A Peer to Peer connection makes a scanner dedicated to a Workstation. This saves time by not having to select a scanner each time you make a scan. Creating a Peer to Peer Connection To create a Peer to Peer connection, follow these steps: 1 In the Scanners Connections tab, click the Pair button on the scanner you wish to pair with the image Acquisition interface. The status (located above the Pair/Unpair button) changes from Connected to Paired and the Pair button changes to Unpair.
The Peer to Peer icon appears in the scanner. Note: If you change the resolution on the Scanner LCD, the change in resolution will only take place upon next scan or image acquisition. Disconnecting a Peer to Peer Connection Disconnecting a Peer to Peer Connection may be done from the image Acquisition interface or from the Scanner interface. To Unpair the Scanner, follow these steps: 1 In the Scanner interface, press The Scanner Settings menu opens. Settings.
Press and press The Pairing status screen shows an Unpaired status. Clearing Scanner Memory The scanner’s storage capacity is 25 images. Images are stored in the scanner’s memory in the following cases: When No ID is selected in the scanner’s Select patient screen. ( When the destination workstation is not available due to network problems and so forth. When the scanner’s memory nears its limit, the scan number icon changes to red. When memory is full, error 071: Memory full.
To clear the scanner’s memory, follow these steps: 1 Start the CS 7600 image Acquisition interface. 2 Open the exam acquisition to which the images belong, or alternatively create a new temporary exam acquisition in which you will save these images. 3 Select the scanner; it is indicated by the following full memory/error state icon: 4 In the Scanner List, select the desired scanner and click The list of No ID images for this scanner is displayed.
66 Chapter 9 Scanner Settings
10 Maintenance Imaging Plate Care If used with care, the imaging plates can be used indefinitely until physically damaged. However, the imaging plates will show wear over time due to continuous use. Inspect the imaging plates regularly and replace them if they are scratched and/or show signs of excessive wear.
Cleaning the Imaging Plate CAUTION: Read and follow the instructions in the Safety Data Sheets (SDS) for Screen Cleaner P/N 1030428. Clean the imaging plate to remove dirt, dust particles, finger prints and so forth that could affect image quality.
Disinfecting the Hygienic Sheath and the Imaging Plates CAUTION: Do not soak the hygienic sheath in any cleaning or disinfecting solutions. Do not autoclave; autoclaved hygienic sheaths must be discarded. Frequency of Disinfection The hygienic sheath is the primary means of infection control and protection from contaminated imaging plates. Disinfect the imaging plate after every use (after the image plate is scanned), before it is inserted in the hygenic sheath for the next use.
Scanner Care Cleaning the Scanner Body Use disinfection wipes for medical equipment to clean the scanner body thoroughly from all sides, including the tray. WARNING: It is the user’s responsibility to disinfect the imaging plates tray daily to prevent cross contamination. Cleaning the Detachable Insertion Panels The CS 7600 makes use of three insertion panel sizes according to the inserted plates size (0, 0–3, and 4). The maintenance procedure is identical for both insertion panels.
3 The scanner displays a prompt to put in place an insertion panel, either a size 0 (left figure in prompt), size 0-3 (middle figure in prompt) or a size 4 (right figure in prompt) insertion panel. 4 Install the cleaned and disinfected insertion panel by positioning the insertion panel at a 45 degree angle to the right side of the insertion panel’s housing (A) and pressing the insertion slot panel in from right to left (B) until it’s completely inserted (C).
If insertion Panel 4 is inserted, the scanner displays the following insertion confirmation. 7 For size 4, press 72 Chapter 10 Maintenance the middle function button to confirm and return to the main screen.
Cleaning the Scanner Rollers Note: Before you perform this procedure, make sure you have the rollers cleaning plate. The scanner rollers should be cleaned in the following cases: When the scanner prompts to clean the rollers (after every 1000 cycles). When there are dirt-related artifacts on the scanned image.
4 Press the middle function button to confirm the Rollers Cleaning message. The Rollers Cleaning screen appears. 5 Peel the adhesive protective sheet from the rollers cleaning plate.
6 Insert gently the rollers cleaning plate into the size 4 insertion slot with the adhesive side first and the black stripe facing downwards until the rollers draw the rollers cleaning plate in. When the cleaning procedure is completed the scanner displays a message that the roller cleaning is done. 7 Remove the roller cleaning plate from the insertion slot. Tip: The cleaning procedure is to be repeated 3 times to ensure the rollers are thoroughly cleaned.
Scan & Go Device Care Cleaning the Scan & Go Use disinfection wipes for medical equipment to clean the Scan & Go device thoroughly from all sides. Replacing the Battery (Scan & Go Generation 1) For Scan & Go Generation 1 models released before June 2012. Scan & Go devices after June 2012 do not have a battery. Important: Replace only with a Carestream Dental genuine BL-5C battery.
2 Open the battery compartment by carefully pressing the clip outwards and lifting out the cover (A, B). A C D B 3 Extract the battery as shown in the figure above (C, D) using the removal depression to lift the battery out. 4 Insert the new battery by first pressing its contacts against the battery housing terminals (A) and then place it in the housing (B). B A D C 5 Press the cover firmly to close it (C, D in above figure). 6 Connect the USB cable to the computer.
7 Charge for 3 hours. 12 9 3 6 3 hrs B A 8 Verify the device operates: press the On/Off button for 1 second and verify that the LEDs function as described in the “Scan & Go Overview” on page 9. 9 Dispose of the old battery in accordance with all applicable federal, state, and local regulations.
11 Troubleshooting Quick Troubleshooting The quick troubleshooting guides you through actions you need to take to correct malfunctions you may encounter and that do not require authorized technical support. For errors and malfunctions not detailed in this section, contact technical support. The following table lists the malfunctions that are displayed on the scanner’s LCD and in the image Acquisition interface, and details the corrective action.
Table 1 Quick Troubleshooting (Continued) Malfunction Scanner Error # 073 Possible Cause Action Imaging plate loading failure. Press the middle function button to eject the imaging plate and then reinsert the imaging plate again. Replace imaging plate if reinsertion does not work. Important: Do not insert any tools into the scanner. If any tool is inserted into the scanner the warranties are null and void.
Table 1 Quick Troubleshooting (Continued) Malfunction Possible Cause Action Either imaging plate inserted incorrectly, or the imaging plate hygienic sheath is inserted in the wrong direction. Scanner Error # 008 Make sure you are using Carestream Dental imaging plates. Make sure you follow the preparations and scanning instructions in the Acquiring Images procedures to prevent incorrect insertion of the imaging plates.
1 Tap the Settings button. 2 Select Move Head Up and press OK. The optical head moves up and the scunner shuts down.
3 Position the scanner so that you have access to the service door and cable compartment. 4 Remove the cable compartment cover.
5 Unscrew the Phillips screw that attaches the service door to the rear cover, using the Phillips PH2 screwdriver and remove the service door. 6 Tilt the scanner backwards, until the plate falls out of the scanner into your hand. Important: If the plate is not in the location shown in the above picture, do not continue to look for the plate. Please call service.
11 Wait for the Insert Plate message. The scanner is ready for use. Important: In case Error 41 appears, check that the rear service cover is closed properly.
86 Chapter 11 Troubleshooting
Appendix: CS 7600 Consumables & Accessories Consumables & Accessories P/N Description Quantity 1014364 CS 7600 hyg. sheaths size 0 200 pcs. 1014372 CS 7600 hyg. sheaths size 1 200 pcs. 1014380 CS 7600 hyg. sheaths size 2 200 pcs. 1014398 CS 7600 hyg. sheaths size 3 200 pcs. 1014406 CS 7600 hyg. sheaths size 4 200 pcs. 1059955 CS 7600 smart plate size 0 4 pcs. 1059963 CS 7600 smart plate size 1 4 pcs. 1059989 CS 7600 smart plate size 2 4 pcs.
Scanner Manager The scanner manager is mainly for use by technicians or under direct supervision of a dental service technician. Note: For information on the Scanner Manager for technicians, see the CS 7600 Service Guide. Accessing the Scanner Manager To access the Scanner Manager, perform the following steps: 1 On the desktop, double-click on the Scanner Manager icon.
“Update Scanner” on page 91– Upgrades the scanner’s embedded software version to the latest version “Collect Info” on page 93 – Enables you to collect the necessary log files from the scanner and from the work station for inspection Scanner Info You may provide this information to your Carestream Dental distributor upon request. To open the Scanner Info screen: 1 In the Settings Screens area, select the Scanner Info checkbox. 2 Click Open Selected Screen. The Scanner Info screen opens.
Scanner Setup To open the Scanner Setup screen: 1 In the Settings Screens area, select the Scanner Setup checkbox. 2 Click Open Selected Screen. The Scanner Setup screen opens. Figure 3 Setup screen Note: Scanner serial number cannot be changed. The Scanner Setup screen enables you to set the following: Scanner Serial Number, Scanner Name, and LCD User Interface language Scanner Date and Time Backup and Restore User settings (File name is - CS7600_UserBackup.tar.gz.
Update Scanner Important: Updating the Scanner is to be performed under direct supervision of a dental service engineer. Note: Not all USB devices are recognized by the scanner. Therefore we suggest that you update the scanner using the scanner manager tool. There are two options for updating the scanner embedded software: Using USB storing device plugged into the USB socket located at the back of the system.
Updating the Scanner using the Scanner Manager Tool The Scanner version (Embedded software) can be updated by the scanner manager tool by selecting the Update Scanner button. To upgrade the scanner perform the following steps: 1 Select the scanner and click Update Scanner. Wait while the scanner updates. The file named Reader_Embedded_xxx.tar is copied from the UPGScanner folder to the scanner The process starts and takes approximately one minute.
4 If the scanner is already up to date, the following message appears, click OK. Collect Info This tool is collecting the necessary log files from the Scanner and from the work station, and then the DSE/User is able to save the zip file on the local PC and send it for inspection. 1 Select the scanner and click Collect Info. 2 Create new folder with the Scanner name (Recommended). A browse window opens. 3 Click on the folder and press OK.
4 Select a destination folder for the information to be stored and click OK. Wait a few minutes as the Information is being collected. Information Collected Successfully 5 When the Collect Information is done message appears, click OK. The information is stored in the selected destination folder. Diagnose Connection Important: Diagnose connection is to be performed under direct supervision of a dental service engineer.