Menu Functions A Delivery must be stopped before changing the cassette message displays. A Delivery Stopped Attention Alarm displays. 4. Press 3. Press to confirm. Or to cancel. After confirmation, a Sending request to the pump message displays. 142 to silence the attention alarm. 5. Press to acknowledge the Pump delivery has stopped Attention Alarm. DKPI-09153-001-Rev. 1.
Menu Functions A Reset the Pump to defaults? Pump will be disconnected! message displays. 6. Press to confirm. Or to cancel. 7. Press 8. Press to silence the attention alarm. . After confirmation, a Sending request to the pump message displays. DKPI-09153-001-Rev. 1.
Menu Functions System Information The System Information displays the Remote, Radio, Command, Supervisor, and Pump Radio software versions. 1. Press the Side Button to wake up the remote. 2. Unlock the screen by pressing each front button from left to right. The remote will display Searching and the last connected pump's serial number until the pump is paired to the remote. See "Pair to a Spare Pump" on page 118. 3. Use the down arrow 4. Press 7. Press 144 . 5. Press the down arrow Information. 6.
to scroll through the software versions and press view more information about the selected software. to Return The Return selection is used to go back to the previous menu. You can also press the side button on the remote to return to the previous menu. 9. Press to return back to the Software Versions list. DKPI-09153-001-Rev. 1.4 145 Menu Functions 8.
Menu Functions This page intentionally left blank. 146 DKPI-09153-001-Rev. 1.
166 Pairing Lost Attention Alarm 167 Pump Idle Attention Alarm 168 Pump Battery Low Attention Alarm 169 About Alarms and Attention Alarms 148 Remote Battery Low Attention Alarm 170 Battery Depleted Alarm 149 Request Refused Attention Alarm 171 Cassette Depleted Alarm 150 Software Version Error Attention Alarm 172 Cassette Removed Alarm 151 Tech Attention Alarm 173 Cassette Problem Alarm 153 Walkaway Attention Alarm 174 Occlusion Alarm 154 Pump Error Alarm 158 Pump Failure Alar
Alarms and Attention Alarms About Alarms and Attention Alarms Alarms and Attention Alarms appear on your remote to inform you of problems you need to address. Alarms have a flashing red and white header, stop delivery, are urgent and should be addressed right away. Attention Alarms have a flashing yellow header, do not interrupt delivery, and should be addressed as soon as possible.
Alarms and Attention Alarms Battery Depleted Alarm When the pump battery is too low for the pump to deliver medication, a Battery Depleted Alarm displays. The top banner of the screen will flash and the remote will beep. 1. Press to silence the alarm. 2. Disconnect the infusion set tubing from your catheter. 3. Press to acknowledge the alarm. 4. Press to acknowledge the Delivery Stopped Attention Alarm. 5. Change the cassette and pump battery. See "Cassette Change" on page 59. DKPI-09153-001-Rev. 1.
Alarms and Attention Alarms Cassette Depleted Alarm When your cassette needs to be changed, a Cassette Depleted Alarm displays. The top banner of the screen will flash and the remote will beep. 1. Press to silence the alarm. 2. Disconnect infusion set tubing from catheter. 3. Press to acknowledge the alarm. 4. Press to acknowledge the Delivery Stopped Attention Alarm. 5. Change the cassette and pump battery. See "Cassette Change" on page 59. 150 DKPI-09153-001-Rev. 1.
Do not remove the pump from the used cassette until after you have disconnected the infusion set tubing from your body. Disconnecting the pump before you disconnect the infusion set tubing from your body could lead to unintended medication delivery. 1. Press to silence the alarm. 2. Disconnect infusion set tubing from catheter. 3. Press When the pump detects that the cassette is removed without being commanded to stop delivery, a Cassette Removed Alarm displays.
Alarms and Attention Alarms b. Wait for the Self Test to complete. c. Reattach the infusion set tubing to the catheter and press and hold the blue pump button or the restart delivery. on the remote to 9. Select No to the New Cassette? question, then press 152 to start delivery. DKPI-09153-001-Rev. 1.
When your cassette is broken, a Cassette Problem Alarm displays. 5. Change the cassette and pump battery starting from the Cassette Change with Remote step"Disconnect Cassette from Pump and Dispose" on page 66. The top banner of the screen will flash and the remote will beep. 1. Press to silence the alarm. 2. Disconnect infusion set tubing from catheter. 3. Press to acknowledge the alarm. 4. Press to acknowledge the Delivery Stopped Attention Alarm. DKPI-09153-001-Rev. 1.
Alarms and Attention Alarms Occlusion Alarm When your pump has detected that it has under delivered by 10 µL or that the pumping chamber hasn't fully emptied for 10 successive attempts, which ever occurs first, an Occlusion Alarm displays. When there is an Occlusion Alarm, disconnect from the pump using the quick disconnect feature of the infusion set before investigating the cause of the Occlusion Alarm. Failure to do so may lead to an unintended Bolus and lead to harm.
3. Press to acknowledge the alarm. DKPI-09153-001-Rev. 1.4 4. Press to acknowledge the Delivery Stopped Attention Alarm. 5. Disconnect the cassette from the pump. 6. Remove and reinsert the pump battery into the pump. 7. Listen for the confirmation beep on the pump. 8. If you found a kink in the infusion set tubing and corrected it you can reuse the infusion set tubing and cassette. Follow Step a through Step f, otherwise skip to Step 9: 155 Alarms and Attention Alarms 1. Press to silence the alarm. 2.
Alarms and Attention Alarms a. Disregard the message on the remote telling you to fill a new cassette with medication, prime and install. You will not need to do that since the Occlusion was a kink. b. Reconnect the pump to the used cassette. c. Wait for the Self Test to complete. d. Reattach the infusion set tubing to the catheter and press and hold the blue pump button or the restart delivery. 156 on the remote to e. Select No when the New Cassette? question appears. f.
Alarms and Attention Alarms 12. Use the down arrow to highlight Yes and press to start delivery. DKPI-09153-001-Rev. 1.
Alarms and Attention Alarms Pump Error Alarm When the pump encounters an error, a Pump Error Alarm displays. The top banner of the screen will flash and the remote will beep. 1. Press to silence the alarm. 2. Disconnect infusion set tubing from catheter. 3. Press to acknowledge the alarm. 7. Listen for the confirmation beep on the pump. 8. Disregard the Fill new cassette with X.X mL, prime, and install message if you want to use your existing cassette. a. Reconnect the pump to the used cassette. b.
When the pump is broken, a Pump Failure Alarm displays. You will need to switch to your spare pump and may need to request a replacement from your specialty pharmacy. The top banner of the screen will flash and the remote will beep. 4. Press to acknowledge the Delivery Stopped Attention Alarm. 5. Disconnect the pump from cassette. 6. Remove the battery from the pump. 7. Gather your spare pump and insert the pump battery. 8. Listen for the confirmation beep on the pump. 9.
Alarms and Attention Alarms Basal Not Started Attention Alarm Adjust Pump Attention Alarm When an idle pump and remote interface have not communicated for some time, the pump needs to check with the remote interface before delivery can be started. If you try to start delivery in this case, the pump makes the attention alarm sound and the remote interface will display a Basal Not Started Attention Alarm. When the pump is having difficulty delivering your dose an Adjust Pump Attention Alarm displays. 1.
The Delivery Stopped Attention Alarm displays to ensure you are aware and reminded that delivery has been stopped. This attention alarm occurs after every alarm and in response to the patient stopping the pump, such as during the Cassette Change process. If a delivery stopped attention alarm appears outside of an alarm, attention alarm, or cassette change condition follow these steps to restart delivery. 1. Press to silence the attention alarm. 2. Press to acknowledge the attention alarm. 3.
Alarms and Attention Alarms Depletes Soon Attention Alarm Excessive Noise Attention Alarm When the cassette will be depleted within 2 hours, a Depletes Soon Attention Alarm displays. When your surrounding environment is excessively noisy, an Excessive Noise Attention Alarm displays to alert you that delivery may be impacted. 1. Press to silence the attention alarm. 2. Press to acknowledge the attention alarm. 3. When the cassette is depleted, change the cassette and pump battery.
When your remote did not receive a response to the last setting change command it sent to the pump, a Message Timeout Attention Alarm displays. This may be caused by wireless communication interference. 1. Press 2. Press to acknowledge the attention alarm. 3. Place the pump and remote within 12 inches of each other until communication is restored. This is indicated when the home screen's top banner is green or blue. 4.
Alarms and Attention Alarms No Communication Attention Alarm When the pump and remote have not communicated with each other for 2 hours, a No Communication Attention Alarm displays with a Verify pump is delivering message. This may be caused by wireless communication interference. 1. Press 2. Press to acknowledge the attention alarm. 3. Press and hold the blue pump button until you hear the four-tone delivering sound: a.
Alarms and Attention Alarms 4. If communication cannot be restored between the pump and remote, gather your spare pump and: a. Pair spare pump with remote and start delivery. See "Pair to a Spare Pump" on page 118. b. Contact your specialty pharmacy to obtain a replacement pump. DKPI-09153-001-Rev. 1.