Constellation® Vision System • • • is constant regardless of the vacuum level. In Linear the limit varies depending on the vacuum level. Phaco Pulse Off Time Up/Down Buttons - Sets the Phaco pulse off time. Phaco Pulse Off Time Progress Bar and Numeric Value - Shows the current Phaco pulse off time.
Constellation® Vision System (pulses per second-pps) and Time On (%) U/S power pulses. The pulse is controlled with the footpedal from detent two to full treadle depression. The actual percent of time U/S power is activated is reflected in the Actual window. The controls for each panel in Pulsed submode have the following functions: Aspiration Panel: Same as Burst submode. Vacuum Panel: Same as Burst submode. Ultrasound Panel: - Sets the power at full treadle.
Constellation® Vision System Vacuum Panel Configuration Button Aspiration Panel Figure 2-59 2.
Constellation® Vision System Ultrasound Panel: • Phaco Power End Up/Down Buttons - Sets the power at full treadle. • Phaco Power Progress Bar and Numerical Value - Shows the current power value. • Ultrasound Configuration Button - Displays the configuration popup. The following controls are available only in the Advanced view: • Smart Pulse - Indicator for smart pulse. Becomes visible when power is greater than 0 and pulse on time is less than 20ms.
Constellation® Vision System • • • Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle. Vacuum Progress Bar and Numerical Value - Displays the current vacuum level Vacuum Configuration Button - Displays the configuration popup. Simplifies large value changes by providing slider controls. Also provides panel specific configuration capabilities such as Drainbag Change. The following controls are available only in the Advanced view: • Flow Limit On/Off Button - Turns flow limit on or off.
Constellation® Vision System • • • Power End Up/Down Buttons - Sets the power at full treadle. The Start value tracks with the End value. Power Progress Bar and Numerical Value - Displays the current power value. Ultrasound Configuration Button - Displays the configuration popup. The following controls are available only in the Advanced view: • Pulse Rate On/Off Button - Turns the Pulse Rate on or off. • Pulse Rate Up/Down Buttons - Sets the pulse rate.
Constellation® Vision System Vacuum Panel: Same as Fixed submode. Flow Panel: Same as Fixed submode. Ultrasound Panel: Same as Fixed submode except that the Start value is always zero instead of tracking with the End value. Fragmentation Mode: MOMENTARY SUBMODE Aspiration Panel Configuration Button Vacuum Panel Figure 2-62 2.
Constellation® Vision System Momentary submode is similar to the Fixed Sub Mode with the exception that ultrasound is only active when the appropriate footswitch button is pressed. Vacuum Panel: Same as Fixed submode. Flow Panel: Figure 2-63 Surgery Screen: Irrigation/Aspiration Mode Same as Fixed submode. Ultrasound Panel: Same as Fixed submode except that there is no On/Off button to turn U/S power Off.
Constellation® Vision System The following controls are available only in the Advanced view: • Flow Limit On/Off Button - Turns flow limit on or off. When the maximum rate is reached, the vacuum level can no longer be increased. If the Flow Limit is turned off, the vacuum level is limited only by the Start and End values. • Flow Limit Up/Down Buttons - Sets the maximum flow rate. • Flow Limit Progress Bar and Numerical Value - Displays the current flow rate.
Constellation® Vision System rate is reached, the flow can no longer be increased. If the Vacuum Limit is turned off, the flow is limited only by the Start and End values. • The following controls are available only in the Advanced view: • Vacuum Limit Up/Down Buttons - Sets the maximum vacuum value. Figure 2-64 • • • Surgery Screen: Extrusion Mode Vacuum Limit Progress Bar and Numerical Value - Displays the current vacuum level. Vacuum Limit Configuration Button - Displays the configuration popup.
Constellation® Vision System • • • • • • off, the vacuum level is limited only by the Start and End values. Flow Limit Up/Down Buttons - Sets the maximum flow rate. Flow Limit Progress Bar and Numerical Value - Displays the current flow rate. Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode. Flow Limit Configuration Button - Displays the configuration popup. Flow Limit Control Drop Down List - Determines whether the flow limit is fixed or linear.
Constellation® Vision System • • Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode. Reflux Label - Read Only display of the current reflux configuration together with its current value. Reflux Configuration Button - Displays the Reflux configuration popup. Cutting Panel: • Cut Rate End Up/Down Buttons - Sets the cut rate. • Cut Rate Progress Bar and Numerical Value - Displays the current cut rate. • Cutting Configuration Button - Displays the configuration dialog.
Constellation® Vision System In “Standard” display mode, only the Vacuum mode controls are available. The controls for each panel in Laser mode have the following functions: Vacuum Panel: • Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle. The Start value is always zero. • Vacuum Progress Bar and Numerical Value - Displays the current vacuum level • Vacuum Configuration Button - Displays the configuration popup. Simplifies large value changes by providing slider controls.
Constellation® Vision System • • • • • • • Exposure Time Up/Down Buttons - Adjust the laser exposure time. Only available in Repeat or Single Shot modes. Interpulse Time Up/Down Buttons - Adjusts the laser interpulse time. Only available in Repeat mode. Aiming Beam Up/Down Buttons - Displays the current relative intensity of the aiming beam. LIO Illumination Intensity Up/Down Buttons - Adjusts the LIO illumination intensity. Shot Count Label - Displays the current shot count.
Constellation® Vision System • Calibrate On/Off Button - Enters Calibration Mode allowing the user select and set the pressure at Start and Close. - When Start is selected, the Start pressure is applied to the port and the Start Up/ Down buttons are active on the Pressure control to allow adjustment of the Start pressure. - When Close is selected, the Close pressure is applied to the port and the Close Up/Down buttons are active on the Pressure control to allow adjustment of the Close pressure.
Constellation® Vision System • • Cut Rate Progress Bar and Numerical Value - Displays the current cut rate. Cutting Configuration Button - Displays the configuration popup. Forceps Panel: • Bimanual On/Off Button - Turning Bimanual mode on allows the control of the pneumatic forceps while the system is in Scissors mode. All controls function the same as when the system is in Forceps mode (see Forceps mode for a description).
Constellation® Vision System set the pressure at Start and Close. - When Start is selected, the Start pressure is applied to the port and the Start Up/ Down buttons are active on the Pressure control to allow adjustment of the Start pressure. - When Close is selected, the Close pressure is applied to the port and the Close Up/Down buttons are active on the Pressure control to allow adjustment of the Close pressure. - Pressing the "Save Calibration Values" button saves these values.
Constellation® Vision System Vacuum for fluid extraction is provided to the syringe proportional to footpedal depression. Max Limit vacuum is reached at full depression. The controls on the Extraction panel have the following functions: • • • Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle. Start value is always zero. Vacuum Progress Bar and Numerical Value - Displays the current vacuum level.
Constellation® Vision System END CASE The End Case Screen, selected by pressing the End Case tab, displays case related metrics as shown in Figure 2-71. It is comprised of a Summary panel area and a tabbed interface area for displaying the various metrics associated with the case. The screen also provides various controls for renaming the case, printing the End Case and Laser forms, and customizing the form’s content and layout via the Setup Form popup.
Constellation® Vision System In computing a step’s cumulative time, it is necessary for work to be done while the step is selected in order for the current elapsed time to be added to the step’s cumulative time. Therefore, a user may access a step and interact with the various touchscreen controls for an extended period of time but, unless the treadle is depressed, the current elapsed time will be discarded and will not be added to the step’s cumulative time.
Constellation® Vision System Setup Form: Editing the Form To modify the form, press the Edit Form button. When pressed, the “smoked glass” panel is retracted allowing access to the various customization controls. The editing controls on the dialog provide the following functionality: • • • • • • • Cell navigation buttons including up, down, left, right. Cell text edit button. Drop list to permit the specification of a cell data fields.
Constellation® Vision System Setup Form: Editing the Form - Creating New Tables To begin the customization process, the user must first select a table on the form by pressing on it. The table is then shown with selection handles and the active cell is shown in red. To insert a new table after the selected table, follow these steps: 1. In the Table section of the Edit Form panel, press the Insert button (Ins).
Constellation® Vision System PROBES AND HANDPIECES Different probes and handpieces are required for each operating mode of the Constellation® Vision System. Following is a representative selection of probes and handpieces with a general description of each. See the Accessories and Parts section of this manual, or consult your Alcon representative, for a complete selection of all probes, handpieces, and handpiece tips available.
Constellation® Vision System Fragmentation Handpiece The Fragmentation handpiece is configured to provide simultaneous vacuum and fragmentation, or vacuum only, depending upon the console setup. The handpiece has a stainless steel shell for improved reliability and durability. Other than attaching and removing the needle and aspiration line, no assembly or disassembly is required.
Constellation® Vision System WARNINGS! Use of an ultrasonic handpiece other than the U/S, or use of a handpiece repaired without Alcon authorization, is not permitted, and may result in patient injury, including potential shock hazard to patient and/or operator. Use of the U/S handpiece in the absence of irrigation flow and/or in the presence of reduced or lost aspiration flow can cause excessive heating and potential damage to the cornea and other tissues. 2.
Constellation® Vision System TurboSonics® Family of Tips U/S tips are made of medical grade titanium alloy, and are attached to an U/S handpiece to deliver mechanical energy to the lens, assisting in its removal by aspiration. Depending on the needs and technique preferred by the surgeon, various styles of tips and tip bevels are available (see Figure 2-77). Various U/S tip styles are color coded. • • • 1.1 mm U/S Tips - The standard ultrasonic tips are the original 1.1 mm TurboSonics® tips.
Constellation® Vision System MicroSmooth™* Infusion Sleeves Infusion sleeves cover the tip of the handpiece to provide irrigation to the anterior chamber of the eye during surgery (see Figure 2-78). Infusion sleeves are used with the Infiniti™* U/S handpiece, and with some Ultraflow™* I/A handpieces. Infusion sleeves used with Infiniti™* U/S handpieces require a BSI (bubble suppression insert). Infusion sleeves must be correctly matched to the specific tip type (see the following descriptions).
Constellation® Vision System Ultraflow™* Handpieces and Tips The Ultraflow™* handpiece is used in I/A mode to maintain chamber pressure with irrigation while removing cortical material via aspiration. (See Figure 2-79 and note the band markings on the tips that identify size of tip aperture.) Some configurations of the Ultraflow™* IT and SP handpieces also use infusion sleeves.
Constellation® Vision System 120° Tip 90° Tip 45° Tip Curved Tip O-Rings Straight Tip Figure 2-79 Ultraflow™* IT handpiece and tips 0.2 mm I/A Tip 0.5 mm I/A Tip O-Rings 0.3 mm I/A Tip Figure 2-80 Threaded Tip Adapter Ultraflow™* IT handpiece with infusion sleeve, reusable I/A tip, and threaded tip adapter O-Ring Extractor Large O-Rings Small O-Rings 2.104 Figure 2-81 Ultraflow™* O-ring tool with large and small O-rings Figure 2-82 Ultraflow™* SP handpiece (handpiece shown with .
Constellation® Vision System Diathermy/Coagulation Handpieces Single use Bipolar Coagulation Brushes are available in a wide variety of configurations: straight, curved, 20-gauge, 23-gauge, tapered, and widestroke. All single use bipolar accessories are available with and without cables. Also available are reusable and single-use bipolar cables. Bipolar Coagulation Forceps are lightweight and ergonomically designed to reduce hand fatigue, as well as to provide precise control and safety.