Operators Manual SmartSystem SHIP L Datalogger
Table of Contents 1 Safety instructions ....................................................................................................................... 4 2 Quick Guide ................................................................................................................................. 5 2.1 Quick Start if configuration is already loaded on the device................................................... 5 2.2 Quick Start if configuration needs first loaded to the device ..............
13.5.2 Upload data manually of read out data to Berlinger SmartView .................................. 19 13.5.3 Return device ................................................................................................................. 19 14 Where to place SmartMonitor SHIP L........................................................................................ 19 15 External Temperature Sensor ....................................................................................................
1 Safety instructions Intended use Temperature • Temperatures outside the operating range may damage the battery. For the operating range, see the information in the technical data sheet. IP Protection • Limited protection against dust ingress and protection against splashing water from all directions. Air Pressure • Excess pressure or vacuum can damage the device. Do not vacuum when used for air cargo. Mechanical force • Avoid violent shocks and impacts.
2 Quick Guide 2.
2.
3 Device view and description No. Name Description 1 Display Multi-information display 2 USB-connector To charge the battery or download the data manually 3 Confirmation/Status LED LED (small lamp) indicates if use of multifunctional button was successful or not 4 Cover for ext.
4 Display Icon Name Description 1 Battery status Indicates the battery status 2 Battery charging Indicates whether external power has been plugged in 3 Alarm status 4 Logging Modus OFF → SmartMonitor SHIP L is ready to start REC → SmartMonitor SHIP L is logging → SmartMonitor SHIP L has stopped 5 7-digit display Information line (see more in section 7-digit display) 6 External readout Device ready for readout via USB or Bluetooth 7 Warning Symbol Additional warning for function 8 Cl
4.1 7-digit display The 7-digit display gives additional information about the status, shipment or other data. In normal logging operation it displays the temperature. 5 Multifunctional button SmartMonitor SHIP L device has one multifunctional button to operate. 1 x long press → Start / Stop the device 3 x short press → Setup – Load configuration 1 x short press → Information • Long press to multifunctional button One long press (> 5 sec.) to multifunctional button will start or stop the device.
6 Confirmation LED Device has an LED which gives an indication if the use of multifunctional button was successful or if a task has been terminated. 1 x short white flash → Confirmation of an action or task 1 sec. green flash → “Cancel Message” Short press has released too early.
8 Humidity and pressure sensor SmartMonitor SHIP L has a built-in humidity and atmospheric pressure sensor which can be enabled or disabled in the Berlinger SmartView software (version 6.0 and higher) configuration. The sensor is located on the left side of the device.
10 Configuration The configuration for the SmartMonitor SHIP L device can be defined via the Berlinger SmartView software (version 6.0 and later). SmartMonitor SHIP L can be ordered pre-configured with a customer desired configuration or with a standard configuration. If a new configuration is to be loaded, it is configured in the Berlinger SmartView software (version 6.0 and higher) and then assigned to the device.
11 Start/Stop 11.1 Start To start the device, press the multifunctional button once >5 seconds or as long till display shows “SETUP” or the green LED lights for at least 2 sec. The following information will be displayed on the screen. While the information “***START***” will be displayed as a ticker, keep pressed the multifunctional button and hold the multifunctional button for at least 5 seconds. As confirmation the green LED will flash for at least 2 sec.
11.2 Stop To stop the device, press the multifunctional button once >5 seconds or as long till display shows “STOP” or the green LED lights for at least 2 sec. As soon as the device has stopped, the 7-segment display shows “STOP”. Additionally, the Stop symbol will be shown . As confirmation the LED will flash 3 times short. Device will start now the synchronisation process by uploading the remaining data.
12 Start delay For SmartMonitor SHIP L device you have the possibility to configure in Berlinger SmartView software (version 6.0 and higher) with a start delay or with a scheduled start. 12.1 Start SmartMonitor SHIP L with start delay If a start delay is configured a counter in minutes:seconds will start counting down to 00:00, after that, the device starts and goes to the Record modus. 12.
13 Usage Process SmartMonitor SHIP L is mainly used for monitoring temperature sensitive products throughout the supply chain. The device has various configuration options. The device states are visualized below and described in more detail in the following chapters. The usage process may vary depending on the configuration. 13.1 Preparation 13.1.1 Battery check Check if the internal battery is charged and has enough capacity for the shipment.
13.2 Setup If no configuration is already loaded on the device, press the multifunctional button quickly 3 times . → The configuration will be loaded. See chapter 5 multifunctional button Configuration of device can be done with Berlinger SmartView software (version 6.0 and higher). 13.2.1 Review/Check configuration To confirm that the loaded configuration is correct, check the configuration on the display by pressing the multifunctional button once . 13.
The device cannot be stopped during start delay phase. This is a safety feature to prevent the device from being stopped by mistake. 13.3.2 Scheduler Start Device has the possibility to configure it with a scheduled start via Berlinger SmartView software (version 6.0 and higher). 13.
• Read-out via Bluetooth: Read out via Bluetooth will be possible with a later version 13.5.2 Upload data manually of read out data to Berlinger SmartView To upload the manually read out data to Berlinger SmartView software (version 6.0 and higher) you have two possibilities: • • Direct upload to Berlinger SmartView software (version 6.0 and higher): Open Berlinger SmartView software (version 6.
15 External Temperature Sensor Use only original Berlinger external sensors for SmartMonitor SHIP L. Make sure that the sensor is properly connected to the device and the connector is secured (tightened by hand). Do not tighten the connector with pliers or any other tool, otherwise the connector may be damaged. The cable must be attached to the shipment in such a way that it does not interfere with normal transport. Berlinger offers several types of external sensor as: 1.5 Meter Temperature Sensor 1.
15.1 Connect an external sensor to SmartMonitor SHIP L device To connect an external Berlinger probe to a SmartMonitor SHIP L remove the grey cover at the right site from device. Remove the grey cover from the SHIPL device Plug in the connector by checking the mark at the connector and the plug of the SHIP L device. Manually tighten the connector.
16 Connect SmartMonitor SHIP L device to SHIP S devices SmartMonitor SHIP L device can be used as a master device which can forward data from SmartMonitor SHIP S devices. Therefore, several SmartMonitor SHIP S can be used in one shipment. Connection of SmartMonitor SHIP S devices must be configured in Berlinger SmartView software (version 6.0 and higher).
17 Troubleshooting Problem Blank display Effect No information visible Device cannot be started Display shows LONG PRESS Solution Connect the device to USB charger to charge the battery. Press the multifunctional button at least 5 second. LOW BATTERY LOW:BATT will be shown on the 7-segnent display and battery symbol is blinking Connect the device to power to charge the battery. BATTERY EXPIRED BAT:EXPIRED will be shown on the 7-segment display and the battery symbol is blinking.
MANUAL SYNC MANUAL SYNC will be shown on the 7-segment display, the cloud symbol crossed out circle is displayed and USB and Bluetooth sign is blinking. Configuration is needed. Manual SYNC Manual SYNC will be shown on the 7-segment display as a ticker information and at the cloud symbol crossed out circle is displayed . Manual upload of data is needed. LOW -.-°C LOW -.-°C will be shown on the 7-segment display and the exclamation mark is blinking.
18 Disposal Electronic devices are recyclable and do not belong in household waste. At the end of its lifetime, dispose of the product in accordance with the law according to your country or return it to Berlinger according Berlinger RMA process. 19 Certification SmartMonitor SHIP L device has following certifications: • • • CE / RED RTCA DO-160 FCC / ISED 19.
• • • • Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/TV technician for help. Radiofrequency radiation exposure Information: This equipment complies with FCC and ISED radiation exposure limits set forth for an uncontrolled environment.