PULSE OXIMETER INSTRUCTION MANUAL V1.
V1.0M122 PULSE OXIMETER INSTRUCTION MANUAL Copyright Our company owns all rights of this unpublished work and intends to maintain it as a confidential work. We may also seek to maintain this work as an unpublished copyright. This publication is to be used solely for the purpose of reference or operation of our device. No part of this work can be disseminated for other purposes.
V1.0M122 PULSE OXIMETER INSTRUCTION MANUAL Content 1 Introduction...........................................................................................3 1.1 Brief Introduction...........................................................................3 1.2 Safety Information .........................................................................3 1.3 Electromagnetism Interference ....................................................5 2 General Descriptions ............................................
1 Introduction 1.1 Brief Introduction Thank you for purchasing the handheld pulse oximeter. The main functions of the device include SpO2 and PR measurements, visual and audible alarm, batteries charging, data storage and review and USB cable&Bluetooth transferring, etc. Please read this manual carefully before using the device.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL z This equipment is for use in the medical field, and measurement results only serve as a reference for any relevant treatment. z Connect the probe correctly; please see the directions for use of any accessories. z Please follow the doctor’s suggestions when starting to monitor the vital sign parameters. z When connecting this device to other peripherals, make sure that you are qualified to operate this device.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL z Application of this device may influence the measuring accuracy in the background of electromagnetic areas such as electro-surgery environment. z SpO2 measurements may be adversely affected in the presence of high ambient light. Shield the probe area (with a surgical towel, for example) if necessary.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL This apparatus complies with the IEC 60601-1-2 international standard. The requirements of this international standard are: CISPR11, GROP1, and CLASS B. 2 General Descriptions The handheld pulse oximeter adopts 2.8 inch TFT screen, which can display the SpO2% and pulse rate value, other indication parameters, such as time, ID number, pulse amplitude bar and battery power status, alarm limits and the connections of Bluetooth and probe, etc. 2.1 Appearance Fig.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL it down to continuously increase the value. Down button: press this button to decrease the value by one decrement. Or press it and hold it down to continually decrease the value. Menu/Select button: for entering main menu, or confirming the selection/setting. Alarm inhibit/off button: Press this button to inhibit alarm sound for 120 seconds. And long press the button, power off the alarm.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL ◆It shows two dashes throughout probe off and finger out conditions. 4:99: SpO2% upper alarm limit indicator 5:90: SpO2% lower alarm limit indicator 6: Pulse amplitude bar. It indicates the dynamic pulse amplitude and rate. As the detected pulse becomes stronger, more bars is illumined with each pulse. The reverse is true for weak pulses.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Keyboard is unlocked Keyboard is locked USB cable is connected USB cable disconnected SpO2 probe is inserted SpO2 probe off Bluetooth activated Bluetooth deactivated 2.3 Power Supply 2.3.1 Powered by batteries Fig.2-4 Batteries Installation : 1) Open the battery cover: Rotate the fixing screw slightly in the rear panel to the position which is marked with “ ” and then push the cover as indicated by arrowhead, as shown in Fig.2-4.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL . And the batteries are locked. Battery life and replacement There are five shapes of the indicator: the centre with 4 bars(full), 3 bars, 2 bars, 1 bar, empty and the frame in red. That the frame of indicator become red means few of battery capacity remains. You should replace the batteries with new ones timely. Or else, the indicator displays with a red frame constantly until battery capacity reaches critical condition (the battery voltage is 3.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL 3. It is not recommended to measure SpO2 during the oximeter is powered by wall outlet or being charged, for the damages and injuries may be caused to the device or users. 4. DO NOT charge when the non-rechargeable batteries are installed, otherwise, the damages or injuries may be caused to the devices or users. 2.4 Product Features Simple to use and easy to operate. Compact, light in weight and convenient to carry.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL The socket is also used for USB cable connected with the PC for data transmission. For more information on data transmission and data management,refer to corresponding software instruction manual. 4 Setting ID, Date and Time Always set the date and time before using the unit for the first time. Set different ID numbers for different users. Check the date and time are correct before using the unit,and reset them if necessary.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig.4-2 Fig.4-3 Pick different sub-items to set and press the Select button to highlight it and then using the Up or Down button to adjust the value. At last, press the Select button to confirm. Set other sub-items of data and time according to the “Year setting” illustrated in the above figures 4-2&4-3. 4.2 ID number Setting Enter the “User ID Setting” from the main menu screen, refer to Fig.4-4.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig. 4-4 5 Take a Measurement Before taking a measurement: z Select the suitable probe in terms of type and dimension. z Plug probe into SpO2 socket on top panel of pulse oximeter. z Clip the patient finger to the rational position of the probe as the illustration as Fig.5-1. Fig.5-1 Placement of the probe Note:If the finger is not in the probe, “Finger Out!!” will be shown.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL integrity and correct positioning and adhesion of the probe. More frequently inspection should be taken depend on different patients if necessary.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig.6-1 Set the low limit for SpO2 as the above steps of SpO2 high limit settings. The range of SpO2 lower limit is 70%-99%. 2) PR alarm setting Also, the limits settings of PR are performed as similarly as the SpO2 limits. The range of High limit is 31bpm-235bpm. And the range of low limit is 30bpm-234bpm.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL are divided by built-in module and can not be changed by user. Assignment of priority: High Medium Paramter SpO2 PR Low Value Red Yellow Alarm lamp Flashing with yellow Flashing with yellow Lamp Frequency 1.5Hz 0.5Hz Audiblesound Di- Di – Di ----- Di - Di Di - Di - Di Alarm cycle 3s 5s 20 s Alarm info SpO2 too high/low PR too high/low Probe off/Finger out Di Notes: 1. The alarm sound will go on until alarm disappears or is turned off.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig.6-3 Press the Up button or Down button select the sub-item to set, and then press the Select button to confirm or Return button to return to the previous screen. 6.2.1 Data Review Pick ”Data Review” sub-item and enter by pressing the Select button. The screen as in fig.6-4 will appear. Press the Select button again, a dialog box will pop up, refer to fig.6-5.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Select button to confirm, you can delete or review all the SpO2 trends saved under the ID. Fig. 6-6 6.2.3 PR Trend Fig. 6-7 The operation is referred to 6.2.2 SpO2 Trend. Fig.6-8 6.2.4 Clear Data Fig.6-9 Pick ”Clear Data” sub-item and enter by pressing the Select button. A dialog box will pop up, refer to fig. 6-10. Press the Up or Down button to pick the Yes or No and then press the Select button to determine whether to delete all the records.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Fig.6-10 6.2.5 In the data management screen, the upper half screen will display the maximum, minimum and average values of SpO2 and PR for the current ID. Refer to fig.6-11. Fig.6-11 6.3 System setting From the main menu, select and enter the System Setting screen, refer to Fig.6-12&6-13. Press the Up or Down button to pick the different sub-items to set and press the Select button to enter. Fig.6-12 Fig.6-13 6.3.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Brightness level: 1-7 Bright time: ON, 15, 30, 45, 60, 75, 90,105,120 seconds. 6.3.2 Bluetooth Enter BlueTooth screen from the System Setting screen, and set the Bluetooth on or off before transferring data to software for reviewing or analyzing data. Notes: For details, refer to attached Software Instruction Manual. When the Bluetooth is free from transferring data for 2 minutes, the Bluetooth will be closed. 6.3.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Enter Default Setup screen from the System Setting screen, and you can restore the default configurations.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL delivered with the accessories. 7.2 Safety Checks Before every use, or after your pulse oximeter has been used for 6 to 12 months, or whenever your pulse oximeter is repaired or upgraded, a thorough inspection should be performed by qualified service personnel to ensure the reliability. Follow these guidelines when inspecting the equipment: z Make sure that the environment and power supply meet the requirements.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL The applied parts touching the patients’ body are required to be disinfected once after each use. The recommended disinfectants include: ethanol 70%, isopropanol 70%, glutaraldehyde-type 2% liquid disinfectants. Disinfection may cause damage to the equipment and is therefore not recommended for this pulse oximeter unless otherwise indicated in your hospital’s servicing schedule. Clean the pulse oximeter before disinfecting it.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL 7.3.3 SpO2 & Pulse Rate Measurement Value Verification a).Connect SpO2 Probe to the SpO2 connector on the oximeter. b).Insert the operator’s finger into the finger probe, the SpO2 measured value of healthy person should be from 95% to 99%, and the pulse rate is same as heart rate, c).If SpO2 Simulator is available, verify the accuracy of Oxygen Saturation Value with probes as follows: Oxygen Saturation Tolerance 96% ±2% 86% ±2% 70% ±3% 7.3.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL a) Our company isn’t responsible for such damage caused by force majeure. For example: fire, thunder flash, flood, cyclone, hail, earthquake, house collapse, commotion, plane failing and traffic accident, deliberate damage, lack of fuel or water, labor and capital bother, strike and stop-working etc.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL professionals or authorized personnel disassemble oximeter and lead to oximeter failure. ●Not carefully read manual and so wrong operation lead to oximeter damage and failure. 7.5.5 User’s Special Request for Guarantee Time Our guarantee constitution for oximeter complies with electronic product after-sale service standard regulated by national laws. The guarantee time of mainboard regulated by our company is one year and all the accessories are three months.
V1.0M122 PULSE OXIMETER OPERATOR' S MANUAL Alarm mode: audible alarm, visual alarm and information Alarm limits range: SpO2 70%-100%, PR 30-235 bpm Default limits: SpO2 High 100%, low 90%; PR High 100 bpm; low 60 bpm SpO2 Display range: 0%~100% Measurement range: 70%~100% Resolution: Accuracy: 1% 70%-100% <70% : ±3 Digit; unspecified Probe LED Specifications: RED IR Wavelength 660±2nm 940±10nm Radiant Power 1.8mW 2.
V1.0M122 IPX1 PULSE OXIMETER OPERATOR' S MANUAL Outline of product Dimension: 125mmX60mmX30mm Weight: 195g (excluding the batteries) Accessories: z Standard accessories: 1. Three AA alkaline batteries. 2. One instruction manual 3. One adult finger probe: Model:M-50E. 4. The key for screwing the screw. 5. Bluetooth adapter 6. Software CD 7. USB Cable z Optional accessories: 1. Binding sensor for pediatric, infant and neonatal: (Pediatric 15-45 Kg, Infant 3-15 Kg). Model: M-50C 2.
V1.
FCC ID: WWIMD300M122 This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.