User's Manual

AI
-
04
Recorder Status cs5recstatus
Page 1 of 3
AG501 Controlserver
Recorder Status
Cs5recstatus helps the user to keep an eye on the most important
parameters during sweep recording. So if obvious errors occur like
active sensors outside the measurement area, deactivated
reference sensors, broken sensors with striking RMS etc. the user
is warned and can possibly solve the problem before sweep
recording is started.
Cs5recstatus starts automatically when the recorder (cs5recorder)
is started and displays with easy to understand symbols and colors
the current recorder-status: green=good; yellow=warning; red=error
Please note: A green recorder-status does not guarantee error-free
sweep recording. There are problems that can occur without
influencing the channel status (detached non-reference sensors,
taxonomic distance limit set to a very high value etc.).
The program has three display-sections:
- Master-LED (Figure 1-1)
- Main-section (Figure 1-2)
- Detail-section (Figure 1-3)
and three display-modes:
- Minimized = only the master-LED is visible
- Normal = master-LED and main-section are visible
- Maximized = all sections are visible
The display-mode at program startup can be configured in the settings dialog (see Detail-section).
The recorder status is set back (reset) when
- a channel is activated / deactivated
- head correction is initialized
- a new recording is started
- the clear status-button in the detail-section is clicked
A. Master-LED
In minimized mode only the master-LED is displayed in the status panel (Figure 2).
The master-LED always shows the worst status since the last reset (red if an error occurred).
A click on the master-LED toggles between minimized and normal mode.
Figure 2: Master-LED in status-panel
Figure 1: Cs5recstatus
Master
-
LED

Summary of content (3 pages)