KODAK DRYVIEW 5800 Laser Imager User’s Guide
Carestream Health, Inc. 150 Verona Street Rochester, New York 14608 © Carestream Health, Inc., 2008 DRYVIEW is a trademark of Carestream Health. KODAK is a trademark of Kodak used under license. Pub No. 2G0733 Rev.
Table of Contents 1 Overview - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 1-1 5800 Laser Imager - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - How the Laser Imager Works - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - System Components - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Film Technical Information - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-1 General Description - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Spectral Sensitivity - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Image Quality - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Environm
Warranty/Special Provisions for Canada - Limited Warranty - - - - - - - - - - - - - - - - - - - - - - - A-13 GARANTIE ET DISPOSITIONS PARTICULIÈRES POUR LE CANADA GARANTIE LIMITÉE A-14 Warranty/Special Provisions for the United States of America and any Other Country - Limited Warranty - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - A-15 January 14, 2008 2G0733 iii
1 Overview 5800 Laser Imager The KODAK DRYVIEW 5800 Laser Imager is a continuous-tone laser imager with an internal photothermographic film processor. Heat, rather than photo chemicals, is used to develop the film. The Imager receives digital images from medical image source devices (modalities) over a network. The format that the Imager accepts is DICOM. The Laser Imager prints images on DRYVIEW Laser Imaging Film. Each film package contains 100 sheets of film.
Overview During normal operation, the Imager requires very little operator attention. The Imager prints automatically in response to print requests from the associated image devices. Information sent along with print requests, such as film size, density and priority, control the print operations. Main operator responsibilities include loading film and monitoring for malfunctions. System Components 4 5 3 2 1 1 2 3 4 5 1-2 Film supply. Your Imager is configured with two film supplies.
Overview Print Sequence When the Imager receives a print request, it determines the requested film size and type and then it selects the appropriate film cartridge. Each time the Imager receives a print request, the following print sequence occurs: 6 5 4 3 2 1 1. Suction cups in the pickup area lift a single sheet of film out of the supply cartridge and feed the film into the transport rollers. 2.
Overview Automatic Image Quality Control An internal densitometer is a key element in the Automatic Image Quality Control (AIQC) process. The densitometer enables the Imager to automatically adjust image processing parameters to produce the best image. The Imager adjusts these parameters each time it prints a calibration film. A calibration film is printed when: • • • The film tray is inserted in the Imager with film of a new lot number. A calibration film is requested from the local panel or web portal.
2 Using and Maintaining the Imager Operator Control of the Imager During normal operation, the Imager receives and automatically prints images sent by modalities over a network. Very little operator control is required. The main responsibilities of the operator are: Turning Imager Power ON and OFF Power switch There is a power switch located at the back right of the Imager. In the event of a power loss, the Imager shuts down.
Using and Maintaining the Imager Local Panel and Display Screen Local Panel Layout 5 6 1 2 7 { 3 4 8 1 2 3 4 5 6 7 Ready LED. This LED, when lit indicates that the Imager is ready for printing. When flashing, the Imager is processing and printing films. When unlit, the Imager is not ready to print. Error LED. When lit, this LED indicates that the Imager has an error. An error code will appear on the display screen indicating what the actual error is. Attention LED.
Using and Maintaining the Imager Display Screen Icons Icon Description This icon indicates that film calibration is required. The Imager is unable to print jobs from this supply until a successful calibration occurs. This icon indicates that a door is open on the Imager. This icon indicates an error with the film tray. An error code will also be shown on the display screen. The film tray cannot be used until the error is corrected.
Using and Maintaining the Imager Icon Description This icon indicates there are prints in the queue that are waiting to be deleted. This icon indicates that a user-initiated system restart is in process. This icon indicates that the Imager has been placed in service mode. This icon indicates that the Imager is warming up. The amount of time remaining until the Imager reaches operating temperature is also shown on the display screen.
Using and Maintaining the Imager Level 3: Local Service - Local Service providers are Carestream Health, Inc., trained and certified self-maintenance customers. Level 3 First Call Service provides access to all Level 1 and 2 features and functions. In addition, First Call activities include preventive maintenance with the ability to reset some imaging parameters. Access to Level 3 requires a Level 3 ID and password. Level 4: Full Service - Full Service providers are Carestream Health, Inc.
Using and Maintaining the Imager To log on to the Web Portal as a Level 2 or Level 3 user, do the following. 1. Click the Log On icon. The following window appears. 2. Enter your User ID and password. NOTE: The Imager comes with a default User ID and password for Level 2 and Level 3 users. • for Level 2: User ID = KeyOperator; Password = DV5800 • for Level 3: User ID = LocalService; Password = DV5800 3. Click Login. The items on the left side of the display will differ according to your level of access.
Using and Maintaining the Imager Operations Unloading and loading the Film Tray IMPORTANT: Leave the Imager powered on while loading or unloading the film tray. When 100 sheets of film have been used, a 0 film count appears on the display screen, and the Imager will automatically unlock the film tray for removal. 1. Remove the film tray from the Imager. Film insert 2. Remove the old film insert from the tray and discard in a manner suitable to local ordinances.
Using and Maintaining the Imager Bag holder Film package Diverter 3. Pull the diverter (plastic bar) and bag holder (silver bar) into the up position. 4. Press the new film package down in the tray with the label facing up. 5. Pull the front flap of the film package forward and place it under the diverter (1). 6. Set the back end of the film package under the bag holder (2). 7. Press down firmly using both hands so the film package sits flat against the bottom of the tray. Bag holder 8.
Using and Maintaining the Imager Tear strip 9. Remove the tear strip from the plastic bag. 10. Swing the diverter of the film tray back in place. IMPORTANT:Do not install the Film Saver onto the film tray. 11. Slide the tray back into the Imager slot. Plastic bag 12. To remove the plastic bag so the Imager can access the film, pull the plastic bag tail firmly and smoothly. The entire film bag will slide out.
Using and Maintaining the Imager Inserting the Film Saver and Removing the Film Tray A film saver must be inserted into a film tray when: • • • the hood is opened. any panel on the Imager is removed. a film tray with film in it is removed. 1. Insert the film saver as shown below. 2. If the film tray has not been unlocked, press the "unlock" button on the local panel. NOTE: The film tray must be removed within 30 seconds after pressing the "unlock" button.
Using and Maintaining the Imager Running a Calibration Print You may occasionally have to request that the Imager print a calibration print to calibrate the Imager. You may be asked to do this if you call for service. You should also run a calibration print if a "Not Calibrated" status message appears on the Web Portal Home screen, or if a calibration error code (2x-624, 2x-631, or 2x-632) appears on the display screen.
Using and Maintaining the Imager 5. Enter the desired density (0.1 to 3.2). 6. Use the drop-down menu to select the film tray. 7. Select the number of copies to be printed. 8. Click Run. Removing Print Jobs From the Undeliverable Jobs Queue When there are jobs in the undeliverable jobs queue that can not be printed, condition code 20703 and an icon show on the display screen. To remove these jobs from the print queue, do the following. 1. Press and hold the Maintenance Reset LED/Button for 5 seconds.
Using and Maintaining the Imager Hood Charcoal filter Front panel 3. Raise the hood. 4. Pull the front panel forward. 5. Remove the charcoal filter by lifting it up and pulling it forward. 6. Install a new charcoal filter. 7. Close the front panel. 8. Close the hood. 9. Install the film trays. 10. Remove the film savers. 11. Turn the Imager ON. 12. Wait for the Imager to complete it’s start-up routine. 13. Clear the preventive maintenance icon and condition code. a.
3 Troubleshooting Error and Alarm Indications The Imager can detect errors and other conditions that require operator action. These errors or abnormal conditions are reported on the display screen and on the Web Portal in the form of condition codes and messages. DICOM Printer Status Messages In response to a DICOM printer N-GET status request from a modality, a printer status message and a printer status info message are returned to the requesting SCU.
Troubleshooting Printer Status Messages The following table describes how the local panel and Web Portal indicate printer status to the user. Table 3-2: Printer Status Messages Printer Status Local Panel Web Portal Display DICOM Status Description Any Film Supply Open • Ready LED off • Supply display is blanked for open supply • Status code: 20702 Not Ready WARNING / COVER OPEN At least one of the film supplies is open (not inserted and latched into place).
Troubleshooting Table 3-2: Printer Status Messages (Cont’d) Printer Status Local Panel Web Portal Display DICOM Status Description Service Mode • Ready LED off • Attention LED on Status code: 20700 Service Mode WARNING / PRINTER OFFLINE The service switch is enabled. The Imager is not "Ready". Warming • Ready LED off • Warming LED on • Number of minutes until warm shown Warming=xx WARNING / PROC INIT The Processor is warming up and will not be ready to print for xx minutes.
Troubleshooting Table 3-3: Film Supply Status Messages (Cont’d) Film Supply State Local Panel Web Portal Display Description Manual Mode • Attention LED on • Status code: 21002 (upper), 23002 (lower) • Unlock button LED on • Cal button LED off • Supply info area displays film type, size, and count No Film Tray • Unlock button LED on • Cal button LED off • Supply info area displays icon for this state Invalid Film Tray There is a film tray in the film supply but it does not contain a liner/RF tag.
Troubleshooting Job Manager Status Messages The following table describes how the local panel and Web Portal indicate job status to the user. Table 3-4: Job Manager Status Messages Job Manager Status Display Screen Web Portal Display Description Active n/a Shows how many jobs having this status are queued. The Imager is accepting DICOM job requests and film is available for all current jobs.
Troubleshooting Condition Codes Condition codes are shown on the display screen in the order in which they are generated. If there is more than one code associated with the current condition of the Imager, the first code is shown on the display screen for 6 seconds, while other codes in the list are displayed for 3 seconds as the list is cycled. The LEDs on the left side of the local panel will be on whenever there is a condition code of that type.
Troubleshooting Table 3-5: Condition Codes (Cont’d) Code 10003 Web Portal Message MIS: Image Buffer Error User Action 1. Shut down the Imager, then start the Imager. 2. If the error persists, call for service. 10015 MIS: Database Error 1. Shut down the Imager, then start the Imager. 2. If the error persists, call for service. 10910 20004 MIS: MCS Communication Failure 1. Shut down the Imager, then start the Imager. USB Failure 1. Shut down the Imager, then start the Imager. 2.
Troubleshooting Table 3-5: Condition Codes (Cont’d) Code Web Portal Message User Action 20706 none 1. Turn the Imager power OFF. 20915 Internal Image Data Transfer Failed 1. Shut down the Imager, then start the Imager. Internal Image Data Render Failed 1. Shut down the Imager, then start the Imager. 21000 or 23000 none None. The film supply requires attention due to an error. The specifice error is reflected by a separate code. 21001 or 23001 none None.
Troubleshooting Table 3-5: Condition Codes (Cont’d) Code 21122 or 23122 Web Portal Message xxxxx Supply: Internal Hardware Failure User Action 1. Cover the film tray with the film saver. Remove the tray from the Imager. 2. Take the film tray to a dark room and remove the film saver. Verify that the film is lying flat. Replace the film saver. 3. Re-insert the film tray and remove the film saver. 4. If this error repeats, try a new film pack. 5. If the error persists, call for service.
Troubleshooting Table 3-5: Condition Codes (Cont’d) Code 21624 or 23624 Web Portal Message xxxxx Supply: Film Calibration Failure User Action 1. Try a different film pack. 2. If the error repeats, shut down the Imager, then start the Imager. 3. If the error persists, call for service. 21631 or 23631 xxxxx Supply: Film Calibration Failure - Dmin Outside Target 1. The minimum density of the film is too high. Calibration results for this film are outside the normal range, but will still be used. 2.
Troubleshooting Table 3-5: Condition Codes (Cont’d) Code 27604 Web Portal Message Optics: Calibration Failed User Action 1. Shut down the Imager, then start the Imager. 2. If the error persists, call for service. 27607 Optics: Calibration Failed 1. Shut down the Imager, then start the Imager. 2. If the error persists, call for service. 27611 27646 27650 27931 28154 28155 28501 28509 Optics: Internal Hardware Failure 1. Shut down the Imager, then start the Imager.
Troubleshooting Table 3-5: Condition Codes (Cont’d) Code 29924 29925 29926 29927 29931 36931 36935 3-12 Web Portal Message User Action Densitometer: Internal Diagnostic Failure 1. Shut down the Imager, then start the Imager. Densitometer: Internal Diagnostic Failure 1. Shut down the Imager, then start the Imager. Densitometer: Internal Diagnostic Failure 1. Shut down the Imager, then start the Imager. Densitometer: Internal Diagnostic Failure 1. Shut down the Imager, then start the Imager.
Troubleshooting Clearing Film Jams CAUTION: Shutting off power to clear a film jam is not required.
Troubleshooting Film Jam - Code 2x-116 1. Remove the film trays (see “Inserting the Film Saver and Removing the Film Tray”). Film tray slots 2. Remove any misplaced films from Area 1. This area can be accessed through the film trays slots. NOTE: If the film is not accessible from the front of the Imager, continue with Step 3.
Troubleshooting Hood Left panel 2 screws 3. Open the hood. 4.
Troubleshooting Pickup access areas 5. Remove any misplaced films seen in the pickup access areas. 6. If Area 1 did not contain misplaced films, take the film trays to a dark room and remove the film saver. Remove any misplaced films and cover the tray with the film saver. 7. Set the left panel in place and tighten the 2 screws. 8. Close the hood.
Troubleshooting Film Jam - Code 2x126 1. Remove the film trays (see “Inserting the Film Saver and Removing the Film Tray”). Film tray slots 2. Remove any misplaced films from Area 1. This area can be accessed through the film trays slots. 3. If Area 1 did not contain misplaced films, take the film trays to a dark room and remove the film saver. Remove any misplaced films and replace the film saver. 4. If misplaced films were not found in the tray, open and close the Imager hood.
Troubleshooting Hood Left panel 2 screws 5. Open the hood. 6. Loosen the 2 screws and remove the left panel.
Troubleshooting 7. Remove any misplaced films seen in the pickup access areas. 8. Set the left panel in place and tighten the 2 screws. 9. Close the hood. Film Jam - Code 26325 IMPORTANT: If this error occurred at the same time as 2x-126, follow the instructions for “Film Jam - Code 2x126”. 1. If this error occurred alone, open and close the hood to clear the error. 2. If this error repeats, restart the Imager. Film Jam - Codes 26326 or 26543 Hood Drum cover Drum cover latches 1. Open the hood. 2.
Troubleshooting 4. Carefully remove any films found in the drum area. 5. Close the drum cover and secure the cover latches. If the film was not found in the drum area, continue with Step 6. Hood Left panel 2 screws 6. Open the hood. 7. Loosen the 2 screws and remove the left panel.
Troubleshooting Film advance knob 8. Use the manual advance knob to move the film through the processor. 9. Set the left panel in place and tighten the 2 screws. 10. Close the hood. If film was not previously removed, the misplaced films will emerge from the IMAGER within 2 minutes. Film Jam - Code 26544 1. Remove any films jammed in the imager film exit. 2. Open the hood and remove any films in Area 2. 3. Close the hood.
Troubleshooting Calling for Support If you cannot correct a condition and need help, call for support. Have the following information ready when you call: • • • Model Number: 5800 K-Number Condition code and message if they are shown on the display screen and Web Portal.
Troubleshooting Table 3-6: Service Phone Numbers (Cont’d) Country Pakistan Palestinian Territory Paraguay Peru Portugal Puerto Rico Qatar Romania Russian Federation Saudi Arabia Spain Sweden Switzerland Syria Tajikistan Turkey Turkmenistan United Arab Emirates United Kingdom United States Venezuela January 14, 2008 2G0733 Phone Number 00 92 21 2638881 00 970 92387421 54-3327-458-195 54-3327-458-163 80051853 351 21 414 7642 18669430227 00 974 4469 830 00 974 4360 820 00 40 1210 3854 7 495 660 5695 00 96
4 Film Technical Information General Description This section describes the characteristics of DRYVIEW Laser Imaging Film, not the operation of the 5800 Laser Imager. DRYVIEW Laser Imaging Film is a high-resolution, infrared-sensitive, photothermographic film designed specifically for the family of DRYVIEW Laser Imagers. DRYVIEW Laser Imaging Film is packaged in daylight-load packages and is available in blue, clear, and mammography 7-mil polyester base.
Film Technical Information Environmental Impact Tests show that DRYVIEW Laser Imaging Film is not considered hazardous to the environment. As a result, you can develop, recycle, and dispose of film with less impact on the environment than if you were using wet-developed silver halide films.
Film Technical Information With DRYVIEW technology, a small amount of final development occurs when the film exits the laser imager and is initially exposed to ambient or view-box lighting. This is virtually undetectable and has no effect on image quality (i.e., typically 0.02 change in density). This small density increase is uniform and permanent upon full exposure of the film under normal handling conditions (i.e., room light or view box).
5 Specifications Location Equipment Specifications Unpacked Packed Height 62 cm (24.25 in.) 95.25 cm (32.0 in.) Width 62 cm (24.25 in.) 81.6 cm (32.125 in.) Depth 66 cm (26.0 in.) without film trays 76 cm (30 in.) with film trays 106.7 cm (42.0 in.) 147.4 kg (325 lbs) Weight 70 kg (155 lbs) Operating Space Requirements The following is the minimum space required around the Imager to perform normal operator functions. Service functions will require more space. 7.6 cm (3.0 in.) 60.
Specifications Other Location Considerations Additional factors that influence where the Imager is located include proximity to patients, flammable materials, liquids, and other equipment. CAUTION: This equipment is not contained in a sealed cabinet. Do not use this equipment in locations where it can come in contact with liquids, including body fluids. IMPORTANT: • This equipment is intended to connect to other medical devices. Only an authorized Service Provider of Carestream Health, Inc.
Specifications Environmental Requirements Temperature Relative Humidity Altitude Surface Condition • • Operating: 15° to 33° C (59° to 91° F) • • Operating: 20% to 80% RH, noncondensing • 30 m (100 ft.) below sea level to 3,048 m (10,000 ft.) above sea level Storage: -40° to 60° C (-40° to 140° F) Storage: 10% to 90% RH, noncondensing The surface must be a level (must be level within 1°) table top or counter capable of supporting the weight of the Imager.
Specifications Power Requirements The Imager power supply has an auto-sensing and auto-switching feature for both voltage and frequency. The current draw is 10 A at 120 V AC, and the power consumption is a maximum of 1275 W. One of the following single-phase, 15 A power sources, with grounding, must be provided within 2.5 m (8 ft.) of the Imager.
Glossary CR CT Cycle power DICOM Dmax Dmin DR GSDF GSM Film model Key operator LED Modality Monotonic MPDB MR N/A PLUT Service image Shutdown SMPTE test pattern Test type January 14, 2008 Computed Radiography, the process of creating digital radiographic images. Computed Tomography, the process of creating digital tomographic images. Cycle power means to shut down and power up the Laser Imager. Digital Imaging and Communications in Medicine.
Glossary TFT Undelivered queue Version G-2 Transfer Function Tables. The queue of jobs that cannot be printed because of problems with the job description. A different rendition of an image. A second version of an image is processed using a second set of image processing parameters.
Appendix A: Warranty and Limitation of Liability Important Notice to Purchaser Carestream Health, Inc. warrants that Carestream Health, Inc. hardware products will be free from defects in parts, materials and manufacture. For defects occurring during the warranty period and about which Carestream Health, Inc. has received notice during the warranty period, Carestream Health, Inc. will provide Customer with free replacement parts and labor to replace warranty-covered items.
Warranty and Limitation of Liability Virus Policy: In accordance with the Health Group Digital Solutions Virus Policy, document 3E4111, Carestream Health, Inc.
Warranty and Limitation of Liability Copyright. The Software is owned by Carestream Health, Inc. or its suppliers and protected by copyright laws and international treaties. You may not copy the Software other than as expressly provided in this license. You may not reverse engineer, decompile, or disassemble the Software.
Warranty and Limitation of Liability cannot be excluded, so the above exclusion may not apply to you. This warranty gives you specific legal rights and you may also have other rights. Limitation of Remedies The remedies set forth above are Carestream Health, Inc.'s only obligation under the warranty. IN NO EVENT WILL CARESTREAM HEALTH, INC.
Warranty and Limitation of Liability YOU UNDER THE TERMS OF THAT LICENSE AGREEMENT. BY INSTALLING, COPYING, DOWNLOADING, ACCESSING OR OTHERWISE USING THE SOFTWARE PRODUCT, YOU AGREE TO BE BOUND BY THE TERMS OF THIS EULA. IF YOU DO NOT AGREE TO THE TERMS OF THIS EULA, MANUFACTURER AND MICROSOFT LICENSING, INC. ("MS") ARE UNWILLING TO LICENSE THE SOFTWARE PRODUCT TO YOU.
Warranty and Limitation of Liability limited to: e-mail, word processing, spreadsheets, database, network browsing, scheduling, and personal finance. – Back-up Copy. If Manufacturer has not included a back-up copy of the SOFTWARE PRODUCT with the COMPUTER, you may make a single back-up copy of the SOFTWARE PRODUCT. You may use the back-up copy solely for archival purposes. Back-up Utility. If the SOFTWARE PRODUCT includes a Microsoft back-up utility, you may use the utility to make the single back-up copy.
Warranty and Limitation of Liability – – – – – – – – – January 14, 2008 (1) Microsoft OS. Upon selection, the one Microsoft OS selected by you will be set up on the COMPUTER, and the other Microsoft OS(s) not selected by you will be automatically and permanently deleted from the hard disk of the COMPUTER. Additional Software.
Warranty and Limitation of Liability – Application Sharing. The SOFTWARE PRODUCT may contain Microsoft NetMeeting, a product that enables applications to be shared between two or more computers, even if an application is installed on only one of the computers. You may use this technology with all Microsoft application products for multi-party conferences.
Warranty and Limitation of Liability 6. LEASED HARDWARE.
Warranty and Limitation of Liability will not directly or indirectly export or re-export the SOFTWARE PRODUCT (or portions thereof), other than to Canada, without first obtaining an export license or determining that a license exception is applicable. For additional information see http://www.microsoft.com/exporting/. 9. NOTE ON JAVA SUPPORT. THE SOFTWARE PRODUCT MAY CONTAIN SUPPORT FOR PROGRAMS WRITTEN IN JAVA.
Warranty and Limitation of Liability LIMITATION OF LIABILITY. To the maximum extent permitted by applicable law, any conditions or warranties imposed or implied by law are hereby excluded. Consumers may nevertheless have the benefit of certain rights or remedies pursuant to the Trade Practices Act and similar state and territory laws in Australia or the Consumer Guarantees Act in New Zealand, in respect of which liability may not be excluded.
Warranty and Limitation of Liability Warranty/Special Provisions for England Scotland, Wales and Ireland Limited Warranty LIMITED WARRANTY.
Warranty and Limitation of Liability INCIDENTAL DAMAGES, THE ABOVE LIMITATION MAY NOT APPLY TO YOU. SPECIAL PROVISIONS: Reverse Engineering: If you acquired the SOFTWARE in the European Community, you may not reverse engineer, decompile, or disassemble the SOFTWARE except to the extent and for the express purposes authorized by applicable law. This Software License Agreement is governed by the laws of England. Warranty/Special Provisions for Canada - Limited Warranty LIMITED WARRANTY.
Warranty and Limitation of Liability LOSS OF BUSINESS INFORMATION, OR ANY OTHER PECUNIARY LOSS) ARISING OUT OF THE USE OF OR INABILITY TO USE THIS PRODUCT, EVEN IF MANUFACTURER HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMAGES. IN ANY CASE, MANUFACTURER'S AND ITS SUPPLIERS' ENTIRE LIABILITY UNDER ANY PROVISION OF THIS AGREEMENT SHALL BE LIMITED TO THE AMOUNT ACTUALLY PAID BY YOU FOR THE SOFTWARE AND/OR MICROSOFT HARDWARE.
Warranty and Limitation of Liability ABSENCE DE RESPONSABILITÉ POUR LES DOMMAGES INDIRECTS.
Warranty and Limitation of Liability WHICH VARY FROM STATE/JURISDICTION TO STATE/JURISDICTION. NO LIABILITY FOR CONSEQUENTIAL DAMAGES.
Index A Agency compliance, 1-4 Altitude, 5-3 Automatic image quality control (AIQC), 1-4 C Calling for support, 3-22 Conventions, 1-4 Correcting film jams, 3-13 D DICOM, 1-1 DICOM printer status messages, 3-1 Display screen icons, 2-3 E End user license agreement, A-2 Environmental requirements, 5-3 Error codes, 3-6 Errors, 3-1 F Film Archiving, 4-3 Correcting jams, 3-13 Environmental impact, 4-2 Jams, 3-13 Loading/unloading, 2-7 Recycling, 4-3 Saver, 2-10 Sizes, 5-4 Spectral sensitivity, 4-1 Storage, 5-4 S
Spectral sensitivity, 4-1 Storing undeveloped film, 4-2 System components, 1-2 T Temperature, 5-3 Troubleshooting, 3-1 W Warranty, A-1 Web portal, 2-4 Accessing, 2-5 Default user passwords, 2-6 Error messages, 3-6 Levels of user access, 2-4 Logging in, 2-6 I-2 2G0733 January 14, 2008
Carestream Health, Inc. 150 Verona Street Rochester, New York 14608 © Carestream Health, Inc., 2008 DRYVIEW is a trademark of Carestream Health. KODAK is a trademark of Kodak used under license. Pub No. 2G0733 Rev.