User's Guide
Bases
Replace the L300 electrode bases if they appear worn or damaged.
To replace the L300 electrode bases:
1. Mark the position of the used L300 electrode bases on the L300 FS Cuff
liner. See Figure 9-6.
Figure 9-6: Marking the position of the used L300 electrode bases.
Figure 9-7: Unsnapping the L300 electrode bases.
2. Unsnap the L300 electrode bases from the plug holes. See Figure 9-7.