SPECIFICATIONS
Dimensions: 2-29/32˝ (73.82 mm) x 2- 7/16˝
(61.93 mm).
Weight: 6 oz. (170 g).
Rated Total Pressure: 50 psig (3.445 bar) surge;
30 psig (2.067 bar) continuous to either pressure
connection.
Ambient Temperature Range: 20°F to 120°F
(-6.67°C to 48.9°C).
Accuracy: ± 5% of FS at 70°F (21.1°C).
Connections: Standard, barbed for 3/16˝ I.D. tubing;
optional,1/8˝ NPT(M).
Housing: Glass-lled nylon, polycarbonate lens.
Finish: Black.
Standard Accessories: (2) 4-40 x 1-5/8˝ mounting
studs, (2) 4-40 hex nuts, (1) .050˝ hex allen wrench,
(1) panel mounting bracket.
Mounting Orientation: Diaphragm in vertical position.
Consult factory for other position orientations.
Agency Approvals: Meets the technical requirements
of EU Directive 2011/65/EU (RoHS II).
Series 2-5000 Minihelic
®
II Differential Pressure Gage
has a clean design, small size, low cost and sufcient
accuracy for all but the most demanding applications.
With housing molded from mineral- and glass-lled
nylon and a lens molded from polycarbonate, this gage
will withstand rough use and exposure, as well as high
total pressure up to 30 psig (2.067 bar). Over-pressure
is accommodated by a blow-out membrane molded in
conjunction with the diaphragm.
INSTALLATION
1. Select a location free from excessive vibration and
where ambient temperature will be between 20°F to
120°F (-6.7°C to 49°C). Sensing lines may be any
length necessary without affecting accuracy.
However, long runs of tubing will dampen readings
slightly and cause a minor increase in response time.
If pulsing pressure or vibration cause excessive
pointer oscillation, please contact factory for ways to
provide additional damping.
2. This gage is calibrated and zeroed in the vertical
position at the factory. If the gage is used in any other
position, it must be re-zeroed each time the position
is changed. Gages with ranges under 5 inches
w.c. (1.24 kPa), or the equivalent, should be used
only in the vertical position unless special calibration
was specied when ordering.
Use only with air or compatible non-
corrosive gases.
CAUTION
[2] Ø5/32 HOLES
IN PANEL FOR SURFAC
E
MOUNT
HIGH PRESSURE
PORT Ø7/16 [11.11]
HOLE IN PANEL
FOR SURFACE MOUNT
SURFACE MOUNT HOLES
LOW PRESSURE PORT
Ø7/16 [11.11] HOLE IN PANEL
FOR SURFACE MOUNT
9/32
[6.96]
1-17/32
[38.79]
2-3/8 [60.38]
Ø2-29/32
[73.81]
1 [25.40]
2-21/64
[59.18]
Ø2-19/32
[65.62]
®
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
dwyer-inst.com
e-mail: info@dwyermail.com