User's Manual Part 2

51
6.3.4 Software Upgrade
The OS-Gemini system has two software image banks; one is a fixed image which is stored in
protected non-volatile memory, the second bank is used by the system administrator to
upgrade the firmware as and when necessary. Figure 21 shows the main software upgrade
webpage.
Figure 21 Software Upgrade
The ‘Fixed’ image is only used if disaster recovery is required, i.e. if an upgrade process is
interrupted or the units are reset to their factory defaults.
These pages are used to update a unit’s operational software. The software image to be
uploaded should be downloaded to local storage from the
Orthogon Systems website. The
software image is delivered by Orthogon Systems as a compressed zip file. Once the zip file
has been downloaded the user should extract the OS-Gemini Software image. This is easily
identifiable by its ‘.dld’ file extension.
The first step (Figure 21) is to use the “Browse” button to locate the software image previously
downloaded to local storage from the
Orthogon Systems website. Once the image is located
the user should press the “Upload image to wireless unit” button to start the software upgrade
process.

Summary of content (50 pages)