Model: 9115, 9115-AT, 9116, 9117 Multi-Range DC Power Supply USER MANUAL
Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument. Before applying power to this instrument: • Read and understand the safety and operational information in this manual. • Apply all the listed safety precautions. • Verify that the voltage selector at the line power cord input is set to the correct line voltage.
Do not use this instrument in an electrical environment with a higher category rating than what is specified in this manual for this instrument. You must ensure that each accessory you use with this instrument has a category rating equal to or higher than the instrument's category rating to maintain the instrument's category rating. Failure to do so will lower the category rating of the measuring system. Electrical Power This instrument is intended to be powered from a CATEGORY II mains power environment.
Do not operate in an explosive or flammable atmosphere Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finelydivided particulates. The instrument is designed to be used in office-type indoor environments. Do not operate the instrument • In the presence of noxious, corrosive, or flammable fumes, gases, vapors, chemicals, or finely-divided particulates. • In relative humidity conditions outside the instrument's specifications.
Precision for repair. Notify B&K Precision of the nature of any contamination of the instrument. Clean the instrument only as instructed Do not clean the instrument, its switches, or its terminals with contact cleaners, abrasives, lubricants, solvents, acids/bases, or other such chemicals. Clean the instrument only with a clean dry lint-free cloth or as instructed in this manual.
Fuse replacement Fuse replacement must be done by qualified service-trained maintenance personnel who are aware of the instrument's fuse requirements and safe replacement procedures. Disconnect the instrument from the power line before replacing fuses. Replace fuses only with new fuses of the fuse types, voltage ratings, and current ratings specified in this manual or on the back of the instrument. Failure to do so may damage the instrument, lead to a safety hazard, or cause a fire.
For continued safe use of the instrument • Do not place heavy objects on the instrument. • Do not obstruct cooling air flow to the instrument. • Do not place a hot soldering iron on the instrument. • Do not pull the instrument with the power cord, connected probe, or connected test lead. Do not move the instrument when a probe is connected to a circuit being tested.
Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems) This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste.
CE Declaration of Conformity The power supply meets the requirements of 2006/95/EC Low Voltage Directive and 2004/108/EC Electromagnetic Compatibility Directive with the following standards.
Safety Symbols Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument. Electric Shock hazard On (Supply). This is the AC mains connect/disconnect switch on the front of the instrument. Off (Supply). This is the AC mains connect/disconnect switch on the front of the instrument.
Table of Contents Safety Summary ...................................................................................................i Compliance Statements ............................................................................................................. vii Safety Symbols ............................................................................................................................ ix 1 General Information ...........................................................................
3.3 Protection Settings ......................................................................................................... 20 Configure Over Voltage Protection (OVP) ............................................................................. 20 Configure Maximum Power Limit ......................................................................................... 21 3.3 SYSTEM Menu ................................................................................................................
4.2 5 Remote Commands ........................................................................................................ 60 Troubleshooting Guide ............................................................................... 61 General .................................................................................................................................. 61 Remote Control .....................................................................................................................
1 General Information 1.1 Product Overview B&K Precision models 9115, 9115-AT, 9116, and 9117 are multi-range single output high power supplies with the capability of producing up to 80 V or 60 A (9115/9115-AT), 150 V or 30 A (9116), and 80 V or 120 A (9117) at a maximum power output of 1200 W (3000 W for 9117). With an easy-to-read VFD display, user-friendly controls and a numeric keypad that allows for easy configurations from the front panel.
contact B&K Precision. 1.3 Product Dimensions The 9115, 9115-AT, and 9116 power supply’s dimensions (WxHxD) are approximately 414.5 mm x 44.5 mm x 483.2 mm (16.29” x 1.75” x 19.02”). It is designed to fit in a standard 19-inch rackmount and is of 1U size. The 9117 power supply’s dimensions (WxHxD) are approximately 436.5 mm x 89.3 mm x 532 mm (17.19” x 3.52” x 20.95”). It is also designed to fit in a standard 19-inch rackmount and is of 2U size.
Figure 1.3 - Front View for 9117 Figure 1.
1.4 Front Panel Overview 1 2 3 4 15 14 6 7 8 9 5 15 14 13 12 8 9 16 5 13 12 11 10 Figure 1.5 - Front Panel for 9115/9115-AT/9116 1 2 3 4 16 6 7 11 10 Figure 1.6 – Front Panel for 9117 Front Panel Description 1 Power On/Off switch 2 VFD display 3 Voltage adjust knob 4 Current adjust knob 5 6 7 8 Up/Down/Left/Right arrow keys Used to adjust cursor location and selecting menu items. On/Off / Lock button Control the output state or locks the front panel button.
to set the instrument back to local mode. Enter / Trigger button Used to confirm setting/parameter changes or to provide a single trigger for list operation. Escape button Used to exit menu settings. 9 10 Numeric keypad 11 12 13 14 15 16 OK I-set / P-max button Used to configure output current or set the maximum power output limit. OVP set button Used to configure over voltage protection (OVP) value.
1.5 Rear Panel Overview 9 10 1 2 3 4 5 8 6 3 7 Figure 1.7 - Rear Panel for 9115/9115-AT/9116 11 4 1 5 6 9 10 8 2 3 Figure 1.
1.6 Display Overview OFF CC CV Rmt Addr Error Trig 80.000V 1 Prot * Shift 12.000A 0.0 W VOVP = 85.000 V 2 3 Figure 1.
2 Getting Started Before connecting and powering up the instrument, please review and go through the instructions in this chapter. 2.1 Input Power and Fuse Requirements Input Power Before connecting to an AC outlet or external power source, be sure that the power switch is in the OFF position and verify that the AC power cord, including the extension line, is compatible with the rated voltage/current and that there is sufficient circuit capacity for the power supply.
three-conductor type with the correct pin connected to earth ground. Connection of this power supply to an AC power source should be made by a qualified electrician or other qualified personnel.
Step 6 Step 4-5 Steps 2-3 Step 1 Figure 2.1 – AC Power cable housing assembly Follow the instructions below to connect the AC power cord to the AC supply mains distribution.
Figure 2.2 – Connecting to Mains Distribution (Top) Phase-to-Phase, (Bottom) Phase-to-Neutral Fuse Requirements An AC input fuse is necessary when powering the instrument. Refer to the table below for the fuse requirements. Model 9115 9115-AT 9116 9117 Fuse Description 15AT / 250V (6 x 32 mm slow blow ceramic tube fuse) 15AT / 250V (6 x 32 mm slow blow ceramic tube fuse) 15AT / 250V (6 x 32 mm slow blow ceramic tube fuse) 20AT / 250V (6 x 32 mm slow blow ceramic tube fuse) Table 2.
2. Use a flat blade screwdriver and turn the fuse capsule counter-clockwise. There is an internal spring that will push it out after turning a few times. 3. Pull out the glass tube fuse inside to check and replace as necessary. 4. Insert the fuse capsule in the same location and use the flat blade screwdriver to turn clockwise. 5. The fuse capsule will now be locked and secured. Note: For model 9117, the fuse capsule is hand fastened, so no flat blade screwdriver is needed. 2.
AWG 6 8 10 12 14 16 18 20 22 Imax(A) mΩ/meter 75 1.3 55 2.1 40 3.3 25 5.2 20 8.3 13 13.2 10 21 7 33.5 5 52.8 2.3 Preliminary Check Complete the following steps to verify that the power supply is ready for use. 1. 2. Verify AC Input Voltage Verify and check to make sure proper AC voltages are available to power the instrument. The AC voltage range must meet the acceptable specification as explained in “2.1 Input Power and Fuse Requirements”.
Output Check Voltage Check Follow the steps below to check basic voltage output with no load connected. 1. Turn on the power supply. The display will show the OFF annunciator above the voltage display. 2. Enable the output by pressing , and the button will be lit. The OFF annunciator will change to CV. 3. Using the numeric keypad or the voltage adjust knob and enter a voltage value. The voltage display will now show the value you entered.
7. Press the power switch to (OFF) position to turn off the power supply and remove the short on the output terminals. Check Model and Firmware Version The model and firmware version can be verified by using the *IDN? query remote command. It can also be found from within the menu system by following these steps: 1. Press and press to enter the menu system. 2. Press the arrow three times until INFO is blinking. Press show the following screen: OK , and the display will Power Info . . .
3 Front Panel Operation 3.1 Menu Options All settings and parameters can be configured from the built-in menu system of the power supply. To access the menu, press and press . The menu system is divided into the following sections and organized as follows: SYSTEM CONFIG LIST1 FUNC2 Initialize Power-On Trigger Memory Buzzer Communication ReturnMeter Reset power supply settings to factory default values. Configure power-on state. Configure Trigger.
1. From the normal display, press and press to enter the menu. 2. The selected item will be blinking. Use keys to move through the menu selections. 3. When the desired menu section is blinking, press 4. Below is the display when SYSTEM is selected. SYSTEM MENU In it i al i ze to access its menu settings. Powe r-O n 5. The selected item will be blinking. Use keys to move through the menu items.
Note: To see the changes to the setting voltage, be sure the display is not showing the measured voltage. This can be checked by looking at the If it is not lit, display is showing setting voltage. button backlight. Setting Current Follow the steps below to set the output current: 1. From the normal front panel display, users can use either the current adjust knob or the numeric keypad to enter the setting current. 2.
Vo+ Vs+ Vs- Vo- Figure 3.1 – Local Sense Connection Diagram DO NOT disconnect the wires if remote sense is not used. Doing so will cause erratic behavior and may damage the power supply under certain conditions. Never connect any power source into any of the four terminals at any time during operation. When output is enabled, DO NOT use your hands to touch the terminals or the screws that are designed to tighten wires to the terminals.
Rear Sense Terminal Vo+ Vs+ Vs- Vo- + DUT - Rear Output Terminal Figure 3.2 – Remote Sense Connection Diagram DO NOT at any time disconnect the wires from the Vs+ and Vs- terminals to the DUT while output is enabled (ON). Doing so may damage the power supply and cause unstable output. 3.3 Protection Settings Configure Over Voltage Protection (OVP) Follow the steps below to set the OVP limit: 1. Press the button. The display will show VOVP on the bottom right.
5. To disable OVP at any time, just press will disappear. twice. When it is disabled, the backlight When OVP protection is tripped, the following screen will display: OFF Prot 80.000VOVP 12.000A 0.0 W To clear the trip status, press once. Configure Maximum Power Limit Follow the steps below to set the maximum power limit: 1. Press and then . The display will show Pmax on the bottom right. Use the voltage or current adjust knob or the numeric keypad to enter the maximum power output limit. 2.
OFF Prot 80.000VOPP 12.000A 0.0 W To clear the trip status, press once. 3.3 SYSTEM Menu All setup procedures and settings explained in this section can be accessed from the SYSTEM menu. To access this menu, press and press . When SYSTEM is blinking, press .
Power-On Load Setup External Control Limit Online Setup Output Rst Off 10v-M Vmin = 0.000V, Vmax = 81.000V Off Disabled Configure Power-On State The initial power-On state of the power supply can be configured by following the steps below: 1. From the SYSTEM menu, select Power-On and press 2. There are two options: Rst(Def) – Factory Default. Sav0 – Settings before last power up. 3. Select the settings you want during power up, and press 4. To exit the menu at any time, press . to save changes. twice.
Save/Recall Instrument Settings The instrument can save up to 100 instrument settings in non-volatile memory. Memory is allocated in 10 different storage groups (group 0 to 9), and each group has 10 memory locations to store settings (0 to 9). The memory group must be selected from the menu first, before settings can be saved within the group. Select Storage Group 1. From the SYSTEM menu, browse and select Memory and press screen will appear. . The following MEMORY Group = 0 2.
Recall Settings 1. Press . Notice the button will be lit and the cursor on the display will disappear. This indicates Recall mode. Instrument settings can only be recalled when the instrument enters this mode. 2. Use the keypad to enter the memory location you want to recall. Enter between 0 to 9. 3. Once entered, the saved settings at the location will be immediately recalled. Note: When in Recall mode, users can recall settings from different locations without having to press additional keys each time.
press mode. and . The RMT indicator will disappear when the instrument is in LOCAL RS-232 Follow the steps below to configure the power supply for RS-232 operation: 1. From the SYSTEM menu, browse and select Communication and press 2. Select RS-232(Def) and press following display will be shown: . to set to RS-232 for remote communication. The RS232 48 00 , 8 , N , 1 , Add r . . . 3. 4800 is the baudrate; 8 is the data bits; N is the parity; 1 is the stop bit; Addr… is for address. 4.
Note: Users who have LabVIEW™ or NI-VISA installed will automatically have this driver in their system. In this case, driver download is not required. GPIB Follow the below instructions to select GPIB interface for remote operation. 1. From the SYSTEM menu, browse and select Communication and press . 2. Select GPIB and press to set to GPIB for remote communication. 3. The display will give a prompt to select an Address. This is the GPIB address to which the power supply will be assigned to. 4.
Note: The default is 4800, 8, N, 1, Address = 0. *Setting the baudrate to 115200 may provide unstable results during remote communication. Select a lower baudrate if communication errors occur. 6. Select Addr… and press . You will be prompted to enter an address. Use current adjust knob or numeric keypad to enter a number between 0 to 31, and then press . 7. For each power supply that you want to control, provide a different address.
Return Meter This option allows users to enable an internal fixed timer delay (5 seconds) for the power supply to automatically switch from setting display to measured display. When enabled, if the power supply output state is ON (enabled) and if the display shows setting voltage and current, it will automatically switch to measured voltage and current display after 5 seconds. The backlight of the button will also be lit.
DO NOT enable this function for battery test applications (i.e. battery charging) and all applications that may have a high charge. If a battery with a low voltage is connected to the power supply while the output is OFF (disabled), the dummy load will drain the battery. If the battery capacity is large and this function is enabled, a large current will flow and damage the power supply and the dummy load. To enable or disable this function, follow the steps below: 1.
2. Select 10v-M or 5v-M so that it is blinking. Use the keys to change between 10vM and 5v-M. – 0 – 5 V scale is used for monitoring 0 – 100% of output. 10v-M – 0 – 10 V scale is used for monitoring 0 – 100% of output. 3. Press 4. Press when finished making the selection. several times to exit the menu. Configure Control Source Follow the steps below to configure the source to use for programming the voltage and current output.
1. From the CONFIG menu, browse and select Ext-Ctrl and press screen will display: Ext – Ctrl Setup 1 0v -M 10v /10 k- P V- P . The following Of f 2. Select the third item V-P or R-P so that it is blinking. Use the keys to change between V-P or R-P. V-P – Use external DC voltage source for programming the output. R-P – Use external resistance for programming the output. 3. Press 4. Press when finished making the selection. several times to exit the menu.
Pin Assignment Below are the pin assignments of the analog interface. Figure 3.
Emergency Shut Off In the case where an emergency shut off to the output of the power supply is required, SHUT_OFF (pin 15) and DGND (pin 1) is used. To shut off, connect/short Pin 15 and Pin 1 together. This will bypass the analog control of the output state and disable (OFF) the output. In this case, pin 14 and pin 1 will need to be disconnected/opened from each other and reconnected to enable output again.
Reference Voltage A 10 VDC reference output is available on REF_10V (pin 21) and Ground (pin 9). Note: These pins are used for reference only and are not designed to power devices. Do not connect a load across it. Configure Voltage Limit The minimum and maximum set voltage can be configured on the power supply to limit the settable voltage range for protection.
Parallel/Series Connection More than one unit of the same model can be connected in a parallel or series configuration. Table 3.2 below shows the total number of units that can be connected in each configuration. The power supplies can be set up in master/slave mode so that the master unit can control all other power supplies in the parallel or series connection.
For parallel connection: Connect each power supplies’ positive (+) terminals together. Do the same for the negative (-) terminals. For series connection: Connect one power supply’s positive (+) terminal to the negative (-) terminal of another. Do the same for all the power supplies. Then, connect all of the power supplies’ Pin 1 of the RS-485 interface together. Do the same for Pin 5. Below illustrates the connection diagram for parallel connection.
- + DUT + - Figure 3.5 - Master/Slave Series Connection Diagram Master/Slave Setup Only one power supply has to be configured as a Master. The rest must be configured as Slave. Up to 3 units can be configured in total. Note: Configure Slave power supplies FIRST, and configure the Master power supply LAST. For remote or front panel operation, only control the Master power supply.
3. Press the key to select Slave or Master. Use the keys to select between the two options. Select Master to set the power supply as a master, or select Slave to set the power supply as a slave. Always set the Slave supplies first and Master supply last. 4. Press the key to select Addr…. Press and the display will prompt to enter an address. Use the numeric keypad to enter an Address, which must be different than all other power supplies that you want to connect together in parallel.
Sequence #1 Step 1.. Step 2.. .. Step 10.. Program #1 Sequence #1 Sequence #2 .. Sequence #10 Save Memory Sequence #10 Step 1.. Step 2.. .. Step 10.. Run Program # Program #10 Sequence #2 Sequence #5 Sequence #7 Sequence #8 Recall Figure 3.6 – List Mode Illustration There are three separate configurations to set up for programming and running a sequence (in order): 1. Configure Sequence Parameters 2. Configure Program (List) 3.
Table 3.3 - Step Parameter Range Parameter Voltage Setting Current Setting Width Slope Adjustable Range 0.000 – < OVP Voltage 0.000 – *max. current setting 0.001 seconds – 86400 seconds 0.002 seconds – 86400 seconds *Maximum current and voltage setting cannot exceed maximum power configured under P-max setting. As an example, follow the steps below to set up a sequence with fixed current limit of 1 A, as illustrated below: V 3 2 1 S 1s 2s 3s 3s 4s Figure 3.7 - Program Sequence Example 1.
SEQ A c ti ve Step: 09876 54321 5. These numbers represent the step numbers of the sequence, where “0” represents step number 10. At this point, use the numeric keypad and try to push any numbers 0 – 9. Notice that the number you press will correspond to the number displayed in Active Step and will toggle to display a “Y” in its place. 6. The “Y” represents a selection of the step number, making it an active step (meaning the step is enabled in the sequence).
9. This is prompting the user to enter the voltage setting for step 1. Use the voltage and current adjust knob or the numeric keypad and enter 1.000. Then press . 10. The display will change to the following, prompting the user to enter the current setting for step 1. EDIT SEQ CURRENT S eq Step 1 Cu r rent = 1.000 A 11. Similar to entering the voltage setting in the previous step, do the same to enter 1.000 as the current setting. Then press .
SAVE SEQ No Yes 16. If stored successfully, the display will show a message Save Success! After a few seconds, it will return to the LIST menu. Configure Program (List) The sequence(s) to execute and run can be selected, as well as its repetitiveness. This means multiple sequences can be run one after another. With 10 steps per sequence and up to 10 sequences configured, a program can run with a maximum of 100 steps.
EDIT LIST POWER L ist Powe r = 120 0.0 W 3. This parameter is configured to limit the maximum power of the program (list) when it is executed. Leave it default to 1200.0 for maximum power output. Press following will display: and the EDIT LIST REPEAT L ist Re pe at = 1 4. The display prompts the user to enter the number of times to repeat the program (list). For this example, enter 5 with the numeric keypad and then press . Note: The maximum number of repeat times is 65535 times. 5.
FILE A c ti ve Seq : 0 987654 3YY 8. Once these two sequences are selected (active), press SEQ REPEAT S eq 1 Repe at : to see the following: 1 9. The display prompts the user to enter the number of times to repeat sequence #1 as part of the program (list). In this example, use the current adjust knob or the numeric keypad to enter 2. Press and the same prompt will immediately follow for sequence #2. Enter 3 and then press . 10. Now it will display the following to confirm saving the program (list).
1. From the LIST menu, use the key to select Recall and press . The display will say RECALL LIST with the prompt Recall List File: to ask to enter a number. 2. This is the location number where the program is stored in. Select a number between 1 to 10 with the current adjust knob or the numeric keypad, then press 3. The display will return back to the LIST menu. . Run Program To run a program sequence after it is recalled, the trigger source must be configured first.
3.6 9115-AT Automotive Test Functions Note: This functionality is only available on model 9115-AT The 9115-AT power supply has a unique automotive testing feature built-in that allows the user to run stress tests based on the DIN 40839 and ISO 16750-2 standards. The functions have the ability to test 12 volt and 24 volt systems. For details on the DIN 40839 and ISO 16750-2 standards, please see www.din.de (DIN standards) and www.iso.org (ISO standards).
Table 3.4 – Parameters for test pulse Follow the steps below to set up the DIN40839 function: 1. Access the main menu by pressing and the 2. Use the key to select FUNC. and press 3. Use the key to select DIN40839 and press . . . The screen below will display: DIN40839 12 V Of f 4. While 12V or 24V is flashing, use keys to toggle between the two options. Set to 12V for the 12V test function and 24V for the 24V test function. 5.
0.000V Trig 0.0 W 0.000A DIN408 39 7. To enable the DIN40839 function, press the key, on the front panel. The proper voltage (12V or 24V) should be metered on the front panel display. 8. To run the DIN40839 function press shift and enter to trigger the test to run. 9. Repeat step 8 to trigger the function to run as many times as needed. ISO 16750-2 Parameters The ISO16750-2 function contains three different test simulations, Short voltage drop, Profile for the reset test, and Starting profile.
Figure 3.10 – Profile for the reset test waveform Figure 3.
Table 3.5 – Starting profile values for 12 V system devices Table 3.6 – Starting profile values for 24 V system devices Short voltage drop Follow the steps below to set up the ISO16750-2 for Short voltage drop function: 1. Access the main menu by pressing and the 2. Use the key to select FUNC. and press 3. Use the key to select ISO16750-2 and press 52 . . .
ISO16750-2 profile S ho rt vol tage d rop 4. While Short voltage drop is flashing, press to access that option. Set to 12V for the 12V test function and 24V for the 24V test function. 5. Use the key to select Off/On and use keys to toggle between the two options. Choose On to be able to run the function. 6. Once both parameters are set, press to confirm the changed settings and press several times to exit the menu.
ISO16750-2 profile P rof il e fo r th e reset test 5. Select Usmin… and press enter. The user can choose a Usmin value between 1 V – 80 V. Once a value is selected, press to confirm. 6. Use the key to select Off/On and use keys to toggle between the two options. Choose On to be able to run the function. 7. Once both parameters are set, press to confirm the changed settings and press several times to exit the menu.
ISO16750-2 profile S tar tin g p rof i le 5. Set to 12V for the 12V test system and 24V for the 24V test system. 6. Use the key to select the profile of 1 / 2 / 3 / 4(12 V only) and use toggle between the four options. 7. Use the key to select Repeat and press . This option allows the user to repeat the profile up to 100 times. Once a value is selected, press 8. Use the key to select Delay and press keys to to confirm. . This option allows the user to delay the repeated profile up to 100 seconds.
3.7 Configure Voltage Slope The power supply has the capability of controlling the slope of the output voltage during changes between voltage settings. Timings can be configured for both the rising and falling edge between voltage output transitions. To illustrate this feature and provide configuration instructions, follow the example below.
2. Trise is the rising edge time for an output voltage transition. Use the numeric keypad or the current adjust knob and enter 5.000 s (5 seconds). 3. Press and Tfall will be displayed. Tfall is the falling edge time for an output voltage transition. 4. Use the numeric keypad or the current adjust knob and enter 10.000 s (10 seconds). 5. Press and the power supply will return to the normal display. 6. Now, set voltage to 5 V using the numeric keypad or the voltage adjust knob.
4 Remote Operation 4.1 Interface Connection RS-232 For RS-232 connectivity, refer to the diagram below for pinout information. The RS-232 is labeled in the rear panel and it is a female DB-9 interface. 5 4 9 3 8 PIN 1 2 3 4 5 6 7 8 9 2 7 1 6 Description Transmit Data Receive Data GND - A straight pin-to-pin DB9 female to DB9 male serial cable is required for using the RS-232 interface. Do not use a null modem or crossover DB9 serial cable.
RS-485 For multi-unit configuration and control, the male DB-9 interface labeled RS-485 in the rear panel is used. The figure below illustrates the connection pins and description. Note: Pin 1 is used as the A pin (-) (non-inverting). Pin 5 is used as the B pin (+) (inverting). SC (reference) pin is not used. 5 4 9 3 8 PIN 1 2 3 4 5 6 7 8 9 2 7 1 6 Description A (-) B (+) - USBTMC The device is SR1, RL1, and DT1 enabled.
4.2 Remote Commands The instrument supports some SCPI commands and some instrument specific commands. These commands enable a computer to remotely communicate and control the power supply over any of the supported remote interfaces: USBTMC, RS-232, GPIB, RS-485, and Ethernet (Model 9117 only). Refer to the programming manual for details, which can be downloaded from www.bkprecision.com.
5 Troubleshooting Guide Below are some frequently asked questions and answers. Please check if any apply to your power supply before contacting B&K Precision. General Q: I cannot power up the power supply. - Check that the power cord is securely connected to the AC input and there is live power from your electrical AC outlet. - Verify that the AC power coming from the mains have the correct voltage. The power supply can accept a specific range of AC input voltages. Refer to section “2.1”.
6 Specifications Note: All specifications apply to the unit after a temperature stabilization time of 15 minutes over an ambient temperature range of 23 °C ± 5 °C. Specifications are subject to change without notice. Environmental Conditions: This power supply is designed for indoor use and operated with maximum relative humidity of 95%.
7 Calibration We recommend a calibration interval of once per year to ensure that the power supply meets specifications. The instrument features closed-case calibration, however it is performed remotely via the remote interface.
SERVICE INFORMATION Warranty Service: Please go to the support and service section on our website at www.bkprecision.com to obtain a RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Please go to the support and service section on our website at www.bkprecision.com to obtain a RMA #.
LIMITED ONE-YEAR WARRANTY B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of one year from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt.
22820 Savi Ranch Parkway Yorba Linda, CA 92887 www.bkprecision.com © 2014, 2015 B&K Precision Corp.