TABLE OF CONTENT Foreword ................................................................................................ 3 Operator Qualifications ....................................................................... 3 Operator Training ................................................................................ 4 Product Description ............................................................................. 5 Safety Instructions ....................................................................
System Setting .................................................................................. 34 Arm Training ...................................................................................... 35 Leg Training ...................................................................................... 46 Spasm Control................................................................................... 47 Emergency Stop ................................................................................
Foreword Thank you for purchasing DQAPT from our company. This manual has been written for the owners and operators of the DQAPT. It contains general information on the instructions for safety, intended use, working principle, operation, maintenance, trouble shooting, and warranty. In order to maximize the use, efficiency, and working life of your unit, please read this manual thoroughly and become familiar with the controls, as well as the accessories, before operating the unit.
clinical environment. Specialists must have the basic physical and cognitive prerequisites such as vision, hearing, and literacy. Also, the basic function of the upper extremities is expected. Operator Training Operators of the DQAPT must have been adequately trained in using this system safely and efficiently before they operate the unit. A documented, cross-functional review must be performed, as many times as necessary, in order to ensure that instructions can be understood by users.
Product Description DQAPT is a rehabilitative training device that suitable for the active and passive movement of a person’s lower and upper extremities. It has four training modes (STRENGTH, ENDURANCE, SYMMETRY and ROM training mode) and spasm control function. The device can connect the heart rate sensor via Bluetooth to real-time display heart rate during training.
Safety Instructions Symbols 1. Symbols on the medical device Symbols Explanation Manufacturer Date of manufacture Correct Disposal of This Product (Waste Electrical & Electronic Equipment) Statement: Contact the local authorities to determine the proper method of disposal of potentially bio-hazardous parts and accessories. Type B applied part complying with IEC 60601-1 This device emits non-ionizing radiation.
2. Symbols on the package Symbols Explanation This side up The transportation package must be vertical and straight up during transportation. Fragile, handle with care The product inside the packaging could be easily damaged if dropped or handled without care and attention. Keep away from rain The product package should keep out of the rain and not to store it in damp conditions. Temperature limitation The product package should be stored at a temperature between -20 and 55 degrees (centigrade).
Precautionary Definitions The precautionary instructions found in this section and throughout this manual are indicated by specific symbols. Understand these symbols and their definitions before operating this equipment. The definition of these symbols is as follows: Text with a “CAUTION” indicator will explain possible safety infractions that could have the potential to cause minor to moderate injury or damage to equipment.
Warnings and Cautions Please read carefully the user manual before using DQAPT and take care of what follows: Read, understand, and practice the precautionary operating instructions. Know the limitations and hazards associated with using the DQAPT. Observe the precautionary and operational decals placed on the unit. Improper installation, operation or maintenance of DQAPT may result in malfunctions of this unit or other devices.
Do not use in the following circumstance: the mixture of the flammable anesthetic gas and air; the mixture of the flammable anesthetic gas and oxygen or nitrous oxide, or other flammable agents. In case of display failure or other obvious defects, switch the device off immediately, and notify a certified service technician. Adjustments or replacement of components may result in the equipment failing to meet the requirements for interference suppression.
During the process of operation, constantly watch patients and instrument case, always pay attention to whether the abnormal conditions occur. Please pull the plug before move the device and try to avoid vibration during moving the device. If the device has not been used for a long time (≥2 weeks), please check and make sure the device and accessories are in good conditions before use. The DQAPT should be kept out of the reach of children.
and scarves that could get caught or tangled in pedal crank must not be worn (especially during leg training). Shoes with shoe laces must not be worn, either. Before starting leg training, the arm trainer has to be swiveled back so that the patient can hold on the handlebar. Before each leg training, you must check whether the patient’s feet are in the correct position on the LE pedals and have been secured with straps. Only put patient’s feet into the LE pedals while seated.
Please dispose of the equipment and other accessories according to local regulations. FCC Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: 1) this device may not cause harmful interference, and 2) this device must accept any interference received, including interference that may cause undesired operation. Bluetooth® ® The Bluetooth word mark and logos are owned by the Bluetooth SIG, Inc. and any use of such marks by Compass Health Brands, Corp. is under license.
EMC Guidance The DQAPT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided. This device has been thoroughly tested and inspected to assure proper performance and operation. Do not use the DQAPT simultaneously with other therapeutic device (such as microwave), to avoid mis-operation. Do not place or use the DQAPT nearby radio, television, copy machine and fax machine.
Cable Information: Item Cable Length Power cord 2.5 m Manufacturer Ching Cheng Wire Material Co., Ltd. Guidance and manufacture’s declaration – electromagnetic emission The DQAPT is intended for use in the electromagnetic environment specified below. The customer of the user of the DQAPT should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment – guidance The DQAPT use RF energy only for its internal function.
Guidance and manufacture’s declaration – electromagnetic immunity The DQAPT is intended for use in the electromagnetic environment specified below. The customer or the user of DQAPT should assure that it is used in such an environment. Immunity IEC 60601 Compliance test test level level Electromagnetic environment guidance Electrostatic ±8 kV contact ±8 kV contact Floors should be wood, discharge ±2,4,8,15 kV ±2,4,8,15 kV concrete or ceramic tile.
s and °,225°,270° °,225°,270° hospital environment. If voltage and 315°) and 315°) the user of the DQAPT variations requires continued on power 0% UT; 1 0% UT; 1 operation during power supply cycle and cycle and mains interruptions, it is input lines 70% UT; 70% UT; recommended that the IEC 25/30 cycles 25/30 cycles DQAPT 61000-4-11 Single Single from an uninterruptible phase: at 0° phase: at 0° power be powered supply or a battery.
Guidance and manufacture’s declaration – electromagnetic immunity (Continued) The DQAPT is intended for use in the electromagnetic environment specified below. The customer or the user of LGT- DQAPT should assure that it is used in such an environment. Immunity test IEC 60601 test Compliance level level 3 Vrms 150 kHz - 80 MHz 3 Vrms 150 kHz - 80 MHz 6 V rms in ISM bands between 150 kHz and 80 MHz 6 V rms in ISM bands between 150 kHz and 80 MHz 80 % AM at 1 kHz 3 V/m 80 MHz - 2.
Table: Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Immunity Test Maximum Band Distance TEST frequency Service a) Modulation b) power (MHz) a) (m) LEVEL (MHz) (W) (V/m) 385 380 to TETRA 400 Pulse 1.8 0.3 27 390 Modulation b): 18Hz 450 430 to GMRS 460, FM c) ± 5 Hz 2 0.3 28 470 FRS 460 deviation 1 kHz sine 710 704 to LTE Band Pulse 0.2 0.3 9 b) 745 787 13, 17 Modulation : 780 217 Hz 810 800 to GSM Pulse 2 0.
5240 5500 5785 5100 to 5800 WLAN 802.11 a/n Pulse Modulation b): 217 Hz 0.2 0.3 9 NOTE: a) For some services, only the uplink frequencies are included. b) The carrier shall be modulated using a 50 % duty cycle square wave signal. c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.
Clinical Instructions Indications Before a training with the Active Passive Trainer for Upper and Lower Limbs, a correct examination and diagnosis should be performed. Please stay current with the latest developments and medical publications on Active Passive Training for details on contraindications and side effects not known at the time of manufacturing.
10) Limbs need to be braked; 11) Mental disorder; 12) Other patients that doctors think not suitable use this device. Adverse Effects You should stop using the device and consult your doctor if you experience adverse reactions from the device. Possible adverse reactions may include the following: partially increased pain; partially too strong muscle tone reduction; skin injuries.
Technical Specifications General Product Name APT Performer Product Model DQAPT Dimensions (W×L×H) 710×760×1072 mm (28×30×42 inch) Standard Weight 45.7 kg (102 lbs) Interface 8" touch screen Safe Working Load UE pedals (total) 5 kg (11 lbs) LE pedals (total) 10 kg (22 lbs) Handlebar 20 kg (44 lbs) Adjustable Range Supporting module 0-150 mm (0-5.90 inch) Handlebar 180° Display (max.) 270°- 360° Rotation Radius UE pedal 100 mm (3.94 inch) LE pedal 67 mm (2.
Spasm Relief Rate 1-5 Timer (duration time) 1-60 min Electrical Power Supply AC100-240V, 50-60Hz Rated Power 160 VA Mode of Operation Continuous System of Protection Class I, Type B Applied Part Ingress Protection IPX0 Fuse (Main part) F2AL250V (Ф5×20mm) Operation and Storage Condition Environmental Conditions of Temperature: 5 to 40°C Operation: Rel.
Working Principle User Interface Speed detection (PWM) Motor drive Arm motor Resistance detection Leg motor MCU Speed detection 26
Inspection of the Goods Unpacking the Unit The unit is generally delivered with the packaging material supplied by the manufacturer. Proceed as follows: Position the transport packaging so that the arrows are pointing upward. Remove the transport packaging upward. Remove the remaining foam material. Inspections Immediately upon unpacking the unit, perform the following steps: 1. Verify the delivery documents to make sure that the delivery is complete. 2.
No. Item Name Amounts Unit 4 Anti-tipper wheelchair hook 1 piece 5 Fuses 4 pieces 6 Certificate of quality 1 piece 7 Warranty card 1 piece 8 User manual 1 piece Other parts of DQAPT are available as accessory on demand. Visit website www.compasshealthbrands.com to obtain more information.
Overview of the Unit Nomenclature A Host 5 Control system 6 LCD Display (Capacitive touch screen) 7 Emergency stop switch 8 USB port: connect with USB flash disk for software update and training data export B Arm trainer C Leg trainer 9 11 3 Upper limb turntable UE pedal (Upper extremity pedal) Calf support 29
4 13 LE pedal (Leg extremity pedal) Lower limb turntable 1 Handlebar 2 Locking screw knob: for adjusting Arm trainer direction 12 Locking screw knob: for adjusting trainer height D Main structure 10 Supporting module 16 Pedestal: transport wheels 14 Power switch, power socket and fuse 17 Leveling glides: rotate to adjust the table balance 15 Accessories 1.
2. UE pedals with forearm support (optional) 1 Hand Grip 2 Connection part with Arm trainer 3 Snap: fix UE pedal to the Arm trainer 4 Adjustment Knob: adjust the position of grip Please purchase the accessories from Compass Health Brands, Corp.
Interface Introduce Self-inspection Insert the power supply; turn on the switch button in the pedestal of the machine, the system will enter into self-inspection interface. The device performs a self-check when it is turned on. DO NOT approach the movement area of the ARM and LEG trainer of the device.
Main Menu Interface A few seconds, the main menu interface appears on. Item Description Bluetooth icon Indicate that the unit has connected with the Heart Rate Sensor. USB icon Indicate that the USB flash disk has been inserted into the unit and recognized. See Training Date Export as details. Information icon Press into a page shows the relevant application information of the training mode. ARM TRAINING LEG TRAINING Press into ARM TRAINING interface. See Arm Training as details.
Item FINE MOTOR TRAINING FAVORITES SYSTEM SETTINGS Description Press into FINE MOTOR TRAINING interface. See Fine Motor Training as details. Press into FAVORITES interface. See Favorites Interface as details. Press into SYSTEM SETTINGS interface. See System Setting as details. System Setting Press SYSTEM SETTINS in the MAIN MENU to enter this interface, the default training parameters can be reset here. Training parameters are maintained even when the device is turned off.
Table 1: System Settings Parameters Range Step Defaults DEFAULT RPM (r/min) 5-60 1 or 10 10 DEFAULT DURATION (min) 1-60 1 or 10 20 DEFAULT RESISTANCE 0-24 1 or 6 0 DEFAULT GAME DIFFICULTY 1-5 1 or 2 2 DEFAULT SPASM LEVEL 1-5 1 or 2 3 DEFAULT SPASM RELIEF RATE 1-5 1 or 2 2 DEFAULT MILEAGE GOAL (mile) 0.1-5.0 0.1 or 1 0.
1. Strength Training STRENGTH training is a mode that system can automatically switch between active mode and passive mode according to the patient's training situation. It is designed to motivate the patient to move actively and try to maintain the state of active movement. Press Information icon to view more application information about this training mode.
NOTE: The number/letter in the frame changes blue, indicating that you have selected the item. rd STEP 3 : Start Training After you have set the training parameters, press START button to start the training. If the speed is set to 30 r/min or above, a prompt will appear on the interface, prompting whether to set the speed to this value, you can press YES to start the training, or press NO to return to the previous interface.
th STEP 4 : During Training Status and Operation Bar Item Heart Rate Description Real time display the remaining time of set duration time Real time display the actual rotation speed when it is greater than the set value Real time display heart rate value Status of Active/Passive Real time display of the active/passive status Remaining Time Rotate Speed *1 38
Item Description Press to change the rotation direction Change Direction FWD REV PAUSE Press to pause the training, enter into setup interface, which you can adjust the training parameters. STOP Press to stop the training NOTE: 1) 2) The heart rate monitoring value is for reference only. If the heart rate sensor is not connected, the heart rate value is displayed 0. Active/Passive Switching The training mode starts from the PASSIVE mode.
Passive Mode Active Mode 40
Adjust the Training Parameters during Training During the training, if you need to change the training parameters, please press PAUSE button, than the system will go to the ARM STRENGTH TRAINING setup interface as picture follows. Adjust the parameters then press RESUM button to start again. 2.
different. In ENDURANCE training game, resistance will decrease when the hungry fish approach, the patient should cycle the pedals faster than the set speed to swim up and over the hungry fish. Press Information icon to view more application information about this training mode. During training process, the remaining mileage is displayed in real time at the upper left corner of the screen, as the picture shows below. 3.
st STEP 1 : Enter the ARM SYMMETRY training mode nd STEP 2 : Setting ARM SYMMETRY training parameters You can set the training parameters as the picture shows. rd STEP 3 : Start Training After you have set the training parameters, press START button to start the training. If the speed is set to 30 r/min or above, a prompt will appear on the interface, prompting whether to set the speed to this value, you can press YES to start the training, or press NO to return to the previous interface.
th STEP 4 : During Training The horizontal bar chart and percentage value shows the relative muscle strength level of the left and right side of the patient during active training.
4. Arm ROM Training The ROM TRAINING, short for range of motion training, is a passive training mode that uses the motor to cycle the pedals, helps to increase the range of movement in a joint and to prevent contractures. Press Information icon to view more application information about this training mode. In the ROM TRAINING mode, you can press to select one of the two training background images for passive training.
Leg Training Press LEG TRAINING in the MAIN MENU to enter this interface. The mode of the Leg Training is the same as that of the Arm Training.
Spasm Control When you get a spasm during passive training, the trainer can identify spasms and prompt "spasm" on the screen with sound prompt. When the spasm protection is activated, the speed of the motor slows down to zero and then change the rotation direction, slowly increased to the original speed to relieve spasm. When the spasm protection is activated more than 5 times, the device will automatically stop the current training. Please stop training and take a break.
The device performs a self-check when release emergency stop. DO NOT approach the movement area of the ARM and LEG trainer of the device. Fine Motor Training The FINE MOTOR TRAINING mode is for only upper excises training to help the patient with cognitive and functional movements using hand, eye coordination. Press Information icon to view more application information about this training mode. NOTE: RATIO indicates the ratio of the number of bugs that appear on the left and right of the screen.
Read the game tips and tell the patient how to train, then press OK button to start training.
Pause and End of Training During the training, if you need to pause the training, please press PAUSE button, if you need to stop the training, please press STOP button. At the end of training, the buzzer will beep, and the sound will stop automatically after three times. The trainer will automatically stop moving and display the training data summary. Training Data Summary After ending training, the unit displays the training data summary, as the following picture shows.
Favorites Interface Press Favorites button on the MAIN MENU to enter the favorites interface. The system can store up to 99 collections. 1. How to Save a Favorite? When the training is finish, the training summary interface appears. Press the SAVE PARAMETERS AS FAVORITE button on the lower right corner of the screen, enter the favorite storage page. Fill the name of this favorite and save. Due to HIPAA guidelines, DO NOT save favorites as patient name.
2. How to Delete a Favorite? In the Favorites Interface, Choose the favorite you want to remove, the favorite name becomes blue, press DELETE button, a confirmation prompt will appear, press OK button to delete. 3. Start Training from the Favorite Interface In the favorite interface, select the desired favorite and press the SELECT button, the system jumps to the setup interface of the training mode, as following picture shows.
Training Date Export The system is equipped a training data export function. How to Export Training Data? In the training summary interface, or before the start of each training program, insert a USB flash disk (FAT32 format) into the USB port, the icon will appear on the upper right corner of the screen, when the icon changes to Symbol USB icon 1 USB icon 2 , the data has been copied.
Operation Guidance Preparation before Training 1) Make sure the trainer is in a stable place. 2) Connect the power cord: connect the male end of the power supply cord to an appropriate electrical outlet, and connect the female end of the power supply cord into the socket on the unit. 3) Press the Power switch to start the unit. Before you plug the DQAPT into a main socket please check that the voltage of the device stated on the marking corresponds with the voltage of the power supply.
Please use the specified heart rate sensor (Polar H10, manufactured by Polar Electro Oy). Secure Patient Position and Adjust Trainer 1. Securing patient position The patient should sit in the center of the wheelchair or chair, at a proper distance from the device. The legs should be slightly angled and not stretched. Improper positioning can cause discomfort in the knee or hip.
through the big screw knob beside the handlebar to fit the patient's training. 4. Adjusting calf support (Leg training) The calf support helps the patient to protect his legs. Use the knob to adjust the height of the calf support so that it can fit the calf. 5. Fixing feet and calves (Leg training) Use the straps to fix the feet and calve to LE pedals or calf supports securely and quickly.
Only put patient’s feet into the LE pedals while seated. Do not put patient’s feet into the LE pedals while standing. Hands are not allowed into the LE pedals. 6. Fixing hands (Arm training) Use hand straps Fix the patient's upper limbs with hand straps. The specific operation is shown in figure below.
Interface Operation and Parameter Settings See Chapter Interface Introduce for details. End of Training 1) At the end of training, there is a buzzer beep, and the sound is automatically stopped after three beeps. The trainer will stop training automatically and display the summary of training data.
data will be cleared when exit this interface. If you want to save the data, please refer to How to Export Training Data? as details. 2) Turn off the power and unplug the power cord. 3) Release the hand / foot straps and remove both hands / feet from the pedals.
Care and Maintenance Cleaning 1. Please turn off and unplug the power plug of the device before the cleaning and disinfection operation. 2. Main unit cleaning: Clean the surface of the main unit with a clean, soft damp cloth. You can disinfect the Handlebar / Pedals by wiping down with 75% alcohol. 3. Straps cleaning: Hand Wash with a mild detergent and do not iron. Cleaning and drying temperature should not exceed 80 °C.
Routine Maintenance Read this maintenance instruction thoroughly before doing any maintenance. Manufacturer will provide circuit diagrams, component part lists, descriptions, calibration instructions to assist to service personnel in parts repair. To preserve product warranty, functionality and product safety we recommend using only original spare parts. The unit and accessories must be checked at regular intervals.
1) Fuse specification: F2AL250V (Ф5 × 20mm) 2) When replace the fuse, please follow the steps: a) Turn off the power switch; b) Unplug the power plug; c) Use a screwdriver to unscrew the fuse shell; d) Remove the bad fuse, replace the new fuse (see figure below); e) The fuse shell loaded back to the original position; f) Plug the power. Please turn off the power switch and unplug the power when replace the fuse to avoid electric shock. And remove the fuse with the screwdriver.
For environmental reasons, do not dispose of the device in the household waste at the end of its useful life. Dispose of the unit at a suitable local collection or recycling point. Dispose of the device in accordance with EC Directive – WEEE (Waste Electrical and Electronic Equipment). If you have any questions, please contact the local authorities responsible for waste disposal.
Troubleshooting NOTE: If the following measures fail to alleviate the problem, please call the distributors. Troubles Possible causes 1. No response 1. Lack of power; after 2. Fuse might be start the device. burned out. Solutions 1. Check whether the power plug and socket of the instrument are plugged or not; 2. Replace the burned fuse with a new one (Specifications: F2AL250V) 2. The pedals UE 1. You had select the 1. Select the Arm training; are Leg training; 2. 2.
Troubles Possible causes Solutions 5. The heart rate 1. The heart rate 1. Check position of the displays sensor was chest strap of heart rate during training. 0 not connected with unit; sensor; 2. The 2. heart rate Check whether the sensor was not put electrode area of heart rate on properly. sensor is moist; 3. After above checking are completed, check whether the Bluetooth icon appears on the unit screen.
Assistance Every intervention on device must be performed by manufacturer. For any assistance intervention and original spare parts please contact the manufacturer at following address: Compass Health Brands, Corp. Add: Toll Free 1.888.549.4945 6753 Engle Road Middleburg Heights, OH 44130 Tel: 800-947-1728 Website: www.compasshealthbrands.
Warranty The Manufacturer warrants that the DQAPT is free of defects in material and workmanship for the main unit. This warranty shall remain in effect for two years (24 months) from the date of original consumer purchase. If this Product fails to function during the two year warranty period due to a defect in material or workmanship, at the Manufacturer's Option, Manufacturer or the authorized dealer will repair this Product without charge.