8/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc Introduction Welcome to the world of ultrasonic scaling as experienced with a Cavitron® ultrasonic scaling system, the original brand of ultrasonic scaling systems. Your decision to add the Cavitron® RF™ Ultrasonic Scaler from DENTSPLY Professional to your practice represents a wise investment in good dentistry.
8/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc Section 1: Indications For Use Ultrasonic procedures: All general supra and subgingival scaling applications. • Periodontal debridement for all types of periodontal diseases. • For patients with a history of sensitivity to ultrasonic scaling. • Endodontic procedures. Section 2: Contraindications and Warnings 2.1 Contraindications • Ultrasonic Systems should not be used for restorative dental procedures involving the condensation of amalgam. 2.
8/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc 3.2 Precautions for Ultrasonic Prophylaxis Procedures • For optimum performance use only inserts manufactured by DENTSPLY® Professional. • Like a toothbrush, ultrasonic inserts “wear” with use. Inserts with just 2 mm of wear lose about 50% of their scaling efficiency. In general it is recommended that ultrasonic inserts be discarded and replaced after one year of use to maintain optimal efficiency and avoid breakage.
8/16/2005 • • • CavitronRF Scaler - DFU Draft 08-01-05.doc A manual shut-off valve on the dental water system supply line should be used so that the water can be completely shut-off when the office is unoccupied. In addition to the water filter supplied, it is recommended that a filter in the dental water system supply line be installed so that any particulates in the water supply will be trapped before reaching the system.
8/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc 5.8 Footswitch Battery Replacement (Picture) • Turn footswitch over and using a Philips screwdriver, carefully remove battery cover screw and battery cover. • Remove used batteries and install two new “AA” batteries as shown. Discard used batteries in accordance with local, state and regional regulations. • Replace battery cover and screw and hand tighten with Philips screwdriver. 5.
8/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc Handpiece Holder Securely holds the system’s Handpiece (with or without insert) when the system is not in use. Also holds cable connector when handpiece is not installed. Lavage Control (Flow Adjustment) See Section 6.3 Foot Control See Section 6.5 6.2 Information Center Graphics Displays and Controls (Picture) Power – (Display) Lights when the main Power ON/OFF Control Switch is ON.
/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc 6.4 Cavitron® 30K™ Ultrasonic Inserts (Picture from current DFU’s) O-Ring Provides a watertight seal when the insert is fully seated in the handpiece. O-ring should be replaced when worn. Connecting Body Transfers and amplifies mechanical motion of stack to insert tip. Insert Tip Shape and size of tip determines access and adaptation. Insert Marking Manufacturer, Date (YYMM), Frequency, Insert Type. E.g.
8/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc Section 8: Techniques for Use 8.1 Patient Positioning (Picture) For optimal access to both the upper and lower arches, the backrest of the chair should be adjusted as for other dental procedures. This assures patient comfort and clinician visibility. Have the patient turn his/her head to the right or left. Also position chin up or down depending upon the quadrant and surface being treated.
8/16/2005 3. 4. 5. 6. CavitronRF Scaler - DFU Draft 08-01-05.doc Set the Power Adjustment knob to the minimum setting. Set the Lavage Control on the handpiece cable to maximum. Hold the Handpiece (without an insert installed) upright over a sink or drain. Activate the Purge button with water flow set to maximum. • The Purge button will light for two minutes indicating proper activation of the purge function.
8/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc 9.2 Weekly Maintenance End of Week Procedures If the system is used with a DualSelectTM Dispensing System, follow the end of the week procedures listed in the DualSelect Dispensing Systems Directions for Use manual. 9.3 Water Line Filter Maintenance When the water line filter becomes discolored, the filter should be replaced to prevent reduced water flow to the Cavitron® RF™ Ultrasonic Scaler.
8/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc 10.2 Technical Support and Repairs For technical support and repair assistance call DENTSPLY Professional Cavitron CareSM Factory Certified Service at 1-800-989-8826 Monday through Friday, 8:00 A.M. to 5:00 P.M. (Eastern Time). For other areas, contact your local DENTSPLY® Professional representative. Section 11: Warranty Period The CavitronE RF™ Ultrasonic Scaler is warranted for TWO YEARS from date of purchase.
8/16/2005 CavitronRF Scaler - DFU Draft 08-01-05.doc Section 14: Disposal of Unit • Accordance with local and state laws.