About this Manual P/N: 01.54.106666-21 Release Date: June 2013 © Copyright EDAN INSTRUMENTS, INC. 2008-2013. All rights reserved. Statement This manual will help you understand the operation and maintenance of the product better. It is reminded that the product shall be used strictly complying with this manual. User’s operation failing to comply with this manual may result in malfunction or accident for which EDAN INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
Terms Used in this Manual This guide is designed to give key concepts on safety precautions. WARNING A WARNING label advises against certain actions or situations that could result in personal injury or death. CAUTION A CAUTION label advises against actions or situations that could damage equipment, produce inaccurate data, or invalidate a procedure. NOTE A NOTE provides useful information regarding a function or a procedure.
Table of Contents Chapter 1 Safety Guidance ........................................................................................................... 1 1.1 Intended Use........................................................................................................................ 1 1.2 Warnings and Cautions........................................................................................................ 1 1.2.1 General Warnings ................................................................
.6 Inspection Before Test ....................................................................................................... 37 5.7 Setting DX12 Transmitter (for Wireless System) ............................................................. 38 5.7.1 Keyboard Locking/Unlocking ................................................................................. 40 5.7.2 Menu Settings .........................................................................................................
.6.5 Analyzing HRV ....................................................................................................... 72 6.6.5.1 Editing the HRV Data on the Analysis Screen .............................................. 73 6.6.5.2 Editing the HRV Waveform in the Waveform Window ................................ 74 6.6.5.3 Previewing HRV ........................................................................................... 75 6.6.6 Analyzing HRT......................................................
.7 Recovery Phase ............................................................................................................... 100 7.8 Exiting the Exercise Test ................................................................................................. 100 7.9 About Analysis Screen .................................................................................................... 101 7.9.1 About Summary Screen........................................................................................
9.2.6 Setting Anti-aliasing .............................................................................................. 125 9.2.7 Selecting QRS Voice ............................................................................................. 125 9.2.8 Selecting Sequence Mode When Sampling .......................................................... 125 9.3 Device Setup ................................................................................................................... 126 9.3.
Chapter 10 Hint Information .................................................................................................... 146 Chapter 11 Cleaning, Care and Maintenance ......................................................................... 148 11.1 Cleaning and Maintaining the Treadmill ....................................................................... 148 11.2 Cleaning and Maintaining the Patient Cable and Reusable Electrodes ........................ 148 11.3 Disinfection ....................
SE-1010 PC ECG User Manual Safety Guidance Chapter 1 Safety Guidance This chapter provides important safety information related to the use of SE-1010 PC ECG. 1.1 Intended Use SE-1010 PC ECG is a PC-based diagnostic tool intended to acquire, process and store ECG signals from adult and pediatric patients undergoing stress exercise test or resting test. SE-1010 PC ECG is intended to be used only in hospitals and healthcare facilities by doctors and trained healthcare professionals.
SE-1010 PC ECG User Manual Safety Guidance WARNING 3. EXPLOSION HAZARD - Do not use the system in the presence of flammable anesthetic mixtures with oxygen or other flammable agents. 4. SHOCK HAZARD - The power receptacle must be a hospital grade grounded outlet. Never try to adapt the three-prong plug to fit a two-slot outlet. 5. Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection can not be guaranteed.
SE-1010 PC ECG User Manual Safety Guidance WARNING 16.SHOCK HAZARD - Don’t connect electrical equipment, which has not been supplied as a part of the system, to the multiple portable socket-outlet supplying the system. 17.Do not connect any equipment or accessories that are not approved by the manufacturer or that are not IEC/EN 60601-1-1 approved to the system.
SE-1010 PC ECG User Manual Safety Guidance WARNING 25.You should purchase computer, printer, treadmill, ergometer, BP monitor and bar code reader from the manufacturer. Otherwise, the manufacturer will not be held responsible for the maintenance of the PC hardware, operating system and other accessories. 26.If multiple instruments are connected to a patient, the sum of the leakage currents may exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard. Consult your service personnel. 27.
SE-1010 PC ECG User Manual Safety Guidance 1.2.3 General Cautions CAUTION 1. Avoid liquid splash and excessive temperature. The temperature must be kept between 5 ºC and 40 ºC during operation, and it should be kept between -20 ºC and 55 ºC during transportation and storage. 2. Do not use the equipment in a dusty environment with bad ventilation or in the presence of corrosive. 3.
SE-1010 PC ECG User Manual Safety Guidance 1.2.5 Preparation and Operation Warnings (for Exercise ECG) WARNING 1. Test the safety stop (mushroom type) and safety stop (cord type) of the treadmill before using the system. 2. During the exercise test, ensure that there are at least 2 experienced physicians present. One of them observes the patient and deals with the emergency. 3.
SE-1010 PC ECG User Manual Safety Guidance 1.2.6 Contraindications (for Exercise ECG) Absolute Contraindications: 1. Acute MI (within 2 days) 2. High-risk unstable angina 3. Hemodynamic compromise caused by uncontrolled cardiac arrhythmia 4. Symptomatic severe aortic stenosis 5. Heart failure with clinic episode uncontrolled 6. Acute pulmonary embolus or pulmonary infarction 7. Acute myocarditis or pericarditis 8. The patient opposes the test. Relative Contraindications: 1. Left main coronary stenosis 2.
SE-1010 PC ECG User Manual Safety Guidance 1.3 List of Symbols Equipment or part of CF type with defibrillator proof Caution Consult Instructions for Use Recycle Part Number Serial Number Date of Manufacture Manufacturer Authorized Representative in the European Community The symbol indicates that the device complies with the European Council Directive 93/42/EEC concerning medical devices.
SE-1010 PC ECG User Manual Safety Guidance Class Ⅱ Transmission Status Indicator of Bluetooth Burglar Lock Power Supply Indicator of DX12 Receiver This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (for DX12 Transmitter) (1) this device may not cause harmful interference, and (for DX12 Receiver) (2) this device must accept any interference received, including interference that may cause undesired operation.
SE-1010 PC ECG User Manual Introduction Chapter 2 Introduction SE-1010 PC ECG has similar functions with an ordinary electrocardiograph. ECG data can be sampled, analyzed and stored in a PC, and it can be saved in PDF, Word, BMP or JPG format. ECG waves can be frozen and reviewed. Auto measurement and diagnosis are available, and the diagnosis template can be edited. SE-1010 PC ECG can also be invocated by Smart ECG Net version 1.3 or above.
SE-1010 PC ECG User Manual Introduction Wired System of SE-1010 PC ECG System 1. Resting ECG of Wired System Patient Patient Cable Resting PC (Manually Configured) ECG Cable DP12 ECG Sampling Box USB Cable Printer (Manually Configured) 2.
SE-1010 PC ECG User Manual Introduction Wireless System of SE-1010 PC ECG System The DX12 device which consists of transmitter and receiver has passed FCC certification. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. 1.
SE-1010 PC ECG User Manual 3. Introduction FCC Statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
SE-1010 PC ECG User Manual Introduction Front Panel Lamp USB Socket Name Explanation Lamp When the ECG sampling box is powered by the PC, the lamp will be lit.
SE-1010 PC ECG User Manual Introduction USB Socket Definitions of corresponding pins: Pin Signal Pin Signal 1 GND 6 GND 2 VCC 7 GND 3 QRS 8 GND 4 GND 9 D- 5 GND 10 D+ WARNING 1. When the computer connected to the USB cable is powered on, do not connect the USB cable to the DP12 ECG sampling box; when the system is powered on, do not disconnect the USB cable from the ECG sampling box. 2.
SE-1010 PC ECG User Manual Introduction Patient Cable Socket Definitions of corresponding pins: Pin Signal Pin Signal Pin Signal 1 C2 / V2 6 SH 11 F / LL 2 C3 / V3 7 NC 12 C1 / V1 or NC 3 C4 / V4 8 NC 13 C1 / V1 4 C5 / V5 9 R / RA 14 RF (N) /RL or NC 5 C6 / V6 10 L / LA 15 RF (N) / RL NOTE: The left side of “/” is European standard, and the right side is American standard.
SE-1010 PC ECG User Manual Introduction WARNING 1. Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configuration shall comply with the valid version of the standard IEC/EN 60601-1-1.
SE-1010 PC ECG User Manual Introduction Front Panel Patient Cable Socket Patient Cable Socket :Applied part of type CF with defibrillator proof :Caution Definitions of corresponding pins: Pin Signal Pin Signal Pin Signal Pin Signal 1 NC 6 C5/V5 11 NC 16 L/LA 2 F/LL 7 NC 12 C2/V2 17 NC 3 NC 8 C4/V4 13 NC 18 R/RA 4 C6/V6 9 NC 14 C1/V1 19 NC 5 NC 10 C3/V3 15 NC 20 N/RL NOTE: The left side of “/” is European standard, and the right side is American standard.
SE-1010 PC ECG User Manual Introduction DX12 Receiver Appearance USB Socket USB Socket - 19 -
SE-1010 PC ECG User Manual Introduction Definitions of corresponding pins: Pin Signal Pin Signal 1 GND 5 GND 2 VCC 6 GND 3 QRS 7 GND 4 GND 8 D- 5 GND 10 D+ Lock 2.4 Features 1. Powerful functions, friendly windows and easy operation 2. 3/6/12-channel ECG waves are displayed and printed simultaneously 3. ECG waves can be frozen and reviewed 4. Supporting auto measurement and diagnosis 5.
SE-1010 PC ECG User Manual Introduction 7. Reports can be printed in PDF, Word, JPG or BMP format 8. Supporting multi-language 9. Updated to be a network electrocardiograph, transmitting ECG data over LAN or WAN or INTERNET 10. Automatic baseline adjustment for optimal printing 11. High performance filters guarantee stable ECG waveforms 12. Real-time analysis, real-time displaying and printing 12-lead simultaneous ECG waveforms 13.
SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System Chapter 3 Assembling SE-1010 PC ECG System 3.
SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System For Resting ECG of wired system, ♦ Insert plug 1 of the patient cable into socket 3 of DP12 ECG sampling box. ♦ Insert plug 8 of the cable into socket 4 of DP12 ECG sampling box. ♦ Insert plug 9 of the cable into the USB socket of the PC. ♦ Connect a printer to the PC. ♦ Insert the Sentinel into the USB socket of the PC. ♦ Make sure that the above parts are properly connected, and then connect the PC, and the printer to the power supply.
SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System 3.
SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System 10 7 6 9 8 DX12 Receiver Exercise ECG Cable 12 11 Burglar Lock Assembly Drawing For Resting ECG of wireless system, 1. Insert plug 1 of the patient cable into socket 2 of DX12 transmitter. 2. Insert plug 4 of the cable into socket 6 of DX12 receiver. 3. Insert plug 5 of the cable into the USB socket of the PC. 4. Connect a printer to the PC. 5. Insert the Sentinel into the USB socket of the PC. 6.
SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System 4. Insert plug 9 of the cable into the USB socket of the PC. 5. Connect plug 10 of the cable to the BP monitor. 6. Connect a treadmill or an ergometer to the PC. 7. Connect a printer to the PC. 8. Insert the Sentinel into the USB socket of the PC. 9. Make sure that the above parts are properly connected, and then connect the PC, treadmill/ergometer and printer to the power supply. WARNING 1.
SE-1010 PC ECG User Manual Installing SE-1010 PC ECG Software Chapter 4 Installing SE-1010 PC ECG Software 4.1 System Running Environment 4.1.
SE-1010 PC ECG User Manual Installing SE-1010 PC ECG Software 4.2 About Installation Window Insert the installation CD into CD-ROM, and double-click on Setup.exe to open the following installation window. Figure 4-1 Installation Window Click on the Install button to install PC ECG. Click on the Next button continually during installation. After installing PC ECG, click on the Install button in the installation window. Then the Environment Detection window pops up.
SE-1010 PC ECG User Manual Preparations Before Operation Chapter 5 Preparations Before Operation 5.1 Preparing the Patient 5.1.1 Instructing the Patient Before attaching the electrodes, greet the patient and explain the procedure. Explaining the procedure decreases the patient’s anxiety. Reassure the patient that the procedure is painless. Privacy is important for relaxation. When possible, prepare the patient in a quiet room or area where others can’t see the patient.
SE-1010 PC ECG User Manual Preparations Before Operation 5.2 Connecting the Electrodes of Wired System WARNING The performance and electric shock protection can be guaranteed only if the original patient cable and electrodes of the manufacturer are used. The patient cable includes main cable and lead wires which can be connected to electrodes according to the colors and identifiers.
SE-1010 PC ECG User Manual Preparations Before Operation 5.3 Connecting the Electrodes of Wireless System The patient cable includes main cable and lead wires which can be connected to electrodes according to the colors and identifiers. Lead Wires Main Cable Connecting to DX12 transmitter Electrode Connector Patient Cable for Resting ECG Connecting to the ECG Sampling Box Connecting to Electrodes Main Cable Lead Wires Patient Cable for Exercise ECG 1.
SE-1010 PC ECG User Manual Preparations Before Operation Table 5-1 Electrodes and Their Identifiers and Color Codes European WILSON FRANK Identifier Right arm Right arm R Left arm Left arm Right leg American Color Code Identifier Color Code Red RA White L Yellow LA Black Right leg N or RF Black RL Green Left leg Left leg F Green LL Red Chest 1 I C1 White/Red V1 Brown/Red Chest 2 E C2 White/Yellow V2 Brown/Yellow Chest 3 C C3 White/Green V3 Brown/Green Ches
SE-1010 PC ECG User Manual Preparations Before Operation 5.4.2 Frank Lead System FRANK lead system is usually adopted when PC ECG is used to produce VCG.
SE-1010 PC ECG User Manual Preparations Before Operation 1. Clean the electrode area which is in a short distance above the ankle or the wrist with alcohol. 2. Daub the electrode area on the limb with gel evenly. 3. Place a small amount of gel on the metal part of the limb electrode clamp. 4. Connect the electrode to the limb, and make sure that the metal part is placed on the electrode area above the ankle or the wrist. 5. Attach all limb electrodes in the same way.
SE-1010 PC ECG User Manual ♦ Preparations Before Operation Attach the disposable electrode to the electrode area at the back midline. The quality of ECG waveform will be affected by the contact resistance between the patient and the electrode. In order to get a high-quality ECG, the skin-electrode resistance must be minimized while connecting electrodes. WARNING 1. Make sure that all electrodes are connected to the patient correctly before operation. 2.
SE-1010 PC ECG User Manual Preparations Before Operation C1 White/Red V1 Brown/Red C2 White/Yellow V2 Brown/Yellow C3 White/Green V3 Brown/Green C4 White/Brown V4 Brown/Blue C5 White/Black V5 Brown/Orange C6 White/Violet V6 Brown/Violet The Precordial Electrodes’ Positions on Body Surface: C1: Fourth intercostal space at the right border of the sternum C2: Fourth intercostal space at the left border of the sternum C3: Fifth rib between C2 and C4 C4: Fifth intercostal space on the l
SE-1010 PC ECG User Manual Preparations Before Operation 5.6 Inspection Before Test In order to avoid safety hazards and get good ECG records, the following inspection procedure is recommended before operation. 1. Environment: 1. Make sure that there is no electromagnetic interference source around the equipment, especially large medical electrical equipment such as electrosurgical equipment, radiological equipment, magnetic resonance imaging equipment etc. Switch off these devices when necessary. 2.
SE-1010 PC ECG User Manual Preparations Before Operation 5.7 Setting DX12 Transmitter (for Wireless System) Switch on DX12 receiver and install batteries on DX12 transmitter. Press to start up DX12 transmitter, and then the company information and the main screen will be displayed. Figure 5-1 Main Screen When the main screen is displayed, press to switch the leads. If the Bluetooth connection icon is not displayed on the main screen, you have to match the device manually.
SE-1010 PC ECG User Manual 1) Press Preparations Before Operation to enter the menu screen. Figure 5-2 Menu Screen 2) Press to display Match Device item in black, and then press screen with a prompt “Inquiring…”.
SE-1010 PC ECG User Manual Preparations Before Operation 3) When the receiver is found, the address of DX12 receiver will be displayed on the screen in 10 seconds. Press to start up the Bluetooth connection. The Bluetooth connection will be displayed on the main screen of DX12 transmitter and the Bluetooth status indicator illuminates in kelly when DX12 transmitter and DX12 receiver are matched successfully.
SE-1010 PC ECG User Manual Preparations Before Operation 5.7.2 Menu Settings Press on the main screen to enter the menu screen (Figure 5-2). Press screen to display an item in black, and then press on the menu to enter the setting screen of this item. Table 5-3 Menu Menu Option Description On Select On to turn on the backlight of LCD screen. Off Select Off to turn off the backlight of LCD screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Chapter 6 Operation Instructions for Resting ECG Double-click on the shortcut icon the desktop icon for SE-1010 PC ECG. on the desktop to display the main screen. is NOTE: Do not use other software when using PC ECG software. Figure 6-1 Toolbar of Main Screen The toolbar contains six buttons. From left to right, they are New Patient, STAT ECG, Data Manager, System Setting, Lead Placement and Exit.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG You can set the unit, print, frequency of AC and the saving path of source data based on your needs. Click on the OK button after setup, the system will enter the main screen automatically. NOTE: You should install the software to the saving path of source data after the uninstallation and reinstallation; otherwise, the software needs a new configuration. 6.1 Viewing Lead Placement Information 1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.2 Selecting a Patient Record to Start a New Test Click on the Data Manager button on the main screen (Figure 6-1) to open the Data Manager screen (Figure 6-3). Figure 6-3 Data Manager Screen 1. Select a search item in the pull-down list on the Data Manager screen. Then all the patient records which meet the search condition are listed in the patient information list. 2.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 3. Or, click on Advanced Search to display the Search Condition window. Enter the search conditions, and click on the Search button, and all the patient records which meet the conditions will be displayed in the patient information list. 4. Click on the patient record in the patient information list and click on the Select button to open the Patient Information window.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.3 Entering New Patient Information 6.3.1 Entering New Patient Information Manually If the patient is a new one, you can click on the New Patient button on the main screen (Figure 6-1) to display the Patient Information window. Then you need to input the patient’s related information. 1. Enter basic information, such as patient ID, name, sex and age.
SE-1010 PC ECG User Manual 3. Operation Instructions for Resting ECG Enter information of doctor and department 1) Enter information of physician, technician or Req. physician Click on the pull-down list , and then click on the Edit button to display the Edit window. Enter the doctor name in the textbox, and then click on the Add button. The doctor name will be displayed in List.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 2) Enter information of department Click on the pull-down list , and then click on the Edit button to display the Edit Department window. Enter the department name in the textbox, and then click on the Add button. The doctor name will be displayed in List. Meanwhile, you can also enter the department name in the textbox of Dept. item, and then click on the OK button in the Patient Information window.
SE-1010 PC ECG User Manual 4. Operation Instructions for Resting ECG Confirm pacemaker information If you select Pacemaker in the Print Setting window, Pacemaker appears in the Patient Information window. Select Pacemaker to detect very small pacemaker pulses. However, when pacemaker pulse enhancer is on, the system is very sensitive, and should not be close to equipment emitting high frequency radiation. High frequency radiation can interfere with pacemaker pulse detection and normal ECG acquisition.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.4 Selecting Sampling Type You can select a sampling type in the Patient information window. 6.5 Sampling Resting ECG After inputting the patient information, click on the OK button in the Patient information window to open the ECG sampling screen. Before sampling, if you do not connect the PC to the ECG sampling box, the following hint will pop up.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG The system begins to pre-sample ECG. Figure 6-5 Pre-Sampling Screen 6.5.1 Specifying Display Mode There are eight display modes including 12*1, 6*2, 3*4, 6*1, 3*1, 3*4+1, 3*4+3 and 6*2+1. When the display mode is set to 12*1, 12-channel ECG waves are displayed on one screen simultaneously. When the display mode is set to 6*2, 12-channel ECG waves are displayed in 2 groups of 6 on one screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG When the display mode is set to 3*1, 3-channel ECG waves are displayed on one screen simultaneously. When the display mode is set to 3*4+1, 12-channel ECG waves are displayed in 4 groups of 3 and one rhythm lead on one screen. When the display mode is set to 3*4+3, 12-channel ECG waves are displayed in 4 groups of 3 and three rhythm leads on one screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.5.5 Recording ECG Data When the pre-sample ECG waves are steady, you can click on the Start button to save the sampled ECG data to the designated directory. For details, please refer Section 9.1.4, “Specifying the Storage Path of the ECG Data”. Figure 6-6 ECG Sampling Screen NOTE: After you click on the Start button, the system will save the sampled ECG data.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-7 Wave Review Window Click on Exit to return to the ECG sampling screen. NOTE: The display modes in the Wave review window are the modes you select on the ECG sampling screen. 3*1 and 6*1 display modes are displayed in 3*4 and 6*2 modes. 6.5.7 Stopping Sampling Data After clicking on the Start button, there are two ways to stop sampling data. 1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.5.8 Printing ECG Waves Click on the Print button on the ECG sampling screen to print the ECG waves on the ECG sampling screen. NOTE: 1. You can set the printer type on the Printer and Faxes screen. For details, please refer to Section 6.6.10, “Printing ECG Reports”. 2. The report can be printed in white- black or color. The report color is defined by setting the printer type and can be observed on the preview screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.1.1 Viewing the Waveform Click on the Waveform tab on the normal ECG analysis screen to open the Waveform window (Figure 6-8). Figure 6-8 Normal ECG - Waveform Window You can choose a speed, a gain and a display mode for the displayed waves. Click on the Measure button in the Waveform window (Figure 6-8). Click on one point on the wave, and then drag the mouse to another point.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG The common parameters are displayed in the circle in Figure 6-8. These parameters are as follows. Designation Description Heart Rate Heart Rate P Duration P-wave duration of the current lead PR Dur. P-R interval of the current lead QRS Dur.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 1. Enter your own opinions in the Diagnosis textbox, and then click on the Save button. 2. Or, double-click on the necessary results required to be added in the Glossary textbox, the selected results will be displayed in the Diagnosis textbox, and then click on the Save button. If the Integration function is activated, the data can be uploaded to Smart ECG Net system after you click on the Save button. 6.6.1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG To Edit Waveform on the Analysis Screen You can press a lead button in to display magnified average templates of this lead. When you press more than one lead button, magnified average templates of these leads will be overlapped with the same central axis. When you press ALL, magnified average templates of all leads will be overlapped with the same central axis. You can set the speed and the gain of average templates.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.1.3 About the Detail Information Window Click on the Detail information tab on the normal ECG analysis screen to open the Detail information window. This window displays lead parameter values as Figure 6-10 shows. Figure 6-10 Normal ECG - Detail Information Window Click on the Export Excel button to export an Excel file. To Edit Diagnosis Result in the Detail Information Window For details, please refer to Section 6.6.1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.1.4 About the Rhythm Wave Window Click on the Rhythm Wave tab on the normal ECG analysis screen to open the Rhythm wave window. This window displays rhythm wave as Figure 6-11 shows. Figure 6-11 Normal ECG - Rhythm Wave Window You can set the gain, the speed and the lead of the displayed ECG waves. You can click on Previous Page or Next Page to display the waves of the previous or next page.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 4. Click on Close to close the normal ECG preview screen and return to the previous screen. Figure 6-12 ECG Wave NOTE: The report title is in one line by default, you can also change the line to two lines. Find the text named PCECG.ini in the installation path, modify LineofTitle=0 to LineofTitle=1, and then save the text to change the line to two lines.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Report Title in Two Lines 6.6.2 Analyzing QT Dispersion Click on the QTD button to open the QT Dispersion screen. QT Dispersion: The difference between the largest QT interval and the shortest QT interval based on the synchronous 12-lead surface ECG. The QT interval is a measurement of the time difference between the start of the Q wave and the end of the T wave.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.2.1 Editing Waveform on the QT Dispersion Screen There are red and blue triangle icons on top of the displayed ECG waves. You can click on the triangle icon to change the color. represents the current chosen R wave represents the position of R wave You can set the speed and the gain of the displayed ECG waves. The lead number and the lead type of the displayed ECG waves can also be chosen. 6.6.2.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.2.3 Previewing QT Dispersion Click on the Preview button on the QT Dispersion screen to open the QT Dispersion preview screen. is the toolbar on the QT Dispersion preview screen. 1. Click on Print(P) to print the report. 2. Click on Next Page/Prev Page to switch to the previous/next preview page. 3. Click on Two Page to preview two pages on one screen simultaneously. 4. Click on Zoom In/ Zoom Out to magnify/minify the preview page. 5.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.3 Analyzing Frequency ECG Click on the FCG button to open the Frequency ECG screen. Frequency ECG displays characteristic waves of ECG signal spectrum. There are two tabs on the Frequency ECG screen: Two leads compare and 12-lead power spectrum. 6.6.3.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Every parameter of Frequency ECG is shown in the following table. Power Spectrum Phase Shift Designation Definition 1/2 If the first peak value is lower than 90% of the second peak value, it is positive (‘+’). Otherwise it is minus (‘-’). 1-N The first peak is too low or disappears. (The frequency value of the first peak’s position * 60 = heart rate. For example, if the frequency value of the first peak’s position is 1.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.3.2 About 12-lead Power Spectrum Window The 12-lead power spectrum window displays the power spectrum of 12-lead ECG waves. Figure 6-16 Frequency ECG - 12-lead Power Spectrum Window 6.6.3.3 Previewing Frequency ECG Click on the Preview button to open the Frequency ECG preview screen. is the toolbar on the Frequency ECG preview screen. 1. Click on Print(P) to print the report. 2.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-17 Two-lead Comparison Report Figure 6-18 12-lead Power Spectrum Report - 69 -
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.4 Analyzing High Frequency ECG In a conventional electrocardiogram, only these ECG signals, of which the amplitude is in the millivolt range and the frequency is less than 100 Hz, are visible. Those ECG signals, of which the amplitude is in the microvolt range and the frequency is between 150Hz and 250Hz, are invisible.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG The ECG wave between the two green lines is the present one. Click on to select another QRS wave. Select the present lead in the pull-down list . You can choose the lead group, the gain and the speed of the waves displayed in the bottom part of the HF ECG analysis screen. is the statistic data of beadings, notches and slurs. You can change a value by double-clicking on the value.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-20 HF ECG Report 6.6.5 Analyzing HRV Click on HRV to display the HRV ECG analysis screen. The HRV ECG analysis screen includes two tabs: Auto diagnosis result and Waveform. NOTE: 1. The HRV sampling time can be set in the Sample Setting window. 2. The HRV analysis lead can be selected in the Sample Setting window.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.5.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG LF Low Frequency HF High Frequency LF/HF Ratio of low frequency to high frequency LF (norm) Standard LF power HF (norm) Standard HF power H- Doctor Diagnosis Field 1. Enter your own opinions in the Auto diagnosis textbox, and then click on the Save button. 2.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG HRV waveform is displayed in the Waveform window (Figure 6-22). 1. You can drag the mouse in the window to choose the wave field to be printed. Then click on the Print button to print the selected wave field. 2. Click on Previous Page or Next Page to display the waves of the previous or next page. 6.6.5.3 Previewing HRV Click on the Preview button to open the HRV preview screen. is the toolbar on the HRV preview screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-23 HRV Preview Screen 6.6.6 Analyzing HRT Click on HRT to display the HRT ECG analysis screen.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG You can select Average Histogram or Single Histogram in Mode View field. Click on Next PVC or Previous PVC to see the Next or Previous PVC wave and the relevant histogram. Click on the Preview button to open the HRT preview screen. is the toolbar on the HRT preview screen. 1. Click on Print(P) to print the report. 2. Click on Next Page/Prev Page to switch to the previous/next preview page. 3.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.7 Analyzing Vector ECG Click on the VCG button on the ECG analysis screen to display the VCG analysis screen. Vector ECG displays 3D image of ECG activity. and the vector loop You can choose the plane on the VCG analysis screen. Plane choices include Frontal (F), Horizontal (H), Sagittal(S) and ALL. Loop choices include P loop, QRS loop, T loop and ALL. 6.6.7.
SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-24 displays Vector ECG with the plane of ALL and the loop of ALL. A- Vector ECG of Frontal (F) B- Vector ECG of Horizontal (H) C- Vector ECG of Sagittal (S) D- Average templates of X, Y and Z leads. Double-click on this figure to display the magnified average template. You can drag these lines marked P1, P2, Q, S, T1 and T2 on the wave. With the change of the line position, the corresponding parameter values change.