User's Manual Part 3

8065751131 3.1
SECTION THREE
OPERATING INSTRUCTIONS
INTRODUCTION
This section details the recommended setup and operation for the
PurePoint™
Laser.
These procedures may be modifi ed to conform to hospital requirements and practices
are performed at various points in the setup procedure to verify instrument operation
must be performed exactly as indicated.
WARNING!
Noncompliance with the instructions contained in this manual may result in
operator injury.
The following procedures (Initial Setup, System Connections, System Power Up, and
Operation) cover preparation for laser treatment involving Slit Lamp, Endoprobe/
Aspirating Endoprobe/Illuminated Endoprobes, or LIO usage.
Any questions pertaining to setup and checkout procedures should be directed to your
Alcon Technical Services representative.
1 INITIAL SETUP
WARNING!
To avoid potential secondary reflections, a qualified laser safety officer should
approve the room used to treat the patient.
1.1 Position the instrument for surgeon’s comfort and preference. Refer to Section
One for "Recommended Laser Room Layout."
1.2 Verify that no combustible materials are adjacent to the laser and its delivery
systems.
2 SYSTEM CONNECTIONS
2.1 Connect footswitch to Footswitch connector.
2.2 Connect Remote Interlock to REMOTE Interlock connection.
2.3 Plug in the power cord to the power cord socket and to a properly grounded
main power outlet (220-240 VAC, 10A minimum; or 100-120 VAC, 15A
minimum).
2.4 Ensure all Doctor/Observer protection fi lters are installed in the optical path of
the slit lamp or microscope, and connected to the Doctor Protection Filter ports
on the rear panel (see Figure 3-1).
2.5 Verify that the power switch on the rear panel is in the ON position.
2.6 Insert key into keyswitch on front panel. Leave in the OFF position.