Obturation System Down Pack User Manual Changzhou Bomedent Medical Technology Co., Ltd.
Thank you very much for choosing the Obturation System Down Pack of Bomedent To give full play to the function of this device, and operate and maintain it correctly and safely, please read this User Manual carefully before using it, and keep this User Manual properly for reference at any time Classification of device safety level: Classification of electric shock protection type: Class II Classification of electric shock protection level: Class B Classification of liquid entry prevention leve
Contents 1. Product Introduction................................................................................................1 1.1 Brief Introduction............................................................................................... 1 1.2 Model.................................................................................................................. 1 1.3 Scope of Application......................................................................................... 1 1.
1. Product Introduction 1.1 Brief Introduction Obturation System Down Pack (hereinafter referred to as “eHeater” or “device”) is used to soften and cut off the gutta percha point by heating. It is used to assist dentists with root canal treatment.
This device shall not be placed in a humid environment or any place where it may come into contact with any type of liquid; Patients are at risk of thermal hazards; Do not expose this device to direct or indirect heat source. This device must be operated and stored in a safe environment; With regard to electromagnetic compatibility (EMC), this device requires special precautions and must be installed and operated in strict accordance with EMC information.
2. Product Configuration 2.1 External Structure of Main Frame 2.2 Main Accessories Heating needle Power adapter Charging base Power cord Wireless foot switch (optional) 2.3 Accessories List Please see the packing list for the device configuration.
3. User Interface Illuminating light Rotating clamp Start button with built-in indicator light LCD Adjust button 3.
【E0】/【E1】In error state, the indicator light flashes in red. Adjust button: Adjust the temperature: press the adjust button to adjust the temperature, including 110℃,140℃,170℃, 200℃ and 230℃; Turn on and off the bluetooth: press and hold the adjust button for more than 1s to turn on the bluetooth; press and hold the adjust button for more than 1s to turn off the bluetooth. 3.
[E0]/[E1]—— Error state: see Section 6 Troubleshooting for details. ③ Temperature indicator: Display the preset temperature, including 110℃,140℃,170℃, 200℃ and 230℃; ④ Bluetooth indicator: : Bluetooth is open normally and can be connected; : Bluetooth has been connected. 4. Product Installation 4.
Please use the original heating needle; The heating needle has not been disinfected and sterilized when they leave the factory. Please clean, disinfect and sterilize the heating needle before each treatment. 4.2 Removal of Heating Needle When the heating needle cools, pull it out in the opposite direction, as shown in the figure below.
5. Product Use 5.1 Operation Start the device: press the switch button of the device for more than 1 second, the device starts up and enters the standby state, and the indicator light is always on in blue. Close the device: in standby state, the device will automatically shut down after 1 minutes without any operation. When the main frame is placed on the charging base, it will shut down automatically.
Heating needle starts cooling: release the start button to cool the heating needle; In cooling state: The indicator light flashes in blue, display real-time temperature, and the cooling time is 8s; The device is in the state of [CD], and waits for the heating needle to cool; After cooling, the device is in the state of [DR] and enters standby mode.
Figure b Caution: When the error code E0 appears during the use, accompanied by a prompt tone, it indicates that the battery power is low and the device should be charged; When the main frame is inserted into the charging base, but it does not enter the charging state, please stop charging it immediately and then contact the local distributor; When charging with the charging base, please place the charging base or the main frame in a dry and safe place where they are not easily knocked over;
to provide doctors with more convenient operation experience, we also provide an optional wireless control solution. You can choose our wireless foot switch to achieve this function. eHeater can be connected wirelessly with our wireless foot switch through bluetooth. After successful connection, the operation of eHeater can be controlled within 5 meters.
6. Troubleshooting When a fault is found, please check the following points before contacting your distributor, and if none of these apply or if the fault is not resolved even after taking action, there may be something wrong with this product. Please contact your distributor. Fault phenomenon The error code E0 appears Fault analysis and Check items handling method The battery power is low.
7. Cleaning, Disinfection and Sterilization Caution: No part of eHeater is sterilized before leaving the factory Warning: Please do not soak the main frame in the ultrasonic cleaning machine; It is recommended to use a soft cloth soaked in alcohol to wipe and disinfect the surface of the main frame of eHeater; Please do not use liquid or spray cleaner directly on the main frame, especially on the display screen; Please do not disinfect the main frame by heating; Part of eHeater that ca
plasma. 7.1 Cleaning and Sterilization of Heating Needle 1) Cleaning Steps 1. Wash Parameters Wash the heating needle with running water for 2 minutes to remove its surface contaminants. Wet the soft clean cloth in the cleaner, and then use it to wipe the surface of the heating 2. Wipe needle thoroughly 5 times. Replace it with a new one after each wipe. If there are still visible contaminants, wipe repeatedly until no contaminants are visible to naked eyes. 3.
8. Storage, Maintenance and Transportation 1. Storage a. This product should be handled with care, kept away from earthquake source, and stored in a dry and ventilated place. b. This product should not be placed together with toxic, corrosive, flammable and explosive items. c. The relative humidity of the storage environment should be 10%~80%, the atmospheric pressure should be 500-1060hPa, and the temperature should be -10℃~ +50℃. 2. Maintenance a.
9. Technical Parameters Manufacturer Changzhou Bomedent Medical Technology Co., Ltd. Product name Obturation System Down Pack Model eHeater Dimensions Main frame: 170*26*26mm Charging base: 153*45*43mm Weight 650g Power supply mode Lithium battery, DC 3.6V, 2600mAh Charging base Input: DC5V/2A Output: DC 3.6V, 2600mAh Power adapter Input: AC100-240V, 50/60Hz 0.
10. Symbol Description Warning: please read this manual Refer to the user manual before use Type B application device Class II device The product complies with the directive on WEEE. This device Serial number must be disposed of as municipal solid waste when abandoned Manufacturer Date of manufacture For indoor use Upward Damp-proof Fragile. Please handle it with care 11.
13. EMC Statement Special precautions regarding electromagnetic compatibility (EMC) should be taken for this Obturation System Down Pack, and it must be installed and used in accordance with the EMC information specified in this manual. Portable and movable RF communication equipment may affect this device.
Guidance and manufacturer’s declaration – electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that is used in such an environment. Immunity test Electrostatic discharge(ESD) EN 61000-4-2 IEC60601 test level Compliance Level Electromagnetic environment - guide Floors should be wood, concrete or ceramic ± 6kV contact ± 6kV contact tile.
Guidance and manufacturer’s declaration – electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that is used in such an environment.
Recommended separation distances between portable and mobile RF communications equipment and the device The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF according to the maximum output power of the communications equipment.
After-sales service company: Changzhou Bomedent Medical Technology Co., Ltd. Tel: 0519-88991980 ChangZhou BoMedent Medical Technology Co.,Ltd. NO.