SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 8 ‘8’ font ring 0.01 (amplitude) The amplitude at 0.01s from QRS loop 0.01 (angle) The angle at 0.01s from QRS loop 0.02 (amplitude) The amplitude at 0.02s from QRS loop 0.02 (angle) The angle at 0.02s from QRS loop 0.03 (amplitude) The amplitude at 0.03s from QRS loop 0.03 (angle) The angle at 0.03s from QRS loop 0.04 (amplitude) The amplitude at 0.04s from QRS loop 0.04 (angle) The angle at 0.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.7.2 Displaying Vector ECG with Frontal Plane and QRS Loop Figure 6-25 Vector ECG - Frontal & QRS loop The percent values of 0.00%, 94.16%, 1.35% and 4.49% in the square represent the area percentages of QRS loop in every quadrant. 20 mm/mV indicates the magnified multiple (gain). The red curve is QRS loop. You can click on the Zoom in button or the Zoom out button to change the gain of the displayed graphics.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Click on Zoom In/ Zoom Out to magnify/minify the picture. Click on Reset to restore the magnified/minified picture to original size. Click on Preview to preview the 3D graph. Click on Print to print the 3D graph. Click on Save to save the graph on the current screen. Click on Return to return to the ECG analysis screen. 6.6.7.4 Previewing Vector ECG Click on the Preview button to open the VCG preview screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-26 VCG Preview Screen (Plane is ALL & loop is ALL) 6.6.8 Analyzing Time Vector ECG Click on the TVCG button on the ECG analysis screen to display the TVCG analysis screen. Time Vector ECG is Vector ECG including time factor. As Figure 6-27 shows, you can observe waves of X lead, Y lead, Z lead, X-Y lead, X-Z lead and Z-Y lead. You can choose the speed and the gain of the displayed waves.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-27 Time Vector ECG Screen Click on the Preview button to display the TVCG preview screen. is the toolbar on the TVCG preview screen. 1) Click on the Zoom In button on the toolbar to magnify the preview page. 2) Click on the Zoom Out button on the toolbar to minify the preview page. 3) Click on the Close button to close the TVCG preview screen and return to the previous screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.9 Analyzing Signal Averaged ECG SAECG is also called Ventricular Late Potential (VLP). Click on SAECG to open the SAECG analysis screen. The SAECG analysis screen includes two tabs: Time domain and Frequency domain. 6.6.9.1 About the Time Domain Window Click on the Time domain tab to open the Time domain window. Figure 6-29 Time Domain Window Every parameter of SAECG is shown in the following table.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Select a filter in the Filter pull-down list . The top of the Time domain window is the window of standard superimposed QRS waveform. The left green line is the starting point of standard unfiltered QRS waveform, and the right green line is the end point. You can respectively drag the two green lines to change QRS duration. The bottom of the Time domain window is the window of standard superimposed QRS waveform after filtering.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.9.3 Previewing Signal Averaged ECG Click on the Preview button to display the SAECG preview screen. is the toolbar on the SAECG preview screen. ♦ Click on Print(P) to print the report. ♦ Click on Next Page/Prev Page to switch to the previous/next preview page. ♦ Click on Two Page to preview two pages on one screen simultaneously. ♦ Click on Zoom In/ Zoom Out to magnify/minify the preview page.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.10 Printing ECG Reports 1. Choose Start > Printers and Faxes, and then right-click on the icon of the printer used, and select Set as Default Printer. Then close the Printers and Faxes window. 2. Click on the Print button on the analysis screen to print an ECG report. 3. Or, click on the Print button on the preview screen to print an ECG report. 6.6.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG If you select Send, the sampled data will be sent by OUTLOOK EXPRESS (Windows XP) or Window Live Mail (Windows 7/Vista) when it is saved to the designated directory. During the saving and sending course, the system will give the hint information. NOTE: In Windows 7/Vista, only if Window Live Mail is installed, can the report be sent by email. 6.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG Chapter 7 Operation Instructions for Exercise ECG The exercise ECG function is optional. It is available only if you purchased this function. 7.1 Viewing Lead Placement Information 1. Click on the Lead Placement button on the main screen to display the Lead Placement window. 2. Click on Exercise ECG lead system to view the lead placement information.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.2 Selecting a Patient Record to Start a New Test 1. You can select a patient record from the database to start a new test. The operation steps are the same as those of resting ECG. For details, refer to Section 6.2, “Selecting a Patient Record to Start a New Test”. 2. Select Exercise ECG in the Patient Information window. 3. Click on OK in the Patient Information window to open the Exercise ECG Setting window.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.2.1 Setting Target HR The system applies the following formulas to calculate the target heart rate. 220 and 85 are default values, and you can modify them in different situations in the Exercise ECG Setting window. 7.2.2 Setting Normal BP Set the normal BP range in the Exercise ECG Setting window. When the patient’s BP exceeds the normal BP range, the system will consider it as abnormal status. 7.2.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.2.5 Setting BP Sampling Mode Select a BP sampling mode from the BP sampling mode list. 7.2.6 Setting BP Triggering Mode You can set the BP triggering mode to Square wave or QRS in the Exercise ECG Setting window. 7.2.7 Setting Auto Printing Select Auto print to print ECG reports automatically. 7.3 Entering New Patient Information If the patient is a new one, 1.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 2. Select Exercise ECG in the Patient information window. 3. Click on OK in the Patient information window to open the Exercise ECG Setting window. After setting the parameters, click on the OK button to open the pre-sampling screen. For details, please refer to Section 7.2, “Selecting a Patient Record to Start a New Test”. 7.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 1. Select a display mode from the display mode pull-down list 2. Select a gain from the gain pull-down list 3. Select a speed from the speed pull-down list 4. Select a lowpass filter from the lowpass filter pull-down list . . . . 7.5 Pretest Phase Click on Print to print the wave of 10 seconds before you click on Print.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.5.1 Viewing the Heart Rate and the Blood Pressure 1. View the heart rate in the heart rate field . The middle number is the current heart rate, and the right number is the target heart rate. The symbol shows the percentage of the current heart rate to the target heart rate, and the blue part indicates the current heart rate.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG NOTE: If the systolic pressure or the diastolic pressure exceeds the normal BP range, the hint The systolic/diastolic BP has exceeded the normal range! will appear on the sampling screen and the background color of the BP field will change from white to yellow. 7.5.2 Viewing Other Information 1. If a treadmill is used, view other information such as the stage time, exercise time, total time, PVC/min, speed/slope, and Max ST/Min ST. 2.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG Click on the Print button to print the current waveform. Click on Exit to return to the ECG sampling screen. 2. Click on the Comment button to display the Comment dialog box. Enter the comment in the Comment dialog box. Click on the OK button, and the comment will be displayed on the ECG waves on the analysis screen. 3. Click on the Template button, and the average waves of 12 leads and the calibration lines will be displayed.
SE-1010 PC ECG User Manual 4. Operation Instructions for Exercise ECG Click on the ST Trend button to display the ST trend. Right-click on the ST trend field to display the lead group menu, and then you can select other leads. 7.5.4 Printing the Pretest Report 1. The pretest report will be printed automatically 14 seconds after the beginning of the pretest phase. 2. Or, you can click on the Print button to print the pretest report. 7.6 Exercise Phase 1.
SE-1010 PC ECG User Manual 7. Operation Instructions for Exercise ECG Click on the Keep button and the system enters the Keep state. The hint Keep will be displayed in the speed/power field. The Next button becomes unavailable. In this state, the system will not follow the previous settings to change the speed and grade of the treadmill, but keep the current speed and grade until this button is pressed again. 7.7 Recovery Phase 1.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9 About Analysis Screen 7.9.1 About Summary Screen 1. Click on Summary to open the Summary screen. Figure 7-2 Summary Screen Click on the Measure button, click on one point on the wave, and then drag the mouse to another point. The distance, amplitude difference and heart rate between the two points will be displayed. 2.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 5. Enter diagnosis results in the Diagnosis textbox, and then click on Save. 6. Or, double-click on the necessary results required to be added in the Glossary textbox, the selected results will be displayed in the Diagnosis textbox, and then click on the Save button. 7. Click on the Preview button to display the Print Set window. Select Summary Report or Current Wave. Click on the OK button to open the preview screen.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG Figure 7-4 Current Wave Report NOTE: The diagnosis result is displayed in the diagnosis field of the summary report. 7.9.2 About ST Analysis Screen 1. Click on the ST analysis button to display the ST analysis screen.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 2. Click on Pretest to display 12-lead ST analysis waves of the pretest phase. 3. Click on Exercise to display 12-lead ST analysis waves of the exercise phase. 4. Click on Recovery to display 12-lead ST analysis waves of the recovery phase. 5. Click on Previous Page/Next Page to display 12-lead ST analysis waves of every 10 seconds. 6. Click on Max ST Ascend to display the Max ST ascending waves. 7.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9.3 About All View Review Screen 1. Click on the ALL View button to display the ALL View screen. Figure 7-6 All View Screen The ALL View screen displays the ECG wave of one lead throughout the whole test, and arrhythmia will be marked in red signs. 2. Select a lead from the lead pull-down list 3. Select a gain from the gain pull-down list to view the ECG wave of the lead. .
SE-1010 PC ECG User Manual 4. Operation Instructions for Exercise ECG Click on one point on the wave and drag the mouse to another point, and then the selected range will be marked with two red lines. Then click on the Zoom in button to display the amplified ECG segment. Drag the bottom scroll bar to view the whole amplified ECG waves. 5. The ALL View screen displays ECG waves of 450 seconds (50 seconds in one line and at least 9 lines in one page).
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9.4 About Trend Screen Click on the Trend button to display the STj Trend screen. Figure 7-7 STj Trend Screen You can observe 12-lead trend of the exercise test on the STj Trend screen. The horizontal coordinate indicates test time, but the unit of the vertical coordinate is different on different trend screens. Click on Zoom In or Zoom Out button to adjust the value on the vertical coordinate.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9.5 About ECG Strip Screen Click on the ECG Strip button to display the ECG Strip screen. Figure 7-8 ECG Strip Screen 1. Click on Pretest to display 12-lead strip waves of the pretest phase. 2. Click on Exercise to display12-lead strip waves of the exercise phase. 3. Click on Recovery to display12-lead strip waves of the recovery phase. 4. Click on Previous Page/Next Page to display strip waves of every 30 seconds. 7.9.
SE-1010 PC ECG User Manual Operation Instructions for Exercise ECG 7.9.7 Printing ECG Reports You can print an ECG report by clicking on the Print button. For details, please refer to Section 6.6.10, “Printing ECG reports”. 7.9.8 Saving ECG Reports You can save an ECG report by clicking on the Report Save button. For details, please refer to Section 6.6.11, “Saving ECG reports”. 7.9.9 Exiting the Analysis Screen Click on the Exit button on the analysis screen to return to the previous screen.
SE-1010 PC ECG User Manual Processing Patient Records Chapter 8 Processing Patient Records Click on the Data Manager button on the main screen (Figure 6-1) to open the Data Manager screen (Figure 8-1). Figure 8-1 Data Manager Screen Click on a patient record in the patient information list, and then all the examination records of the patient will be displayed in the examination record list.
SE-1010 PC ECG User Manual Processing Patient Records 2. Or, select a search item in the pull-down list , enter the corresponding information in the right textbox, and then click on the Search button. All the patient records which meet the conditions will be displayed in the patient information list. 3. Or, click on Advanced Search to display the Search Condition window, and then enter the search conditions.
SE-1010 PC ECG User Manual 4. Processing Patient Records Or, you can click on Physician or Examination , and then choose the doctor name or examination types, all the patient records which meet the conditions will be displayed in the patient information list. 5. Select The patients with no examination, and then the patient records which are registered but not examined, will appear in the patient information list. 8.
SE-1010 PC ECG User Manual Processing Patient Records 8.3 Deleting Records NOTE: The deletion of records is permanent, and you can’t restore the records deleted. Please use this operation cautiously. 8.3.1 Deleting Patient Records Click on a patient record in the patient information list on the Data Manager screen, and then click on the Delete button to delete the patient record from the patient information list. At the same time, all the examination records of the patient will be deleted.
SE-1010 PC ECG User Manual Processing Patient Records 8.5 Merging Examination Records Click on one or more examination records in the examination record list on the Data Manager screen, and then click on the Merge/Assign button to display the Patient Information window. Input a patient ID and click on the OK button to assign the examination record selected to this patient. 8.
SE-1010 PC ECG User Manual Processing Patient Records When Template is selected, the templates of the two examination records will be displayed in the window. You can press Print button to print the current window. Click on the Ruler button on the Compare window. Click on one point on the wave, and then drag the mouse to another point. The distance, amplitude difference and heart rate between the two points will be displayed.
SE-1010 PC ECG User Manual Processing Patient Records 8.7 Importing ECG Data into the Data Manager Screen Click on the Import button on the Data Manager screen (Figure 8-1) to open the following window. Select the data to be imported and click on the Select button to import the data into the Data Manager screen.
SE-1010 PC ECG User Manual Processing Patient Records If the data to be imported exists on the Data Manager screen, the following hint will pop up. If you press the Yes button, the imported record will replace the file with the same name. If you press the No button, the system will hint you a failure operation occurs. NOTE: Only ECG data in DAT format can be imported.
SE-1010 PC ECG User Manual Processing Patient Records 8.8 Exporting ECG Data from the Data Manager Screen Select examination records and click on the Export button on the Data Manager screen (Figure 8-1) to open the following window. Assign the file name, saving path and export file format (SCP, FDA-XML, DICOM, dat, pdf), and then click on the OK button to export the data into the selected path. At the same time, the patient information of these records will be exported.
SE-1010 PC ECG User Manual Processing Patient Records 8.9 Viewing an Examination Record Click on a patient record in the patient information list, and then all the records of the patient will be displayed in the examination record list. Select All records and all the examination records will be displayed in the examination record list. Double-click on an examination record in the examination record list on the Data Manager screen (Figure 8-1).
SE-1010 PC ECG User Manual Configuring the System Chapter 9 Configuring the System Click on the System Setting button on the main screen (Figure 6-1) to open the System Setting window. There are eight tabs in the System Setting window: Basic Information, Sample Setting, Device, Print Setting, Output File, Data Maintenance, GDT and Others. After you modify some information in the System Setting window, 1. Click on the OK button to save these modifications and exit. 2.
SE-1010 PC ECG User Manual Configuring the System 9.1.1 Setting Basic Information Enter information in the Hospital Name, User-Defined 1 or User-Defined 2 textbox in the Basic Information window (Figure 9-1). When you fill in the User-Defined 1/2 textbox, the corresponding items in the Patient Information window will change into what is filled.
SE-1010 PC ECG User Manual Configuring the System NOTE: Click on the New Patient button on the main screen to open the Patient Information window as the above figure shows. Select Memory in Basic Information window, the content of User-Defined 1 in the Patient Information window will be saved.
SE-1010 PC ECG User Manual Configuring the System 9.2 Sample Setup Figure 9-2 Sample Setup Window 9.2.1 Setting Filter Set filters in the Sample Setting window (Figure 9-2).
SE-1010 PC ECG User Manual Configuring the System DFT Filter DFT filter greatly reduces the baseline fluctuations without affecting ECG signals. There are two options: Weak and Strong. NOTE: If DFT filter is set to Strong, the ECG data displayed on the screen is 0.85 seconds later than the real-time ECG data; if DFT filter is set to Weak, the ECG data displayed on the screen is 1.8 seconds later than the real-time ECG data. EMG Filter EMG filter suppresses the disturbance caused by strong muscle tremor.
SE-1010 PC ECG User Manual Configuring the System 2. HR Calculation You can set the HR calculation to one of the 12 standard leads: І, ІІ, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6. 3. HR Analysis Lead You can set the HR Analysis Lead to one of the 12 standard leads: І, ІІ, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6. 9.2.
SE-1010 PC ECG User Manual Configuring the System 9.3 Device Setup Figure 9-3 Device Setup Window 9.3.1 Setting Sampling Device Select a sample device from the Sampling Device pull-down list in the Device window (Figure 9-3).
SE-1010 PC ECG User Manual Configuring the System When DX12 wireless sampling box is selected, you can click on the Address button to view the receiver address. You can match the transmitter and receiver based on the address from the Sampling Device pull-down list. 9.3.2 Setting Device Type/Mode Select a device type/mode from the Device Type/Device pull-down list . 9.3.
SE-1010 PC ECG User Manual Configuring the System 9.3.4.1 Creating a New Protocol Click on the New button in the Device window (Figure 9-3) to enter the Edit the protocol window. 1. Enter the protocol name in the Protocol Name textbox. 2. Right-click on the grid to display the menu , click on Add Stage, and enter information in the grid. 3. Set the auto printing time of 12-lead ECG reports in the 12-lead ECG Report (First) and 12-lead ECG Report (Repeat) columns. 4.
SE-1010 PC ECG User Manual Configuring the System 9.3.4.2 Modifying a Protocol 1. Select a protocol from the Protocol pull-down list, and then click on the Modify button to display the Edit Protocol window. 2. Double-click on an option of the phase row to be modified in the protocol list, and the pull-down list the corresponding grid. will be displayed. Select a phase and it will be displayed in 3.
SE-1010 PC ECG User Manual Configuring the System 9.3.4.3 Deleting a Protocol Select a protocol from the Protocol pull-down list, and then click on the Delete button to delete the protocol. NOTE: You can’t delete default protocols Bruce and IsoPower. 9.3.4.4 Restoring the default protocol Click on the Restore the default button to resume the default protocol. 9.3.5 Advanced Setup Click on Advanced Setting, input the correct password in the pop-up textbox, and then click on the OK button. 9.3.5.
SE-1010 PC ECG User Manual Configuring the System Only if Enable System Integration is selected, can Device ID and Output Path be set. Device ID and Output Path should be set to the same values as those of Smart ECG Net. Otherwise, data of this system can not be uploaded to Smart ECG Net system normally.
SE-1010 PC ECG User Manual Configuring the System 3. Click on Device Manager on the Computer Management screen, and then click on Human Interface Devices to view port information. 4. In the Barcode Setting window, set Device to the port you view on the Computer Management screen, and then click on the OK button to confirm. 5. Restart the PC ECG software.
SE-1010 PC ECG User Manual Configuring the System 9.4 Print Setup Figure 9-4 Print Setting Window 9.4.1 Choosing Patient Information to be Printed The default items of the patient information are Department, Room No and Physician. You can also select the additional information, such as FirstName/LastName, Date of Birth, height, weight, BP, race, medication, Pacemaker, Technician and Ref-Physician.
SE-1010 PC ECG User Manual Configuring the System If you select First Name/Last Name, the Patient Name textbox in the Patient Information window will change into the First Name and Last Name textboxes. If you select D.O.B, the D.O.B textbox appears in the Patient Information window, and the Age textbox becomes unavailable. 9.4.2 Choosing Diagnosis Information to be Printed The diagnosis information is displayed on the preview screen and in the report printed out.
SE-1010 PC ECG User Manual Configuring the System 9.4.3 Setting Rhythm Lead The rhythm lead can be one of 12 standard leads: І, ІІ, ІІІ, aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6. When the printing mode is set to 3×4+1 or 6×2+1, the rhythm lead selected in the Rhythm1 list box will be printed out. When the printing mode is set to 3×4+3, 3 rhythm leads selected in the Rhythm1, Rhythm2 and Rhythm3 list boxes will be printed out. 9.4.4 Defining Printing Format 1.
SE-1010 PC ECG User Manual Configuring the System 9.5 Output File Setup Figure 9-5 Output File Setup Window 9.5.1 File Naming The default file name is Exam Time-ID, and a “-” exists between every two fields. You can also modify the file name manually, such as Name-ID-Exam Time. Each field includes ID, name, exam time, age and sex items, you can randomly use these items to combine a file name, but you should select at least one item as a file name.
SE-1010 PC ECG User Manual Configuring the System 9.5.2 Setting PDF/JPG Select Output PDF when making diagnosis, the system will automatically output files in PDF format when making diagnoses. Select Output JPG when making diagnosis, the system will automatically output files in JPG format when making diagnoses. 9.5.3 Setting SCP Select Data Compression, the system will automatically output files in SCP format of data compression.
SE-1010 PC ECG User Manual Configuring the System 9.5.5 Setting DICOM Select Output when sampling finishes, the system will automatically output files in DICOM format when sampling finishes. Deselect Output when sampling finishes, the system will not output files in DICOM format when sampling finishes. Select Output when making diagnosis, the system will automatically output files in DICOM format when making diagnoses.
SE-1010 PC ECG User Manual Configuring the System 9.6.1 Database Rebuild You can rebuild the database to avoid losing data because of damaged data files. Click on the Rebuild Database button in the Data Maintenance window, and then you can select the path to rebuild database. You cannot enter the path manually. After rebuilding the path successfully, hint information will be displayed. 9.6.2 Database Backup You can also make a backup for the data files to avoid your data files being damaged or missing.
SE-1010 PC ECG User Manual Configuring the System Click on the OK button in the Backup Setting window, the system will begin backup. You can check the backup status according to the progress bar or you can stop backup at any times. The system will back up the data files for the latest time if you do not change the path.
SE-1010 PC ECG User Manual Configuring the System You can also set the reminded backup period to remind you back up the data files at any time. The default system reminded period is None, you can also select 7 days, 14 days, 30 days or you can define it manually. 9.
SE-1010 PC ECG User Manual Configuring the System Select GDT to enable GDT features. Click on the Browse button, and then appoint the path to exchange files with EDP. Fill in the Input file name textbox to set command file name assigned by EDP to the PC ECG system. Fill in the Output file name textbox to set data file name that is used by the PC ECG system to transmit the result data to EDP. Enter ECG ID to set GDT ID assigned to the system which will be entered in field 8315 or 8316 in the GDT protocol.
SE-1010 PC ECG User Manual Configuring the System 9.8.1 Setting Unit and Color Set the height unit to cm, inch or mm. Set the weight unit to Kg, g or Pound. Set the BP unit to kPa or mmHg. Set the speed unit to Km/h or mph. Set the grade unit to % or degree. Set the time mode to 24Hour or 12Hour. Set the date mode to MM-DD-YYYY, DD-MM-YYYY OR YYYY-MM-DD. Set the color of the background, waves, grid (5mm), grid (1mm), mark and text.
SE-1010 PC ECG User Manual Configuring the System 9.9 Modifying the Glossary Click on Edit Diagnosis Template from the Tool (F) pull-down list on the main screen (Figure 6-1), and then the Edit Diagnosis Template window appears. 1. Adding an item Enter a diagnosis item, such as aa in the textbox, and then click on the Add button. The added item will be displayed in the Edit Diagnosis Template window.
SE-1010 PC ECG User Manual 2. Configuring the System Adding a subitem Click on the item you wanted to add a subitem, enter a diagnosis subitem, such as bb in the textbox, and then click on the Add button. The added subitem will be displayed under aa. 3. Deleting an item Click on the item you wanted to delete from the Edit Diagnosis Template window, and then click on the Delete button to delete this item. 4. Save the settings Click on the Save button to save these modifications.
SE-1010 PC ECG User Manual Hint Information Chapter 10 Hint Information Hint information and the corresponding causes provided by the system are listed as follows. Table 10-1 Hint Information and Causes Hint Information Causes Lead off: X Electrodes fall off the patient or the patient cable falls off the ECG sampling box. It is pre-sampling now, please click on During the pre-sampling course ‘Start’ to begin recording.
SE-1010 PC ECG User Manual Hint Information Sampling Device is in sleep mode, please DX12 transmitter is in sleep mode. press “Power” to wake it up. Overload The direct current offset voltage on an electrode is too high. Sorry, Can not Connect to the Database! MSDE 2000 or SQL Server 2005 Express is not started up. Fail to create database! The system fails to create database.
SE-1010 PC ECG User Manual Cleaning, Care and Maintenance Chapter 11 Cleaning, Care and Maintenance CAUTION Turn off the system power and drag the power cable out from the socket before cleaning or disinfection. 11.1 Cleaning and Maintaining the Treadmill Daily Cleaning and Maintenance 1. Wipe the treadmill with a clean soft cloth to remove dust, moisture and sweat stain. 2. Wipe the handrail of the treadmill with a clean soft cloth damped in non-caustic neutral detergent. 3.
SE-1010 PC ECG User Manual Cleaning, Care and Maintenance ♦ Do not drag or twist the patient cable with excessive stress while using it. Hold the connector plugs instead of the cable when connecting or disconnecting the patient cable. ♦ Align the patient cable to avoid twisting, knotting or crooking at a closed angle while using it. ♦ Store the lead wires in a big wheel. ♦ Once damage or aging of the patient cable is found, replace it with a new one immediately.
SE-1010 PC ECG User Manual Cleaning, Care and Maintenance a) Inspect the equipment and accessories for mechanical and functional damage. b) Inspect the safety related labels for legibility. c) Inspect the fuse to verify compliance with the rated current and circuit-breaking characteristics. d) Verify that the device functions properly as described in the instructions for use. e) Test the protection earth resistance according to IEC/EN 60601-1: Limit: 0.1 ohm.
SE-1010 PC ECG User Manual Accessories Chapter 12 Accessories WARNING Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection can not be guaranteed. Table 12-1 Standard Accessory List for Wired System Accessory Part Number DP12 ECG Sampling Box 02.01.210039 Sentinel / USB 01.18.047225 Resting ECG External USB Cable 01.13.036134 DP12 Patient Cable / European Standard 01.57.
SE-1010 PC ECG User Manual Accessories Disposable electrodes 11.57.471046 DX12 Belt 01.57.106750 Portable Bag 11.56.078136 Table 12-3 Optional Accessory List Accessory Part Number Sentinel / USB 01.18.047229 Patient Cable for Exercise ECG / European Standard 01.57.109850 (only for Wired ECG System) Patient Cable for Exercise ECG / American Standard 01.57.109851 (only for Wired ECG System) Resting ECG External USB Cable 01.13.036134 Disposable Frosting Film for Skin Preparation 01.57.
SE-1010 PC ECG User Manual Warranty & Service Chapter 13 Warranty & Service 13.1 Warranty EDAN warrants that EDAN’s products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within warranty period. The warranty is void in cases of: 1. damage caused by mishandling during shipping. 2. subsequent damage caused by improper use or maintenance. 3. damage caused by alteration or repair by anyone not authorized by EDAN. 4.
SE-1010 PC ECG User Manual Recommended Optional Accessories Chapter 14 Recommended Optional Accessories Treadmill: Model: TM-400 Manufacturer: EDAN INSTRUMENTS, INC. China CE marking Model: Valiant Manufacturer: Lode B.V. The Nettherlands CE marking Model: h/p/cosmos (all medical models) with coscom interface Manufacturer: Full Vision Inc. USA CE marking Model: mercury med 4.0, mercury 4.
SE-1010 PC ECG User Manual Recommended Optional Accessories STRESS BP: Model: Tango+ Manufacturer: SunTech Medical Inc. USA CE Certificate and FDA 510(k) clearance Isolating Transformer: Model: ES710 Manufacturer: BenDer Inc. Deutschland Electrical Outlet: Power Consumption: no less than 4500VA Special use for medical equipment Printer: Model: HP2010, HP2035 Manufacturer: HP Company, USA Model: CANON iP1980 Manufacturer: CANON Company, Japan WARNING 1.
SE-1010 PC ECG User Manual Technical Specifications Appendix 1 Technical Specifications A1.
SE-1010 PC ECG User Manual Technical Specifications A1.3 Physical Specifications DP12 ECG sampling box: 148 mm (L) ×100 mm (W) × 40 mm (H) (5.8in×3.9in×1.6in) Dimensions DX12 transmitter: 63mm(L)×107mm(W) ×23mm(H) (2.5in×4.2in×0.9in) DX12 receiver: 155mm(L)×100mm(W)×30mm(H) (6.1in×3.9in×1.2in) DP12 ECG sampling box: Approx. 210g Weight DX12 transmitter: Approx. 113g (not including battery) DX12 receiver: Approx. 173g A1.
SE-1010 PC ECG User Manual Technical Specifications Paper Speed: 5 mm/s, 10 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s(±3%) Record message: Date, Time, Printing Speed, Filter, Symbol, Heart Rate, Patient ID, Sex, Age, Lead Mark, Lead Wave, Average Template Wave or Rhythm Lead Wave, Measurement Result and Interpretation Information Result (option) etc.
SE-1010 PC ECG User Manual Technical Specifications Work Frequency Filter DFT Filter: weak/strong LOWPASS Filter: 25 Hz / 35 Hz / 45 Hz / 75 Hz / 100 Hz / 150 Hz DP12 ECG sampling box≥110 dB CMRR DX12 transmitter≥100 dB Pacemaker Detection DP12 ECG sampling box: ±2 to ±700 mV Amplitude DX12 transmitter: ±2 to ±500 mV Width 0.1 to 2.0 ms Sampling Frequency 10,000 /sec/channel NOTE: Test the accuracy of input signal reproduction according to the methods described in clause 4.2.7.
SE-1010 PC ECG User Manual EMC Information Appendix 2 EMC Information Guidance and manufacture’s declaration - electromagnetic emissionsfor all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration - electromagnetic emission SE-1010 PC ECG is intended for use in the electromagnetic environment specified below. The customer or the user of SE-1010 PC ECG should assure that it is used in such an environment.
SE-1010 PC ECG User Manual EMC Information Guidance and manufacture’s declaration - electromagnetic immunity for all EQUIPMENT and SYSTEMS Guidance and manufacture’s declaration - electromagnetic immunity SE-1010 PC ECG is intended for use in the electromagnetic environment specified below. The customer or the user of SE-1010 PC ECG should assure that it is used in such an environment.
SE-1010 PC ECG User Manual EMC Information SE-1010 PC ECG be powered from an uninterruptible power supply or a battery. 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5 sec NOTE UT is the a.c. mains voltage prior to application of the test level.
SE-1010 PC ECG User Manual EMC Information recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations.
SE-1010 PC ECG User Manual EMC Information Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM – for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and SE-1010 PC ECG SE-1010 PC ECG is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
SE-1010 PC ECG User Manual Abbreviation Appendix 3 Abbreviation Abbreviation Statement ECG Electrocardiograph/Electrocardiogram VCG Vector ECG TVCG Time Vector ECG QTD QT Dispersion FCG Frequency ECG HRV Heart Rate Variability HFECG High Frequency ECG HRT Heart Rate Turbulence SAECG Signal Averaged ECG TO Turbulence Onset TS Turbulence Slope VLP Ventricular Late Potential HR Heart Rate P Dur P-wave Duration PR Dur P-R Interval QRS Dur QRS Complexes Duration QT/QTc Q-T In
SE-1010 PC ECG User Manual Abbreviation HF (norm) Standard HF Power aVF Left Foot Augmented Lead aVL Left Arm Augmented Lead aVR Right Arm Augmented Lead LA Left Arm R Right RA Right Arm RL Right Leg ID Identification AC Alternating Current USB Universal Serial Bus - 166 -