User Manual Part 2

TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-35
When a detection window becomes satised, the pulse generator begins
calculating for sudden Onset in a two-stage sequence.
Stage 1 measures the ventricular intervals prior to the start of the episode
and locates the pair of adjacent intervals (pivot point) where the cycle
length decreased the most. If the decrease in cycle length is equal to or
greater than the programmed Onset value, stage 1 declares sudden Onset.
Stage 2 then compares additional intervals. If the difference between the
average interval before the pivot point and 3 out of the rst 4 intervals
following the pivot point is equal to or greater than the programmed Onset
Threshold, stage 2 declares sudden Onset.
If both stages declare the rhythm sudden, therapy will be initiated. If either
stage indicates a gradual onset, initial ventricular therapy will be inhibited in the
lowest zone. Therapy will not be inhibited by Onset if:
The rate accelerates to a higher ventricular zone
Information from the atrial lead determines that the RV rate is faster than
theatrialrate(VRate>ARateprogrammedtoOn)
The SRD timer expires
Onset is measured using RV intervals only. It can be programmed as a
percentage of cycle length or as an interval length (in ms). It is limited to the
lowest therapy zone of a multi-zone conguration. The selected Onset value
represents the minim um difference that must exist between interva ls that are
above and below the lowest programmed rate threshold. The pulse generator
performs Onset calculations (even when the feature is programmed to Off) for
all episodes e xcept induced or commanded episodes. The m easured Onset
results from a two-stage calculation are stored in therapy history. This stored
data may be used to program an appropriate Onset value.
Sustained Rate Du r atio n (SRD)
Sustained Rate Duration allows delivery of the programmed ventricular therapy
when a tach ycardia is sustained for a programme d period beyond Duration, but
the programmed therapy inhibitors (Vector Timing and Correlation, AFib Rate
Threshold, Onset, a n d/or Stability) indicate to withhold the rapy (Figure 3-21 on
page 3-36).
- DRAFT -

Summary of content (100 pages)