MEGAM.O.L.E.® 20 Hardware & Software Help Guide MEGAM.O.L.E.® 20 Hardware & Software Help Guide A47-6342-00 Rev 2.
1.0 Introduction ............................................................................................................. 3 1.1 Copyright................................................................................................................... 3 1.2 Declaration of Conformity.......................................................................................... 5 1.3 How to use this Users Help Guide............................................................................. 6 1.
5.0 Software ................................................................................................................. 53 5.1 Computer Hardware Requirements......................................................................... 53 5.2 Installation ............................................................................................................... 54 5.2.1 Starting the Software ..................................................................................... 56 5.2.
1.0 Introduction 1.1 Copyright ECD World Headquarters China Electronic Controls Design, Inc. 4287B S.E. International Way Milwaukie, Oregon 97222-8825 U.S.A. Telephone: +(1) 800.323.4548 +(1) 503.659.6100 Electronic Technology Shanghai China LTD. Email: ecd.china@ecd.com Internet: http://www.ecdchina.co m FAX: +(1) 503.659.4422 Technical Support: +(1) 800.323.4548 Email: ecd@ecd.com Internet: http://www.ecd.com Asia ECD Asia/Pacific Covent Garden Post Office P.O.
©2008 ECD. All Rights Reserved. Printed in USA. US Products of ECD are covered by foreign and US Patents and Patents Pending. Information in this publication supersedes all previously published information. This Publication may not be translated and/or reproduced or stored in data retrieval system, or transmitted in any form or by any means without the express written permission of ECD. Specification and price change privileges reserved. The trapezoidal ECD logo®, and M.O.L.E.
1.2 Declaration of Conformity Electronic Controls Design, Inc. 4287-B S.E. International Way Milwaukie, Oregon U.S.A. 97222-8825 (503) 659-6100 / (800) 323-4548 FAX: (503) 659-4422 Declaration of Conformity Product: MegaM.O.L.E.® Profiler Part number: E47-6342-45 MegaM.O.L.E.® 20 I/O Module E47-6342-50 MegaM.O.L.E.® Power Pack Battery E47-6342-30 The above product is in conformity with the following standards or other normative documents: • EN 301 489-1 V1.6.1 (2005-09) • EN 300 328 V1.7.
1.3 How to use this Users Help Guide This Users Help Guide explains how to use ECD Thermal Profiling instruments and software. This Users Help Guide is written for users of varied experience. If a section covers information you already know, feel free to skip to the next section. You do not need to be a computer expert to use this Users Help Guide or the software. • The Users Help Guide assumes you are familiar with Microsoft® Windows® .
1.4 Terms Used In 1986 ECD introduced our original Thermal Profiling instrument called the M.O.L.E.® (Multichannel Occurrent Logger Evaluator). Over the years ECD has produced several models of the M.O.L.E.® for use in a wide variety of applications. In this Users Help Guide, we may refer to all of our Thermal Profiling instruments as the M.O.L.E.® Profiler. The M.O.L.E.® Profiler is a registered trademark of ECD.
1.5 Fonts Used This manual uses a specific font to indicate terms or words that can be found directly on the display of the computer. For example: On the File menu, click Preferences to configure the software global settings. This font indicates the words File and Preferences are actually found on the computer display. MEGAM.O.L.E.
2. Safety 2.1 Operators Safety Information The safety information in this section is for the benefit of operating personnel. Warnings and Cautions will also be found throughout the manual where they apply. Hardware changes or modifications to the M.O.L.E. Profiler or components are not expressly approved by ECD and could void the warranty. The warranty will not cover damage caused by neglect or abuse of any ECD products.
• NEVER make any changes or modifications to the Wireless RF components. Such changes are not expressly approved by ECD and could void the user's authority to operate the equipment. MEGAM.O.L.E.® Profiler is designed to be operated at a distance of greater than 20 centimeters from the head or torso of the user. When removing M.O.L.E. Profiler from any temperature environment, be careful of extreme temperatures and use protective gloves. MEGAM.O.L.E.
2.2 Battery Warnings Warnings: • Charge “Rechargeable” Power Pack batteries using only the ECD approved charger. • Always dispose of used batteries promptly and properly. • Keep all batteries away from children. • The batteries may explode if mistreated. • The batteries contain electrolytes. Replace the M.O.L.E. Profiler batteries with same type only: • MEGAM.O.L.E.: Rechargeable Power Pack battery, ECD Part No.
2.3 Equipment Maintenance Maintain your equipment to prevent future problems and unwanted costs. Keep your M.O.L.E. Profiler at peak performance by taking care of the system components and keeping it calibrated. Typical maintenance should include the following: • Wipe the exterior of the instrument enclosure or components with a cloth dampened with water or IPA alcohol. IPA alcohol is the only type of solvent that is acceptable to clean Thermal Profiling instruments or any of the system components.
2.4 Product Symbols The following symbols may be present on the M.O.L.E. Profiler and/or system components: ON/OFF Start Data Run Recording Battery Level Operating Temperature Universal Serial Port (USB) Battery Charging Indicator Radio Frequency (RF) Activity Radio Frequency (RF) Jack Waste Electrical and Electronic Equipment (WEEE) ♦13♦ MEGAM.O.L.E.
Direct Current (DC) Power Alternating Current (AC) Power RoHS / RoHS Compliant CAUTION: Whenever this internationally recognized symbol is used on the product, additional information concerning that particular feature or function appears in the manual. European Conformity FCC Federal Communications Commission (FCC) Power transformer is protected throughout by double or reinforced insulation. Indoor use only.
3.0 Equipment 3.1 MEGAM.O.L.E.® The MEGAM.O.L.E.® Profiler is the most compact and durable thermal profiling instrument ECD offers. It is powered by a removable Power Pack battery and includes a "Smart" I/O Pod that interfaces with the test assembly. Features/functions: • I/O Pod:This pod connects to the MEGAM.O.L.E. Profiler and has four input connections for attaching up to a maximum of twenty thermocouple sensors. • • Main Module: This is the main controller which stores collected data runs.
3.1.1 I/O Pod This I/O Pod is considered a "Smart" pod as it is capable of storing all of the configured machine, assembly and process documentation. It also locks the data run (How did we do it) documentation information and can remain with the test assembly for quick setup time. I/O Pod Features/functions: • Thermocouple/Inputs: This is where Type “K” Thermocouples sensors are connected.
3.1.2 Main Module The Main module is the controller unit of the MEGAM.O.L.E. Profiler. This module stores the recorded data runs that can be downloaded to a computer through the Data Port located on the Power Pack battery. To configure and download data from the Main Module, the Power Pack battery must be connected to the unit. Refer to topic Power Pack Battery for more information. ♦17♦ MEGAM.O.L.E.
Features/functions: • Buttons: ON/OFF, Record & OK. Button: ON/OFF Record OK Action: Turns Main module "ON/OFF". Starts/Stops Main module recording data. Invokes "OK" process where the last recorded profile is compared to preconfigured criteria resulting in a "Pass" or "Fail" mode. • Activity indicators: These are LED's that indicate what state the M.O.L.E. Profiler is in. Refer to the illustration and table below.
• Interface Connectors: There are two interface connectors in the Main module. The front connector interfaces with the I/O Pod and is where the converted digital data from the I/O pod is transferred to the Main module. The rear connector interfaces with the Power Pack battery and power and signal data is transferred. • Alignment Sockets: These are where the mating I/O Pod and Power Pack battery alignment pins are inserted. • RF Antenna Connector: This is where the Optional RF antenna is connected.
3.1.3 Power Pack Battery The MEGAM.O.L.E. Profiler is powered by a Power Pack battery and it is important to make sure it is charged and operating properly before each data run. When connected to a computer, the software displays the voltage of the Power Pack battery on the Status bar. Refer to topic Status Bar for more information. When the Power Pack needs to be charged use the transformer charger or USB computer interface cable supplied with the kit.
Features/functions: • Data/Charging Port: This port is located on the backside of the battery. This is used to transfer data through the USB computer Interface cable to the computer. It is also used to charge the battery using the supplied transformer. When the Power Pack battery is connected to a powered computer USB port, the power from the USB port will charge the Power Pack battery. • Activity indicator: This LED illuminates when the Power Pack battery is charging.
3.1.4 Sensors Up to 20 sensors can be attached to the MEGAM.O.L.E. Profiler per data run. Four sets of five 0.005” K-Type thermocouple sensors are included in the Kit. Bottom View Top View ECD also offers a wide selection of sensors and adaptors to meet the needs of many applications. For more information about optional sensors for specific applications, refer to topic How to Get Additional Help to contact ECD. Plug the sensor connector into the desired MEGAM.O.L.E. Profiler channel block.
3.1.5 USB Computer Interface Cable Two USB cables are included in the hardware package. These cables connect the MEGAM.O.L.E. Profiler to a computer USB port and power pack battery charger. ♦23♦ MEGAM.O.L.E.
3.1.6 Barrier & Yellow Jacket The Uni-Barrier with Yellow Jacket is the standard barrier in all our Lead-Free profiling kits, including the wireless RF kit. A thermal barrier is the protection the M.O.L.E. needs to ensure longevity through harsh environments. With piano-hinged stainless steel construction, sheathed in cool-to the touch high-temperature fabric this thermal barrier is ideal for lead-free processes. This barrier is rated to protect the MEGAM.O.L.E. for 10 minutes at 200C.
3.1.7 MEGAM.O.L.E.® to Standard T/C Adaptor This optional thermocouple adaptor is attached to the thermal barrier and converts the MEGAM.O.L.E.® profiler nano connectors to standard connectors. This is especially useful when a user has an experimental assembly pre-wired with standard thermocouples saving considerable setup time. ♦25♦ MEGAM.O.L.E.
3.1.8 Wireless RF Option The wireless RF option for the MEGAM.O.L.E.® profiler produces real time profiles fast for immediate data collection and transmission to a computer. The Transmitter is an built-in component that when used with the antenna sends data to the RF transceiver which connects to a USB port on the computer. Wireless RF Transceiver: Front Side Back Side Antenna: MEGAM.O.L.E.
4.0 Basics 4.1 Setup The Setup topic offers a brief description of the system hardware configuration. 4.1.1 Charging the Power Pack Battery Because the M.O.L.E. profiler is powered by a rechargeable Power Pack battery, it is important that it is charged and operating properly prior to performing every experiment. A spare Power Pack battery may be ordered so the one battery is charging while the other one is being used.
The Power Pack battery can be charged continuously whenever the M.O.L.E. profiler is not being used, however, if the M.O.L.E. profiler is going to sit idle for five days or more, you may want to remove it from the charger. MEGAM.O.L.E.
4.1.2 Communications Setup To connect the M.O.L.E. Profiler: The software must be installed prior to communications setup. (Refer to Software Installation for more information). 1) Insert the USB computer interface cable into a computer USB port and the other end into the Power Pack Battery Data Port. ♦29♦ MEGAM.O.L.E.
The first time a M.O.L.E. Profiler is connected to a computer two drivers will be installed. One is a device driver for the M.O.L.E. Profiler and the other is for USB communication. 2) During installation of the device driver, when prompted to select the location of the device driver, select "Install the software automatically". Follow the remaining wizard instructions closely.
4) After M.O.L.E. Profiler is connected to a computer, the software must be configured so they can communicate. Start the software program by either double-clicking the MAP software icon or selecting it from the ECD program submenu. 5) On the M.O.L.E. menu, click the Select Instrument command. 6) Select the desired instrument from the list box. If there are none listed, click the Scan for Instruments command button to detect all available instruments.
4.1.3 Wireless RF Option Setup To connect the wireless RF option: Prior to wireless RF communication, the M.O.L.E. Profiler must first be connected to a computer through the USB port. This allows the software to register the M.O.L.E. Profiler ensuring that other local computers cannot read RF data from this profiler. (Refer to Communications Setup for more information) 1) Insert the Wirless RF Transceiver into an available USB port and the antenna to the M.O.L.E. Profiler.
The M.O.L.E. Profiler is designed to work with with included antenna ECD P/N: E47-6342-11 only. Never make any changes or modifications to the Wireless RF components. Such changes are not expressly approved by ECD and could void the user's authority to operate the equipment. 2) After the Transceiver is connected to a computer, it needs to be detected by the software to properly communicate. ♦33♦ MEGAM.O.L.E.
3) Start the software program by either double-clicking the MAP software icon or selecting it from the ECD program sub-menu. 4) On the M.O.L.E. menu, click the Select Instrument command. 5) Select the desired instrument from the list box. If there are none listed, click the Scan for Instruments command button to detect all available instruments.
7) Click the OK command button to accept or Cancel to quit the command. The system is now ready for Wireless RF communication. ♦35♦ MEGAM.O.L.E.
4.2 Operation This section guides the user through a typical collection process. Portions of this section will require referral to some software sections of the Users Help Guide for additional information. The M.O.L.E. Profiler is dependent on the software to control how it collects and interprets data. Several kinds of data runs may need to be performed to achieve desired information, or the same data run may be performed repeatedly over time to monitor one process.
4) On the File menu, click New/Start. A message box appears with the three workflow wizard options. 6) On the Start dialog box, click the Fresh Start command button and the workflow wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps. Current ♦37♦ Completed Remaining MEGAM.O.L.E.
7) Select the desired instrument from the list box to make active. If there are none listed, click the Scan for Instruments command button to detect all available instruments. MAP software allows multiple instruments to be connected to a computer at one time. Selecting the Scan for Instruments command button will detect all instruments and display them in the list. If no instrument is detected, the default Demonstration MEGAM.O.L.E.® profiler is displayed. 8) Click the Next command button. MEGAM.O.L.E.
9) Select a machine from the Machine drop down list. If the desired machine does not appear in the list click the New command button to create a new machine. Refer to topic Software>Menus>Machine>Create new Machine for more information. ♦39♦ MEGAM.O.L.E.
10) Set the machine recipe settings such as Conveyor Speed, Zone Temperatures and Temperature units. The software includes features to save and load machine recipe setting files. These files are helpful so the user can quickly recall machine information and ensure it is always the same. 11) Click the Next command button. If no value is specified for the conveyor speed or the default oven temperature values, the software will remind the user to set them. MEGAM.O.L.E.
12) Enter the assembly information such as part number, board size, sensor locations and a product image. If a product image has been selected, clicking the Enlarge command button displays the Set Sensor Locations dialog box where the user can specify the locations of each sensor. To move sensor locations, drag the sensor markers to the approximate location where the sensors are attached. ♦41♦ MEGAM.O.L.E.
13) When finished, click the OK command button to accept or Cancel to return without making any changes. 14) Click the Next command button. 15) Attach sensors to a test assembly. When soldering a T/C sensor to a component with high temperature solder, use Kester SN10 (or equivalent) for 183°C eutetic solder or Kester SN5 (or equivalent) for lead free soldering. MEGAM.O.L.E.
16) Unwind the sensor leads and attach the connectors to the M.O.L.E. Profiler. Make sure to connect them to the same channel(s) that were configured in the wizard workflow. For example, if configuring the M.O.L.E. Profiler to have Channels 1 through 5 active, be sure the sensors are attached to the same five channels. Kapton® tape, aluminum tape and Temprobe™ can also be used to hold T/C wire to the test assembly. ♦43♦ MEGAM.O.L.E.
17) Select your process specification. The user can select a Paste from the database or a pre-defined Target 10 file. Then select a Profile Type (Ramp-Soak-Spike or Ramp-to-Spike). If your Paste does not appear in the database list click the New command button to create a new paste. Refer to topic Software>Menus>Process>Create new Paste for more information. Once a paste is selected the specifications are displayed on the graph.
18) Click the Next command button ♦45♦ MEGAM.O.L.E.
19) Set Recording Parameters such as the instrument name, recording interval, start parameters and stop parameters. This step is where the user can also turn a sensor channel ON or OFF, set the sensor location description and sensor type. Refer to topic Software>Menus>M.O.L.E.>Set Recording Parameters for detailed information for each setting. 20) Click the Next command button. MEGAM.O.L.E.
21) Click the Next command button to send the recording parameters to the instrument. ♦47♦ MEGAM.O.L.E.
22) Click the Next command button. MEGAM.O.L.E.
Step 2: Perform Data Run 1) Place the M.O.L.E. Profiler in the appropriate thermal barrier making sure the Thermocouple and/or Sensor wires are not damaged. Never permit the M.O.L.E. Profiler to exceed the absolute maximum warranteed internal temperature, as permanent damage may result. The warranty will not cover damage caused by exceeding the maximum specified internal temperature. 2) After the oven stabilizes, turn the M.O.L.E. Profiler on and press the record button.
It is highly recommended that protective gloves are used when retrieving the thermal barrier from the oven and when opening the thermal barrier. 4) As the test assembly and M.O.L.E. Profiler emerge from the machine, carry the test assembly with sensors attached and the M.O.L.E. Profiler in the Thermal barrier to a table or flat surface. If a sensor is removed before the M.O.L.E. Profiler has stopped collecting data, the data for that channel might become distorted.
Step 3: Read Data Run 1) Restore the software and click the Finish command button to read the data run from the M.O.L.E. Profiler. 2) When the data run has been downloaded, the software will suggest a default file name or prompt the user to specify a new file name. ♦51♦ MEGAM.O.L.E.
I 3) When finished, click the Save command button. 4) The software then prompts the user if they want to enter Prediction mode. Entering prediction mode enables the user to change a zone temperature values or the conveyor speed and predict the outcome of that change on the data run profile. Refer to topic Software>Menus>Tools>Prediction for more information. 5) Click the Predict command button to enter Prediction mode or Done to complete the workflow wizard .
5.0 Software 5.
5.2 Installation Autorun Install: 1) Insert the CD in the drive and the MAP software autorun menu appears. If the autorun menu does not appear, proceed to Manual Install for instructions. 2) Select the Install Software command button to start the installation. 3) Closely follow the setup instructions provided with the software. The user must have administrator permissions for the computer to install this software. MEGAM.O.L.E.
Manual Install: 1) Insert the CD in the drive. 2) Select Run from the Start menu. 3) Select the Browse command button and navigate to the software CD. Doubleclick the installation (.EXE) file. 4) Select the OK command button to start the installation. 5) Closely follow the setup instructions provided with the software. ♦55♦ MEGAM.O.L.E.
5.2.1 Starting the Software Prior to starting, click the README icon from the MEGAM.O.L.E.® program sub-menu to read the latest release notes After the software is installed, start the software program by double-clicking the MEGAM.O.L.E.® icon from the desktop. Once the software installation is complete, it is important to start the software and configure the software to communicate with the M.O.L.E. Profiler. Refer to topic Basics>Setup for more information. MEGAM.O.L.E.
5.2.2 Software Authorization The MEGAM.O.L.E.® software is a fully functional 31-day trial version. Once the trial period is over, you may purchase the software from an ECD Sales person. For Sales contact information refer to topic Service>How to Get Additional Help. To authorize: 1) Start the software program by by double-clicking the MEGAM.O.L.E.® icon from the desktop and the Software Authorization dialog box appears.
5.3 Features This section presents an overview of the software window. When the software is started, it automatically defaults to the previously selected working directory. Refer to topic Software>Menus>File>Open Working Directory for more information. After installation, the software is started a with the a sample working directory with files will be selected for users to familiarize themselves with the program.
Page Tab Views: These tabs are used to gain access to each Page Tab View. Split Bar: This bar slides the Horizontal Scroll bar to the left or right so all or part of the page tabs can be viewed. Status Bar: This bar on the bottom of the display, shows the status of the M.O.L.E. Profiler Power Pack battery, Internal operating temperature, connected COM port, available Help information, mouse pointer X-Y position, current date and time. Scroll Bars: These bars scroll the display horizontally and vertically.
5.3.1 Page Tabs There are five standard Page Tabs. These tabs are located on the bottom left of the display. To display the contents of a Tab, use the mouse pointer. The tab will then become highlighted, and the worksheet will now be visible. The keyboard does not allow access to the Tabs. The only way to select a Tab is by using the mouse pointer. MEGAM.O.L.E.
5.3.2 Split Bar The Split bar lets the user slide the resize the Horizontal scroll bar to the left or right, so all of the Tabs can be viewed. This feature is always located on the left edge of the Horizontal scroll bar. Hidden Tabs may also be displayed by using the Tab Scroll Arrows located on the left side of the Tabs. ♦61♦ MEGAM.O.L.E.
5.3.3 Status Bar This bar is located on the bottom of the software display. It shows the status of the M.O.L.E. Profiler Power Pack battery, Internal operating temperature, connected COM port, available Help information, mouse pointer X-Y position, current date and time. Using the Preferences command, the user can decide which M.O.L.E. Profiler status bar items will be displayed. Refer to topic Software>Menus>File>Preferences>M.O.L.E. for more information.
M.O.L.E. Profiler Connection: This indicator displays a symbol for the computer communication connection type such as Serial, USB or wireless RF. Instrument User Name: This indicator displays the user configured name for the M.O.L.E. Profiler. Help Information: This indicator displays the action that button performs when the mouse pointer is placed over a Toolbar button or Menu command.
5.3.4 Scroll Bars All Tab Views have both Horizontal and Vertical screen scroll bars so the non-visible areas of the display can be scrolled into view. The Horizontal scroll bar is located in the lower right corner and can be scrolled left or right by pressing the left or right arrows located on each end of the scroll bar. The Vertical scroll bar located on the right side of the screen has the same features as the Horizontal scroll bar except it scrolls the display up and down.
5.4 Page Tab Descriptions The Page Tabs offer four different ways to view the data recorded by the M.O.L.E. Profiler. Each page tab focuses on specific data and offers different features and functions. Along with the Welcome page tab they formulate a complete Profiling report. 5.4.1 Welcome Page Tab Welcome is the introductory tab. It contains an introductory illustration and text box for entering a company or profiling report name.
5.4.1.1 Menus & Toolbar • Menus: File, Edit, Wizards, M.O.L.E.®, Machine-Oven, Assembly-Board, Process-Paste, Profile, Tools and Help. • Toolbar Buttons: Start, Open Working Directory, Save, Print, and Help. MEGAM.O.L.E.
5.4.1.2 Company/Report Name The text box located on the bottom half of the Welcome page tab allows the user enter a company or report name. To enter a name: 1) Using the mouse pointer, click in the text box. 2) Type a desired name and then hit the [enter] key to accept or [esc] to cancel. ♦67♦ MEGAM.O.L.E.
5.4.2 Summary Page Tab The Summary Page Tab is where individual data runs are viewed in summarized page format. Summary Tab features: Menus and Toolbar Summary Template MEGAM.O.L.E.
5.4.2.1 Menus & Toolbar • Menus: File, Edit, Wizards, M.O.L.E.®, Machine-Oven, Assembly-Board, Process-Paste, Profile, Tools and Help. • Toolbar Buttons: Start, Open Working Directory, Save, Print, Help, First (data run of the data set), Back (to previous data run), Forward (to the next data run), and Last (data run of the data set). ♦69♦ MEGAM.O.L.E.
5.4.2.2 Summary Template The Summary Page Tab is built using a template file (*.TSU) overlaid on a cell grid. The Summary template is automatically loaded every time the software is started. This template file is specified on the Summary page tab of the Preferences dialog box. Refer to topic Software>Menus>File>Preferences>Summary for more information. To display Template commands: 1) Move the mouse pointer over a template cell. 2) Using the right mouse button, right-click and a shortcut menu appears.
3) You can now select from the following template commands. ♦71♦ MEGAM.O.L.E.
5.4.2.2.1 Add & Edit Content To add or edit template content, the software includes a wizard to guide the user through the related content options. The template allows five different calculation categories to be displayed. Add & Edit Content wizards: Text Temperature Value (Y) Time Value (X) Slope (dX/dY) Temperature (Y) Delta Special Values This wizard contains all the related steps to add or edit content to the template.
5.4.2.2.1.1 Text To add or edit Text content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps. Current Completed Remaining 3) Click Text. 4) Select the Next command button. ♦73♦ MEGAM.O.L.E.
5) Select desired text formatting options. 6) Select the Next command button. MEGAM.O.L.E.
7) Select desired cell border options. 8) Select the Finish command button to complete the wizard and display the new calculation data in the selected template cell. ♦75♦ MEGAM.O.L.E.
5.4.2.2.1.2 Temperature Value (Y) To add or edit Y-Axis Values content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps.
5) Select a Temperature (Y) Axis Value. If Temperature at Time Reference calculation is selected, the software requires the user to select an established Time (X) Reference line. If one is not established the software automatically creates one on the Profile Page Tab Data Graph. 6) Select the Next command button. ♦77♦ MEGAM.O.L.E.
7) Select the calculation constraints. These options are the specified area on the Time (X) Axis where the values are to be extracted from. If the Within Magnified Window constraint is selected and the Magnify tool is used to zoom in on a portion of the Data Graph, the Data Table displays the statistics for those values within the magnified window. 8) Select the Next command button. MEGAM.O.L.E.
9) Select desired text formatting options. 10) Select the Next command button. ♦79♦ MEGAM.O.L.E.
11) Select desired cell border options. 12) Select the Next command button. MEGAM.O.L.E.
13) Select Specification Limits and Units. If these values are violated colored bars will appear in the formatted template cell. Refer to topic Software>Page Tabs>Summary>Template>Specification Limit Indicators for more information. 14) Select the Finish command button to complete the wizard and display the new calculation data in the selected template cell. ♦81♦ MEGAM.O.L.E.
5.4.2.2.1.3. Time Value (X) To add or edit X-Axis Values content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps. Current Completed Remaining 3) Click Time Value (X) and which channel to derive the data from.
5) Select a Time (X) Axis Value. If any Temperature Reference (Y) calculation is selected, the software requires a Temperature (Y) Reference Line to be established. Refer to topic Add Temperature (Y) Reference Lines. 6) Select the Next command button. ♦83♦ MEGAM.O.L.E.
7) Select the calculation constraints. These options are the specified area on the Time (X) Axis where the values are to be extracted from. If the Within Magnified Window constraint is selected and the Magnify tool is used to zoom in on a portion of the Data Graph, the Data Table displays the statistics for those values within the magnified window. 8) Select the Next command button. MEGAM.O.L.E.
9) Select desired text formatting options. 10) Select the Next command button. ♦85♦ MEGAM.O.L.E.
11) Select desired cell border options. 12) Select the Next command button. MEGAM.O.L.E.
13) Select Specification Limits and Units. If these values are violated colored bars will appear in the formatted template cell. Refer to topic Software>Page Tabs>Summary>Template>Specification Limit Indicators for more information. 14) Select the Finish command button to complete the wizard and display the new calculation data in the selected template cell. ♦87♦ MEGAM.O.L.E.
5.4.2.2.1.4. Slope (dX/dY) To add or edit Slope Value content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps. Current Completed Remaining 3) Click Slope (dX/dY) and which channel to derive the data from.
5) Select a Slope Value. If Slope Between Time References calculation is selected, the software requires the user to select an established Time (X) Reference line. If one is not established the software automatically creates one on the Profile Page Tab Data Graph. 6) Select the Next command button. ♦89♦ MEGAM.O.L.E.
7) Select the calculation constraints. These options are the specified area on the Time (X) Axis where the values are to be extracted from. 8) Select the Next command button. MEGAM.O.L.E.
9) Select desired text formatting options. 10) Select the Next command button. ♦91♦ MEGAM.O.L.E.
11) Select desired cell border options. 12) Select the Next command button. MEGAM.O.L.E.
13) Select Specification Limits and Units. If these values are violated colored bars will appear in the formatted template cell. Refer to topic Software>Page Tabs>Summary>Template>Specification Limit Indicators for more information. 14) Select the Finish command button to complete the wizard and display the new calculation data in the selected template cell. ♦93♦ MEGAM.O.L.E.
5.4.2.2.1.5. Temperature (Y) Delta To add or edit Temperature (Y) Delta content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps.
5) Select a Y-Axis value delta calculation and which channels to you wish to be included in this calculation. 6) Select the Next command button. ♦95♦ MEGAM.O.L.E.
7) Select desired text formatting options. 8) Select the Next command button. MEGAM.O.L.E.
9) Select desired cell border options. 10) Select the Next command button. ♦97♦ MEGAM.O.L.E.
11) Select Specification Limits and Units. If these values are violated colored bars will appear in the formatted template cell. Refer to topic Software>Page Tabs>Summary>Template>Specification Limit Indicators for more information. 12) Select the Finish command button to complete the wizard and display the new calculation data in the selected template cell. MEGAM.O.L.E.
5.4.2.2.1.6. Special Value To add or edit Special Value content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps. Current Completed Remaining 3) Click Special Value. 4) Select the Next command button.
5) Select a Special Value type. 6) Select the Next command button. MEGAM.O.L.E.
7) Select desired text formatting options. 8) Select the Next command button. ♦101♦ MEGAM.O.L.E.
9) Select desired cell border options. 10) Select the Finish command button to complete the wizard and display the new calculation data in the selected template cell. MEGAM.O.L.E.
5.4.2.2.2. Specification Limit Indicators Parameters displayed on the Summary Page Tab can have both Lower and Upper specifications applied. If a specification limit is violated, the software displays a red or blue indicator on the left edge of the Data Table cell. If a USL has been exceeded, that parameter indicator will appear in red (indicating it is above the specification limit).
5.4.3. Spreadsheet Page Tab The Spreadsheet Page Tab contains data that is collected by the M.O.L.E. profiler put into standard spreadsheet format. Each row in the spreadsheet represents one data run. The columns after the Data Run and User defined Parameter Groups include SPC Flags so the user can select which parameters are to be displayed on the SPC Page Tab.
5.4.3.1. Menus & Toolbar • Menus: File, Edit, Wizards, M.O.L.E.®, Machine-Oven, Assembly-Board, Process-Paste, Profile, Tools and Help. • Toolbar Buttons: Start, Open Working Directory, Save, Print, and Help. ♦105♦ MEGAM.O.L.E.
5.4.3.2. Spreadsheet Template The Spreadsheet Page Tab is built using a template file (*.TSH) overlaid on a cell grid. Columns after the Data Run and User Parameter Groups allow the user to define parameters using the Template commands. To view the template use the horizontal scrollbar to slide the template left. The Spreadsheet template is automatically loaded every time the software is started. This template file is specified on the Spreadsheet Page Tab of the Preferences dialog box.
3) You can now select from the template commands. ♦107♦ MEGAM.O.L.E.
5.4.3.2.1. Add & Edit Content To add or edit template content, the software includes a wizard to guide the user through the related content options. The template allows four different calculation categories to be displayed. Add & Edit Content wizards: Temperature Value (Y) Time Value (X) Slope (dX/dY) Temperature (Y) Delta Special Values This wizard contains all the related steps to add or edit content to the template.
5.4.3.2.1.1. Temperature Value (Y) To add or edit Y-Axis Values content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps.
6) Select the Next command button. 7) Select the calculation constraints. These options are the specified area on the Time (X) Axis where the values are to be extracted from. If the Within Magnified Window constraint is selected and the Magnify tool is used to zoom in on a portion of the Data Graph, the Data Table displays the statistics for those values within the magnified window. MEGAM.O.L.E.
8) Select the Next command button. 9) Select desired text formatting options. ♦111♦ MEGAM.O.L.E.
10) Select the Next command button. 11) Select Specification Limits and Units. If these values are violated colored bars will appear in the formatted template cell. Refer to topic Software>Page Tabs>Spreadsheet>Template>Specification Limit Indicators for more information. MEGAM.O.L.E.
12) Select the Finish command button to complete the wizard and display the new calculation data in the selected template column. ♦113♦ MEGAM.O.L.E.
5.4.3.2.1.2. Time Value (X) To add or edit X-Axis Values content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps. Current Completed Remaining 3) Click Time Value (X) and which channel to derive the data from.
6) Select the Next command button. 7) Select the calculation constraints. These options are the specified area on the Time (X) Axis where the values are to be extracted from. If the Within Magnified Window constraint is selected and the Magnify tool is used to zoom in on a portion of the Data Graph, the Data Table displays the statistics for those values within the magnified window. ♦115♦ MEGAM.O.L.E.
8) Select the Next command button. 9) Select desired text formatting options. MEGAM.O.L.E.
10) Select the Next command button. 11) Select Specification Limits and Units. If these values are violated colored bars will appear in the formatted template cell. Refer to topic Software>Page Tabs>Spreadsheet>Template>Specification Limit Indicators for more information. ♦117♦ MEGAM.O.L.E.
12) Select the Finish command button to complete the wizard and display the new calculation data in the selected template column. MEGAM.O.L.E.
5.4.3.2.1.3. Slope (dX/dY) To add or edit Slope Value content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps. Current Completed Remaining 3) Click Slope (dX/dY) and which channel to derive the data from.
6) Select the Next command button. 7) Select the calculation constraints. These options are the specified area on the Time (X) Axis where the values are to be extracted from. MEGAM.O.L.E.
8) Select the Next command button. 9) Select desired text formatting options. ♦121♦ MEGAM.O.L.E.
10) Select the Next command button. 11) Select Specification Limits and Units. If these values are violated colored bars will appear in the formatted template cell. Refer to topic Software>Page Tabs>Spreadsheet>Template>Specification Limit Indicators for more information. MEGAM.O.L.E.
12) Select the Finish command button to complete the wizard and display the new calculation data in the selected template column. ♦123♦ MEGAM.O.L.E.
5.4.3.2.1.4. Temperature (Y) Delta To add or edit Temperature (Y) Delta content: 1) Right-click a template cell and a shortcut menu appears. 2) Select Add Content or Edit Content from the shortcut menu and the Add or Change a Calculation wizard appears. When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps.