IDEAL LIFE Gluco-Manager TM Blood Glucose Management System Model GMM00010001 INSTRUCTION MANUAL Read before use May be used with IDEAL LIFE Pod™ (sold separately) for use with IDEAL LIFE Health Management System™ Products may vary from images shown *NOTE: The IDEAL LIFE Gluco-Manager™ is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm,the calf and the thigh.
Table of Contents Table of Contents Table of Contents......................................2 Introduction.................................................3 Preface..........................................................3 Contraindications......................................3 Precautions..................................................4 Important Information..............................4 About Alternative Site Testing (AST)....6 Package Contents......................................
Introduction Introduction Preface Thank you for purchasing the IDEAL LIFE Gluco-Manager™, or GMM 0001 and IDEAL LIFE Pod™, or ILP 0001. This completely automatic system measures your blood glucose, captures subjective information and manages information automatically so that you can better manage your blood glucose with your care provider.
Precautions Precautions • Do not drop this unit. Please avoid strong impact. Disassembly or modification of the unit may lead to permanent reading error. • The device might not meet its performance specifications if stored or used beyond the specified temperature and humidity ranges. • Operating conditions: 50°F to 104°F (10°C to 40°C), below 80% RH. • System storage: 39.2°F to 86°F (4°C to 30°C), below 80% RH. • Be sure to keep this manual for future reference.
Important Information Important Information • Severe dehydration and excessive water loss may cause false low results. If you believe you are suffering from severe dehydration, consult a healthcare professional immediately. • Test results below 60 mg/dL (3.3 mmol/L) indicate hypoglycemia, or low glucose. Test results higher than 240 mg/dL (13.3 mmol/L) indicate hyperglycemia, or high glucose. Consult your physician when one of these results is shown repetitively.
About Alternative Site Testing (AST) About Alternative Site Testing (AST) IMPORTANT: There are limitations for doing alternative site testing (AST).Please consult your healthcare professional before you do AST. What is AST? Alternative site testing (AST) means that people use parts of the body other than fingertips to check their blood glucose levels. This system provides you to test on your forearm, upper arm, calf, or thigh with the equivalent results to fingertip testing when used at appropriate times.
Package Contents Package Contents How to increase the accuracy? Stimulating blood perfusion by rubbing the puncture site prior to blood extraction has a significant influence on the glucose value obtained. Blood from the site without rubbing exhibits a measurably different glucose concentration than blood from the finger. When the puncture site was rubbed prior to blood extraction, the difference was significantly reduced.
Additional Items Additional Items (*sold separately) There are 3 levels for control solutions. Please call 1-866-IDEAL-41 (866-433-2541) for availability. 1 CLEAR CAP LANCING DEVICE 25 TEST STRIPS 25 STERILE LANCETS CONTROL SOLUTION ILP 0001 IDEAL LIFE POD™ AC ADAPTER NOTE: Check your system to be sure that it is unopened prior to use and that it contains all parts listed above. If either of these conditions occurred, please return your system to the place of purchase.
Location/Interference Location/Interference The IDEAL LIFE Manager™ may be used with a communication gateway such as the IDEAL LIFE Pod™ or access point such as a cellular phone or computing device. The IDEAL LIFE Pod™ uses wireless radio waves to transfer information from your IDEAL LIFE Manager™.
Before You Start Before You Start TM Your IDEAL LIFE Gluco-Manager Test Port Blue LED LCD Display Left “Soft”Button Right “Soft”Button START/ESCAPE Button AAA Batteries Battery Cover Battery Cover 10 IDEAL LIFE Gluco-Manager™
Your IDEAL LIFE PODTM Your IDEAL LIFE Pod™ (*An optional accessory sold separately) LED Display AC Adapter Jack USB Port Ethernet Cable RJ45 Jack Connection Telephone Cord RJ11 Jack For Telephone Connection (sealed for cellular internet connectivity) (Sealed with an antenna for cellular Internet connectivity) Telephone Connection Telephone Cord RJ11 Jack For Line Connection Battery Cover * Note: There is only a battery compartment for telephone line connectivity Instruction Manual 11
Test Strip Description Test Strip Description (Model: AGS-1112) • Your system measures the amount of sugar (glucose) in whole blood. Blood is applied to the absorbent hole of the test strip and is automatically drawn into the reaction cell where the reaction takes place. • The test strip consists of the following parts: Contact Bars Insert this end of the test strip into the meter. Push it in firmly until it will go no further Test Strip Handle Hold this part to insert the test strip into the slot.
Powering Up Your IDEAL LIFE Gluco-Manager™ Powering Up Your IDEAL LIFE Gluco-Manager™ INSTALLING THE BATTERIES (INCLUDED) • Remove the battery compartment door on your IDEAL LIFE Gluco-Manager™ by lifting the battery door using the notch. • Place the batteries as instructed on the device. • Replace the battery compartment door and click into place. • NOTE: Replacing the battery does not affect previous test results stored in memory of the meter.
First Time Power Up And Activation When Using An IDEAL LIFE POD™ First Time Power Up And Activation When Using An IDEAL LIFE POD™ (*an optional accessory, sold separately) To use the IDEAL LIFE Gluco-Manager™with your IDEAL LIFE Pod™you must call an IDEAL LIFE customer representative to activate. Make sure to call IDEAL LIFE before you begin.
About Control Solution About Control Solution (sold separately) Control solutions contain a known amount of glucose that reacts with test strips. By comparing your control solution test results with the expected range printed on the test strip vial label, it is able to check that the meter and the test strips are working together as a system and that you are performing the test correctly. It is very important that you do this simple check routinely to make sure you get accurate results.
How To Do A Control Solution Test? How To Do A Control Solution Test? Step 1. Power up Device From the off position, press the power button to turn on the display. From the default screen, press the power button once more. The display will instruct you to insert a strip at this time. Insert a test strip with contact bars end first and facing up into the test slot. (Contact bars must be inserted all the way into the meter or you may get an inaccurate test result.
How To Do A Control Solution Test? CAUTION! To avoid contaminating the control solution with the content of the test strip, you have to place a drop of control solution on a clean surface or on your finger tip first. Do not directly apply control solution into a strip. Step 3. Display test result Compare the result with the range printed on the test strip vial. The result should fall within this range.
Functions Functions (when using the IDEAL LIFE POD™, an optional accessory sold separately) Main Screen The Main Screen is your default screen. This is where you will be directed to after completing functions and where you will find status indicators for the functioning of your IDEAL LIFE Gluco-Manager™.
Taking Your Blood Glucose Reading Taking Your Blood Glucose Reading Be sure to read this section and any inserts found in the test strip box carefully before taking a reading. Wash and dry your hands before starting. Getting a Drop of Blood Step 1. Use your lancing device (sold separately) to obtain a blood sample. Follow the instructions included in your lancing device. WARNING! To reduce the chance of infection: • Never share a lancet or the lancing device with anyone.
Taking a Reading Taking a Reading Step 1. Insert the Test Strip Insert a test strip with contact bars end first and facing up into the test slot. (Contact bars must be inserted all the way into the meter or you may get an inaccurate test result.) The meter turns on automatically and shows the following messages: Step 2. Apply Sample Obtain a drop of blood using the adjustable lancing device. The blood a sample must be at least 0.7 microliter in volume or you may get an inaccurate test result.
Looking At Previously Taken Readings Looking At Previously Taken Readings Your IDEAL LIFE Gluco-Manager™ will store up to 575 of your latest readings. Press Menu to view the feature list. SCROLL to highlight READINGS (it should already be highlighted as it is the first line item in the list). Press SELECT to enter your selection. You can now scroll through your readings, screen by screen using the SCROLL button.
Looking At Messages Looking at Messages Your IDEAL LIFE Gluco-Manager™ is equipped for 2-way, interactive communication when using a communication gateway or access point such as a cellular phone, computing device or the IDEAL LIFE Pod™. Messages can be sent remotely to your IDEAL LIFE Gluco-Manager™ and responded to directly. When messages are transmitted to your IDEAL LIFEGluco-Manager™, they are stored for automatic display before your next blood glucose measurement.
Looking at Graphics Looking at Graphics When using a communication gateway or access point such as a cellular phone, computing device or the IDEAL LIFE Pod™ your IDEAL LIFE Gluco-Manager™will store different types of graphics downloaded from the IDEAL LIFE website. (All the graphics can be altered through the IDEAL LIFE website, refer to www.ideallifeonline.com for details). Press Menu to view the feature list. • SCROLL to highlight GRAPHICS. • Press SELECT to enter your selection.
Comparing Meter And Laboratory Results Comparing Meter And Laboratory Results The meter provides you with whole blood equivalent results.The result you obtain from your meter may different somewhat from your laboratory result due to normal variation. Meter results can be affected by factors and conditions that do not affect laboratory results in the same way (see test strip package insert for typical accuracy and precision data and for important information on limitations).
Care and Maintenace Care and Maintenance • Handle the meter with care. Dropping or throwing the meter may cause damage to the device. • Don’t expose the Blood Glucose Monitoring system, Test strip, and Control solution, to extreme conditions, such as high humidity, heat, freezing cold or dust. Storage conditions for the blood glucose Monitoring system (meter and test strips) are 39.2°F ~86°F Humidity <80% RH. Storage conditions for the meter are -4°F ~131°F, Humidity <80%RH.
Cleaning and Disinfection Cleaning and Disinfection The cleaning and disinfection is absolutely necessary for the test procedure. Cleaning can insure that the meter works well (for example, the display will be clear to see after cleaning); and disinfection can avoid infection. We strongly advise that the glucometer and the lancing device be cleaned and disinfected following every use. We suggest you use the following product: CaviWipes (EPA registration number: 46781-8).
Cleaning and Disinfection Note: 1. Each disinfection step requires a pre-cleaning step. 2. The users should wash hands thoroughly with soap and water after handling the meter, lancing device, or test strips. 3. If you have any questions you can contact an IDEAL LIFE customer representative at 1-866-IDEAL-41 (866-433-2541) to resolve the issue. 4.
Specifications Specifications IDEAL LIFE Gluco-Manager™ Model: GMM0001 IDEAL LIFE Gluco-Manager™ Display: LCD Digital Display Power Source: Two 1.5V AAA alkaline batteries Battery Life: Over 200 determinants OperatingTemperature/Humidity: 50°F to 104°F (10°C to 40°C), below 80%R.H. (non-condensing) Meter Storage Temperature/Humidity: -4°F to 131°F (-20°C to 55°C), below 80%R.H. Weight: Dimension: 77.3 mm(l) x 60.
Limitations of Use Limitations of Use • The blood glucose monitoring system is not intended for use on neonates. • The blood glucose monitoring system is not intended for use on artery blood, neonates’ serum and plasma. • IDEAL LIFE Gluco-Manager™ should be used with IDEAL LIFE test strip. • The system can be used up to an altitude of 10744 feet.
Problem Solving Guide Problem Solving Guide The following is a summary of potential problems in operation to help troubleshoot various issues. Improper use may cause an inaccurate result without producing an error message or a symbol. In the event of a problem, refer to the information under action. If you follow the recommended instructions but a problem still exists, please contact an IDEAL LIFE™ customer representative at 1-866-IDEAL-41 (866-433-2541) to resolve the issue.
Problem Solving Guide Error Message MESSAGE LOW BATTERY WHAT IT MEANS ACTION Appears when the batteries can not provide enough power for a test. Replace the batteries immediately. Appears when inserting a used test strip. Test with a newtest strip. Remove the strip after applying blood to the absorbent hole. Re-test with a new test strip. Problem with the meter. Review the instructions and re-test with a newtest strip.
Summary of Operation Summary of Operation This summary is intended only for quick reference and not as a substitute for the instruction manual. Please read the entire manual before you begin testing. Step 1. Insert test strip The meter is automatically turned on and displays the “ ” symbol Step 2. Apply sample Select and clean the puncture site. Obtain a drop of blood by lancing the puncture site.
Federal Communications Commission (FCC) Statement Federal Communications Commission (FCC) Statement 15.21 You are cautioned that changes or modifications not expressly approved by the part responsible for compliance could void the user’s authority to operate the equipment. 15.105(b) This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules.
Customer Information Customer Information This equipment complies with Part 68 of the FCC rules and the requirements adopted by the ACTA. On the bottom of this equipment is a label that contains, among other information, a product identifier of [US: 5CQMM00BILP0001]. If requested, this number must be provided to the telephone company. If this equipment [ILP 0001] caused harm to the telephone network, the telephone company will notify you in advance that temporary discontinuance of service may be required.
Service Requirements Service Requirements In the event of equipment malfunction, all repairs should be performed by our Company or an authorized agent. It is the responsibility of users requiring service to report the need for service to our Company or to one of our authorized agents. Please refer to the location you purchased or received your equipment for details of where service can be facilitated through.
IDEAL LIFE US INC. Between 9:00 am and 8:00 pm (EST) Tel 1-866-IDEAL-41 1-866-433-2541 www.IDEALLIFEonline.com At all other times, you should contact your healthcare professional for assistance. © 2011 - V1 IDEAL LIFE INC. All rights reserved. IDEAL LIFE is a registered trademark in the United States for weight scales, blood glucose meters, blood pressure cuffs and meters, heart monitors.