User's Manual Part 3

2.50 8065751025
Constellation
®
Vision System
Detailed Setup: Probe
The Probe setup panel shown in Figure 2-38 displays the steps required to properly
connect the various vit probes to the system.
Select Probe - Provides a selection of probes and, when a probe is selected, shows
where to connect to the front panel. When the system detects that the operator
has plugged in the probe, the handpiece type is automatically selected, the
“connected” icon appears next to the selection list, and the selection list is locked.
Select Guage - When a probe is selected, the Select Guage section offers three
options: 20, 23, and 25 guage. Once selected, the selected guage size appears
under the selected probe.
Connect Probe - This section shows an illustration of the connector panel and
highlights the connections for the selected probe.
Skip Probe Prime/Test - Indicates whether the operator wants to skip priming the
selected vitreous probe. When the checkbox is checked, the system will not prime or
test the vitreous probe when the user requests Prime & Test.
1.
2.
3.
Figure 2-38 The Detailed Probe Setup Panel