Pluse Oximeter Instruction Manual MD300C318
Fingertip Pulse Oximeter V2.0C318 Copyright Our company owns all rights to this unpublished work and intends to maintain this work as confidential. Our company may also seek to maintain this work as an unpublished copyright. This publication is to be used solely for the purpose of reference, operation, maintenance or repair of our equipment. No part of this can be disseminated for other purposes.
Fingertip Pulse Oximeter V2.0C318 Safety Information Please read this manual thoroughly before using the pulse oximeter! Keep it in hand for future reference. Warnings alert the user to potential serious outcomes, such as injury or adverse events to the patient or user. Cautions alert the user to exercise care necessary for the safe and effective use of the pulse oximeter. Notes contain important information that may be overlooked or missed. Warnings! ● DO NOT strike or needle the battery.
Fingertip Pulse Oximeter V2.0C318 ● Prolonged use or the patient’s condition may require changing the measurement site periodically. Change measurement site and check skin integrity, circulatory status, and correct alignment at least every 4 hours. ● Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, including batteries. ● Only the authorized service personnel can replace the battery or repair this device.
Fingertip Pulse Oximeter ¾ Venous pulsations may cause inaccurate readings. V2.0C318 ¾ Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line may cause inaccurate readings. ¾ The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia may cause inaccurate readings. ¾ Operation of the pulse oximeter may be affected by the use of an electrosurgical unit (ESU).
Fingertip Pulse Oximeter V2.0C318 Contents CHAPTER 1 General Information...................................................................................................................................................... 7 1.1 Measuring principle................................................................................................................................................................. 7 1.2 Product Features .............................................................................
Fingertip Pulse Oximeter V2.0C318 CHAPTER 1 General Information The pulse oximeter is integrated with Bluetooth® technology allowing the user to transfer measurement data any time and anywhere. The oximeter is designed with the measurement, storage, review, audible and visible alarms, vibration alert and data transmission function (optional). Note:The illustrations used in this manual may differ slightly from the appearance of the actual product. 1.
Fingertip Pulse Oximeter V2.0C318 1.2 Product Features ¾ Compact and light weight ¾ Four display modes ¾ Low power consumption ¾ Battery low indicator ¾ Rechargeable lithium-ion battery ¾ Visible, audible and vibration alarms. ¾ Automatically power off after finger out for 8 seconds. ¾ Bluetooth® and USB modes for data transmission. 1.
Fingertip Pulse Oximeter V2.0C318 1.4 Appearance Introduction Fig.2 Fig.3 Description: 1: Display screen: OLED display 2: Power switch & function button Short press: press this button for about 0.5 second to power on or change the select item. Long press: Press this button for about 1.5 seconds to confirm your selection and enter into the submenu. 3:USB cable port: It is used to connect the USB cable with the pulse oximeter.
Fingertip Pulse Oximeter V2.0C318 1.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 2 Date, Time and ID setting Always set the ID number, date and time before using the oximeter for the first time. Set the ID number for different users. Make sure that the date and time are correct before using the unit, reset them if necessary. The ID number, date and time are important indicators when a measurement is taken. 2.1 ID setting Press the button to power on the oximeter, and then the initial screen is as follows: Fig.4 Fig.
Fingertip Pulse Oximeter In figure7 long press the press the V2.0C318 button to set the ID number, the range is 01 to 99. After finishing setting, short button to select the “Return” item and long press the Fig.7 button to return to the previous menu. Fig.8 2.2 Date and Time setting In the main menu short press the button to select the Date and Time icon, refer to Fig.9, and then long press the button to enter into the “Date and time” setting screen, refer to Fig.10. Fig.9 Fig.
Fingertip Pulse Oximeter V2.0C318 The displaying mode: Date: Year-Month-Day Time: Hour:Minute:Second The setting range: Year: 2008~2028 Hour: 0~23 Month: 1~12 Minute: 0~59 Day: 1~31 Second: 0~59 CAUTION: In the Date and Time setting screen, BE SURE TO select the “Return” item and long press the button to return to the previous menu. Otherwise, the settings will not be saved.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 3 Take a measurement After the ID, date and time setting. Return to the main menu and then short press the button to select the “Return” item and long press the button to confirm your selection to return to the measuring screen. Open the clamp and insert a finger into the oximeter as illustrated in the figure 11 before releasing the clamp. Caution: When your finger is plugged into the Oximeter, your nail surface must be upward. Fig.
Fingertip Pulse Oximeter V2.0C318 The measurement screen refers to figure 13: Fig.13 Description of figure 13: 1—Time display: The current time is 22:10. 2—SpO2% area of display 3—Measured SpO2% 4—The SpO2% Plethysmograph 5—Measured pulse rate. 6—Pulse amplitude indicator. 7—Pulse rate unit . 8—Battery power indicator: When battery power is low the icon will be empty and turn to red. 9—Data transmission mode area of display The current transmission mode is Bluetooth®.
Fingertip Pulse Oximeter V2.0C318 During the measuring, each time you press the button the oximeter will switch to another display mode, there are 4 display modes shown as follows: Fig.14 Fig.15 Fig.16 Fig.17 NOTE: Please use the medical alcohol to clean the rubber touching the finger inside of Oximeter with a soft cloth dampened with 70% isopropyl alcohol, and clean the test finger using alcohol before and after each measurement.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 4 Battery Charge 4.1 Power Supply Battery Model: SP080 Battery Type: One 3.7V Lithium Ion Rechargeable Battery 4.2 Battery charge 4.2.1 Connect the oximeter with attached the USB cable, as shown in Fig.18. 4.2.1 Connect the other end of the USB cable with the attatched charger or a computer’s USB port. Fig.18 Fig.19 Charging temperature: 0℃~40℃ Charging Voltage: 110Vac~240 Vac, 50Hz~60Hz. It will take about 4hours to complete charging not more than 6 hours.
Fingertip Pulse Oximeter V2.0C318 Warning! Keep away from source of fire and/or heat. Avoiding the strongly impacting to the oximeter. Avoid the oximeter exposure straightly to the strongly sunlight. DO not leave the oximeter in the sealed car. DO NOT open or disassemble the device, opening and disassembly may cause damage to the oximeter. ONLY USE the attacched battery charger, DO NOT USE other chargers. If not that may cause damage even danger to the Oximeter or person.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 5 Data Manage In the main menu (shown as Fig.20), select the “Data Manage” icon and long press the Fig.21 to review, clear the data and look over the used space for data storage. button to enter into Fig.20 Fig.21 Fig.22 Fig.23 Data review: In figure 20 long press the button to review the stored data, refers to the figure 21. Short press the button to turn page up or down. If there is no stored data the screen will be the figure 22.
Fingertip Pulse Oximeter V2.0C318 Clear Data: Short press the button to select the “Clear Data” item and then long press the selection, and then the screen as shown in Fig.26 will be appear. Short press the “No” and then long press the button to confirm your selection. Fig.25 After finishing “Data Manage”, short press the button to confirm the selection. Fig.26 button to select the “Return” item and then long press the Fig.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 6 Alarm set The Alarm limits display allows you to adjust the upper and lower saturation and pulse rate limits. When the measured values exceed the setting limits, the visible, audioble or vibratory alarms will occur according to the alarm setting. In the main menu short press the button to select the “Alarm Set” icon, and long press the confirm your selection and enter into the Alarm Set screen, refer to figure 29. Fig.28 button to Fig.
Fingertip Pulse Oximeter V2.0C318 Low alarm limit setting: Fig.30 The SpO2% low alarm limit range is 70% to 98%. The upper value of the SpO2% low alarm limit is limited to the SpO2 %upper alarm limit. The SpO2% low alarm limit cannot be set equal to or higher than the SpO2% upper alarm limit. button to select the “SpO2 ALMLO” and then long press the SpO2 ALMLO (low alarm limit): Short press the button to decrease the low alarm limit. If you long press the button, the number will be cycled displaying.
Fingertip Pulse Oximeter V2.0C318 Fig.31 PR ALMHI (Upper alarm limit): Short press the button to select the “PR ALMHI” and then long press the button, the number will be cycled displaying. button to increase the upper alarm limit. If you long press the Low alarm limit setting: The pulse rate low alarm limit range is 30 to 234. The upper value of the pulse rate low alarm limit is limited to one number lower than the pulse rate low alarm limit.
Fingertip Pulse Oximeter V2.0C318 Alarm mode setting: Mode: There are three alarm modes for selection: OFF, ON and Sound Off. “OFF”: The alarm function is off, the audible, visible and vibratory alarms are ineffective. “ON”: The alarm function is on, the audible, visible and vibratory alarms are effective. “Sound Off”: The audible alarm will be ineffective, but the visible and vibratory alarms is still effective. Short press the button to select the “ Mode” and then long press the refer to the figure 33.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 7 System Set In System Set menu, you can set the brightness of backlight, BlueTooth® mode, vibration alert and so on. In the main menu short press the button to select the “System Set” item refer to figure35, and then button to confirm your selection and enter into the sub menu of System Set, refer to long press the figure 36. Fig.35 Brightness: Adjust the backlight brightness, in figure 36, long press the the backlight. The brightness level is 1 to 7.
Fingertip Pulse Oximeter V2.0C318 Power Save: “ON” or “OFF” setting. When setting “Power save: ON”, the “BlueTooth” will be set “OFF” automatically and the beep is inefective. button return to NOTE:If the “Power Save” is on, the display screen turns black. Long press the the displaying screen, short press the button the following screen will appear showing the oximeter is in process of measuring. Fig.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 8 Data transmission (optional) Please be sure to register the pulse oximeter before transmission. Contact our company for registration. The measurement results saved in the pulse oximeter can be uploaded to a computer for review or management. The modes of Data transmission are Bluetooth® transmission and USB calbe transmission. Please refer to the “software user manual” for the details of data transmission. The data transimission screen is shown in Fig.38.
Fingertip Pulse Oximeter CHAPTER 9 Specifications Display: Display type: OLED display SpO2 display range: 0~99% PR display range: 30~235 BPM PR display mode: Amplitude Bar Data update period: <15 s Resolution: SpO2%: ±1% Pulse rate: ±1BPM Measurement Accuracy: SpO2: 70%--99% ±3%; ≤69% unspecified. PR: 30~99 bpm ±2 bpm; 100~235 bpm ±2%. Antenna Information: Antenna Type/Patten: Internal Frequency Range: 2402 to 2480 MHz LED Wavelengths and Out put Power: Red: approximately 660nm @0.
Fingertip Pulse Oximeter V2.0C318 Physical Characteristics: Dimensions: 56mm x 38mm x 28mm (2.2”x 1.5”x 1.1”) Weight: 51g (0.11 lbs.)(including the lithium ion battery) Battery life: Typical lithium ion battery: 300 cycles.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 10 Maintenance And Calibration 10.1 Please charge the battery timely when the power low indicator flashes. 10.2 Clean the surface of the fingertip Oximeter before it is used in diagnosis for patients. 10.3 It is best to preserve the product in the environment specified in this manual. 10.4 It is recommended that the product should be kept in a dry environment anytime. A wet ambient might affect its lifetime and even might damage the device.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 11 Troubleshooting Problem Cause Solution SpO2% or pulse rate does not display. 1. Finger is not plugged correctly. 2. Patient’s SpO2 value is too low to be measured. 1.Retry by plugging the finger. 2.There is excessive illumination. 3.Measure other patients to make sure that no problem exists in the product. Go to a hospital in a timely manner for an exact diagnosis. SpO2% or PR is shown unstably. 1.
Fingertip Pulse Oximeter V2.0C318 CHAPTER 12 Declaration FCC-ID: WWIMD300C318 This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
Fingertip Pulse Oximeter V2.0C318 EMC of this product comply with IEC60601-1-2 standard.
Fingertip Pulse Oximeter 34 V2.
Fingertip Pulse Oximeter V2.
Fingertip Pulse Oximeter 36 V2.
Fingertip Pulse Oximeter V2.
Fingertip Pulse Oximeter V2.0C318 Manufacturer: Beijing Choice Electronic Technology Co., Ltd. Address: Rm. 1127-1128, Fuxing Road, A36, Bailangyuan Building B, 100039, Beijing, People’s Republic of China. Tel: (+86)10- 88204188、68221006 、88203520 Fax: (+86)10-88204632 www.choicemmed.