Introduction Remote Front The Front Button Screen Labels displayed on the screen are executed by pressing the coinciding Front Button below the label. DKPI-09153-001-Rev. 1.4 To unlock the remote press the Front Buttons in order from left to right. Once unlocked, the remote screen will stay active for the programmed Screen Timeout minutes before going into sleep mode. See "Screen Timeout" on page 129.
Introduction Front Button Screen Labels Scroll Up / Increase a Value Repeat a Function Scroll Down / Decrease a Value Cancel or Exit a Function Next Screen Access Help Information Silence Alarm Edit a Value Previous Screen Pump Settings Unlock Screen Acknowledge or Accept 36 DKPI-09153-001-Rev. 1.
Introduction Remote Back The remote includes a bay on the back to store and charge a spare pump battery. This battery does not power the remote. Contact your local authorities to determine the proper method for disposing of the durable components when they are no longer needed. These components have electronics that contain lead and Lithium Polymer batteries. The Pump Battery Status Light on the bottom of the remote indicates the pump battery’s charging status when in the bay.
Introduction Home Screen The Home Screen displays details about your pump's medication delivery and any conditions you need to address, such as Alarms or Attention Alarms. See "Alarms and Attention Alarms" on page 147. The color of the Status Bar indicates the remote’s connection to the pump and the current delivery: Connected and Delivering Connected and NOT Delivering NOT Connected Alarm Attention Alarm 38 DKPI-09153-001-Rev. 1.
The Pump connects to a Cassette and pumps medication under your skin at the delivery rate you set using the remote. The pump will continue to deliver medication even when it is not communicating with the remote. Introduction Pump Overview The pump is powered by a rechargeable pump battery that is inserted into the battery bay located on the bottom of the pump. Every three days you must replace the pump battery with a fully charged pump battery.
Introduction Pump with Cassette Attached Remove the Clear Dust Cover The pump is shipped with a Clear Dust Cover attached to protect the pump. 1. Remove the clear dust cover by holding the dust cover in one hand and rotating the pump counterclockwise with the other hand. 2. Separate the pump from the dust cover. 40 DKPI-09153-001-Rev. 1.
How to Wear Your Pump Whenever the pump is not being used, store the pump with the clear dust cover in place. Your Pump can be worn a few different ways: 1. Align the pump-bump with the unlock symbol on the clear dust cover. 2. Holding the clear dust cover in one hand and the pump in the other, push and rotate to the lock symbol on the clear dust cover. DKPI-09153-001-Rev. 1.
Introduction Belt Clip 1. The pump slides into the Belt Clip with the blue button facing inward. 42 2. Slide the belt clip onto your belt or other location on your body that will firmly support the pump. DKPI-09153-001-Rev. 1.
Introduction In Your Pocket Insert the pump into your pocket making sure the infusion set tubing is not pinched. DKPI-09153-001-Rev. 1.
Introduction This page intentionally left blank. 44 DKPI-09153-001-Rev. 1.
Set Up Set Up Contact your specialty pharmacy for assistance with any of the set up instructions. Charge the Pump Batteries 46 Charge the Remote and One Pump Battery 48 Pair Remote and Pump for the First Time 52 Enter Delivery Rate 57 DKPI-09153-001-Rev. 1.
Set Up Charge the Pump Batteries You must charge the Pump Batteries before you can use the pump. A fully charged pump battery can power the pump for 3 days (72 hours). If the pump batteries have been subjected to the extremes of the specified storage temperatures, before charging or using the batteries allow them to cool off / warm up for 20 minutes. Each time you replace a cassette, you should also replace the pump battery with a fully charged battery.
Set Up Avoid touching the electrical contacts on the battery charger when it is connected to a power source. 4. Amber battery charger status light(s) indicate the batteries are charging. Color 3. Insert two rechargeable batteries into the battery charger by aligning the contacts. DKPI-09153-001-Rev. 1.
Set Up Charge the Remote and One Pump Battery You should charge the remote before use. The remote and pump battery in the remote will typically fully charge in 6 hours or less when connected to the supplied dual USB AC adapter. The charging time for the remote may be longer if connected to power sources other than the supplied dual USB wall adapter. 5. Wait until the pump batteries are fully charged before use. A pump battery is fully charged when the battery charger status light(s) are green.
Set Up When the system is used in a loud environment the remote should be kept in a pant or shirt pocket so that the vibratory aspect of an alarm notification can be felt. If the remote has been subjected to the extremes of the specified storage temperatures, before charging or using the remote, allow the device to cool off/warm up for one hour. 1. Plug the second USB cable into the bottom of the remote and the other end of the USB cable into the dual USB wall adapter.
Set Up 2. Remove the clear cover on the back of the remote by pushing down on the tab and pulling forward. While the remote is charging you are also charging the pump battery (indicated by the amber light on the back of the remote). Color Status Green Battery charging is complete Amber Battery is charging Blinking Red Charging Fault 4. Insert the clear cover into the back of the remote. 3. Insert a rechargeable pump battery into the back of the remote making sure to align the contacts.
Set Up 5. To check the remote's battery charge, press and hold the wake up button on the side of the remote. 6. Check the top right of the display screen for the charging symbol on top of the battery symbol, which indicates that the remote is charging. DKPI-09153-001-Rev. 1.4 7. When the battery symbol is full green on the remote screen, the remote is fully charged.
Set Up Pair Remote and Pump for the First Time Pairing the remote with the pump enables the remote to control the pump settings, including the medication delivery rate. 1. Press the side button to wake up the remote. The remote will search for a pump when you power on. Cell phones, Bluetooth devices, RFID readers, and Wi-Fi equipment may affect the communication between the Pump and Remote or interrupt operation of the Pump. Keep the Pump and Remote at least 0.8 m [3 ft] away from these items.
Set Up 2. Unlock the screen by pressing each front button from left to right to unlock the remote screen. 3. Remove the clear dust cover by holding the dust cover in one hand and rotating the pump counterclockwise with the other hand. 4. Lift the dust cover off and store it with your other Remunity supplies. Once unlocked, the screen will stay active for 2 minutes before going into sleep mode. The pump is shipped with a Clear Dust Cover attached to protect the pump. DKPI-09153-001-Rev. 1.
Set Up 7. Place the pump and remote within 2 in (50 mm) of each other during the pairing process. 5. Insert a fully charged pump battery into the pump's battery bay. The pump will be in pairing mode and sound a ready tone (one beep). If you hear a different tone, See "Troubleshooting Pump" on page 189. 8. If your pump's serial number does not show up on screen after a minute, press and hold the pump’s blue button for 10 seconds to activate pairing mode.
Set Up 9. On the remote, use to highlight the pump's serial number and press DKPI-09153-001-Rev. 1.4 . 10. After selecting your pump serial number you will be prompted to confirm the selected pump. Press to confirm.
Set Up A Pairing To... message displays. 11. Upon successful pairing the Paired Pump If pairing fails, see "Pairing Failed Attention Alarm" on page 166 message displays. Press . 12. If the remote is paired to an incorrect pump, press and repeat the pairing instructions from Step 6. 56 DKPI-09153-001-Rev. 1.
Cell phones, Bluetooth devices, RFID readers, and Wi-Fi equipment may affect the communication between the Pump and Remote or interrupt operation of the Pump. Keep the Pump and Remote at least 0.8 m [3 ft] away from these items. See Table 28 on page 223 and See Table 30 on page 226 for more information. Set Up Enter Delivery Rate 1. The remote will prompt you to set a delivery rate (the amount of medication the pump will deliver per hour of use).
Set Up 3. The remote displays a Fill new cassette with X.X mL, prime, and install message. However you will not need to follow the steps to fill the cassette with the amount displayed on the Remote for the Pharmacy-Filled cassettes. 4. Wash your hands, gather cassette change supplies, and follow the cassette change steps starting from "Prepare Priming Aid" on page 70.
Cassette Change with Remote 60 Supplies Needed for Cassette Change with Remote 61 Stop Delivery with Remote 63 Disconnect Infusion Site 65 Disconnect Cassette from Pump and Dispose 66 Replace Pump Battery 68 Prepare Priming Aid 70 Connect and Prime Infusion Set 73 Stop Priming and Remove Priming Aid 75 Connect Cassette to Pump 78 Connect to Catheter 80 Start Delivery 81 Cassette Change Without Remote 85 Replace Pump Battery 87 Prepare Priming Aid 89 Connect and Prime Infusion S
Cassette Change Cassette Change with Remote The Cassette Change is a step-by-step process for replacing the pump battery, the infusion set, and an empty cassette with a filled cassette. 1. Before you start, wash your hands with soap and water. Follow the step-by-step procedure in the order listed to ensure a safe and efficient replacement of the battery, infusion set, and cassette when the pump and remote are in communication.
Cassette Change Supplies Needed for Cassette Change with Remote The following supplies are needed to complete a Cassette Change. Table 4: Components for Cassette Change Remote with One Fully Charged Battery DKPI-09153-001-Rev. 1.
Cassette Change Table 5: Disposables for Cassette Change Disposable Kit Infusion Sets Use only one of the following Infusion Sets • Medtronic Quick-set Infusion Set (MMT-392, MMT393) • Medtronic Silhouette Infusion Set (MMT-373) • Smiths Medical Cleo 90 Infusion Set (21-7230-24, 21-7220-24) Always refer to your specific infusion set instructions. 62 DKPI-09153-001-Rev. 1.
Cassette Change Stop Delivery with Remote Cell phones, Bluetooth devices, RFID readers, and Wi-Fi equipment may affect the communication between the Pump and Remote or interrupt operation of the Pump. Keep the Pump and Remote at least 0.8 m [3 ft] away from these items. See Table 28 on page 223 and See Table 30 on page 226 for more information. 1. Press the side button to wake up the remote. Your delivery rate and the hours remaining on the cassette displays. DKPI-09153-001-Rev. 1.4 2.
Cassette Change 3. Press to select Change Cassette. A Delivery must be stopped before changing the cassette message displays. 4. Press to stop medication delivery. A Sending request to the pump message displays. 64 DKPI-09153-001-Rev. 1.