® In Scan Tray Count Out Bucket SmartWand Power Entry For use with SmartSponge® Surgical Sponges Lift Serial Number 100417 Rev.
Apply Plug in power cord, . flip switch to between sheets, directly below surgical site. (May be left in ON position when not in use) ON Press SCAN IN END WAND to enter COUNT MODE. Place Count Out Bucket. into Press “ to scan packs in by holding them on the . Open and use sponges.
Ensure System is in “COUNTING OUT” Mode; verified by display. Discard all sponges directly into the Count Out Bucket. Ensure that all sponges have been accounted for by placing them into the Count Out Bucket before ending the case. ● If the ”Change Bag” System Alert appears, remove the sponges or discard the full bag and replace with a new liner. Sponge counts will not change. Do not fill the Count Out Bucket with more than 50 sponges at a time.
Press the button to view the final sponge counts. Make sure all columns have a green check mark. Press the button to save the case data and power down the device. Press the “STANDBY” button to retain case data and temporarily power down the device. If the is requested and the sponge counts are not reconciled, will be an needed to end the case.
Cover the with a sterile cover while passing it into the sterile field. Press the button to activate the wand. The green LED on the wand will illuminate when activated. WAND Hold the wand over the surgical site above the location of the . When the SmartTag is detected, the on-screen figure will change to green as shown. Slowly scan the patient holding the wand inches above, at a rate no faster than inches a second while following the onscreen directions.
System Messages are displayed on-screen as System Alerts, Warnings, or Failures. SYSTEM MESSAGE CAUSE ACTION Pack Not Verified – Retry Pack System unable to scan pack. Flip or Rotate pack and Rescan. If rescanning does not work, discard pack. Discard Pack System has detected a problem with pack. Discard pack and resume scanning with a new pack. Multiple Packs Detected – Remove and Scan One Pack at a Time System unable to scan more than one pack at a time.