User Notice 1.Thanks for purchasing Urine Analyzer! Please read the User Manual carefully before using this product. The User Manual which describes the operating procedures should be followed strictly. 2.This manual detailed introduce the steps must be noted when using the product, operation which may result in abnormal.
Statement This user manual contains proprietary information, which is protected by copyright. All rights reserved. Reproduction, adaption or translation, for any part of the manual without prior written permission, is prohibited. Our company assumes no responsibility for incidental or consequential damage in use caused by any error appear in this document. In case of product upgrades, the information contained in this manual is subject to change without notice.
Serial number. Manufacturer. This item is compliant with In Vitro Diagnostic Medical Device Directive 98/79/EC of Dec.27,1998, a directive of the European Economic Community. FCC WARNING This device complies with part 15 of the FCC rules. Operation is subje ct to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
- Increase the separation between the equipment and receiver. -Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -Consult the dealer or an experienced radio/TV technician for help RF Exposure When using the device, ensure that the antenna of the device is as least 20 cm away from all persons.
Content User Notice ................................................................................... I Statement ..................................................................................... II Content ........................................................................................ V Chapter 1 Overview ...................................................................... 1 1.1 Summarization ................................................................ 1 1.2 Precautions for Use .......
4.3 Clean the Paper Tray .................................................... 23 4.4 Disinfection .................................................................. 24 4.5 Waste Disposal ............................................................. 24 4.6 Troubleshooting ............................................................ 24 Appendix ....................................................................................
Chapter 1 Overview 1.1 Summarization Urine test is the most common method for checking disease in clinical test, and checking with urine test strip is the most effective method. urine test strip and urine sample produce a chemical reaction, the color of each test color-area will change after reaction, the corresponding test results can be obtained according to the color change.
1.2 Precautions for Use Note Before using, please read the Manual carefully and strictly operate according to it. Please don't use the accessories not provided by manufacture. Please don't use the analyzer in condition that the test strip is expired or the device is damaged. Please transport, install and operate the analyzer following the User Manual.
1.
Dimension 126mm(L)*73.
Chapter 2 Installation 2.1 Operation Environment As with all precise electronic instruments, the urine analyzer should avoid placing in high temperature and humidity environment for a long time. To get optimal result, please keep relative stable temperature(10℃~30℃) and humidity(≤80%), and the tabletop to be placed the analyzer should be level.
2.3 Appearance and Structure 2.3.1 Front View Figure 2-1 Front view ① LCD display: man-machine conversation window ② Mini USB socket: AC adapter power supply socket, interface of data transmission. ③ Keyboard: 5-key touch control keyboard, operate the analyzer with the buttons. ④ Test paper tray: place the strip to be tested on it.
2.3.2 Back View Figure 2-2 Back view ① Battery cover: it's need to open the battery cover when install and remove the lithium battery; ② Reset key: if necessary, press this button to reset the urine analyzer. 2.4 Power Supply The device can use the built-in rechargeable battery as power supply. Connect the urine analyzer to the AC adapter when the built-in battery power is low or it needs to use the external power.
Warning AC power outlet must be well connected to ground (zero grounding voltage<5V). The AC power must be stable, avoid using the same power together with high-power appliance, and a manostat is recommended to configure. Please turn off the power supply immediately and contact with the maintenance center, when fog, peculiar smell or strange sound was found in device. Hold the adapter itself when you unplug it, rather than the data cable.
Chapter 3 Operation 3.1 Buttons Figure 3-1 Buttons ① ON/OFF button: Long press this button to turn on/off the device. ② Menu button: In the main interface, short press this button to switch current test user; In the main interface, long press this button for 2 seconds to enter the setup interface; In other interface, execute the return operation. OK button: Confirm the current operation; In the ③ main interface, press this button to start test.
⑤ Down direction button: In the main interface, long press this button to move the tray down; in the menu interface, press this button to move the cursor down; in history interface, long press this button to quickly move the cursor. 3.
When the device prompts to calibrate the clock after the self-test is completed, users need to set time themselves. 3.3 Start Test Preparation before test: Urine test strips that matched the device; Urine sample that stored no more than 4 hours; Absorbent paper for sucking residual urine; Protective gloves for preventing contamination.
3) Begin 60-seconds countdown. After finishing the countdown or once again, the device begin to test data. If press OK button there are error messages during the test, follow the prompt, then 4) to continue the measurement; press OK button Measurement is completed, display and store the test result. As figure 3-3. Note The sample number starting from 1 after the first power-on every day, after testing 1 sample, the sample number automatically add 1.
Figure 3-3 Test is completed 3.4 Power Off Recommended to power off as follows: 1) for 3 Under the main interface, long press the direction button seconds to make the paper tray back to the device storehouse, then long press the direction button make the paper tray out of the device. 2) for 3 seconds again to Remove the tray and rinse the bracket with clear water, then blot up the liquid above and below the tray with absorbent paper.
3.5 Setup Menu Figure 3-4 Menu Figure 3-4 shows the device menu interface, it includes historical records, unit, clock, language, factory settings. Select menu option through direction button , press OK button submenu, press the menu button to enter next in any interface to return to the previous interface. 3.5.1 Records Historical records menu interface, as shown in Figure 3-5.
Figure 3-5 Historical records All records Enter the all records query interface, each page shows 10 historical data at most, user can page to see more historical records through , after selecting a record, press OK button direction button see the corresponding record. As shown in Figure 3-6.
Figure 3-6 Historical records User list Query the historical test data of the selected user. Date list Query the historical test data of the selected date. Send The urine analyzer upload all test results to PC through Mini USB interface or bluetooth(apply to BC401BT). Note If transmit data through Mini USB interface, users need to use the data cable which is specified by the manufacturer. Use the PC software which is specified by the manufacturer(optional). Delete Delete all historical data.
The system clock always has some cumulative error, the user should calibrate once every two months. Urine analyzer manages test reports according to the sample number, date and time of test report, please input the date and time according to the facts, otherwise it will lead to disorder of test report management. 3.5.4 Language Set the language of device interface, the device supports both Chinese and English. 3.5.5 Factory Settings Figure 3-7 The interface of inputing password User password: 0000.
Figure 3-8 Factory settings Adjust sensitivity Adjust the sensitivity for the currently selected test strip. During using, the sensitivity can be adjusted when the user wants to increase or decrease the sensitivity of the analyzer. When you set the sensitivity, must be careful, a valid setting is recommended, you can use urine quality control materials or homemade known content quality control substitutes.
After understanding the above, you can securely adjust the sensitivity. The setup interface of adjusting sensitivity is shown in figure 3-9. Figure 3-9 Adjust sensitivity Select the item which need modified, enter the sensitivity adjusting menu of each grads in this item, as shown in figure 3-10.
Figure 3-10 Adjust sensitivity Press direction button press OK button to select grads value which need modified, to enter the modification of current grads value, then modify the corresponding value through direction button , to confirm the modification, press menu button press OK button to cancel the modification. After finishing modification, press menu button to exit the modification of current item. After modifing all item, press menu button settings menu.
3.6.2 Operating Steps of Bluetooth Communication 1. Power On Turn on bluetooth function of computer, mobile phone or other intelligent device(hereinafter referred to as PC), and the Urine Analyzer is ON. 2. Matching PC searchs and matching the bluetooth urine analyzer. Under normal circumstance, it doesn't need to input password during matching(the default bluetooth password is 0000). 3. Connecting After a successful matching, PC can communicate with urine analyzer. 3.6.
Chapter 4 Maintenance 4.1 Maintenance 1. 2. 3. 4. After daily use, the test tray should be took out for cleaning, and the remained urine should be cleaned with absorbent paper or cotton swab in time, to avoid inaccurate result for cross-pollution. Often clean the analyzer with soft cloth to keep it clean. If the surface of the analyzer is very dirty, then wipe it with clean water or neutral cleaning fluid. Do not clean with gasoline, paint dilutions, benzene compounds, alcohol and other organic solvents.
When installing, insert the clean tray into the device from underside, hold the paper tray gently with hand, at this time, press the button for three seconds, then the paper tray will move direction back to the device storehouse. 4.
Please do not clean the calibration comparator block with any solvent. Please do not contact the calibration comparator block with NaOH. 4.4 Disinfection 1) 2) 3) 4) According to one of the following three methods to configure disinfectant: 2% glutaraldehyde solution; 0.05% sodium hypochlorite solution -----1:100 dilution: add 1mL sodium hypochlorite solution(concentration:5%) to 99ml water; Isopropanol (70% -80%), without dilution.
1. Abnormal Motor 2. 1. Abnormal light source 2. Motor rotation is abnormal, check whether there is debris on the paper tray, causing the motor stuck. Low battery, restart after connecting the AC adapter. The light source is strong, weak or damaged. The calibration comparator block is contaminated, please clean it. Please confirm whether the device work under Abnormal strong light source, re-arrange the work backlight environment of the device according to requirements of the manual.
Appendix Grads Table Item cod e URO 1 BLD 2 BIL 3 KET 4 LEU 5 GLU 6 PRO 7 PH 8 Gra ds code 0 1 2 3 0 1 2 3 4 0 1 2 3 0 1 2 3 0 1 2 3 4 0 1 2 3 4 5 0 1 2 3 4 0 1 Special unit Internatio nal unit Conventio nal unit Norm +1 +2 >=+3 ++1 +2 +3 +1 +2 +3 +1 +2 +3 ++1 +2 +3 ++1 +2 +3 +4 ++1 +2 >=+3 5 6 3.3umol/l 33umol/l 66umol/l 131umol/l 10/ul 25/ul 50/ul 250/ul 0umol/l 17umol/l 50umol/l 100umol/l 0mmol/l 1.5mmol/l 4.0mmol/l 8.0mmol/l 15cells/ul 70cells/ul 125cells/ul 500cells/ul 0mmol/l 2.
NIT 9 SG 10 VC 11 2 3 4 0 1 0 1 2 3 4 5 0 1 2 3 4 7 8 9 +1 <=1.005 1.010 1.015 1.020 1.025 >=1.030 ++1 +2 +3 7 8 9 18umol/l <=1.005 1.010 1.015 1.020 1.025 >=1.030 0mmol/l 0.6mmol/l 1.4mmol/l 2.8mmol/l 5.6mmol/l 7 8 9 0.12mg/dl <=1.005 1.010 1.015 1.020 1.025 >=1.030 0mg/dl 10mg/dl 25mg/dl 50mg/dl 100mg/dl Note: 1. The parameter between in the table and the test strip may be different, please refer to Strip Instruction for details. 2.