Instruction Manual

www.Fisher.com
Supplement to HART
R
Communicating Fisher
R
FIELDVUE
R
Instrument Instruction Manuals
Audio Monitor for HART
R
Communications
The frequencies used for HART
communications
are in the audible range—you can hear them! The
1200 and 2200 Hz waveforms sound similar to the
sounds you hear from a FAX machine’s phone
modem. This document outlines how to assemble a
device to monitor these audible waveforms and a
few suggestions on how to use it for troubleshooting.
The audio monitor is shown in figure 1. The cable
can also be used as an input to a cassette recorder
if recordings of the HART messages are desired.
HART
R
Protocol—A Little
Background
The HART communication protocol has a “question-
answer” format; the HART master asks a question
and the HART slave answers back. The slave will
not communicate unless a master first asks a
question. With the audio monitor powered up and
attached to a FIELDVUE
instrument which is
communicating with either a 375 Field
Communicator or a PC running AMS
ValveLink
Software or AMS Suite: Intelligent Device Manager,
a series of “squawks” are heard. Often the pattern
heard is “dot dot dash ...dot dot dash...dot dot
dash...” that occur at intervals of about once a
second with a slight pause in between. The first “dot”
is the master asking the question and the remaining
“dot dash” is the instrument’s response.
It is possible to have two HART masters
communicating simultaneously with one HART
slave. This is done by having one master be
“primary” and the second master be “secondary”.
The HART protocol provides time for each master to
BANANA PLUGS
MINI AMPLIFIER/SPEAKER
INPUT JACK
0.1 μF DISC
CERAMIC CAPACITOR
470K OHM RESISTOR
WIRE TIE
PC BOARD
1/8-INCH
PHONE PLUG
W6943/IL
Figure 1. Audio Monitor for HART
Communications
alternate making “question-answer” exchanges with
the slave. The pattern heard would then be “dot dot
dash dot dot dash dot dot dash dot dot dash” with no
pauses in between exchanges. Assuming the
communications are within HART waveform
specifications, no communication errors should
normally occur.
Some analog output circuits used to generate the 4
to 20 milliamp control signal are adversely affected
by the HART communications waveforms. The usual
result is that the impedance of the output circuit is
substantially reduced during communications. Good
HART communications depend on the master device
creating the waveform in a circuit where the load
from the FIELDVUE instrument and the current
source are roughly equal such that the signal is
uniform throughout the loop. If the output circuit’s
impedance drops, the HART waveform is unevenly
split (as in a voltage divider) and the signal received
by the FIELDVUE instrument is reduced. The HART
filter blocks the HART waveform from being detected
by the analog output circuitry and prevents the
source impedance from dropping.
Another thing to understand is the 375 Field
Communicator has a broadcasting mode called the
Instruction Manual Supplement
D103265X012
February 2009
FIELDVUE
Instruments

Summary of content (4 pages)