Operators Manual SmartSystem SITE L Datalogger
Table of Contents 1 Safety instructions ....................................................................................................................... 4 2 Quick Guide ................................................................................................................................. 5 2.1 Quick Start if configuration is already loaded on the device................................................... 5 2.2 Quick Start if configuration needs first loaded to the device ..............
13.5.2 Upload data manually of read out data to Berlinger SmartView software (version 6.0 and higher)..................................................................................................................................... 18 13.5.3 Return device ................................................................................................................. 18 14 Where to place SmartMonitor SITE L ........................................................................................
1 Safety instructions Intended use Temperature • Temperatures outside the operating range may damage the battery. For the operating range, see the information in the technical data sheet. IP Protection • Limited protection against dust ingress and protection against splashing water from all directions. Air Pressure • Excess pressure or vacuum can damage the device. Do not vacuum when used for air cargo. Mechanical force • Avoid violent shocks and impacts.
2 Quick Guide 2.
2.
3 Device view and description No. Name Description 1 Display Multi-information display 2 USB-connector To charge the battery or download the data manually 3 Confirmation/Status LED LED (small lamp) indicates if use of multifunctional button was successful or not 4 Cover for ext.
4 Display Icon Name Description 1 Battery status Indicates the battery status 2 Battery charging Indicates whether external power has been plugged in 3 Alarm status 4 Logging Modus OFF → SmartMonitor SITE L is ready to start REC → SmartMonitor SITE L is logging → SmartMonitor SITE L has stopped 5 7-digit display Information line (see more in section 7-digit display) 6 External readout Device ready for readout via USB or Bluetooth 7 Warning Symbol Additional warning for function 8 Cl
4.1 7-digit display The 7-digit display gives additional information about the status or other data. In normal logging operation it displays the temperature. 5 Multifunctional button SmartMonitor SITE L device has one multifunctional button to operate. 1 x long press → Start / Stop the device 3 x short press → Setup – Load configuration 1 x short press → Information • Long press to multifunctional button One long press (> 5 sec.) to multifunctional button will start or stop the device.
6 Confirmation LED Device has an LED which gives an indication if the use of multifunctional button was successful or if a task has been terminated. 1 x short white flash → Confirmation of an action or task 1 sec. green flash → “Cancel Message” Short press has released too early.
8 Humidity and pressure sensor SmartMonitor SITE L has a built-in humidity and atmospheric pressure sensor which can be enabled or disabled in the Berlinger SmartView software (version 6.0 and higher) configuration. The sensor is located on the left side of the device.
10 Configuration The configuration for the SmartMonitor SITE L device can be defined via the Berlinger SmartView software (version 6.0 and later). SmartMonitor SITE L can be ordered pre-configured with a customer desired configuration or with a standard configuration. If a new configuration is to be loaded, it is configured in the Berlinger SmartView software (version 6.0 and higher) and then assigned to the device.
11 Start/Stop 11.1 Start To start the device, press the multifunctional button once >5 seconds or as long till display shows “SETUP” or the green LED lights for at least 2 sec. The following information will be displayed on the screen. While the information “***START***” will be displayed as a ticker, keep pressed the multifunctional button and hold the multifunctional button for at least 5 seconds. As confirmation the green LED will flash for at least 2 sec.
11.2 Stop SmartMonitor SITE L device can only be stopped via Berlinger SmartView software (version 6.0 and later). As soon as the stop command has been entered via SmartView, the device will be stopped during the next synchronization between SmartView and SITE L and the information DONE will be displayed on the 7-segment display on the device. 12 Start delay For SmartMonitor SITE L device you have the possibility to configure in Berlinger SmartView software (version 6.
12.2 Start SmartMonitor SITE L with scheduled start If a scheduled start is configured a counter in days:hours:minutes (e.g., 01d14:32 = start in 1 day 14 hours and 32 sec) will be displayed on the 7-segment display. By a short press to the multifunctional button, the information SCHEDULED START will show as ticker on the display with above information about the time till the device will start recording.
13 Usage Process SmartMonitor SITE L is mainly used for monitoring the storage of temperature-sensitive products. The device has various configuration options. The device states are visualized below and described in more detail in the following chapters. The usage process may vary depending on the configuration. 13.1 Preparation 13.1.1 Display check To check the display, press once short the multifunctional button, the display will show the status of the device. 13.1.
13.2.1 WIFI reset If device has no WIFI connection, press the multifunctional button quickly 5 times to reset the connection . 13.2.2 Download Configuration If no configuration is already loaded on the device, press the multifunctional button quickly 3 times . The configuration will be loaded. See chapter 5 multifunctional button Configuration of device can be done with Berlinger SmartView software (version 6.0 and higher). 13.2.
13.4.1 Assess state of goods and information User has the possibility to asset state of the monitoring and get information on the device display by pressing the multifunctional button once . 13.5 Stop SmartMonitor SITE L device can only be stopped via Berlinger SmartView software (version 6.0 and later).
14 Where to place SmartMonitor SITE L Berlinger recommends placing the SmartMonitor SITE L as close as possible to the product to be monitored so that the temperature can be recorded as accurately as possible. If the temperature is to be measured in a refrigerator or freezer, it is recommended to do this with an external sensor. This allows the temperature to be read from the outside on the display. Furthermore, the WIFI connection inside a fridge is poor or not possible at all.
15 External Sensor Use only original Berlinger external sensors for SmartMonitor SITE L. Make sure that the sensor is properly connected to the device and the connector is secured (tightened by hand). Do not tighten the connector with pliers or any other tool, otherwise the connector may be damaged. The cable of the external sensor should be placed in such a way that it does not interfere with the daily handling of the personnel or the cable can be disconnected.
15.1 Connect an external sensor to SmartMonitor SITE L device To connect an external Berlinger probe to a SmartMonitor SITE L remove the grey cover at the right site from device. Remove the grey cover from the SHIPL device Plug in the connector by checking the mark at the connector and the plug of the SHIP L device. Manually tighten the connector.
15.2 Multi Sensor Adapter With SITE L device it is possible to connect a total of 4 external sensors with the help of the Multi Sensor Adapter. For this, connect the Multi Sensor Adapter to the SITE L device and make sure that the connector makes contact in a secure way. The configuration for this Multi Sensor Adapter is done by Berlinger SmartView (version 6.0 and higher).
16 Troubleshooting Problem Blank display Effect No information visible Device cannot be started Display shows LONG PRESS Solution Connect the device to USB charger to charge the battery. Press the multifunctional button at least 5 second. LOW BATTERY LOW:BATT will be shown on the 7-segnent display and battery symbol is blinking Connect the device to power to charge the battery. BATTERY EXPIRED BAT:EXPIRED will be shown on the 7-segment display and the battery symbol is blinking.
MANUAL SYNC MANUAL SYNC will be shown on the 7-segment display, the cloud symbol crossed out circle is displayed and USB and Bluetooth sign is blinking. Configuration is needed. Manual SYNC Manual SYNC will be shown on the 7-segment display as a ticker information and at the cloud symbol crossed out circle is displayed . Manual upload of data is needed. LOW -.-°C LOW -.-°C will be shown on the 7-segment display and the exclamation mark is blinking.
17 Disposal Electronic devices are recyclable and do not belong in household waste. At the end of its lifetime, dispose of the product in accordance with the law according to your country or return it to Berlinger according Berlinger RMA process. 18 Certification SmartMonitor SITE L device has following certifications: • • • CE / RED RTCA DO-160 FCC / ISED 18.
• • • • Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. Radiofrequency radiation exposure Information: This equipment complies with FCC and ISED radiation exposure limits set forth for an uncontrolled environment.