INSTRUCTION MANUAL MANUAL DE INSTRUCCIONES MODEL 4011A MODELO 4011A 5 MHz FUNCTION GENERATOR with DIGITAL DISPLAY 5MHz GENERADOR DE FUNCIONES CON DIGITOS INDICADOR 1
TEST INSTRUMENT SAFETY WARNING Normal use of test equipment exposes you to a certain amount of danger from electrical shock because testing must sometimes be performed where exposed voltage is present. An electrical shock causing 10 milliamps of current to pass through the heart will stop most human heartbeats. Voltage as low as 35 volts dc or ac rms should be considered dangerous and hazardous since it can produce a lethal current under certain conditions.
Instruction Manual for MODEL 4011A 5 MHz FUNCTION GENERATOR with DIGITAL DISPLAY 22820 Savi Ranch Parkway • Yorba Linda, CA 92887-4610 3
TABLE OF CONTENTS page page TEST INSTRUMENT ..................SAFETY inside front cover MAINTENANCE ........................................ 15 INTRODUCTION . ............................................................. 5 Fuse Replacement ........................................ 15 SPECIFICATIONS ............................................................. 6 Instrument Repair Service............................. 15 CONTROLS AND INDICATORS ......................................
INTRODUCTION The heart of the function generator is a VCG (voltagecontrolled generator) that produces precision sine, square, or triangle waves over the 0.5Hz to 5MHz range. This encompasses subaudible, audio, ultrasonic, and RF applications. A continuously variable dc offset allows the output to be injected directly into circuits at the correct bias level. The B+K Precision Model 4011A Function Generator is a versatile signal source which combines several functions into one unit.
SPECIFICATIONS FREQUENCY CHARACTERISTICS Attenuation -20 dB ±1 dB DC Offset Preset: Variable: Waveforms Sine, Square, Triangle, ± Pulse, ± Ramp Range 0.5Hz to 5MHz in 7 ranges Resolution 4 digits Tuning Range Coarse: 10:1, Fine: ±5% of Coarse Setting Variable Duty Cycle 15:85:15 Continuously Variable Operating Modes Normal, VCG (Voltage Controlled Generator) Frequency Stability Output frequency will not change more than 0.09% in 15 minutes after 1 hour warmup. ±0.
SPECIFICATIONS TTL OUTPUT Level Rise Time Duty Cycle CMOS OUTPUT Max Frequency Level Rise Time Time Base Accuracy ±10 PPM (23 °C ±5 °C) Display 4 digit LED > 2.4V ≤20nS 50% typical POWER SOURCE 120 / 230VAC ±10%, 50 / 60Hz, internal jumper selectable 2MHz 4V to 14V ±0.5V p-p, Continuously variable ≤120nS DIMENSIONS (H x W x D) 10-3/8" x 3-3/8" x 11-7/16" (26.4cm x 8.6cm x 29.
CONTROLS AND INDICATORS FRONT PANEL (Refer to Fig. 1) 1. POWER Switch. Turns power on and off. 2. RANGE Switch. Selects output frequency range. Seven ranges from 5Hz to 5MHz. Switch indicates maximum frequency of range and is adjusted with COARSE FREQUENCY control to 0.1 times the maximum. For example, if the 500kHz range is selected, the output frequency can be adjusted from 50kHz to 500kHz. 3. FUNCTION Switch. Selects sine, square, or triangle waveform at OUTPUT jack. 4. OUTPUT LEVEL Control.
Figure 1. Model 4011A Controls and Indicators. 13. CMOS LEVEL Switch. When engaged, changes the TTL signal to CMOS signal at the TTL/CMOS jack. 14. DUTY CYCLE Switch. When engaged, enables operation of DUTY CYCLE control (10). 15. FINE FREQUENCY Control. Vernier adjustment of the output frequency for ease of setting frequency. 16. COARSE FREQUENCY Control. Coarse adjustment of the output frequency from 0.1 to 1 times the selected range. 17. COUNTER DISPLAY.
OPERATING INSTRUCTIONS The B+K Precision Model 4011A Function Generator is a versatile instrument, capable of producing a variety of output waveforms over a broad range of frequencies. To gain a working familiarity with the unit, it is recommended that it be connected initially to an oscilloscope, so that the effects of the various controls on the output waveforms can be observed. Use this manual as required for reference until becoming accustomed to the operating procedures.
OPERATING INSTRUCTIONS 7. A superimposed DC component can be added to the output signal by engaging the DC OFFSET switch (12) to enable operation of the DC OFFSET control (5). Rotation of this control adds a positive or negative DC component to the output signal. The DC component introduced is independent of the OUTPUT LEVEL control and can be varied by ±10 volts open circuited or ±5 volts into 50Ω. The DC Offset does not affect the TTL/CMOS output jack. The effect of DC OFFSET is shown in Fig. 3.
OPERATING INSTRUCTIONS DUTY CYCLE CONTROL The DUTY CYCLE control can be used to alter the symmetry of the output waveform, to produce waveshapes such as those shown in Fig. 4. For a square wave, symmetry variation amounts to changing the duty cycle (ratio of "high" to "low" time), effectively converting the instrument into a pulse generator. For a triangle wave, the result is a ramp, and with a sine wave, a distorted waveshape called a slewed sine is produced.
OPERATING INSTRUCTIONS TTL/CMOS OUTPUT The TTL/CMOS output jack provides a fast rise time square wave output. Either a fixed TTL or a variable CMOS output level is available. The output is positive with respect to ground and can be used as an external sync pulse for oscilloscopes or as a variable frequency signal source for exercising logic circuits. Because of the fast rise time of this output, cable length should be minimized to limit ringing and overshoot. 1.
OPERATING INSTRUCTIONS 2. If in doubt about the safety of a signal injection point, measure the voltage present at the intended point of signal injection before connecting the function generator output to that point. 3. When applying the main output of the function generator to a circuit point containing a dc level, adjust the DC OFFSET control so that the dc level at the main output matches the circuit voltage. 4. Connect the TTL output only to TTL-level circuits.
MAINTENANCE WARNING INSTRUMENT REPAIR SERVICE The following instructions are for use by qualified service personnel only. To avoid electrical shock, do not perform servicing other than contained in the operating instructions unless you are qualified to do so. Because of the specialized skills and test equipment required for instrument repair and calibration, many customers prefer to rely upon B+K PRECISION for this service.
CUSTOMER SUPPORT 1-800-462-9832 B+K Precision offers courteous, professional technical support before and after the sale of their test instruments. The following services are typical of those available from our toll-free telephone number: Technical advice on the use of your instrument. Technical advice on special applications of your instrument. Technical advice on selecting the best instrument for a given task. Information on optional accessories for your instrument.
Service Information Warranty Service: Please return the product in the original packaging with proof of purchase to the address below. Clearly state in writing the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Return the product in the original packaging to the address below. Clearly state in writing the performance problem and return any leads, probes, connectors and accessories that you are using with the device.
Limited Two-Year Warranty B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of two years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt. To obtain warranty coverage in the U.S.A.
TEST INSTRUMENT SAFETY (continued from inside front cover) 6. Some equipment with a two-wire ac power cord, including some with polarized power plugs, is the “hot chassis” type. This includes most recent television receivers and audio equipment. A plastic or wooden cabinet insulates the chassis to protect the customer. When the cabinet is removed for servicing, a serious shock hazard exists if the chassis is touched.
SEGURIDAD DE EL INSTRUMENTO DE PRUEBA PRECAUCIONES Use normal de equip de probe tee espouse a caret cantina de peligro por un choque eléctrico porque revisiones son algunas veces hechas donde hay alto voltaje descubierto. Un choque eléctrico que cause 10 milliamps pasar a través del corazón pararía la mayoría de los corazones humanos. Voltaje tan bajo hasta 30 voltios dc o ac rms podría ser considerado peligroso porque puede producir una corriente letal bajo ciertas condiciones.
MANUAL DE INSTRUIONES MODELO 4011A 5MHz GENERADOR DE FUNCIONES CON DIGITOS INDICADOR 22820 Savi Ranch Parkway • Yorba Linda, CA 92887-4610 21
TABLA DE CONTENDIDO SEGURIDAD DEL INSTRUMENTO DE PRUEBA Operación de voltaje controlada frecuencia ..........................31 INTRODUCION ...................................................................... 23 Consideraciones de protección de salida...............................31 ESPECIFICACIONES ............................................................. 25 Guía de aplicaciones del generador de funciones..................32 CONTROLESE INDICADORES ............................................
INTRODUCCIÓN triangulares desde los 0.5Hz hasta 5MHz rango. Esto en compasa subaudible, sonido, untrasonido, ultrasonico y RF aplicaciones. Un continuamente variable compensación de cd permite a la salida ser injectada dentro de circuitos al el correcto nivel medio. El generador de barrido ofrece linear y logarítmico barrido con un rango variable de barrido y tiempo de barrido ajustable. El tablero de control delantero permite el ajuste de comienzo y alto de frecuencias.
ESPECIFICACIONES CARACTERÍSTICAS DE FRECUENCIA Atenuación: -20db+/-1db DC compensación : Preset: ±0.1V typico Variable ±10V circuito abierto ,±5V Dentro 50Ω Formas de onda: seno,cuadrada,triangular±-Impulso,±rampa Rango: 0.5Hz a 5MHz en 7 rangos Resolución 4 dígitos Afinación rango: gruesco 10:1, Fino+/--5% de posición gruesa Variable ciclo de servicio: 15:85:15 continuamente variable Modos de operación: normal,barrido,VCG,AM,FM estallido Estabilidad de frecuencia: la salida combiara menos que 0.
ESPECIFICACIONES SALIDA TTL Nivel Tiempo de elevación Ciclo de servicio: CMOS SALIDA Max frecuencia: Nive: Tiempo de elevación 0.8V a 4V +0.5Vp-p ≤20ns 50% tipico 2Mhz 4V a 14 V +/-0.
CONTROLES E INDICADORES PANEL DELANTERO (Vea a al Fig 1) 1. INTERRUPTOR DE ENCENDIDO. Enciende y apaga el poder. 2. SWITCH DE RANGO. Selecciona el rango de la frecuencia de salida. 7 rangos de 5Hz a 5MHz. El switch indica la máxima frecuenta de rango y es ajustado con el control grueso de frecuencia a 0.1 veces el máximo. Por ejemplo, si el rango de 500KHz es seleccionado, la salida de frecuencia puede ser ajustada de 50KHz a 500KHz 3. SWITCH DE FUNCIONES.
CONTROLES E INDICADORES 16. Control GRUESO DE FRECUENCIA. Ajuste grueso de la frecuencia de salida de 0.1 a 1 veces el rango seleccionado. 17. PANTALLA DE EL CONTADOR. Muestra la frecuencia de formas de onda generadas internamente. 18. LED PUERTA. Indica cuando la pantalla de el contador cambia. Cuando los 50KHz a 5MHz rangos son seleccionados, el LED relampaguea 10 veces por segundo (cada 0.1 segundos).
INSTRUCCIONES DE OPERACION El B&K Precison Modelo 4011A Barrido/Funcion generador es un instrumento versátil, capaz de producir una variedad de formas de onda a través un ancho rango de frecuencias. Para ganar una buena familiaridad con esta unidad, es recomendable que este conectada inicialmente a un osciloscopio, para que los efectos de los varios controles en las formas de onda que son generadas puedan ser observadas.
INSTRUCCIONES DE OPERACION 5. Un CD componente puede ser sumado at la senal de salida oprimiendo en la compensación de CD switch (12) para activar la cooperación del COMPENSACION CD control (5). Rotación de este control suma un positivo o negativo compensación CD componente a la senal de salida. El componente CD introducido es independiente de el control de el NIVEL DE SALIDA y puede ser variado por +/- 10 voltios circuito abierto o +/- voltios atraves de 50 ohms.
INSTRUCCIONES DE OPERACION CONTROL DE CICLO DE SERVICIO El CICLO DE SERVICIO control puede ser usado para alterar la simetría de la forma de onda de salida, para producir ondas de forma tal como esas mostradas en Fig. 4. Por una onda cuadrada, la cantidad de variación de simetría a cambiar el ciclo servicio (factor de “alto” a “bajo” tiempo), efectivamente convirtiendo el instrumento en un generador de pulsos.
INSTRUCCIONES DE OPERACION TTL/CMOS SALIDA El TTL/CMOS jack de salida provee un rapido tiempo de subido onda cuadrada de salida. Sea un fijo TTL o un variable CMOS nivel de salida esta disponible. La salida es positiva con respecto a tierra y puede ser usada como un pulso sync para osciloscopios o como una senal variable de frecuencia para ejercitando circuitos de logica. Por el rapido tiempo de subida de esta salida, lo largo de el cable debe de ser minimizado para limitar riesonaciony sobretiro. 1.
INSTRUCCIONES DE OPERACION 2. Si en duda acerca de la seguridad de un punto de injeccion de senales, mida el voltaje presente en el punto de injeccion de senales intentado antes de connectar la salida de el generador de funciones a este punto. 3. Cuando aplique la salida principal de el generador de senales a un punto de un circuito que contenga un nivel de cd, ajuse el CD AJUSTE control para que el nivel de cd a la salida principal iguale el voltaje de el circuito. 4.
MANTENIMIENTO REEMPLAZO DE FUSIBLES PRECAUCION 1. 2. Las siguientes instrucciones son para uso solo por personal de servico calificado. Para evitar choque electrico, no haga servicio otro que el contenido en las instrucciones de operacion a menos que este calificado para hacerlo. Localice el sostenedor de fusible en la linea de entrada. Quite el sostenedor del fusible y dubstituya el fusible por un fusible de valor egual.
SOPORTE AL CLIENTE 1-800-462-9832 b+k Precision ofrece soporte técnico profesional y cortés antes y después de la venta de sus equipos de prueba.
Información de Servicio Servicio de Garantía: Por favor regrese el producto en el empaquetado original con prueba de la fecha de la compra a la dirección debajo. Indique claramente el problema en escritura, incluya todos los accesorios que se estan usado con el equipo. Servicio de No Garantía: Por favor regrese el producto en el empaquetado original con prueba de la fecha de la compra a la dirección debajo.
Garantía Limitada de Dos Anos B&K Precision Corp. Autorizaciones al comprador original que su productos y componentes serán libre de defectos por el periodo de dos anos desde el día en que se compro. B&K Precision Corp. sin carga, repararemos o sustituir, a nuestra opción, producto defectivo o componentes. Producto devuelto tiene que ser acompañado con prueba de la fecha del la compra en la forma de tres recibo de las ventas. Para obtener cobertura en los EE.UU.
1. 2. 3. 4. (continuacion desde el ) Algun equipo con un dos-conductores ca cordon de poder, incluyendo algunos con polarizados tapones de poder, es el “chasis caliente” tipo. Esto incluye la mayoria de los mas recientes receptores de television y el equipo de sonido. Un cabinete de plastico o de madera insula el chasis para proteger al cliente. Cuando el cabinete es removido para servicio, un serio peligro de choque existe si el chasis es tocado.
Declaration of CE Conformity according to EEC directives and NF EN 45014 norm Responsible Party Alternate Manufacturing Site Manufacturer’s Name: B&K Precision Corporation B&K Taiwan 0574 Manufacture’s Address: 22820 Savi Ranch Pkwy.
22820 Savi Ranch Parkway • Yorba Linda, CA 92887-4610 © 2005 B+K Precision 480-844-9-001 Printed in U.S.A.