User instructions ERGOcom 3 Always on the safe side.
Vertrieb / Distribution: KaVo Dental GmbH & Co. KG Bahnhofstraße 20 D-88447 Warthausen Tel.: ++ 49 / 73 51 / 56 - 0 Fax: ++ 49 / 73 51 / 56 16 27 Hersteller / Manufacturer: Kaltenbach & Voigt GmbH & Co.
ERGOcom 3 Table of contents Table of contents Table of contents ............................................................................................................................................3 1 User Notes ....................................................................................................................................................5 1.1 User guidelines ..........................................................................................................................
ERGOcom 3 Table of contents 5.3 Operation, with display interface (comfort/excellence) ..........................................................................35 5.3.1 Operational mode .........................................................................................................................35 5.3.2 VGA OUT - set image source .......................................................................................................37 5.3.3 VIDEO OUT 1 - set image source .........................
ERGOcom 3 1 User Notes | 1.1 User guidelines 1 User Notes 1.1 User guidelines Prerequisite Please read these instructions before using the product to avoid operator error and damage. 1.1.1 Abbreviations Abbreviation UI SI EI SFC IEC Meaning User instructions Setup instructions Engineer's instructions Safety checks International Electrotechnical Commission 1.1.
ERGOcom 3 1 User Notes | 1.2 Target audience 1.2 Target audience This document is intended for use by dentists and other dental practice employees.
ERGOcom 3 1 User Notes | 1.3 Service 1.3 Service Please send all questions about the product, services and maintenance to the following addresses. Always quote product serial number in any correspondence. KaVo Dental GmbH & Co. KG After Sales Service Bahnhofstraße 20 D-88445 Warthausen 07351-56 2700 07351-18218 kunden-service@kavo.de www.kavo.
ERGOcom 3 1 User Notes | 1.4 Provisions of guarantee 1.
ERGOcom 3 1 User Notes | 1.5 Transport and storage 1.5 Transport and storage 1.5.1 The German packaging ordinance, 28 August 1998 Note Applies only to the Federal Republic of Germany. KaVo transport packaging is disposed of and recycled by local waste management and recycling companies under Germany's Dual System. For more information about waste management and recycling, and for up-to-date lists of local waste management and recycling companies, visit the following sites: http://www.umweltdatenbank.
ERGOcom 3 1 User Notes | 1.5 Transport and storage Outside Germany Note KaVo shall not be liable for damage caused in transit. Check the shipment immediately upon delivery! If the outer packaging is noticeably damaged upon delivery, the following procedure must be adhered to: 1. The recipient must record the loss or damage on the notice of receipt. The recipient and transport company employee delivering the product must both sign the notice of receipt.
ERGOcom 3 1 User Notes | 1.5 Transport and storage Stacking limitation. Temperature limitations.
ERGOcom 3 2 Safety | 2.1 Description of safety instructions 2 Safety 2.1 Description of safety instructions 2.1.1 Hazard warning symbol Hazard warning symbol 2.1.2 Structure DANGER The introduction describes the type and source of the danger. This section shows what could happen if the instructions are not followed. ▶ The optional action shows what measures to take to avoid danger. 2.1.
ERGOcom 3 2 Safety | 2.2 Intended purpose 2.2 Intended purpose 2.2.1 General The operator is required to make sure that the device is in a fully functional and safe condition before commencing use. During use, the handler must comply with state stipulations; in particular: ▪ the industrial safety regulations that are in place. ▪ the regulations that are in place for the prevention of industrial accidents.
ERGOcom 3 2 Safety | 2.3 Safety instructions 2.3 Safety instructions 2.3.1 General WARNING CAUTION CAUTION Injuries or harm caused by damaged functional components Damaged functional components can cause personal harm or injury. ▶ If functional components are damaged: Stop work,remedythe problem or inform the service engineer. Malfunction due to electromagnetic fields This product meets all the requirements that are in effect relating to electromagnetic fields.
ERGOcom 3 3 Product description | 3.1 Product name 3 Product description 3.1 Product name ERGOcom 3 is supplied in classic, comfort and excellence models. These models differ with regard to the components built in, among other things. Model classic comfort excellence Components No display interface No radio master (ERGOremote) With display interface No radio master (ERGOremote) With display interface With radio master (ERGOremote) Every component can be individually upgraded.
ERGOcom 3 3 Product description | 3.2 ERGOcom 3 3.
ERGOcom 3 3 Product description | 3.3 Connections 3.
ERGOcom 3 3 Product description | 3.4 Rating plate 3.4 Rating plate SN: REF Type: Serial number Material number Device typeERGOcom 3 HF emission Note: consult accompanying documents.
ERGOcom 3 3 Product description | 3.5 Technical data 3.5 Technical data ERGOcom 3 Weight 3 kg Housing dimensions Width 260 mm Depth 260 mm Height 46.5 mm Electricity supply AC voltage 100 -240 V Rated frequency 50 - 60 Hz Power consumption 70 VA Ambient conditions Permissible operating temperature 0 - 40° C Connection of interoral camera Type BF Compatible displays Max.
ERGOcom 3 4 Start-up | 4.1 Start-up 4 Start-up 4.1 Start-up 4.1.1 Start-up CAUTION Malfunction due to incorrect use An increase in operating temperature due to poor ventilation can lead to malfunction or failure. ▶ Do not set up near to a heat source. ▶ Do not cover the product's vents. A minimum gap of 50 mm must be left in front of and behind the unit. ▶ Operate the product only when it is horizontal. ▶ Connect theERGOcom 3 to a properly earthed power source that is protected by a fuse. 4.1.
ERGOcom 3 4 Start-up | 4.1 Start-up ▶ Select the required channel using navigation. Note If there is no input for 8 seconds, the ERGOcom 3 switches back to operational mode. This saves the last setting on the status display. Learn channel The ERGOremote 's radio code is synchronised with the channel configured onERGOcom 3 . ▶ Hold down the Channel Selection button on ERGOcom 3 for more than 2 seconds. Learn Channel Mode appears on the status display.
ERGOcom 3 4 Start-up | 4.1 Start-up ▪ xx = image number 01 to 99 ▪ y = image number 1 to 4 A maximum of 8 images may be saved. When a ninth image is saved, image 1 is automatically overwritten. A KaVo company logo is stored on the USB drive at the time of delivery. Connect PC Prerequisite ● Windows XP (Service Pack 1) or Windows 2000 must be installed on the PC. ● The PC must have a free USB 2 interface. ▶ Install Windows patch.
ERGOcom 3 4 Start-up | 4.1 Start-up ▶ Save this logo/image in bitmap format, giving it the filename USERLOGO.bmp. ▶ Copy the file USERLOGO.bmp to the USB drive using the drag&drop function. The USB drive reports to the PC using the name ERGOCOM3. Note The ERGOcom 3 must not be switched off during transfer of the image (takes approx. 20 seconds). You can set which logo/image is to be displayed in the OSD menu display interface. See also: 5.4.
ERGOcom 3 5 Operation | 5.1 Operation, general 5 Operation 5.1 Operation, general 5.1.1 Switch on ▶ Switch on main switch. These indicator lamps appear on the status display after open-shop testing: Note If the device is ready for operation, the current monitor image source is shown in the display. 5.1.2 Standby ▶ Press the standby button. The standby LED lights up (blue). The status display and display are switched off and all device settings are retained.
ERGOcom 3 5 Operation | 5.1 Operation, general ▪ ▪ ▪ ▪ ▪ ▪ ▪ System camera Camera holder USBpict.store Store format Mouse Z-Axis Firmware version Factory reset? Note The last 6 menu points are reserved for engineers. ▶ Press the OSD menu button once again in order to return to operationalmode. 5.1.4 OSD menu, operate ERGOcom 3 ▶ Select the required menu point using navigation. When you leave a menu point, its status display will be saved automatically. ▶ Select the required setting using navigation.
ERGOcom 3 5 Operation | 5.1 Operation, general Various setting parameters for image functions are offered on account of the various technical systems in the intraoral cameras ERGOcam 3 and ERGOcam 4 offered by KaVo. For optimum setting, the KaVo intraoral camera built into the treatment device should be selected here. On: When the internal / external camera is taken out of the holder, the camera image is automatically displayed on the monitor. Off: Automatic operations are not available.
ERGOcom 3 5 Operation | 5.2 Operation, no display interface (classic) 5.2 Operation, no display interface (classic) Equivalent to ERGOcom 3 , classic level. In all subsequent descriptions, please note the following: to view the selected option on Display 1, the relevant source must be active. 5.2.1 Operational mode In operational mode, it is possible to switch between VGA or VIDEO image sources, including audio, shown on KaVo Display 1. VGA IN - set image source ▶ Press the VGA IN button.
ERGOcom 3 5 Operation | 5.2 Operation, no display interface (classic) See also: 5.4.5 Menu point, other, Page 49 VIDEO IN - set image source Prerequisite ● A grabber card has to be installed in the PC in order to show video sources on Display 1. ● VIDEO OUT 1 or VIDEO OUT 2 has to be connected to the grabber card. ▶ Press the VIDEO IN button. If a VGA image source was selected before, the last VIDEO IN image source selected will be shown in the status display.
ERGOcom 3 5 Operation | 5.2 Operation, no display interface (classic) ▶ Select the required VIDEO input by pressing the VIDEO IN button. See also: 5.4.5 Menu point, other, Page 49 5.2.2 VGA OUT - set image source ▶ Hold down the VGA OUT button for < 2 seconds. The last image source selected is shown in the status display. The status display returns to operational mode after 5 seconds.
ERGOcom 3 5 Operation | 5.2 Operation, no display interface (classic) See also: 5.2.1 Operational mode, Page 27 ▶ Hold down the VGA OUT button for > 2 seconds. The status display flashes and setting mode is activated. ▶ Use VGA OUT to select the required VGA input. Note If there is no input for 8 seconds, the ERGOcom 3 switches back to operational mode. This saves the last setting on the status display. 5.2.
ERGOcom 3 5 Operation | 5.2 Operation, no display interface (classic) ▶ Hold down the VIDEO OUT 1 button for < 2 seconds. The last image source selected is shown in the status display. The status display returns to operational mode after 5 seconds. See also: 5.2.1 Operational mode, Page 27 ▶ Hold down VIDEO OUT 1 for > 2 seconds. The status display flashes and setting mode is activated. ▶ Select the required VIDEO input by pressing the VIDEO OUT 1 button.
ERGOcom 3 5 Operation | 5.2 Operation, no display interface (classic) 5.2.4 VIDEO OUT 2 - set image source ▶ Hold down the VIDEO OUT 2 button for < 2 seconds. The last image source selected is shown in the status display. The status display returns to operational mode after 5 seconds. See also: 5.2.1 Operational mode, Page 27 ▶ Hold down the VIDEO OUT 2 button for > 2 seconds. The status display flashes and setting mode is activated. ▶ Select the required VIDEO input by pressing the VIDEO OUT 2 button.
ERGOcom 3 5 Operation | 5.2 Operation, no display interface (classic) Note If there is no input for 8 seconds, the ERGOcom 3 switches back to operational mode. This saves the last setting on the status display. 5.2.5 Generate still Prerequisite ● A video signal is shown on Display 1. ● TheERGOcom 3 is connected to a PC fitted with a grabber card by means of a USB cable. ● On this PC, the KaVo Display Interface System software is installed. ▶ Hold down the footswitch for < 2 seconds.
ERGOcom 3 5 Operation | 5.2 Operation, no display interface (classic) 5.2.7 Audio signals The audio signals will be linked synchronously to the relevant image signals.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) 5.3 Operation, with display interface (comfort/excellence) Equivalent to ERGOcom 3 , comfort or excellence level. In addition to excellence level, instructions for operation for the radio master (ERGOremote) can be found in the separate UI. See also: UIERGOremote In all subsequent descriptions, please note the following: to view the selected option on Display 1, the relevant source must be active. 5.3.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) ▶ Select the required VGA input by pressing the VGA IN button. Note If no image source is connected to VIDEO IN 1/2, the logo set in the OSD menu display interface will be displayed. See also: 5.4.5 Menu point, other, Page 49 VIDEO IN - set image source ▶ Press the VIDEO IN button. If a VGA image source was selected before, the last VIDEO IN image source selected will be shown in the status display.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) ▶ Select the required VIDEO input by pressing the VIDEO IN button. Note If no image source is connected to VIDEO IN 1/2, the logo set in the OSD menu display interface will be displayed. See also: 5.4.5 Menu point, other, Page 49 5.3.2 VGA OUT - set image source ▶ Hold down the VGA OUT button for < 2 seconds. The last image source selected is shown in the status display.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) See also: 5.2.1 Operational mode, Page 27 ▶ Hold down the VGA OUT button for > 2 seconds. The status display flashes and setting mode is activated. ▶ Use VGA OUT to select the required VGA input. Note If there is no input for 8 seconds, the ERGOcom 3 switches back to operational mode. This saves the last setting on the status display. 5.3.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) ▶ Hold down the VIDEO OUT 1 button for < 2 seconds. The last image source selected is shown in the status display. The status display returns to operational mode after 5 seconds. ▶ Hold down VIDEO OUT 1 for > 2 seconds. The status display flashes and setting mode is activated. ▶ Select the required VIDEO input by pressing the VIDEO OUT 1 button.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) 5.3.4 VIDEO OUT 2 - set image source ▶ Hold down the VIDEO OUT 2 button for < 2 seconds. The last image source selected is shown in the status display. The status display returns to operational mode after 5 seconds. See also: 5.2.1 Operational mode, Page 27 ▶ Hold down the VIDEO OUT 2 button for > 2 seconds. The status display flashes and setting mode is activated.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) Note If there is no input for 8 seconds, the ERGOcom 3 switches back to operational mode. This saves the last setting on the status display. 5.3.5 Audio signals The audio signals will be linked synchronously to the relevant image signals. 5.3.6 Generate still Prerequisite A video signal is shown on Display 1. ▶ Hold down the Freeze button for < 2 seconds. A still is generated on Display 1 from the current video image.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) or ▶ Press the footswitch again to return to live image display and activate the next image (1-4) as a live image. 5.3.7 Save still Prerequisite ● TheERGOcom 3 is connected to a PC fitted with a grabber card by means of a USB cable. ● On this PC, the KaVo Display Interface System software is installed. ▶ Hold down the Freeze button for > 2 seconds. The successful save is indicated on Display 1 with save.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) Note The image number of the active image is shown in white. ▶ Press the Full Screen / Quad Mode button again to return to Full Screen mode. 5.3.9 Select image Prerequisite The image selected must be a still. ▶ Select the required image using the 4 navigation buttons. The still selected is retained as a still. ▶ Press the Freeze button. The still selected is deleted or turned into a live image.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) 5.3.10 Delete image / images Note Possible in Full Screen and Quad mode. ERGOcom 3 Prerequisite The image selected must be a still. ▶ Hold down the Delete image / images button for < 2 seconds. The still is deleted and the logo selected in the OSD menu display interface is shown. See also: 5.4.5 Menu point, other, Page 49 ▶ Hold down the Delete image / images button for > 2 seconds.
ERGOcom 3 5 Operation | 5.3 Operation, with display interface (comfort/excellence) 5.3.11 Change image source Prerequisite A still has to be generated. See also: 5.3.6 Generate still, Page 41 ▶ Put the camera in the holder. ▶ Press the footswitch. TheERGOcom 3 switches to the image source selected before the camera was removed.
ERGOcom 3 5 Operation | 5.4 Operation, OSD menu display interface 5.4 Operation, OSD menu display interface The overview diagrams were structured according to the type of image source and the connected function: VGA main menu VIDEO main menu Menu point, other Overview of main menu with VGA IN 1 or VGA IN 2 image source. Overview of main menu with CAMERA, VIDEO IN 1 or VIDEO IN 2 image source. This menu point is the same for all image sources. 5.4.
ERGOcom 3 5 Operation | 5.4 Operation, OSD menu display interface Navigation Function Change settings. Save changed setting. Reset the selected menu point. 5.4.3 VGA main menu There is a separate main menu for every VGA IN image source. This allows the settings on Display 1 to be defined and saved for every image source. Note The requisite explanations of the individual menu points or settings are inserted directly on Display 1.
ERGOcom 3 5 Operation | 5.4 Operation, OSD menu display interface 5.4.4 VIDEO main menu There is a separate main menu for every VIDEO IN image source. This allows the settings on Display 1 to be defined and saved for every image source. Note The requisite explanations of the individual menu points or settings are inserted directly on Display 1.
ERGOcom 3 5 Operation | 5.4 Operation, OSD menu display interface 5.4.5 Menu point, other This menu point is available in all main menus (VGA and VIDEO). However, it is not saved separately for all menus, but only once, covering all menus. Note The requisite explanations of the individual menu points or settings are inserted directly on Display 1.
ERGOcom 3 5 Operation | 5.
ERGOcom 3 6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning 6 Preparation methods DIN EN ISO 17664 6.1 Cleaning 6.1.1 Exterior cleaning by hand The wide variety of drugs and chemicals used in dental practice means that damage may occur to painted surfaces or plastics if anything is allowed to drip onto them. Tests have shown that surfaces cannot be protected entirely against all substances available on the market.
ERGOcom 3 6 Preparation methods DIN EN ISO 17664 | 6.2 Disinfection 6.2 Disinfection CAUTION Damage caused by fluids. Faults in electrical components. ▶ Protect device openings from the penetration of fluids. ▶ Wipe surface with a soft, fluff-free cloth to disinfect them. Note Contaminated parts must be disinfected after every patient.
ERGOcom 3 6 Preparation methods DIN EN ISO 17664 | 6.3 Maintenance 6.3 Maintenance The product does not require regular user maintenance.
ERGOcom 3 7 Safety checks 7 Safety checks If an EC3 Display is integrated with an ERGOcom 3 , safety checks (SFC) must be performed every 2 years. A detailed description of SFC can be found in the engineer's instructions.
ERGOcom 3 8 Troubleshooting 8 Troubleshooting Fault No status display. Cause Remedy ERGOcom 3 switched to standby. ▶ Leave Standby mode. See also: 5.1.2 Standby, Page 24 ▶ ERGOcom 3 on. ERGOcom 3 off. Mains fuse burnt out. ▶ Test mains fuse and replace where necessary. ▶ Ensure good ventilation for the ERGOcom 3 switches itself off and Overheating. cannot be switched back on immeERGOcom 3 . diately. See also: 4 Start-up, Page 20 Everything has stopped working. Mains failure / overvoltage.
ERGOcom 3 9 Glossary 9 Glossary Term Display 1 Description Display 1 is the display connected via the VGA Monitor interface and installed on the lamp mounting rod. Display 2 Display 2 is the display connected via the VGA out interface. Full screen mode The image is displayed across the entire display area. Quad mode Four independent images are shown in the display. OSD menu display inter- The OSD menu of the display interface. This is conface trolled viaERGOcom 3 and shown in Display 1.
1.003.1609 · Ma · 08.
User instructions ERGOremote Always on the safe side.
Sales and Distribution: KaVo Dental GmbH Bismarckring 39 D-88400 Biberach Tel.
User instructions ERGOremote Table of contents Table of contents Table of contents ..............................................................................................................................................1 1 User Notes .....................................................................................................................................................3 1.1 User guidelines ....................................................................................................
User instructions ERGOremote Table of contents 6 Preparation methods DIN EN ISO 17664 ....................................................................................................22 6.1 Cleaning .................................................................................................................................................22 6.1.1 Exterior cleaning by hand .............................................................................................................22 6.2 Disinfection .
User instructions ERGOremote 1 User Notes | 1.1 User guidelines 1 User Notes 1.1 User guidelines Prerequisite Please read these instructions before using the product to avoid operator error and damage. 1.1.1 Abbreviations Abbre‐ viation UI SI STI SFC IEC Meaning User instructions Setup instructions Service Technician's instructions Safety checks International Electrotechnical Commission 1.1.
User instructions ERGOremote 1 User Notes | 1.2 Target audience 1.2 Target audience This document is intended for use by dentists and other dental practice employees.
User instructions ERGOremote 1 User Notes | 1.3 Service 1.3 Service Please send all questions about the product, services and maintenance to the fol‐ lowing addresses. Always quote product serial number in any correspondence. KaVo Dental GmbH After Sales Service Bahnhofstraße 20 D-88445 Warthausen 0049 - (0) 7351-56 2700 0049 - (0) 7351-18218 multimedia@kavo.de www.kavo.
User instructions ERGOremote 1 User Notes | 1.4 Provisions of guarantee 1.
User instructions ERGOremote 1 User Notes | 1.5 Transport and storage 1.5 Transport and storage Note When transporting used units that are already covered in dust, completely seal the extraction vent at the rear of the unit (e.g. using suitable adhesive tape). 1.5.1 The German packaging ordinance, 28 August 1998 Note Applies only to the Federal Republic of Germany. KaVo transport packaging is disposed of and recycled by local waste management and recycling companies under Germany's Dual System.
User instructions ERGOremote 1 User Notes | 1.5 Transport and storage Outside Germany Note KaVo shall not be liable for damage caused in transit. Check the shipment immediately upon delivery! If the outer packaging is noticeably damaged upon delivery, the following procedure must be adhered to: 1. The recipient must record the loss or damage on the notice of receipt. The reci‐ pient and transport company employee delivering the product must both sign the notice of receipt.
User instructions ERGOremote 2 Safety | 2.1 Description of safety instructions 2 Safety 2.1 Description of safety instructions 2.1.1 Hazard warning symbol Hazard warning symbol 2.1.2 Structure DANGER The introduction describes the type and source of the danger. This section shows what could happen if the instructions are not followed. ▶ The optional action shows what measures to take to avoid danger. 2.1.
User instructions ERGOremote 2 Safety | 2.2 Intended purpose 2.2 Intended purpose 2.2.1 General The operator is required to make sure that the device is in a fully functional and safe condition before commencing use. The appropriate, comprehensive guidelines for this product and/or national laws, national regulations and technlogical rules for starting up and operating have to be applied and fulfilled in line with the specified, intended purpose of the KaVo product.
User instructions ERGOremote 2 Safety | 2.3 Safety instructions 2.3 Safety instructions 2.3.1 General This KaVo product is not permitted for use in areas where there is a risk of explosion. WARNING CAUTION CAUTION CAUTION Injuries or harm caused by damaged functional components. Damaged functional components can cause personal harm or injury. ▶ If functional components are damaged: Stop work, remedy the problem or inform the Service Technician. Malfunction due to electromagnetic fields.
User instructions ERGOremote 3 Product description | 3.1 ERGOremote 3 Product description 3.
User instructions ERGOremote 3 Product description | 3.2 Rating plate 3.2 Rating plate Type Purchase order no. Serial no. DL123+++ ERGOremote 1.003.
User instructions ERGOremote 3 Product description | 3.3 Technical data 3.3 Technical data Wave band ISM 2.
User instructions ERGOremote 4 Start-up | 4.1 Insert battery 4 Start-up 4.1 Insert battery Note A DL 123 battery is supplied with the equipment. Only use original batteries. We cannot provide a functional guarantee for other batteries. ▶ Remove screw to open battery compartment. ▶ Insert the battery correctly in the compartment (in line with pole labels).
User instructions ERGOremote 4 Start-up | 4.2 Learn channel 4.2 Learn channel The ERGOremote 's radio code is synchronised with the channel configured onERGOcom 3 . ▶ Hold down the Channel Selection button on ERGOcom 3 for more than 2 se‐ conds. Learn Channel Mode appears on the status display. Note If the channels do not synchronise within approx. 30 seconds, ERGOcom 3 swit‐ ches back to operational mode. ▶ Hold down the Wake Up/Learn button on the ERGOremote for more than 3 se‐ conds.
User instructions ERGOremote 5 Operation | 5.1 Controls/displays 5 Operation 5.1 Controls/displays 5.1.1 Synchronise the ERGOremote to ERGOcom 3 Note Channel 1 is set as a default. If there are a number of different ERGOremote and ERGOcom 3 instruments in use at the practice, a different channel must be set and the associated learning procedure completed in every room of the practice. ▶ Switch on Learn mode on ERGOcom 3 . See also: 4.2 Learn channel, Page 16 5.1.
User instructions ERGOremote 5 Operation | 5.1 Controls/displays ▶ Press the Standby button. The right, left, up and down button functions are reset. Note ERGOcom 3 and the ERGOcom 3 display switch to Standby mode. 5.1.6 Delete image in Video mode You can use the cursor buttons to navigate from one image to another. A brighter image number indicates which image is active. ▶ Press delete. The active image in Quad or Full Screen mode is deleted. 5.1.
User instructions ERGOremote 5 Operation | 5.1 Controls/displays 5.1.9 OSD Display Interface ▶ Press the OSD button. The On Screen Display appears. See also: ERGOcom 3 Instructions for Use 5.1.10 Cursor keys Joystick Prerequisite The corresponding PC is connected via the USB interface of ERGOcom 3 . The joystick can be used to move the cursor on the PC interface. If the z axis in ERGOcom 3 is activated, you can perform click and double-click functions by placing axial pressure on the joystick.
User instructions ERGOremote 5 Operation | 5.1 Controls/displays Down In VGA mode, the down button is used to scroll through inactive menus. In Video mode and OSD (Display Interface) mode, it can be used to navigate bet‐ ween the four images. 5.1.11 Video Input 2 to Display 1 This button is used to switch the incoming image signal at Video Input 2 Directly to Display 1. 5.1.
User instructions ERGOremote 5 Operation | 5.1 Controls/displays If the cell voltage falls below the level required for safe operation, the red LED will flash for approximately 3 seconds after wake-up indicating that the battery needs replacing.
User instructions ERGOremote 6 Preparation methods DIN EN ISO 17664 | 6.1 Cleaning 6 Preparation methods DIN EN ISO 17664 6.1 Cleaning 6.1.1 Exterior cleaning by hand The wide variety of drugs and chemicals used in dental practice means that damage may occur to painted surfaces or plastics if anything is allowed to drip onto them. Tests have shown that surfaces cannot be protected entirely against all substances available on the market.
User instructions ERGOremote 6 Preparation methods DIN EN ISO 17664 | 6.2 Disinfection 6.2 Disinfection CAUTION Damage caused by fluids. Faults in electrical components. ▶ Protect product openings from the penetration of fluids. ▶ Wipe surface with a soft, fluff-free cloth to disinfect them.
User instructions ERGOremote 6 Preparation methods DIN EN ISO 17664 | 6.3 Maintenance 6.3 Maintenance The product does not require regular user maintenance.
User instructions ERGOremote 7 Glossary 7 Glossary Term Display 1 Description Display 1 is the display connected via the VGA Monitor interface and installed on the lamp mounting rod. Display 2 Display 2 is the display connected via the VGA out in‐ terface. Full screen mode The image is displayed across the entire display area. Quad mode Four independent images are shown in the display. OSD menu display inter‐ The OSD menu of the display interface.