Part number 20068, Rev 0 Pharmacy Refrigerator and Freezer Combination ABT-RFC-7P OWNER’S INSTRUCTIONS This manual describes how to operate and care for your appliance to get the best, most efficient, performance. READ THIS BOOK! Note to Customer: This merchandise was carefully packed and thoroughly inspected before leaving our plant. Responsibility for its safe delivery was assumed by the carrier upon acceptance of the shipment.
Part number 20068, Rev 0 TABLE OF CONTENTS RECEIVING............................................................................................... 3 INSTALLATION INSTRUCTIONS ................................................................ 4 REFRIGERATOR OPERATION.................................................................... 6 ELECTRONIC TEMPERATURE CONTROL ................................................... 7 QUICK TROUBLESHOOTING GUIDE .......................................................
Part number 20068, Rev 0 RECEIVING Your unit was built, packaged, and inspected with extreme care. We shipped it to you using carriers we trust with a proven track record of careful handling, good customer service, and on time delivery. Unfortunately, regardless of all of these efforts sometimes accidents happen and occasionally those accidents result in shipping damage. When the carrier picked up the merchandise from us, they assumed responsibility for its condition en route to you.
Part number 20068, Rev 0 INSTALLATION INSTRUCTIONS UNCRATING Move your refrigerator as close to the final location as possible before unpacking. Remove the carton and the styrofoam strips. The location should be as close as possible to the power outlet so no extension cord is needed. LEVELING You can level your unit with the screwtype adjustable leveling legs on the front corners of the unit. Turn counterclockwise to raise the corner; turn clockwise to lower it. Leveling legs are required in front only.
Part number 20068, Rev 0 STARTING There are no compressor shipping bolts to loosen or valves to open. All that is necessary after the unit has been properly leveled is to plug the service cord into an electrical outlet When starting this new appliance, allow the cabinet to operate a minimum of four hours or until it has started cycling normally before placing product in the cabinet.
Part number 20068, Rev 0 REFRIGERATOR OPERATION After the unit is properly installed and power is applied, it will take some time before the system is cooled down to temperature and cycling normally. You should wait 4 hours on the first startup before beginning to add product to the unit. This ensures that the unit is installed and operating properly before being put into service. After this wait time, the unit should be cycling in the desired temperature band.
Part number 20068, Rev 0 ELECTRONIC TEMPERATURE CONTROL Description and Functions The Electronic Temperature Control is a digital microprocessor based temperature control, indication, and alarm system. It continuously monitors the temperature of the refrigerator/freezer through its sensor assembly. When the temperature warms to the point where it reaches the upper temperature limit, the controller sends a signal to turn on the compressor, which cools the unit.
Part number 20068, Rev 0 Programming Mode Note: The programming mode gives access to several functions that are not intended for user adjustment. The procedures below provide direction to the functions that you can adjust. Please do not attempt to change any of the other functions in the programming area or your unit may be damaged or become unstable. Locking and Unlocking the Control The control keys can be locked out to prevent inadvertent operation.
Part number 20068, Rev 0 Hysteresis is set at 3, then the compressor will cycle from 3 to 6 degrees Celsius. To change this value, press SET and DOWN keys for 3 seconds to enter programming mode. “Hy” will display meaning Hysteresis. Press Set again and the current value will be displayed. Use the UP and/or DOWN arrow keys to adjust the value and press SET again to lock it in. Press SET and UP to return to Operating Mode.
Part number 20068, Rev 0 REFRIGERATOR OPERATION TEMPERATURE CONTROL NOTE: The control is preset at the factory to provide the desired air and product temperatures inside of the unit and requires no further adjustment. Please contact the manufacturer’s Technical Support Department before making any adjustments to determine if adjustment is necessary and, if so, to make sure it is performed properly. This unit has a manual thermostat which controls the compressor. It is located inside the compartment.
Part number 20068, Rev 0 NORMAL OPERATION After the unit is properly installed and power is applied, it will take some time before the system is cooled down to temperature and cycling normally. You should wait 8 hours on the first startup before beginning to add product to the unit. This ensures that the unit is installed and operating properly before being put into service. On subsequent startups (after cleanings, for example), this wait time can be reduced to about 3 hours.
Part number 20068, Rev 0 QUICK TROUBLESHOOTING GUIDE Check these items before calling for service PROBLEM: Unit does not run POSSIBLE CAUSE / SOULTIONS: Unit does not maintain at the proper temperature Appliance runs too long Temperature of external wall surface is warm Compressor noises Electrical circuit is not 110-120V 60Hz. The power cord is not plugged in. No power at electrical outlet. Check to make sure breaker is not tripped or fuse is not blown.
Part number 20068, Rev 0 PROBLEM: POSSIBLE CAUSE / SOULTIONS: Door gasket is not sealing properly. Check for debris, cracks, and items passing through door at the gasket. The refrigerator or freezer is facing a doorway or is underneath of air conditioning vent. Relocate the unit or redirect air vent. Too many door openings. Minimize time door is open. Hot, humid weather increases condensation. Make sure there is a water trap (U-shaped loop) in the drain tube near the compressor.
Part number 20068, Rev 0 MAINTENANCE AND CLEANING These units require very little maintenance. They should be cleaned periodically to keep them running efficiently. Use the cleaning agents and suggestions below for best results.
Part number 20068, Rev 0 Keep your equipment clean- Use proper cleaners at recommended strength. Clean frequently to avoid build-up of hard stubborn stains. Rinse thoroughly - If chlorinated cleaners are used, you must rinse and wipe dry immediately. The sooner you wipe off standing water, the better. After wiping the unit down, allow it to air dry. The oxygen helps maintain the stainless steels passivity film. Never use hydrochloric acid (muriatic acid) on stainless steel.
Part number 20068, Rev 0 PRODUCT WARRANTY INFORMATION HORIZON SCIENTIFIC, INC. PRODUCT WARRANTY Horizon Scientific, Inc. warrants to the original purchaser every new Horizon Scientific, Inc. refrigerated unit, the cabinet and all parts thereof, to be free from defects in material or workmanship, when such unit is installed, used, and maintained in accordance with provided instructions, for a period of two (2) years.
Part number 20068, Rev 0 OR NOT INSURED, SHALL NOT EXCEED THE COST OF THE PRODUCT(S) GIVING RISE TO THE CLAIM OR LIABILITY. WARRANTY IS NOT TRANSFERABLE. This warranty is not assignable and applies only in favor of the original purchaser/user to whom delivered. Any such assignment or transfer shall void the warranties herein made and shall void all warranties, express or implied, including any warranty of merchantability or fitness for a particular purpose.