Cordless Prophy System DIRECTIONS FOR USE Mode d’emploi Instrucciones de uso Gebrauchsanweisung Istruzioni per l’uso Указания по применению Please read carefully and completely before operating unit. Prière de lire attentivement et complètement avant la première utilisation de l’appareil. Por favor lea cuidadosamente y en su totalidad antes de operar la unidad. Bitte vor Inbetriebnahme der Einheit sorgfältig und vollständig durchlesen.
TABLE OF CONTENTS OVERVIEW . . . . . . . . . . . . . 1. INDICATIONS FOR USE . 2. CONTRAINDICATIONS . 3. WARNINGS . . . . . . . . . . . . . . . .3 . . . . . . .3 . . . . . . .3 . . . . . . .3 4. PRECAUTIONS . . . . . . . . . . . . 5. ADVERSE REACTIONS . . . . . . 6. MIDWEST® RDH FREEDOM™ CORDLESS PROPHY SYSTEM DESCRIPTION . . . . . . . . . . . . 7. UNPACKING THE SYSTEM . . . 8. INFECTION CONTROL PROCEDURES . . . . . . . . . . . 9. SYSTEM SETUP . . . . . . . . . . . 10.
Overview The MIDWEST® RDH Freedom™ Cordless Prophy System offers a cordless design that eliminates cord drag and allows clinicians easy access and comfort during prophylaxis procedures. Quieter than traditional hygiene and low speed handpieces, the system offers all-day battery life while delivering the same performance you would expect from a corded handpiece. The system includes an autoclavable outer sheath for infection control. 1. Indications for Use 2.
• • There are no user-serviceable items in the inner module, power supply, outer sheath, foot pedal or charging base. Opening any of these units may result in unsafe operation and will void the warranty. • Under Industry Canada regulations, this radio transmitter may only operate using an antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry Canada.
6.
7.
Instructions for Sterilizing The Outer Sheath And Handpiece Cradle WARNINGS These instructions are for use ONLY on the outer sheath and handpiece cradle. All other parts of the system should be disinfected according to the procedures in the “Disinfection” section. The outer sheath and handpiece cradle for the MIDWEST® RDH Freedom™ Cordless Prophy System are not sterile upon receipt and must be sterilized prior to use in accordance with the following instructions.
Instructions For Sterilizing The Outer Sheath And Tray Cradle, Cont. MAINTENANCE Visually inspect to ensure that all contamination has been removed. Check for distortion, damage or wear. Discard damaged, worn or corroded instruments. STORAGE To maintain sterility, instruments should remain bagged until ready for use. MANUFACTURER CONTACT In the United States, contact DENTSPLY Professional Customer Service Technical Support at 800-989-8826.
Instructions For Disinfecting All Other Parts (Charging Base, Inner Module, Foot Pedal and Power Supply), Cont. DRYING When cleaning, wipe surfaces dry with a clean cloth. To achieve disinfection, allow surfaces to air dry. MAINTENANCE Visually inspect to ensure that all contamination has been removed. Visually inspect power supply and cords for damage.
.3 Place the inner module or handpiece on the charging base, aligning the charging contacts on each. Charging Contacts 10.4 Refer to the figure and table below to determine the charging progress for the inner module. Indicator Lights Degree of Charge Scrolling orange Less than 50% Scrolling green 50-90% Solid green Greater than 95% NOTE: • After one week of non-use, both the handpiece and the foot pedal enter an enhanced battery-saving mode.
.5 Invert the inner module and place it on the charging base as shown. The LED shown below will blink green if the position is correct for synchronization. LED 12. Preparation for Use 12.1 Position the foot pedal for use, ensuring that the floor is level. 12.2 Ensure that the outer sheath has been sterilized according to the Infection Control Procedures (section 8). 12.3 When it is sufficiently charged, insert the inner module into a Disposa-Shield®, piercing the shield just enough to expose the tip.
13. Operation 13.1 The handpiece powers up when it detects motion, i.e. when the user picks it up. If not in use, it will shut down again after a minute. 13.2 Load the DPA with prophy paste and depress the foot pedal slowly to avoid splatter. The DPA (motor) will only rotate in one direction. 13.3 Adjust pressure on the foot pedal to control the rotation speed of the DPA throughout the procedure. 13.4 Follow standard prophylaxis procedures as you would with any corded prophy device. 13.
15. Specifications Power Supply manufacturer: Power Supply model number: Ault (SL Power) MW170KB0502B3 AC Input Continuous (100-240 VAC) AC Input Current Less than 0.7A AC Input Phases Single AC Input Frequency 50-60 Hz DC Output Power 8W DC Output Voltage and Current +5VDC at 0-1.
17. Symbol Identification The following standard symbols appear on the device label. Class II Equipment Class II Equipment Type B applied part MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL-2601-1/60601-1, CAN/CSA C22.2 NO.601.
18.Disposal of Unit U.S. - Dispose of the system components in accordance with state and local laws. EU - Dispose of in accordance with the Waste Electrical and Electronic Equipment Directive 2002/96/EC of the European Parliment and the Council of the European Union. 19. Troubleshooting Problem Solution DPA is not revolving 1. Ensure the outer sheath and the DPA are snapped together securely. 2. Ensure that the Inner Module and the Outer Sheath are snapped together securely. 3.
Difficulty removing outer sheath from inner module 1. Check the components for gross debris 2. Hold the handpiece outer sheath securely and twist the inner module 3. Inspect parts for wear 4. Call Technical Support if necessary Charging base does not drain liquids Clear any debris from the hole at the bottom of the charging base. Handpiece does not charge 1. Clean the charge contacts on the handpiece and charging base, using one of the approved cleaning solutions described in section 8. 2.
20.
21. Limited Warranty The DENTSPLY Professional MIDWEST® RDH Freedom™ Cordless Prophy System is designed exclusively for dental use and this warranty is not applicable to other uses. This warranty extends to Midwest RDH Freedom system purchased from an authorized DENTSPLY distributor, and only to the original purchaser. The system is made up of four significant assemblies, the Inner Module, Metal Outer Sheath, Charging Base and Wireless Foot Pedal.
Manufactured by: DENTSPLY Professional DENTSPLY International 1301 Smile Way York, PA 17404-1785 USA DENTSPLY DeTrey GmbH De-Trey-Str. 1 78467 Konstanz Germany Imported and Distributed by: DENTSPLY Canada Woodbridge, Ontario L4L 4A3 PN: 90714 Rev.