Liquiline Mobile CML18 Table of contents Table of contents 1 About this document . . . . . . . . . . . 4 1.1 1.2 1.3 Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Symbols on the device . . . . . . . . . . . . . . . . 5 2 Basic safety instructions . . . . . . . 6 2.1 2.2 2.3 2.4 2.5 Requirements for personnel . . . . . . . . . . . Designated use . . . . . . . . . . . . . . . . . . . . . . Workplace safety . . . . .
About this document Liquiline Mobile CML18 1 About this document 1.1 Warnings Structure of information LDANGER Causes (/consequences) If necessary, Consequences of noncompliance (if applicable) ‣ Corrective action LWARNING Causes (/consequences) If necessary, Consequences of noncompliance (if applicable) ‣ Corrective action LCAUTION Causes (/consequences) If necessary, Consequences of noncompliance (if applicable) ‣ Corrective action NOTICE Meaning This symbol alerts you to a dangerous situation.
Liquiline Mobile CML18 1.
Basic safety instructions 2 Basic safety instructions 2.1 Requirements for personnel Liquiline Mobile CML18 • Installation, commissioning, operation and maintenance of the measuring system may be carried out only by specially trained technical personnel. • The technical personnel must be authorized by the plant operator to carry out the specified activities. • The electrical connection may be performed only by an electrical technician.
Liquiline Mobile CML18 Basic safety instructions 2. Ensure that electrical cables and hose connections are undamaged. 3. Do not operate damaged products, and protect them against unintentional operation. 4. Label damaged products as defective. During operation: ‣ If faults cannot be rectified: products must be taken out of service and protected against unintentional operation. 2.5 Product safety 2.5.
Product description Liquiline Mobile CML18 3 Product description 3.
Liquiline Mobile CML18 3.1.1 Product description Measuring parameters The mobile device is designed for digital Memosens sensors with an inductive plug-in head and fixed cable sensors with the Memosens protocol and no external power supply: • pH • ORP • pH/ORP combination sensors • Conductive conductivity • Inductive conductivity • Dissolved oxygen (optical/amperometric) • Temperature The measuring range is adapted to the individual sensor type.
Incoming acceptance and product identification Liquiline Mobile CML18 4 Incoming acceptance and product identification 4.1 Incoming acceptance 1. Verify that the packaging is undamaged. Notify the supplier of any damage to the packaging. Keep the damaged packaging until the issue has been resolved. 2. Verify that the contents are undamaged. Notify the supplier of any damage to the delivery contents. Keep the damaged goods until the issue has been resolved. 3.
Liquiline Mobile CML18 Incoming acceptance and product identification Obtaining information on the product 1. Go to www.endress.com. 2. Call up the site search (magnifying glass). 3. Enter a valid serial number. 4. Search. The product structure is displayed in a popup window. 5. Click on the product image in the popup window. A new window (Device Viewer) opens. All of the information relating to your device is displayed in this window as well as the product documentation.
Incoming acceptance and product identification 4.4.3 Liquiline Mobile CML18 ISED Canada This device complies with ISED’s license-exempt RSSs. Operation is subject to the following two conditions: (1) This device may not cause interference; and (2) This device must accept any interference, including interference that may cause undesired operation of the device. This device complies with the safety requirements for RF exposure in accordance with RSS-102 Issue 5 for portable use conditions.
Liquiline Mobile CML18 Installation Installation 5.1 Installation conditions 5.1.1 Dimensions 47 (1.85) 84 (3.31) 5 54 (2.
Electrical connection Liquiline Mobile CML18 6 Electrical connection 6.1 Connecting the sensor 6.1.1 Connecting the Memosens sensor directly 1. 2. A0040973 3 Sensor connection 1. Insert the sensor into the Memosens connection. 2. Click the Memosens connection into place. 6.1.2 Connecting the Memosens sensor with the M12 fixed cable connection 3. 1. 2. A0040974 1. Remove the protection cap. 2. Insert the M12 fixed cable. 3. Screw on the M12 fixed cable.
Liquiline Mobile CML18 6.1.3 Electrical connection Connecting the sensor via the Memosens M12 cable 3. 1. 5. 2. 4. A0040974 4 1. Remove the protection cap. 2. Attach the M12 cable. 3. Screw on the M12 cable. 4. Insert the sensor into the Memosens connection. 5. Click the Memosens connection into place. 6.2 A0040972 A0041005 Cable connection Charging the device Only use Qi-certified chargers (Baseline Power)! For more information, visit: www.wirelesspowerconsortium.
Electrical connection Liquiline Mobile CML18 2. 1. A0040975 5 Inductive charging 1. Connect the charger to the power source. 2. Place the device with the charging side on the charger. Charging begins and the charging state is indicated on the display. 6.3 Ensuring the degree of protection Only the mechanical and electrical connections which are described in these instructions and which are necessary for the required, designated use, may be carried out on the device delivered.
Liquiline Mobile CML18 Operation options 7 Operation options 7.1 Overview of operation options Operation and settings via: • Internal operating menu with keys • SmartBlue App → 20 7.1.
Operation options 7.1.
Liquiline Mobile CML18 Operation options Display structure 2 3 4 1 5 A0041052 7 1 2 3 4 5 Schematic representation of the display structure Menu path/title of measuring screen Bluetooth status Battery level, charging information NAMUR indicator * Measuring screen 1) NAMUR indicator Status OK The device and sensor are working reliably. F Failure of device or sensor. F status signal as per NAMUR NE107 M Device or sensor requires maintenance.
Operation options Liquiline Mobile CML18 Structure of the measurement window The measurement window has 3 measuring screens, which the user can scroll through: Measuring screen (1 of 3) Measuring screen (2 of 3) Measuring screen (3 of 3) Main value Main and secondary measured value All measured values of the sensor input 7.1.3 LED status indicator The status LED is used for the quick visualization of the sensor status.
Liquiline Mobile CML18 Operation options A0029747 9 SmartBlue App icon Bluetooth must be enabled on both devices. Enable Bluetooth → 25 CML18_M41073505G11 A0041291 10 SmartBlue App Livelist The Livelist displays all of the devices that are within range. ‣ Tap the device to select it. To be able to use the device with SmartBlue, the Bluetooth connection must be confirmed by entering a user name and password. 1. User name >> admin 2.
Operation options Liquiline Mobile CML18 The current measured values are displayed in the Home view. The device information (device tag, serial number, firmware version, order code) is also displayed. 1 CML18_M41073505G11 M41073505G11 2 M41073505G11 01.01.
Liquiline Mobile CML18 Operation options CML18_M41073505G11 1 2 3 4 A0041294 12 1 2 3 4 Main menus of the SmartBlue App Guidance Diagnostics Application System Menu Function Guidance Contains functions involving a self-contained sequence of activities, e.g. for calibration (= "Wizard", guided operation). Diagnostics Contains information regarding operation, diagnostics and troubleshooting, as well as the configuration of the diagnostic behavior.
Commissioning 8 Commissioning 8.1 Preparatory steps Liquiline Mobile CML18 Charge the device.→ 14 Connect the sensor.→ 14 8.2 Function check LWARNING Connection errors The safety of people and of the measuring point is at risk! ‣ Put the device into operation only if you can answer yes to all the following questions.
Liquiline Mobile CML18 8.4 Commissioning Setting the display language 1. Navigate to: Display language Main menu >> System >> Display language 2. Press to scroll through the predefined values. Description of setting Configuration options Change the language of the operating menu. • Deutsch • English 8.5 Configuring the measuring device 8.5.1 Bluetooth connection 1. Navigate to: Bluetooth Main menu >> System >> Bluetooth 2. Press to scroll through the predefined values.
Commissioning 2. Press Liquiline Mobile CML18 to scroll through the predefined values. The following energy settings are available: • Power save w. charger (power saving with charger) • Power save w/o charger (power saving without charger) • Power-off w. charger (power off with charger) • Power-off w/o charger (power off without charger) The power save mode is activated after the set time if there is no user interaction.
Liquiline Mobile CML18 Commissioning Power-off w/o charger (power off without charger) Function description Configuration options Set the time until the device switches off automatically if it is running on the battery. • • • • • • • 8.6.3 1 min 5 min 15 min 30 min 1h 2h Never Signal sounds 1. Navigate to: Signal sounds Main menu >> System >> Signal sounds 2. Press to scroll through the predefined values.
Commissioning 8.6.7 Liquiline Mobile CML18 Data logger Defining the log interval The log interval can only be changed if the data logger is deactivated. 1. Navigate to: Log interval Main menu >> Application >> Data logger >> Log interval 2. Press to scroll through the predefined values. Description of setting Configuration options Set the time until the next measured value is saved automatically.
Liquiline Mobile CML18 9 Operation 9.1 Reading measured values Operation Measuring screens are shown on the display when a sensor is connected. For each sensor, there are 3 measuring screens with different measured variables→ 20. To scroll through the measuring screens: ‣ Press . After the last measuring screen, the display returns to the first measuring screen. 9.1.1 Saving the individual measured value (Grab Sample) ‣ In the measurement window, press .
Operation Liquiline Mobile CML18 3. A0042257 3. Select Guidance. 4. A0042258 4. 30 Select Data transfer.
Liquiline Mobile CML18 Operation 5. A0042261 5. Select Next to continue. 6. 8. 10.
Operation Liquiline Mobile CML18 6. Select Data source. Select Grab sample logger for saved individual measured values. Select Cont. data logger for data records of the data logger. 7. Press Ok to confirm. Select to discard changes and close the selection menu. 8. Select File name. Click the text line to enter an individual name for the generated data package. 9. Press Ok to confirm. Select to discard changes and close the selection menu. 10. Select Next to continue.
Liquiline Mobile CML18 Operation 12. 13. A0042265 12. Use Share ZIP file... to save the exported data records locally or to send them. 13. Select Finish to complete the export. 9.1.5 Deleting saved measured values ‣ Navigate to: Erase data Main menu >> Application >> Data logger >> Erase data The data are divided into 2 categories: • Erase continuous logs Selects all data logger entries for deletion. • Erase grab values Selects all grab values (individual measured values) for deletion.
Operation 9.1.6 Liquiline Mobile CML18 Switching off the device 1. Navigate to: Power-off Main menu >> Power-off 2. Press 34 to switch off the device.
Liquiline Mobile CML18 10 Diagnostics and troubleshooting 10.1 General troubleshooting 10.1.1 Service interface Diagnostics and troubleshooting The service interface may only be used by the manufacturer's service staff. The device can be connected to a computer for service purposes via the M12 connection. This enables data transmission and wired device charging. 10.2 Diagnostics information via the onsite display 10.2.1 Opening the diagnostics list 1.
Maintenance Liquiline Mobile CML18 11 Maintenance 11.1 Maintenance tasks 11.1.1 Cleaning ‣ Only clean with a damp cloth and commercially available cleaning agents. The device is resistant to: • Ethanol (for a short time) • Soap-based household cleaning agents • Dishwashing detergent NOTICE Cleaning agents not permitted Damage to the housing surface or housing seal ‣ Do not use concentrated mineral acids or alkaline solutions for cleaning.
Liquiline Mobile CML18 12 Repair 12.1 Return Repair The product must be returned if repairs or a factory calibration are required, or if the wrong product was ordered or delivered. As an ISO-certified company and also due to legal regulations, Endress+Hauser is obliged to follow certain procedures when handling any returned products that have been in contact with medium. To ensure the swift, safe and professional return of the device: ‣ Refer to the website www.endress.
Technical data 14 Technical data 14.1 Input 14.1.1 Input power Liquiline Mobile CML18 Wireless charging 5W M12 connection 5 V; 0.6 A 14.1.2 • • • • • • Measured variables pH ORP pH/ORP Oxygen Conductivity Temperature 14.1.3 Measuring range → Documentation of the connected sensor 14.1.4 Type of input Memosens connection for sensors with Memosens technology M12 connection for digital measuring cables for sensors with Memosens technology Memosens sensors CLS50D and CLS54D 14.2 Output 14.
Liquiline Mobile CML18 14.3.4 Technical data Overvoltage protection IEC 61 000-4-4 with 0.6 kV IEC 61 000-4-5 with 2.0 kV 14.3.5 Sensor connection Sensors with Memosens technology 14.3.6 Cable specification Digital measuring cable CYK10-Axx2+x Digital measuring cable CYK20-AAxxC1 14.4 Environment 14.4.
Technical data 14.5 Mechanical construction 14.5.1 Materials Liquiline Mobile CML18 Components Material Housing PBT Display window, light guide PMMA Buttons, cap TPE M12 connection CuZn, nickel-plated 14.5.2 Impact loads The product is designed for mechanical impact loads of 1 J (IK06) as per the requirements of EN61010-1. 14.5.3 Weight Liquiline Mobile CML18 40 155 g (5.
Liquiline Mobile CML18 Index Index A Accessories . . . . . . . . . . . . . . . . . . . . . 37 Ambient temperature . . . . . . . . . . . . . . . 39 Approvals . . . . . . . . . . . . . . . . . . . . . . 11 B Battery life . . . . . . . . . . . . . . . . . . . . . 38 Bluetooth connection . . . . . . . . . . . . . . . 25 C Cable specification . . . . . . . . . . . . . . . . . Certificates . . . . . . . . . . . . . . . . . . . . . Charging the device . . . . . . . . . . . . . . . . Cleaning . . . . . . . . .
Index Liquiline Mobile CML18 Saving the measured value Data logger . . . . . . . . . . . . . . . . . . 29 Grab Sample . . . . . . . . . . . . . . . . . 29 Scope of delivery . . . . . . . . . . . . . . . . . . 11 Sensor Connection . . . . . . . . . . . . . . . . . . 39 Sensor connection . . . . . . . . . . . . . . . . . 14 Settings . . . . . . . . . . . . . . . . . . . . . . . 25 Sounds Signal sounds . . . . . . . . . . . . . . . . . 27 State-of-the-art technology . . . . . . . . . . . .