selection establishes the TRL/LRM LINE/MATCH standard as the characteristic impedance. CAL Z0: SYSTEM Z0 allows you to modify the characteristic impedance of the system for TRL/LRM calibration. CALIBRATE MENU leads to the calibration menu, which provides several accuracy enhancement procedures ranging from a simple frequency response calibration to a full two-port calibration.
Hardkey/Softkey Reference Analyzer Functions CH3 DATA LIMIT LN selects channel 3 data trace and limit line for display color modification. CH3 MEM selects channel 3 memory trace for display color. CH3 MEM [ ] brings up the printer color selection menu. The channel 2 data trace default color is green for color prints. CH4 DATA [ ] brings up the printer color selection menu. The channel 4 data trace default color is blue for color prints.
Hardkey/Softkey Reference Analyzer Functions CHANNEL POSITION configures multiple-channel displays so that the auxiliary channels are adjacent to or beneath the primary channels. CHOP A and B measures A and B inputs simultaneously for faster measurements. CLEAR BIT when the parallel port is configured for GPIO, 8 output bits can be controlled with this key. When this key is pressed, “TTL OUT BIT NUMBER” becomes the active function.
Hardkey/Softkey Reference Analyzer Functions CONTINUE ECal is displayed during ECal dual module operation and during manual thru operation as a pause in the process of measurement. After you have inserted the next module or connected the thru, press this softkey to continue the process. CONTINUE SEQUENCE resumes a paused sequence.
Hardkey/Softkey Reference Analyzer Functions D2/D1 to D2 on OFF this math function ratios channels 1 and 2, and puts the results in the channel 2 data array. Both channels must be on and have the same number of points. Refer to the "Making Measurements" chapter in the user’s guide for information on how to use this function to make gain compression measurements. DAC NUM HIGH BAND sets the source tune DAC for frequencies above 20.05 GHz.
Hardkey/Softkey Reference Analyzer Functions DEFAULT PLOT SETUP resets the plotting parameters to their default values. DEFAULT PRNT SETUP resets the printing parameters to their default values. DEFINE DISK-SAVE leads to the define save menu. Use this menu to specify the data to be stored on disk in addition to the instrument state. DEFINE PLOT leads to a sequence of three menus. The first defines which elements are to be plotted and the auto feed state.
Hardkey/Softkey Reference Analyzer Functions DIRECTORY SIZE lets you specify the number of directory files to be initialized on a disk. This is particularly useful with a hard disk, where you may want a directory larger than the default 256 files, or with a floppy disk you may want to reduce the directory to allow extra space for data files. The number of directory files must be a multiple of 8. The minimum number is 8, and there is no practical maximum limit.
Hardkey/Softkey Reference Analyzer Functions DO SEQUENCE jumps to a softkey position, not to a specific sequence title. Whatever sequence is in the selected softkey position will run when the DO SEQUENCE command is executed. This command prompts the operator to select a destination sequence position. DONE 1-PORT CAL finishes one-port calibration (after all standards are measured) and turns error correction on.
Hardkey/Softkey Reference Analyzer Functions EACH SWEEP Power meter calibration occurs on each sweep. Each measurement point is measured by the power meter, which provides the analyzer with the actual power reading. The analyzer corrects the power level at that point. The number of measurement/correction iterations performed on each point is determined by the NUMBER OF READINGS softkey. This measurement mode sweeps slowly, especially when the measured power is low.
Hardkey/Softkey Reference Analyzer Functions END SWEEP HIGH PULSE sets the TTL output on TEST SEQ BNC or the test set interconnect to normally high with a 10 microseconds pulse high at the end of each sweep. END SWEEP LOW PULSE sets the TTL output on TEST SEQ BNC or the test set interconnect to normally low with a 10 µs pulse low at the end of each sweep. ENHANCED RESPONSE provides access to the series of menus used to perform an enhanced response calibration. ENH. REFL.
Hardkey/Softkey Reference Analyzer Functions EXTENSION INPUT B adds electrical delay to the input B reference plane for any B input measurements including S-parameters. EXTENSION PORT 1 extends the reference plane for measurements of S11, S21, and S12. EXTENSION PORT 2 extends the reference plane for measurements of S22, S12, and S21. EXTENSIONS on OFF toggles the reference plane extension mode.
Hardkey/Softkey Reference Analyzer Functions FIXED MKR STIMULUS changes the stimulus value of the fixed marker. Fixed marker stimulus values can be different for the two channels if the channel markers are uncoupled using the marker mode menu. To read absolute active marker stimulus values following a MKR ZERO operation, the stimulus value can be reset to zero. FIXED MKR VALUE changes the response value of the fixed marker. In a Cartesian format this is the y-axis value.
Hardkey/Softkey Reference Analyzer Functions FORMAT INT MEMORY clears all internal save registers and associated cal data and memory traces. FORMAT: LIF causes subsequent disk initialization to use the LIF disk format. FORMAT: LIF is the default setting. FORWARD: OPENS provides access to the menu for selecting an open calibration type when the cal kit defines more than one open standard. FREQ OFFS MENU leads to the frequency offset menu (Option 089 only).
Hardkey/Softkey Reference Analyzer Functions G+jB MKR G/n displays the complex admittance values of the active marker in rectangular form. The active marker values are displayed in terms of conductance (in Siemens), susceptance, and equivalent capacitance or inductance. Siemens are the international units of admittance, and are equivalent to mhos (the inverse of Ωs). The Smith chart graticule is changed to admittance form. giga/nano (109 / 10−9). Used to terminate numeric entries.
Hardkey/Softkey Reference Analyzer Functions GRAPH FMT [ ] sets the format of the graphic file when FILETYPE: GRAPHIC is selected. The only graphic selection currently available is the JPEG format. GRAPHICS on OFF specifies whether or not to store display graphics on disk with the instrument state. GRATICULE [ ] brings up the graticule print color definition menu. The graticule default print color is cyan. GRATICULE selects the display graticule for color modification.
Hardkey/Softkey Reference Analyzer Functions IF LOOP < > COUNTER 0 prompts the user to select a destination sequence position (SEQUENCE 1 through 6). When the value of the loop counter is no longer zero, the sequence in the specified position will run. IMAGINARY displays only the imaginary (reactive) portion of the measured data on a Cartesian format. This format is similar to the real format except that reactance data is displayed on the trace instead of impedance data.
Hardkey/Softkey Reference Analyzer Functions KIT DONE (MODIFIED) terminates the cal kit modification process, after all standards are defined and all classes are specified. Be sure to save the kit with the SAVE USER KIT softkey, if it is to be used later. LABEL CLASS leads to the label class menu, to give the class a meaningful label for future reference during calibration. LABEL CLASS DONE finishes the label class function and returns to the modify cal kit menu.
Hardkey/Softkey Reference Analyzer Functions LIMIT TEST on OFF turns limit testing on or off. When limit testing is on, the data is compared with the defined limits at each measured point. Limit tests occur at the end of each sweep, whenever the data is updated, when formatted data is changed, and when limit testing is first turned on. Limit testing is available for both magnitude and phase values in Cartesian formats.
Hardkey/Softkey Reference Analyzer Functions LIST provides a tabular listing of all the measured data points and their current values, together with limit information if it is turned on. At the same time, the screen menu is presented, to enable hard copy listings and access new pages of the table. 30 lines of data are listed on each page, and the number of pages is determined by the number of measurement points specified in the stimulus menu.
Hardkey/Softkey Reference Analyzer Functions LO MENU leads to the LO menu. Allows you to configure the external source for frequency offset. LO SOURCE ADDRESS shows the GPIB address of the LO source. LOAD defines the standard type as a load (termination). Loads are assigned a terminal impedance equal to the system characteristic impedance Z0, but delay and loss offsets may still be added. If the load impedance is not Z0, use the arbitrary impedance standard definition.
Hardkey/Softkey Reference Analyzer Functions LOOP COUNTER displays the current value of the loop counter and allows you to change the value of the loop counter. Enter any number from 0 to 32767 and terminate with the x1 key. The default value of the counter is zero. This command should be placed in a sequence that is separate from the measurement sequence. For this reason: the measurement sequence containing a loop decision command must call itself in order to function.
Hardkey/Softkey Reference Analyzer Functions MAN ’ L THRU on OFF allows the insertion of a different thru during the ECal correction routine. When turned ON, the routine will pause for the insertion of the thru and display a menu that contains a continue and abort softkey. MANUAL TRG ON POINT waits for a manual trigger for each point. Subsequent pressing of this softkey triggers each measurement.
Hardkey/Softkey Reference Analyzer Functions MARKER → DELAY adjusts the electrical delay to balance the phase of the DUT. This is performed automatically, regardless of the format and the measurement being made. Enough line length is added to or subtracted from the receiver input to compensate for the phase slope at the active marker position. This effectively flattens the phase trace around the active marker, and can be used to measure electrical length or deviation from linear phase.
Hardkey/Softkey Reference Analyzer Functions MARKER 2 turns on marker 2 and makes it the active marker. If another marker is present, that marker becomes inactive and is represented on the display as ∆. MARKER 3 turns on marker 3 and makes it the active marker. MARKER 4 turns on marker 4 and makes it the active marker. MARKER 5 turns on marker 5 and makes it the active marker.
Hardkey/Softkey Reference Analyzer Functions MAXIMUM FREQUENCY is used to: 1) define the highest frequency at which a calibration kit standard can be used during measurement calibration. In waveguide, this is normally the upper cutoff frequency of the standard. or 2) set the maximum frequency of the selected frequency band when setting up ripple test parameters. MAXIMUM RIPPLE sets the maximum ripple allowed of the selected frequency band. The maximum allowable ripple is 100 dB.
Hardkey/Softkey Reference Analyzer Functions MINIMUM FREQUENCY is used to: 1) define the lowest frequency at which a calibration kit standard can be used during measurement calibration. In waveguide, this must be the lower cutoff frequency of the standard, so that the analyzer can calculate dispersive effects correctly (see OFFSET DELAY ). or 2) set the minimum frequency of the selected frequency band when setting up ripple test parameters.
Hardkey/Softkey Reference Analyzer Functions MODULE A b selects the module to be used in the ECal operation. Select the module according to frequency range and connector type of the DUT. In some cases the Module B selection will not be used. MODULE INFO presents a text window that contains information about the selected module. MORE RANGES provides access to more power ranges. (ES only) N 50Ω 85054 selects the 85054 cal kit. N 50Ω 85032F selects the 85032F cal kit.
Hardkey/Softkey Reference Analyzer Functions The possible values that can be entered for number of points are 3, 11, 26, 51, 101, 201, 401,801, and 1601. The number of points can be different for the two channels if the stimulus values are uncoupled. In list frequency sweep, the number of points displayed is the total number of frequency points for the defined list (see “Sweep Types” in the “Operating Concepts” chapter of the user’s guide).
Hardkey/Softkey Reference Analyzer Functions ONE-PATH 2-PORT leads to the series of menus used to perform a high-accuracy two-port calibration without an S-parameter test set. This calibration procedure effectively removes directivity, source match, load match, isolation, reflection tracking, and transmission tracking errors in one direction only. Isolation correction can be omitted for measurements of devices with limited dynamic range.
Hardkey/Softkey Reference Analyzer Functions PARALL IN IF BIT L while creating a sequence, this softkey inserts a command to jump to another sequence if the single input selected is in a low state. PARALLEL sets the printer or plotter port to parallel. PARALLEL [COPY/GPIO] toggles the parallel output port between the copy and GPIO output modes. PARALLEL OUT ALL allows you to input a number (0 to 255) in base 10, and outputs it to the bus as binary, when the parallel port is in GPIO mode.
Hardkey/Softkey Reference Analyzer Functions PHASE OFFSET adds or subtracts a phase offset that is constant with frequency (rather than linear). This is independent of MARKER → DELAY and ELECTRICAL DELAY . PHASE (Option 010 only) displays a Cartesian format of the phase portion of the data, measured in degrees. This format displays the phase shift versus frequency.
Hardkey/Softkey Reference Analyzer Functions PLTR PORT SERIAL configures the analyzer for a plotter that has a serial (RS-232) interface. PLTR TYPE [PLOTTER] selects a pen plotter such as the HP 7440A, HP 7470A, HP 7475A, or HP 7550B as the plotter type. PLTR TYPE [HPGL PRT] selects a PCL5 compatible printer, which supports HP-GL/2, such as the LaserJet III or LaserJet 4 for a monochrome plotter type, or the DeskJet 1200C for a color plotter type. POLAR displays a polar format.
Hardkey/Softkey Reference Analyzer Functions This feature may be used to compensate for attenuation non-linearities in either a directional coupler or a power splitter. Up to 12 segments may be entered, each with a different frequency and power loss value. POWER MTR toggles between 436A or 438A/437 . These power meters are GPIB compatible with the analyzer. The model number in the softkey label must match the power meter to be used.
Hardkey/Softkey Reference Analyzer Functions PRINT ALL MONOCHROME when displaying list values, prints the entire list in monochrome. When displaying operating parameters, prints all but the last page in monochrome. The data is sent to the printer as ASCII text rather than as raster graphics, which causes the printout to be faster. PRINT: COLOR sets the print command to default to a color printer. The printer output is always in the analyzer default color values.
Hardkey/Softkey Reference Analyzer Functions PWR LOSS on OFF turns on or off power loss correction. Power loss correction should be used when the power output is measured by a directional coupler. Enter the power loss caused by the coupled arm with the LOSS/SENSR LISTS softkey submenus described below. PWR RANGE AUTO man toggles the power range mode between auto and manual. Auto mode selects the power range based on the power selected.
Hardkey/Softkey Reference Analyzer Functions RAW OFFSET On Off selects whether sampler and attenuator offsets are ON or OFF. By selecting raw offsets OFF, a full two port error correction can be performed without including the effects of the offsets. It also saves substantial time at recalls and during frequency changes. Raw offsets follow the channel coupling. This softkey is used with "Take4" mode. Refer to the examples in the programmer’s guide.
Hardkey/Softkey Reference Analyzer Functions RECALL COLORS recalls the previously saved modified version of the color set. This key appears only when a color set has been saved. RECALL KEYS accesses two recall keys which allows you to set the recall keys menu as the initial menu displayed when Save/Recall is pressed or select specific registers to recall. RECALL KEYS MENU provides access to the recall keys menu where specific registers can be recalled.
Hardkey/Softkey Reference Analyzer Functions REFL: REV S22 (B/R) defines the measurement as S22, the complex reflection coefficient (magnitude and phase) of the test device output. REFLECT AND LINE measures the reflection and thru paths of the current calibration standard. REFLECTION leads to the reflection calibration menu. REFLECTION 1-PORT leads to the reflection calibration menu (ET only). In the ECal menu, this softkey performs the reflection 1-port calibration.
Hardkey/Softkey Reference Analyzer Functions • When in the response and isolation menu, RESPONSE ISOL'N leads to the menus used to perform a response and isolation measurement calibration, for measurement of devices with wide dynamic range. This procedure effectively removes the same frequency response errors as the response calibration. In addition, it effectively removes the isolation (crosstalk) error in transmission measurements or the directivity error in reflection measurements.
Hardkey/Softkey Reference Analyzer Functions REV TRANS THRU is used to enter the standard numbers for the reverse transmission (thru) calibration. (For default kits, this is the thru.) REVERSE: OPENS provides access to the menu for selecting an open calibration type when the cal kit defines more than one open standard. RF > LO adjusts the source frequency higher than the LO by the amount of the LO (within the limits of the analyzer).
Hardkey/Softkey Reference Analyzer Functions RIPL VALUE [ ] displays the ripple value of the selected frequency band. The ripple value can be displayed in two ways or turned off. Selecting OFF removes the displayed ripple value from the display. Selecting ABSOLUTE or MARGIN displays the ripple value. The ripple value is preceded on the display by an indicator of the selected band. For example, when the ripple value is preceded by “B2”, this indicates that the ripple value shown is for Band 2.
Hardkey/Softkey Reference Analyzer Functions S11B is used to enter the standard numbers for the second class required for an S11 1-port calibration. (For default cal kits, this is the short.) S11C is used to enter the standard numbers for the third class required for an S11 1-port calibration. (For default kits, this is the load.) S11 REFL SHORT measures the short circuit TRL/LRM calibration data for PORT 1. S11/S21 ENH. RESP.
Hardkey/Softkey Reference Analyzer Functions SAVE USER KIT stores the user-modified or user-defined kit into memory, after it has been modified. SAVE USING ASCII selects ASCII format for data storage to disk. SAVE USING BINARY selects binary format for data storage. Save/Recall provides access to all the menus used for saving and recalling instrument states in internal memory and for storing to, or loading from the internal or external disk.
Hardkey/Softkey Reference Analyzer Functions SEARCH: OFF turns off the marker search function. SEARCH: TARGET searches for the user-specified target point on the trace. SEGMENT specifies which limit segment in the table is to be modified. A maximum of three sets of segment values are displayed at one time, and the list can be scrolled up or down to show other segment entries. Use the entry block controls to move the pointer > to the required segment number.
Hardkey/Softkey Reference Analyzer Functions SELF DIAGNOSE prompts the analyzer to run a series of tests to determine a problem. Seq accesses a series of sequencing menus. These allow you to create, modify, and store up to 6 sequences which can be run automatically. SEQUENCE 1 SEQ1 activates editing mode for the segment titled "SEQ1" (default title). SEQUENCE 2 SEQ2 activates editing mode for the segment titled "SEQ2" (default title).
Hardkey/Softkey Reference Analyzer Functions SET MONTH allows you to set the month in the analyzer's internal clock. SET REF: REFLECT sets the measurement reference plane to the TRL/LRM REFLECT standard. SET REF: THRU sets the measurement reference plane to the TRL/LRM THRU standard. SET YEAR allows you to set the year in the analyzer's internal clock. SET Z0 sets the characteristic impedance used by the analyzer in calculating measured impedance with Smith chart markers and conversion parameters.
Hardkey/Softkey Reference Analyzer Functions SINGLE POINT sets the limits at a single stimulus point. If limit lines are on, the upper limit value of a single point limit is displayed as ∨ and the lower limit is displayed as ∧. A limit test at a single point not terminating a flat or sloped line tests the nearest actual measured data point. A single point limit can be used as a termination for a flat line or sloping line limit segment.
Hardkey/Softkey Reference Analyzer Functions SMOOTHING APERTURE lets you change the value of the smoothing aperture as a percent of the span. When smoothing aperture is the active function, its value in stimulus units is displayed below its percent value in the active entry area. Smoothing aperture is also used to set the aperture for group delay measurements. Note that the displayed smoothing aperture is not the group delay aperture unless smoothing is on.
Hardkey/Softkey Reference Analyzer Functions SPUR AVOID On Off selects whether spur avoidance is ON or OFF. Selecting spur avoidance OFF, along with selecting raw offsets OFF, saves substantial time at recalls and during frequency changes. Spur avoidance is always coupled between channels. STANDARD DONE returns to the define standard menu. Start is used to define the start frequency of a frequency range. When the Start key is pressed it becomes the active function.
Hardkey/Softkey Reference Analyzer Functions STIMULUS VALUE sets the starting stimulus value of a segment, using entry block controls. The ending stimulus value of the segment is defined by the start of the next line segment. No more than one segment can be defined over the same stimulus range. STIMULUS OFFSET adds or subtracts an offset in stimulus value. This allows limits already defined to be used for testing in a different stimulus range. Use the entry block controls to specify the offset required.
Hardkey/Softkey Reference Analyzer Functions The system controller mode can be used without knowledge of GPIB programming. However, the GPIB address must be entered for each peripheral device. This mode can only be selected manually from the analyzer's front panel, and can be used only if no active computer controller is connected to the system through GPIB. If you try to set system controller mode when another controller is present, the message ANOTHER SYSTEM CONTROLLER ON GPIB is displayed.
Hardkey/Softkey Reference Analyzer Functions TERMINAL IMPEDANCE is used to specify the (arbitrary) impedance of the standard, in Ωs. TEST OPTIONS is used to set configurations before running the service tests. TESTPORT 1 2 is used to direct the RF power to port 1 or port 2. (For non-S parameter inputs only.) TESTSET I/O FWD is used to support specialized test sets, such as a test set that measures duplexers.
Hardkey/Softkey Reference Analyzer Functions TITLE presents the title menu in the softkey labels area and the character set in the active entry area. These are used to label the active channel display. A title more menu allows up to four values to be included in the printed title active entry, active marker amplitude, limit test results, and loop counter value. TITLE SEQUENCE allows the operator to rename any sequence with an eight character title.
Hardkey/Softkey Reference Analyzer Functions TITLE TO PRNTR/GPIB outputs a title string to any device with an GPIB address that matches the address set with the analyzer Local SET ADDRESSES ADDRESS: PRINTER commands. This softkey is generally used for two purposes: • Sending a title to a printer for data logging or documentation purposes. • Sending commands to a printer or other GPIB device. TRACE TYPE [DATA] is located in the ECal confidence menu and ECal service menu.
Hardkey/Softkey Reference Analyzer Functions TRANSMISSION leads to the transmission menu. TRIGGER MENU presents the trigger menu, which is used to select the type and number of the sweep trigger. TRIGGER: TRIG OFF turns off external trigger mode. TRL 3.5 mm 85052C selects the 85052C cal kit. TRL*/LRM* 2-PORT leads to the TRL*/LRM* 2-port calibration menu. TRL/LRM OPTION selects the TRL/LRM Option Menu, under the modify cal kit menu.
Hardkey/Softkey Reference Analyzer Functions USE MEMORY ON off (Option 010 only) remembers a specified window pulse width (or step rise time) different from the standard window values. A window is activated only for viewing a time domain response, and does not affect a displayed frequency domain response.
Hardkey/Softkey Reference Analyzer Functions VOLUME NUMBER specifies the number of the disk volume to be accessed. In general, all 3.5 inch floppy disks are considered one volume (volume 0). For hard disk drives, such as the HP 9153A (Winchester), a switch in the disk drive must be set to define the number of volumes on the disk. WAIT x pauses the execution of subsequent sequence commands for x number of seconds. Terminate this command with x1 .
Hardkey/Softkey Reference Analyzer Functions If a delta marker or fixed marker is on, it is used as the reference point from which the bandwidth amplitude is measured. For example, if marker 1 is the delta marker and is set at the passband maximum, and the width value is set to −3 dB, the bandwidth search finds the bandwidth cutoff points 3 dB below the maximum and calculates the 3 dB bandwidth and Q.
6 Error Messages 6-1
Error Messages Error Messages Error Messages This chapter contains the following information to help you interpret any error messages that may be displayed on the analyzer LCD or transmitted by the instrument over GPIB: • An alphabetical listing of all error messages, including: ❏ An explanation of the message ❏ Suggestions to help solve the problem • A numerical listing of all error messages NOTE 6-2 Some messages described in this chapter are for information only and do not indicate an error condition.
Error Messages Error Messages in Alphabetical Order Error Messages in Alphabetical Order ABORTING COPY OUTPUT Information Message This message is displayed briefly if you have pressed Local to abort a copy operation. If the message is not subsequently replaced by error message number 25, PRINT ABORTED (or PLOT ABORTED) the copy device may be hung. Press Local once more to exit the abort process and verify the status of the copy device.
Error Messages Error Messages in Alphabetical Order ARGUMENT OUT OF RANGE Error Number 206 The argument for a programming command is out of the specified range. Refer to the programmer’s guide for a list of programming commands and argument ranges. ASCII: MISSING 'BEGIN' STATEMENT Error Number 193 The CITIfile you just downloaded over the GPIB or via disk was not properly organized. The analyzer is unable to read the “BEGIN” statement.
Error Messages Error Messages in Alphabetical Order BATTERY FAILED. STATE MEMORY CLEARED Error Number 183 The battery protection of the non-volatile CMOS memory has failed. The CMOS memory has been cleared. Refer to the service guide for battery replacement instructions. See Chapter 8 , “Preset State and Memory Allocation” for more information about the CMOS memory. BATTERY LOW! STORE SAVE REGS TO DISK Error Number 184 The battery protection of the non-volatile CMOS memory is in danger of failing.
Error Messages Error Messages in Alphabetical Order CANNOT FORMAT DOS DISKS ON THIS DRIVE Error Number 185 You have attempted to initialize a floppy disk to DOS format on an external disk drive that does not support writing to all 80 tracks of the double density and high density disks. The older single-sided disks had only 66 tracks and some disk drives were limited to accessing that number of tracks. To format the disk, either choose another external disk drive or use the analyzer's internal disk drive.
Error Messages Error Messages in Alphabetical Order CH1 (CH2, CH3, CH4) TARGET VALUE NOT FOUND Error Number 159 Your target value for the marker search function does not exist on the current data trace. CONTINUOUS SWITCHING NOT ALLOWED Error Number 10 Your current measurement requires different power ranges on channel 1 and channel 2. To protect the attenuator from undue mechanical wear, test set hold will be activated.
Error Messages Error Messages in Alphabetical Order CORRECTION TURNED OFF Error Number 66 Critical parameters in your current instrument state do not match the parameters for the calibration set, therefore correction has been turned off. The critical instrument state parameters are sweep type, start frequency, frequency span, and number of points. CURRENT PARAMETER NOT IN CAL SET Error Number 64 Correction is not valid for your selected measurement parameter.
Error Messages Error Messages in Alphabetical Order DISK IS WRITE PROTECTED Error Number 48 The store operation cannot write to a write-protected disk. Slide the write-protect tab over the write-protect opening in order to write data on the disk. DISK MEDIUM NOT INITIALIZED Error Number 40 You must initialize the disk before it can be used. DISK MESSAGE LENGTH ERROR Error Number 190 The analyzer and the external disk drive aren't communicating properly.
Error Messages Error Messages in Alphabetical Order ECal: CHECKING ORIENTATION Information Message This message is displayed while the network analyzer verifies that the ECal module is properly connected in the RF path. ECal: CONFIDENCE STATE Information Message This message is displayed when the confidence state is initiated. The confidence state response will be displayed. ECal FAILED Error Number 224 This is a serious ECal failure.
Error Messages Error Messages in Alphabetical Order ECal: SELECT OTHER MODULE Error Number 221 This error occurs when two ECal modules are connected. The selected ECal module may not be able to measure the selected frequency range. Check the frequency range of the other module and connect in the RF path. ECal: THRU STATE Information Message This message is displayed when the ECal module is set to the thru state.
Error Messages Error Messages in Alphabetical Order FREQ OFFSET ONLY VALID IN NETWORK ANALYZER MODE (ES Option 089 Only) Error Number 140 You can only make frequency offset measurements in the network analyzer mode. FREQS CANNOT BE CHANGED, TOO MANY POINTS Error Number 204 The number of points selected for setting the low pass transform frequencies is too high. Reduce the number of points so that the low pass criteria is met.
Error Messages Error Messages in Alphabetical Order GPIB COPY IN PROGRESS, ABORT WITH LOCAL Error Number 169 An GPIB copy was already in progress when you requested the GPIB for another function. To abort the first copy, press Local , otherwise the GPIB is unavailable until the first copy is completed. IF BW KEY DISABLED, EDIT LIST MODE TBL Information Message When list IF bandwidth has been enabled and swept list mode is on, you will not be able to change the IF bandwidth using the IF BW key.
Error Messages Error Messages in Alphabetical Order INSUFFICIENT MEMORY, PWR MTR CAL OFF Error Number 154 There is not enough memory space for the power meter calibration array. Increase the available memory by clearing one or more save/recall registers, or by reducing the number of points. INVALID KEY Error Number 2 You pressed an undefined softkey. ISOL AVERAGES < SWP AVERAGES Error Number 223 The isolation averages are less than the instrument sweep averages.
Error Messages Error Messages in Alphabetical Order MEMORY FOR CURRENT SEQUENCE IS FULL Error Number 132 All the memory in the sequence you are modifying is filled with instrument commands. MORE SLIDES NEEDED Error Number 71 When you use a sliding load (in a user-defined calibration kit), you must set at least three slide positions to complete the calibration. NO CALIBRATION CURRENTLY IN PROGRESS Error Number 69 The RESUME CAL SEQUENCE softkey is not valid unless a calibration is already in progress.
Error Messages Error Messages in Alphabetical Order NO MEMORY AVAILABLE FOR INTERPOLATION Error Number 123 You cannot perform interpolated error correction due to insufficient memory. NO MEMORY AVAILABLE FOR SEQUENCING Error Number 126 You cannot modify the sequence due to insufficient memory. NO MODULE DATA FOR ISOLATION STD Information Message This message is displayed in ECal service when there is no isolation data in the ECal module.
Error Messages Error Messages in Alphabetical Order OK TO ALTER CORRECTION CONSTANTS? Error Number 186 This message is displayed as a prompt during operations when the correction constants will be changed. Correction constants may be change in the adjustment tests in the service menu. ONLY LETTERS AND NUMBERS ARE ALLOWED Error Number 43 You can only use alpha-numeric characters (and underscores) in disk file titles or internal save register titles.
Error Messages Error Messages in Alphabetical Order OVERLOAD ON INPUT R, POWER REDUCED Error Number 57 You have exceeded approximately +14 dBm at one of the test ports. The RF output power is automatically reduced to −85 dBm. The annotation P⇓ appears in the left margin of the display to indicate that the power trip function has been activated. When this occurs, reset the power to a lower level, then toggle the softkey to switch on the power again.
Error Messages Error Messages in Alphabetical Order PLOTTER: not on, not connect, wrong addrs Error Number 26 The plotter does not respond to control. Verify power to the plotter, and check the GPIB connection between the analyzer and the plotter. Ensure that the plotter address recognized by the analyzer matches the GPIB address set on the plotter itself. PLOTTER NOT READY-PINCH WHEELS UP Error Number 28 The plotter pinch wheels clamp the paper in place.
Error Messages Error Messages in Alphabetical Order PRINT ABORTED Error Number 25 When you press the Local key, the analyzer aborts output to the printer. print color not supported with EPSON Error Number 178 You have defined the printer type as EPSON-P2. Color print is not supported with this printer. The print will abort. PRINTER: busy Error Number 176 The parallel port printer is not accepting data. PRINTER: error Error Number 175 The parallel port printer is malfunctioning.
Error Messages Error Messages in Alphabetical Order PRINTER: power off Error Number 174 The power to the printer at the parallel port is off. PRINTER: reset in progress Information Message If the printer takes longer than a half-second to reset, this message will be displayed until printer is finished with reset.
Error Messages Error Messages in Alphabetical Order SEGMENT #n POWER OUTSIDE RANGE LIMIT Information Message The selected power range does not support the power level of one or more segments in the swept list table. This message appears when swept list mode is not on and reports the first segment that is out of range. Change the segment power or change the power range.
Error Messages Error Messages in Alphabetical Order SOURCE POWER DISABLED, EDIT LIST MODE TBL Information Message When list power has been enabled and swept list mode is on, you will not be able to change the power level using the Power key. To change the power level, edit the swept list table. SOURCE POWER TURNED OFF, RESET UNDER POWER MENU Information Message You have exceeded the maximum power level at one of the inputs and power has been automatically reduced.
Error Messages Error Messages in Alphabetical Order SYST CTRL OR PASS CTRL IN LOCAL MENU Error Number 36 The analyzer is in talker/listener mode. In this mode, the analyzer cannot control a peripheral device on the bus. Use the local menu to change to system controller or pass control mode. TEST ABORTED Error Number 113 You have prematurely stopped a service test.
Error Messages Error Messages in Alphabetical Order VALID ONLY FOR BILATERAL DEVICES Information Message This message is displayed when an enhanced reflection calibration is initiated. If the device tested is not a bilateral device, the enhanced reflection calibration will cause errors in the measurement results. A bilateral device has similar forward and reverse transmission characteristics. Examples of bilateral devices are passive devices (filters, attenuators, and switches).
Error Messages Error Messages in Numerical Order Error Messages in Numerical Order Error Number Error 1 OPTIONAL FUNCTION; NOT INSTALLED 2 INVALID KEY 3 CORRECTION CONSTANTS NOT STORED 4 PHASE LOCK CAL FAILED 5 NO IF FOUND: CHECK R INPUT LEVEL 6 POSSIBLE FALSE LOCK 7 PHASE LOCK FAILURE 8 PHASE LOCK LOST 9 LIST TABLE EMPTY 10 CONTINUOUS SWITCHING NOT ALLOWED 11 SWEEP TIME INCREASED 12 SWEEP TIME TOO FAST 13 AVERAGING INVALID ON NON-RATIO MEASURE 14 FUNCTION NOT VALID 15 NO MAR
Error Messages Error Messages in Numerical Order Error Number Error 33 SYNTAX ERROR 34 BLOCK INPUT ERROR 35 BLOCK INPUT LENGTH ERROR 36 SYST CTRL OR PASS CTRL IN LOCAL MENU 37 ANOTHER SYSTEM CONTROLLER ON GPIB 38 DISK: not on, not connected, wrong addrs 39 DISK HARDWARE PROBLEM 40 DISK MEDIUM NOT INITIALIZED 41 NO DISK MEDIUM IN DRIVE 42 FIRST CHARACTER MUST BE A LETTER 43 ONLY LETTERS AND NUMBERS ARE ALLOWED 44 NOT ENOUGH SPACE ON DISK FOR STORE 45 NO FILE(S) FOUND ON DISK 46
Error Messages Error Messages in Numerical Order Error Number Error 63 CALIBRATION REQUIRED 64 CURRENT PARAMETER NOT IN CAL SET 65 CORRECTION AND DOMAIN RESET 66 CORRECTION TURNED OFF 67 DOMAIN RESET 68 ADDITIONAL STANDARDS NEEDED 69 NO CALIBRATION CURRENTLY IN PROGRESS 70 NO SPACE FOR NEW CAL.
Error Messages Error Messages in Numerical Order Error Number Error 132 MEMORY FOR CURRENT SEQUENCE IS FULL 133 THIS LIST FREQ INVALID 140 FREQ OFFSET ONLY VALID IN NETWORK ANALYZER MODE 144 NO LIMIT LINES DISPLAYED 145 SWEEP TYPE CHANGED TO LINEAR SWEEP 150 LOG SWEEP REQUIRES 2 OCTAVE MINIMUM SPAN 151 SAVE FAILED / INSUFFICIENT MEMORY 152 D2/D1 INVALID: CH1 CH2 NUM PTS DIFFERENT 153 SEQUENCE MAY HAVE CHANGED, CAN'T CONTINUE 154 INSUFFICIENT MEMORY, PWR MTR CAL OFF 157 SEQUENCE ABORT
Error Messages Error Messages in Numerical Order Error Number Error 181 BAD FREQ FOR HARMONIC OR FREQ OFFSET 182 LIST MODE OFF: INVALID WITH LO FREQ 183 BATTERY FAILED.
Error Messages Error Messages in Numerical Order Error Number Error 212 ANALOG BUS DISABLED IN 6 KHZ IF BW 213 RANGE CAUSED POWER LVL CHANGE IN LIST 214 CORRECTION ON: AUX CHANNEL(S) RESTORED 215 CAUTION: CORRECTION OFF: AUX CHANNEL(S) DISABLED 218 CAUTION: FLOPPY DISK IS FULL 219 ECal MODULE NOT IN RF PATH 220 SELECTED MODULE OUTSIDE START-STOP FREQ RANGE 221 ECal SELECT OTHER MODULE 222 ECal MODULE NOT RESPONDING 223 ISOL AVGS < SWP AVGS 224 ECal FAILED 6-31
7 Options and Accessories 7-1
Options and Accessories Using This Chapter Using This Chapter This chapter contains information on the following subjects: • “Analyzer Options Available” on page 7-3 • “Accessories Available” on page 7-5 7-2
Options and Accessories Analyzer Options Available Analyzer Options Available Option 1D5, High Stability Frequency Reference Option 1D5 offers ±0.05 ppm temperature stability from 0 to 55 °C (referenced to 25 °C), and aging rate of ±0.5 ppm per year (typical). Option 004, Source Attenuator (ET Only) This option adds a 55 dB step attenuator that extends the output power range by allowing lower power levels.
Options and Accessories Analyzer Options Available Additionally, there is an external reference input that allows the external amplifier's frequency response and drift to be ratioed out, and there are internally controlled step attenuators between the couplers and samplers to prevent overload.
Options and Accessories Accessories Available Accessories Available For accessories not listed in this section, refer to the configuration guide for your analyzer or refer to the following Internet site: www.agilent.com/find/8720 Measurement Accessories Accessories are available in these connector types: 2.4-mm, 3.5-mm, 7-mm, 50 Ω type-N, and 7-16. Test port cables are used to connect to the device under test.
Options and Accessories Accessories Available • 85134F set, flexible: 3.5-mm to 2.4-mm, 53-cm (21-in) each. For the 8722ET/ES. Test-Port Cables: 7mm • 85132D set, semi-rigid: 7-mm to 3.5-mm, 53-cm (21-in) each. For the 8719ET/ES or 8720ET/ES. • 85132E single, flexible: 7-mm to 3.5-mm, 96-cm (38-in). For the 8719ET/ES or 8720ET/ES. • 85132F set, flexible: 7-mm to 3.5-mm, 53-cm (21-in) each. For the 8719ET/ES or 8720ET/ES. • 85135C single, semi-rigid: 7-mm to 2.4-mm, 81-cm (32-in). For the 8722ET/ES.
Options and Accessories Accessories Available — Option 001 Deletes 3.5-mm to 7-mm adapters. Recommended if the kit will not be used with the 8753 network analyzer. • 85033E 50 Ω 3.5-mm calibration kit (30kHz to 9 GHz) Contains a female and male open and short, female and male load standard, and a torque wrench. • 85050B 7-mm standard calibration kit (0.045 to 18 GHz) Contains fixed loads, open and short circuits, and terminations. • 85050C 7-mm TRL calibration kit (0.
Options and Accessories Accessories Available RF electronic calibration (ECal) modules and PC software This product family provides electronic calibration (ECal) capability. With ECal, the usual calibration kit standards are replaced by one solid-state calibration module. A full two-port calibration can be done with a single connection, with reduced errors and wear on connectors.
Options and Accessories Accessories Available — Option 00M substitutes type-N (m) to type-N (m) Microwave ECal module. — Option 001 adds a 30 kHz to 9 GHz RF ECal module. — Option 00A adds: — type-N (f) to type-N (f) adapter — type-N (m) to type-N (m) adapter • 85092B RF ECal module (30 kHz to 9 GHz), 50 Ω type-N (m) to 50 Ω type-N (f) connectors — Option 00F substitutes type-N (f) to type-N (f) RF ECal module. — Option 00M substitutes type-N (m) to type-N (m) RF ECal module.
Options and Accessories Accessories Available Adapters • 11904A 2.4-mm (m) to K (m) adapter • 11904B 2.4-mm (f) to K (f) adapter • 11904C 2.4-mm (m) to K (f) adapter • 11904D 2.4-mm (f) to K (m) adapter • 11904S 2.4-mm to K adapter set • 11906A 7-16 to 7-16 adapter kit Contains one 7-16(m) to 7-16(m) adapter, one 7-16(f) to 7-16(f) adapter, and two 7-16(m) to 7-16(f) adapters.
Options and Accessories Accessories Available Test Configuration Accessories Power Meters For more accurate control of leveled test-port power. Requires an 8480-series power sensor. • E4418B single-channel power meter • E4419B dual-channel power meter Power Sensors For more accurate control of leveled test-port power. Requires a 436A, 437B, 438B, or E4419B power meter and an GPIB cable for connection to the network analyzer. • 8481B 10 MHz to 18 GHz, type-N (m), 25 watt • 8482B 100 kHz to 4.
Options and Accessories Accessories Available Keyboard Template The analyzer is designed to accept most PC-AT-compatible keyboards with a mini-DIN connector. The keyboard can be used for control or data input, such as titling files. The information found on the analyzer keyboard template (part number 08753-80220) is also listed in Table 7-1.
8 Preset State and Memory Allocation 8-1
Preset State and Memory Allocation Preset State Preset State When the Preset key is pressed, the analyzer reverts to a known state called the factory preset state. This state is defined in Table 8-1. There are subtle differences between the preset state and the power-up state. These differences are documented in Table 8-3. If power to non-volatile memory is lost, the analyzer will have certain parameters set to default settings. The affected parameters are shown in Table 8-4.
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions Preset Conditions Preset Value Analyzer Mode Analyzer Mode Network Analyzer Mode Frequency Offset Operation (Opt. 089) Off Offset Value 0 High Power (Opt. 085) External R Channel Off Attenuator A 0 dB Attenuator B 0 dB Stimulus Conditions Sweep Type Linear Frequency Step Sweep Off Step Sweep (Opt. 085) On Display Mode Start/Stop Trigger Type Continuous External Trigger Off Sweep Time 515.
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Source Power (8722, Opt. 007) −5 dBm Start Power (8719/20) −15.0 dBm Start Power (8719/20, Opt. 007) −10.0 dBm Start Power (8719/20, Opt. 400) −20.0 dBm Start Power (8722) −20.0 dBm Start Power (8722, Opt. 007) −15.0 dBm Start Power (8722, Opt. 400) −25.
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Conversion Off Format Log Magnitude (all inputs) Display Data Color Selections Same as before Preset Dual Channel Off Active Channel Channel 1 Auxiliary Channel Disabled Frequency Blank Disabled Retrace Power Standard Test Set Switch (ES) Continuous Test Set Switch (Opt.
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value System Z0 50 Ohms Velocity Factor 1 Extensions Off Port 1 0s Port 2 0s Input A 0s Input B 0s Chop A and B (ES) On Chop RFL & TRN (ET) On Power Meter Calibration Off Number of Readings 1 Power Loss Correction Off Sensor A/B A Interpolated Error Correction On Electronic Calibration (ECal) Module A (information may not be loaded) Omit Isolation On Isolation A
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Marker Width Value −3 dB; Off Marker Tracking Off Marker Stimulus Offset 0 Hz Marker Value Offset 0 dB Marker Aux Offset (Phase) 0 Degrees Marker Statistics Off Polar Marker Lin Mkr Smith Marker R+jX Mkr Limit Menu Limit Lines Limit Lines Off Limit Testing Off Limit List Empty Edit Mode Upper/Lower Limits Stimulus Offset 0 Hz Amplitude Offset 0 dB Limit Type
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Transform Span 4 nanoseconds Gating Off Gate Shape Normal Gate Start −500 picoseconds Gate Span 500 picoseconds Demodulation Off Window Normal Use Memory Off System Parameters GPIB Addresses Last Active State GPIB Mode Last Active State Clock Time Stamp On Preset: Factory/User Last Selected State Copy Configuration Parallel Port Last Active State Plotter Type L
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Formatted Data Array Off Graphics Off Data Only Off Directory Size Defaulta Save Using Binary Select Disk Internal Memory Disk Format LIF Sequencingb Loop Counter 0 TTL OUT High Service Modes GPIB Diagnostic Off Source Phase Lock Loop On Aux Input Resolution Low Analog Bus Node Off Plot Plot Data On Plot Memory On Plot Graticule On Plot Text On Plot Mark
Preset State and Memory Allocation Preset State Table 8-1 Preset Conditions (Continued) Preset Conditions Preset Value Ch2/Ch4 Memory 6 Ch1/Ch3 Graticule 1 Ch2/Ch4 Graticule 1 Ch1/Ch3 Text 7 Ch2/Ch4 Text 7 Ch1/Ch3 Marker 7 Ch2/Ch4 Marker 7 Line Type: Ch1/Ch3 Data 7 Ch2/Ch4 Data 7 Ch1/Ch3 Memory 7 Ch2/Ch4 Memory 7 Print Printer Mode Last Active State Auto-Feed On Printer Colors Ch1/Ch3 Data Magenta Ch1/Ch3 Mem Green Ch2/Ch4 Data Blue Ch2/Ch4 Mem Red Graticule Cyan Warni
Preset State and Memory Allocation Preset State Table 8-2 Preset Conditions Format Table Scale Reference Position Value Log Magnitude (dB) 10.0 5.0 0.0 Phase (degree) 90.0 5.0 0.0 Group Delay (ns) 10.0 5.0 0.0 Smith Chart 1.00 – 1.0 Polar 1.00 – 1.0 Linear Magnitude 0.1 0.0 0.0 Real 0.2 5.0 0.0 Imaginary 0.2 5.0 0.0 SWR 1.00 0.0 1.0 Table 8-3 Power-On Conditions (versus Preset) GPIB MODE Talker/listener.
Preset State and Memory Allocation Preset State Table 8-4 Results of Power Loss to Non-Volatile Memory GPIB ADDRESSES are set to the following defaults: ANALYZER 16 USER DISPLAY 17 PLOTTER 5 PRINTER 1 POWER METER 13 DISK 0 DISK UNIT NUMBER 0 DISK VOLUME NUMBER 0 POWER METER TYPE is set to 438A/437 INTERNAL REGISTER TITLESa are set to defaults: REG1 through REG32 EXTERNAL REGISTER TITLESa (store files) are set to defaults: FILE1 through FILE 5 PRINT TYPE is set to default: MONOCHROME PRINTIN
Preset State and Memory Allocation Memory Allocation Memory Allocation The analyzer is capable of saving complete instrument states for later retrieval. It can store these instrument states into the internal memory, to the internal disk, or to an external disk.
Preset State and Memory Allocation Memory Allocation NOTE Even though calibration data is stored in non-volatile memory, if the associated instrument state is not saved, you will not be able to retrieve the calibration data after a power cycle.
Preset State and Memory Allocation Memory Allocation Determining Memory Requirements Table 8-5 shows the memory requirements of calibration arrays and memory trace arrays to help you approximate memory requirements. For example, add the following memory requirements: • a full 2-port calibration with 801 points (58 k) • the memory trace array (4.9 k) • the instrument state (6 k) The total memory requirement is 68.9 kbytes. There is sufficient memory to store 29 calibrations of this type.
Preset State and Memory Allocation Memory Allocation Table 8-5 Memory Requirements of Calibration and Memory Trace Arrays Variable Approximate Totals (Bytes) Data Length (Bytes)a 401 pts 801 pts 1 chan 1601 pts 1 chan 2 chans Calibration Arrays Response N × 6 + 52 2.5 k 5k 10 k 19 k Response and isolation N × 6 × 2 + 52 5k 10 k 19 k 38 k N × 6 × 3 + 52 7k 14 k 29 k 58 k N × 6 × 12 + 52 29 k 58 k 115 k 230 k (Nc × 2 × number channelsd) +208 1k 1.8 k 3.4 k 6.
Preset State and Memory Allocation Memory Allocation Storing Data to Disk You can use the internal disk drive or connect an external disk drive for storage of instrument states, calibration data, measurement data, and plot files. (Refer to the “Printing, Plotting, and Saving Measurement Results” chapter in the user’s guide for more information on saving measurement data and plot files.) The analyzer displays one file name per stored instrument state when you list the disk directory.
Preset State and Memory Allocation Memory Allocation Table 8-6 Suffix Character Definitions Char 1 Definition I, P Instrument statea W Four-channel instrument state G Graphics D R F Error corrected data Raw data Formatted data Char 2 Definition 1 Display graphics 0 Graphics index 1 Channel 1 2 Channel 2 3 Channel 3 4 Channel 4 1 to 4 Channel 1/3, raw arrays 1 to 4b 5 to 8 Channel 2/4, raw arrays 5 to 8 1 Channel 1 2 Channel 2 3 Channel 3 4 Channel 4 C Cal K Cal ki
Preset State and Memory Allocation Memory Allocation If correction is on at the time of an external store, the calibration set is stored to disk. (Note that inactive calibrations are not stored to disk.) When an instrument state is loaded into the analyzer from disk, the stimulus and response parameters are restored first. If correction is on for the loaded state, the analyzer will load a calibration set from disk that carries the same title as the one stored for the instrument state.
9 Understanding the CITIfile Data Format 9-1
Understanding the CITIfile Data Format Using This Chapter Using This Chapter The descriptions and examples shown in this chapter demonstrate how CITIfile may be used to store and transfer both measurement information and data. The use of a single, common format will allow data to be more easily moved between instruments and computers.
Understanding the CITIfile Data Format The CITIfile Data Format The CITIfile Data Format Description and Overview CITIfile is a standardized data format, used for exchanging data between different computers and instruments. CITIfile is an abbreviation for “Common Instrumentation Transfer and Interchange file”. This standard has been a group effort between instrument designers and designers of computer-aided design programs.
Understanding the CITIfile Data Format The CITIfile Data Format A CITIfile Package A typical package is divided into two parts: The first part, the header, is made up of keywords and setup information. The second part, the data, usually consists of one or more arrays of data. Example 1 shows the basic structure of a CITIfile package: Example 1, A CITIfile Package The “header” part CITIFILE A.01.00 NAME MEMORY VAR FREQ MAG 3 DATA S RI The “data” part BEGIN -3.54545E-2, -1.38601E-3 0.23491E-3, -1.
Understanding the CITIfile Data Format The CITIfile Data Format A CITIfile package may contain more than one array of data. Arrays of data start after the BEGIN keyword, and the END keyword will follow the last data element in an array. A CITIfile package does not necessarily need to include data arrays; for instance, CITIfile could be used to store the current state of an instrument. In that case the keywords VAR, DATA, BEGIN, and END would not be required.
Understanding the CITIfile Data Format The CITIfile Data Format NAME DATA #NA REGISTER 1 VAR FREQ MAG 10 DATA S[1,1] RI SEG_LIST_BEGIN SEG 1000000000 4000000000 10 SEG_LIST_END BEGIN 0.86303E-1,-8.98651E-1 8.97491E-1,3.06915E-1 -4.96887E-1,7.87323E-1 -5.65338E-1,-7.05291E-1 8.94287E-1,-4.25537E-1 1.77551E-1,8.96606E-1 -9.35028E-1,-1.10504E-1 3.69079E-1,-9.13787E-1 7.80120E-1,5.37841E-1 -7.78350E-1,5.
Understanding the CITIfile Data Format The CITIfile Data Format #NA DUPLICATES 0 #NA ARB_SEG 1000000000 1000000000 1 #NA ARB_SEG 2000000000 3000000000 3 VAR_LIST_BEGIN 1000000000 2000000000 2500000000 3000000000 VAR_LIST_END BEGIN 1.12134E-3,1.73103E-3 4.23145E-3,-5.36775E-3 -0.56815E-3,5.32650E-3 -1.85942E-3,-4.07981E-3 END BEGIN 2.03895E-2,-0.82674E-2 -4.21371E-2,-0.24871E-2 0.21038E-2,-3.06778E-2 1.20315E-2,5.99861E-2 END BEGIN 4.45404E-1,4.31518E-1 8.34777E-1,-1.33056E-1 -7.09137E-1,5.58410E-1 4.
Understanding the CITIfile Data Format CITIfile Keywords CITIfile Keywords Keyword Explanation and Examples CITIFILE CITIFILE A.01.01 identifies the file as a CITIfile, and indicates the revision level of the file. The CITIfile keyword and revision code must precede any other keywords. The CITIfile keyword at the beginning of the package assures the device reading the file that the data that follows is in the CITIfile format. The revision number allows for future extensions of the CITIfile standard.
Understanding the CITIfile Data Format CITIfile Keywords SEG_LIST_BEGIN SEG_LIST_BEGIN indicates that a list of segments for the independent variable follow. Format for the segments is: [segment type] [start] [stop] [number of points]. The current implementation only supports a single segment. If there is more than one segment, the VAR_LIST_BEGIN construct is used. CITIfile revision A.01.00 supports only the SEG (linear segment) segment type.
Understanding the CITIfile Data Format CITIfile Keywords CONSTANT CONSTANT [name] [value] allows for the recording of values which don’t change when the independent variable changes. CONSTANTs are part of the main CITIfile definition. Users must not define their own CONSTANTs. Use the #KEYWORD device specification to create your own KEYWORD instead. The #NA device specification is an example of this. No constants were defined for revision A.01.00 of CITIfile. CITIfile revision A.01.
Understanding the CITIfile Data Format Useful Calculations Useful Calculations This section contains information on computing frequency points and expressing CITIfile data in other data formats. Computing Frequency Points In CITIfile, the frequency data is not listed point by point, only the start and stop values are given. If you are using a spreadsheet program, you can create a new frequency column to the left of the data pairs.
Understanding the CITIfile Data Format Useful Calculations Expressing CITIfile Data in Other Data Formats CITIfile data is represented in real and imaginary pairs. Equations can be used to express this information in logarithmic magnitude, phase, polar, and Smith chart formats. Refer to the following table for these equations.
Understanding the CITIfile Data Format Useful Calculations Example Data This example shows how the following CITIfile data for a three-point trace can be expressed in other data formats. CITIFILE A.01.00 #NA VERSION HP8753E.07.12 NAME DATA VAR FREQ MAG 3.0000 DATA S[11] RI SEG_LIST_BEGIN SEG 1550000000 1570000000 3.
10 Determining System Measurement Uncertainties 10-1
Determining System Measurement Uncertainties Introduction Introduction In any measurement, certain measurement errors associated with the system add uncertainty to the measured results. This uncertainty defines how accurately a device under test (DUT) can be measured. This chapter describes how the various network analyzer measurement error sources contribute to uncertainties in the magnitude and phase measurements of both transmission and reflection.
Determining System Measurement Uncertainties Sources of Measurement Errors Sources of Measurement Errors Measurement errors are made up of systematic errors, random errors, and drift errors. Each of these measurement error types is discussed in this section. Sources of Systematic Errors The residual (after measurement calibration) systematic errors result from imperfections in the calibration standards. All measurements are affected by dynamic accuracy.
Determining System Measurement Uncertainties Sources of Measurement Errors Sources of Random Errors The random error sources are noise, connector repeatability and interconnecting cable stability. There are two types of noise in any measurement system: low level noise (noise floor) and high level noise (trace noise). Low level noise is the broadband noise floor of the receiver which can be reduced through averaging or by changing the IF bandwidth.
Determining System Measurement Uncertainties Determining Expected System Performance Determining Expected System Performance Improper connection techniques and contact surfaces can degrade measurement accuracy. Proper connection techniques include using a torque wrench with proper torque limits, ensuring that the connector pin depths meet specifications, ensuring that the center conductor of sliding loads is properly set, and observing proper handling procedures for beadless airlines.
Determining System Measurement Uncertainties Determining Cable Stability Terms (CR1, CR2, CTM1, CTM2, CTP1, CTP2) Determining Cable Stability Terms (CR1, CR2, CTM1, CTM2, CTP1, CTP2) Cable stability is dependent on the cable used and the amount of cable movement between calibration and measurement.
Determining System Measurement Uncertainties Determining Cable Stability Terms (CR1, CR2, CTM1, CTM2, CTP1, CTP2) Figure 10-2 Cable Stability with a Short Connected Figure 10-2 and Figure 10-3 demonstrate the concepts useful in determining cable transmission stability. A short is connected to the free end. The DATA/MEM feature provides an indication of the two-way cable transmission stability.
Determining System Measurement Uncertainties Measurement Uncertainty Equations Measurement Uncertainty Equations Any measurement result is the vector sum of the actual test device response plus all error terms. The precise effect of each error term depends on its magnitude and phase relationship to the actual test device response. When the phase of an error response is not known, phase is assumed to be worst case (–l80° to +180°). Forward Reflection Uncertainty Equation 10-1.
Determining System Measurement Uncertainties Measurement Uncertainty Equations Forward Transmission Uncertainty Equation 10-3.
Determining System Measurement Uncertainties Measurement Uncertainty Equations Reverse Reflection Uncertainty Equation 10-5.
Determining System Measurement Uncertainties Measurement Uncertainty Equations Reverse Transmission Uncertainty Equation 10-7.
Index A accessories available, 7-5 keyboard template, 7-12 measurement accessories, 7-5 adapters, 7-10 allocation, memory, 8-13 Analyzer panels front, 3-2 analyzer display, 3-5 analyzer functions, 5-4 analyzer options available, 7-3 direct access receiver configuration, Option 012, 7-3 four-sampler test set, Option 400, 7-4 frequency offset mode, Option 089, 7-4 high power system, Option 085, 7-3 high stability frequency reference, Option 1D5, 7-3 mechanical transfer switch, Option 007, 7-3 rack mount flan
Index P pads, minimum loss, 7-10 PC software, 7-8 performance, determining, 10-5 power meters, 7-11 preset state, 8-2 uncertainty, measurement, 10-1 V verification kit, 7-9 volatile memory, 8-13 R rack mount flange kit with handles, Option 1CP, 7-4 rack mount flange kit without handles, Option 1CM, 7-4 random errors, 10-2, 10-4 raw error terms, 10-2 rear panel features and connectors, 3-10 repeatability, connector, 10-5 repeatable errors, 10-2 residual errors, 10-2, 10-3 reverse reflection uncertainty, 1