User's Manual Part 4

2.76 8065751025
Constellation
®
Vision System
is constant regardless of the vacuum level. In Linear the limit varies depending on
the vacuum level.
Phaco Pulse Off Time Up/Down Buttons - Sets the Phaco pulse off time.
Phaco Pulse Off Time Progress Bar and Numeric Value - Shows the current Phaco
pulse off time.
Phaco Pulse Off Time Limit Control Drop Down List - Determines whether
Figure 2-58 Surgery Screen: Phaco Mode - Pulsed Submode
Aspiration Panel
Vacuum Panel
Configuration Button
the pulse off time pulse type limit is decreasing, fixed, or increasing. In Fixed
the limit is constant regardless of the vacuum level. In Linear the limit varies
depending on the vacuum level.
Phaco Pulse Configuration Button - Displays the configuration dialog. Simplifies
large value changes by providing slider controls.
Phaco Mode: PULSED SUBMODE
In the Phaco Custom submode fixed U/S power (Power %) can be adjusted for Rate