User's Manual Part 4

8065751131 4.1
INTRODUCTION
This section of the manual is designed to inform the operator of basic care and
maintenance of the instrument. It is recommended to verify the calibration annually.
In the event that recalibration is required, it is also recommended that the procedure
to recalibrate the system be performed by Alcon Technical Services personnel. If
a problem occurs on the instrument, call the Alcon Technical Services department
and give details of the circumstances and effects. From these elements, a specialized
WARNINGS!
Maintenance on any part of the laser system must be performed with the laser
off and the main power plug disconnected.
When keyswitch power is on, all individuals in the laser room must wear laser
protective eyewear, OD 4 or above at 532 nm.
CAUTION
There are no operator replaceable parts other than the fuse. Contact Alcon Technical
Services for all servicing issues.
Care and Cleaning
WARNING!
A qualified technician must perform a visual inspection of the following components
every twelve months:
• Warning labels (see Section One)
• Power Cord
• Fuses
In case of a deficiency, do not use the system; call Alcon Technical Services.
A qualified technician must check ground continuity and both polarities for leakage
current every twelve months to ensure they are within the applicable standards
(for example: EN60601-1/IEC601-1). Values must be recorded. If they are above the
applicable standards, or 50% above initial measurement, do not use the system;
call Alcon Technical Services.
The following tips are recommended for proper care of the
PurePoint™
Laser
system:
Turn off the system correctly after each use with the rear panel switch.
Cover the slit lamp with the plastic cover.
Cover the fi ber optic connector with the dust cover.
Cover the fi ber port with the dust cover.
Clean the exterior portion of the equipment with a dry, lint-free cloth or tissue. No other
products can be used.
Use care not to damage or scratch the laser apertures or fi ber optic connector.
Place the system into its traveling case when moving to another location.
SECTION FOUR
CARE AND MAINTENANCE

Summary of content (12 pages)