Patient Safeguard System™ Installation Tool Directions for Use For Use With FAS1543 (PSS Installation Tool) DOC8022 Revision A
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. Table of Contents Reference Documents ................................................................................................................................................3 Abstract .......................................................................................................................................................................3 Revision History ........................................................
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. 5.3 Turning the Display Back On ......................................................................................................................... 11 5.4 Beacon Communications Testing .................................................................................................................. 11 5.5 Testing a Beacon’s Bentley Communications ................................................................................
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. 1.0 Introduction to the Patient Safeguard System (PSS) The Patient Safeguard System is a state of the art wireless hand hygiene reminder system designed to enable best practices for proper hand hygiene in a healthcare facility.
Patient Safeguard System Installation Tool Directions for Use System A System B Page 5 Ecolab, Inc.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. 2.0 Purpose of the PSS Installation Tool The purpose of the PSS Installation Tool (FAS1543) is to ease the tasks of installing and maintaining the Patient Safeguard System™ (PSS). The Tool displays the address of all Beacons, but for more modern Beacons (System B), it can also display the firmware revision, battery level, range setting, and the bed / stretcher setting.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. Menu Key: The Menu key is used to enter the Tool’s Menu. Select Key: The Select key has three functions. The first function is to select a Menu item when in the Menu. The second function is to turn the display on to read the information on the Display when in the “Beacon”, “Bentley”, and “Range Adjust” modes. The Select key is also used to begin a test in “Badge” mode.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. 4. Read the Tool’s Display; a. If the Display reads “Both Pass”, the AC Sensor Fails. b. If the Display reads “Antenna Pass”, the AC Sensor Fails. c. If the Display reads “AC Sensor Pass”, the AC Sensor is good. 4.6.2 Test the Antenna (FAS1540) To test an Antenna, please use the following steps 1. Place the Tool in the Peripheral Test Mode. 2. Connect the Antenna to the Tool. 3. Read the Tool’s Display; a.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. in some manner. The yellow LED will light if the 125 KHz signal is found to be correct. The red LED will light when the Beacon and Tool correctly complete a communications cycle (both the 125 KHz and 2.4 GHz radios have worked correctly). The Tool will then display the Beacon’s Address. If the radios are active, the Tool’s display may be too dim to see.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. Badge Battery Percentage Badge Address Tool Mode Results of Badge Communications Test Figure 6 – Badge Mode Display, System A 4.9.7 Network Mode Network mode is used to test the integrity of the Bentley Network in a facility. When in this mode, pressing the Select key will cause the Tool to transmit a PSS Event over the Bentley Network, the green LED will light letting the user know that a message has been sent.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. 5.0 Operation with System B 5.1 Boot Post When the power is turned on the Tool will display the message “Booting”, all three indicator LEDs will flash in sequence as the Tool performs an internal power on self-test, and the audible alert will beep for about 1 second. This is normal and provides feedback that the display, all three LEDs, and the audible alert are functioning properly.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. used to turn the display on allowing the Beacon’s current Bed/Stretcher mode to be read. The Up and Down keys also turn the display on allowing the new mode to be read. 5.8 Testing Peripherals The Tool can be used to verify the operation of the AC Sensor (FAS1542). To test the AC Sensor, please use the following steps. 1. Place the Tool in the Peripheral Test Mode. 2. Connect the AC Sensor that is to be tested directly to the Tool.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. have worked correctly). The Tool will then display the Beacon’s address, the Beacon’s battery level as a percentage, and the Beacon’s firmware revision. If the radios are active, the Tool’s display may be too dim to see. To correct this press the Select key and the Tool will turn off the radios and turn the display on for 5 seconds. 5.11.
Patient Safeguard System Installation Tool Directions for Use Beacon Address Ecolab, Inc. Beacon Mode Tool Mode Figure 12 – Display in Bed/Stretcher Mode, System B 5.11.5 Peripheral Test Mode The Peripheral Test Mode is used to test the AC Sensor. Test Results Tool Mode Figure 13 – Display of a Passing AC Sensor, System B To test the AC Sensor, connect it to the Tool. If it passed, the Tool will display ”AC Sensor Pass”. If the AC Sensor failed, the top line of the display will remain blank. 5.11.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. Tool Battery Level Tool Firmware Revision Figure 15 – Display in Tool Info Mode (First Screen), System B Antenna Resonant Frequency Capacitor Used Axis Identification Figure 16 – Display in Tool Info Mode (Second Screen), System B Selecting Tool Info mode will allow the viewing of information about the Tool. The first screen displays the Tool’s firmware revision and battery level as a percentage.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. Appendix A - Certification and Safety Approvals FCC Statement NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc. Appendix B - Warranty This device is warranted against defective materials and workmanship for two years from the date of purchase. Equipment covered by this warranty will be repaired or replaced in the United States and Canada, WITHOUT CHARGE, except for shipping and handling, by our Factory Service Center.