Keysight 85309B LO/IF Distribution Unit Use this manual with the Keysight Antenna Selection Guide, part number 5968-6759E.
Notices © Keysight Technologies 2013 - 2015 Manual Part Number No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies as governed by United States and international copyright laws.
Printing Copies of Documentation from the Web To print copies of documentation from the Web, download the PDF file from the Keysight web site: • Go to http://www.keysight.com. • Enter the product model number (Ex: 85309B) in the Search function. • Click Search. • Click the Manuals hyperlink. • Click the hyperlink title for the document you want to print - this downloads the PDF • Print the document after the PDF has fully downloaded.
amplify LO drive power through RF cables to the mixers in the antenna test configuration. The high output power allows the mixers to be located more than seven meters from the 85309B. There are also IF amplifiers located in the 85309B, which serve as a preamplifier for the receiver, reducing the overall system noise figure significantly. The 85309B LO/IF Distribution Unit can be installed in a variety of configurations for antenna testing.
Option 1CP - Rack Mount Kit With Handles Rack mount kit with handles. To obtain this item after receiving the 85309B, order part number 5063-9221.
Contents 1 Getting Started STEP 1. Check the Shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 STEP 2. Meet Electrical and Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3 Step 3. Connect the 85309B with Other System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents Declaration of Conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5 Statement of Compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6 General Safety Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Getting Started Service Guide 85309-90082 1-1
Getting Started STEP 1. Check the Shipment LO/IF Distribution Unit 85309B STEP 1. Check the Shipment 1. Unpack the contents of the shipping container. Keep the packaging material in case the analyzer needs to be returned to the factory. 2. Carefully inspect the 85309B to make sure that it was not damaged during shipment. If damaged, refer to “Contacting Keysight” on page 1-12. 3. Use the Contents List in the shipping container to verify the completeness of your shipment.
LO/IF Distribution Unit 85309B Getting Started STEP 2. Meet Electrical and Environmental Requirements STEP 2. Meet Electrical and Environmental Requirements CAUTION This instrument has autoranging line voltage input. Be sure the supply voltage is within the specified range. 1. Ensure the available AC power source meets the requirements found in “General Specifications” on page 3-2. 2. Ensure the operating environment meets the requirements found in “General Specifications” on page 3-2. 3.
Getting Started Step 3. Connect the 85309B with Other System Components LO/IF Distribution Unit 85309B Step 3. Connect the 85309B with Other System Components Your antenna test system configuration includes the 85309B and a PNA network analyzer. The need for other components is determined by the type of measurements required. To learn about the various system configurations, refer to the Keysight Antenna Test Selection Guide, part number 5968-6759E.
LO/IF Distribution Unit 85309B Getting Started Step 3.
Getting Started Step 3.
LO/IF Distribution Unit 85309B Figure 1-3 Getting Started Step 3.
Getting Started Step 4. (Option 40x Only): Set the 85309B Band Selection LO/IF Distribution Unit 85309B Step 4. (Option 40x Only): Set the 85309B Band Selection The Band Selection knob on the front panel of the 85309B Option 400/401/402 models allows the user to switch between the low band, high band, and external selections. For more information, refer to “85309B Front Panel Features” on page 2-3.
LO/IF Distribution Unit 85309B Getting Started Step 5: Power On the 85309B Step 5: Power On the 85309B Power on the 85309B by pressing the front panel On/Off switch. If the 85309B will not power on, make sure the instrument is plugged into an operating AC power outlet, and check the fuse.
Getting Started Step 6. Adjust the LO Power LO/IF Distribution Unit 85309B Step 6. Adjust the LO Power The procedure below is applicable to all 85309B models, and assumes the user has basic knowledge on the antenna test system. To learn about the various antenna test systems, refer to the Keysight Antenna Test Selection Guide, part number 5968-6759E. See “Printing Copies of Documentation from the Web” in the first section of this manual.
LO/IF Distribution Unit 85309B Getting Started STEP 7. (Optional) Run the 85309B Operator’s Check STEP 7. (Optional) Run the 85309B Operator’s Check The operator’s check should be performed when you first receive your 85309B, and any time you wish to have confidence that it is working properly. The operator’s check does not verify performance to specifications, but should give you a high degree of confidence that the 85309B is performing properly.
Getting Started Getting Help with Your 85309B LO/IF Distribution Unit 85309B Getting Help with Your 85309B Keysight Antenna Test Selection Guide Use the Keysight Antenna Test Selection Guide (part number 5968-6759E) to: • Learn about interface requirements between system components. • Understand issues related to selecting the equipment required to make antenna measurements.
2 Operating Information Service Guide 85309-90082 2-1
Operating Information Electrostatic Discharge Protection LO/IF Distribution Unit 85309B Electrostatic Discharge Protection Protection against electrostatic discharge (ESD) is essential while removing assemblies from or connecting cables to the 85309B. Static electricity can build up on your body and can easily damage sensitive internal circuit elements when discharged. Static discharges too small to be felt can cause permanent damage.
LO/IF Distribution Unit 85309B Operating Information 85309B Front Panel Features 85309B Front Panel Features Figure 2-2 85309B Standard/Option 001/002 Front Panel Figure 2-3 85309B Option 400/401/402 Front Panel DETECTOR VOLTAGE Display The reference mixer has a detector that measures the amount of LO power coming into its LO input. This voltage is sent to the 85309B where it is displayed in the DETECTOR VOLTAGE display. The voltage is directly proportional to the amount of LO power.
Operating Information 85309B Front Panel Features LO/IF Distribution Unit 85309B LO POWER OUT OF RANGE When lit, this LED indicates that the LO source is not supplying the amount of LO power requested by the LO POWER ADJUST control. The LO POWER OUT OF RANGE circuit compares the LO POWER ADJUST setting with the voltage coming back from the reference detector. The amount of expected detector voltage varies depending on the LO POWER ADJUST setting.
LO/IF Distribution Unit 85309B Operating Information 85309B Rear Panel Features 85309B Rear Panel Features CAUTION The 85309B contains static sensitive devices. Never touch the center conductor of any connector. When cleaning connectors, wear a grounded wrist strap. Figure 2-4 shows 85309B Option 402 rear panel features, including two additional test channels and an active J15 “High/Low Band Select” port. The rear panel on your 85309B model may not include all of the ports shown.
Operating Information 85309B Rear Panel Features LO/IF Distribution Unit 85309B power. The allowable input voltage range for this connector is 0 to -200 mV. Test Mixers Connector Group The three connectors in this group, J3 TEST 1 LO/IF (standard), J11 TEST 2 LO/IF (option 001 or 401), and J13 TEST 3 LO/IF (option 002 or 402), are connected to the test mixers. A test mixer only requires one connection.
LO/IF Distribution Unit 85309B Operating Information 85309B Rear Panel Features J7 and J8 (Using Non-Diplexed Mixers). J7 and J8 allow you to use a non-diplexed test mixer (a mixer that has separate connections for the LO and IF signal). This is done by removing the Type-N jumper that normally connects J7 and J8.
Operating Information 85309B Block Diagrams LO/IF Distribution Unit 85309B 85309B Block Diagrams Block diagrams are located on the following two pages.
LO INPUT J1 W24 FL1 85309-61607 DPDT A22 POWER SUPPLY +15, +15 -15V +24V W10 W21 W52 B1 FAN W12 F5 J15 15V F3 5V F1 F2 7 J6 -15V W54 Regulator 15V to 12V Regulator 24V to 20V 15V 15V -15V F4 24V J7 15V DC-DC 15V to 5V J16 J12 A10 0955-0791 85309-60135 7 5087-7768 -15V 5V J1 W17 J4 To J3, J4, J5, J17, J18 J18 J17 J3 DC-DC 15V to 10V To J3,J4,J17,J18 J5 J8 A14 A2 REG/ALC BOARD To J8, J9, J13, J14 To J3,J4,J5,J8,J9,J11,J13 ,J14,J15,J17,J18 10V Reg Ref 15
CPLD programming J6 + W24 J4 Remote APPL 85309-61607 FL1 33311-60069 SW3 + W21 W52 B1 FAN Switch Control F5 J12 J15 15V F4 24V J7 15V F3 F1 F2 7 J6 -15V W54 W42 Regulator 15V to 12V J1 5V -15V W17 J4 To J3, J4, J5, J17, J18 J18 DC-DC 15V to 10V LOAD LOAD J17 J3 To J3,J4,J17,J18 J5 J8 AT11 1dB W3 85309-60134 J9 A40 A39 A42 J13 85309-61601 J14 1 4 A37 A41 ALC J11 0955-2249 2 J10 A18 AT8 1dB LO POWER ADJUST A46 0955-2249 W48 0955-2249 A38 A
LO/IF Distribution Unit 85309B Operating Information Cable Length Limits Cable Length Limits The following figure defines the allowable maximum cable lengths from the LO Source to the 85309B, and the cable connecting the 85309B and the external mixers. The cable lengths are provided when using standard RF cable assemblies and with MicroCoax (UFB311A RF) low-loss cable assemblies. The standard mixers used in Keysight antenna measurement systems are the 85320A test mixer and the 85320B reference mixer.
Operating Information Cable Length Limits Figure 2-6 2-12 LO/IF Distribution Unit 85309B Cable Length Configurations (1 of 2) Service Guide 85309-90082
LO/IF Distribution Unit 85309B Figure 2-7 Operating Information Cable Length Limits Cable Length configurations (2 of 2) Service Guide 85309-90082 2-13
Operating Information Performing the Operator’s Check LO/IF Distribution Unit 85309B Performing the Operator’s Check Function The Operator’s Check verifies that the 85309B functions properly before permanent installation in the antenna test system. It may also be used as a stand-alone check anytime that a problem is suspected with the 85309B.
LO/IF Distribution Unit 85309B Operating Information Performing the Operator’s Check LO Power Output Test NOTE The PNA input must be attenuated to avoid causing receiver compression or damaging the receiver. An external 30 dB pad attenuator is required if the PNA does not have an internal source step attenuator. The instructions for this test use PNA Port 1 to generate the LO source, and use PNA Port 2 as receiver B. NOTE All keystrokes in the following procedure are for the PNA.
Operating Information Performing the Operator’s Check LO/IF Distribution Unit 85309B Checking the 85309B Test 1 LO/IF (J3) Channel 16. Connect a test cable between the PNA Port 1 and the 85309B rear panel port LO Input (J1). 17. Connect a test cable between the PNA Port 2 and the 85309B rear panel port Test 1 LO/IF (J3). 18.
LO/IF Distribution Unit 85309B Operating Information Performing the Operator’s Check power traces with the limits across the entire frequency range in Table 2-5 on page 2-18. 34. Set the PNA Port 1 source power to 6 dBm. 35. Repeat Step 32 or Step 33, as appropriate. For Option 002 and 402 Models Only: Checking the 85309B Test 3 LO/IF (J13) Channel 36. Connect a test cable between the PNA Port 1 and the 85309B rear panel port LO Input (J1). 37.
Operating Information Performing the Operator’s Check LO/IF Distribution Unit 85309B Table 2-4 85309B Options 40x Output Power Limits for HIGH Band LO Output Power (+6 dBm Input) Frequency Range Power Level 0.3 – 0.5 GHz > +21.3 dBm 0.5 – 3 GHz > +22.4 dBm 3 – 6.2 GHz > +24.4 dBm 6.2 – 18 GHz > +22 dBm LO Output Power (0 dBm Input) 0.3 – 0.5 GHz > +21.3 dBm 0.5 – 3 GHz > +22.4 dBm 3 – 6.2 GHz > +24.4 dBm 6.
LO/IF Distribution Unit 85309B Operating Information Performing the Operator’s Check 6. Connect a thru adapter between the two test cables. 7. Normalize the PNA trace. If necessary, get instructions by searching the embedded Help System of the PNA for “normalize.” After normalizing, the trace should be flat at 0 dBm 8. Remove the thru adapter between the two test cables. Checking the J10 Reference IF Out 9. Connect the test cable on PNA Port 1 to the 85309B rear panel port Reference IF In (J4). 10.
Operating Information Performing the Operator’s Check LO/IF Distribution Unit 85309B Figure 2-9 Typical Test IF Response Trace For Option 00x and 40x Models Only: Checking the J12 Test 2 IF Out 1. Connect a test cable between the PNA Port 1 and the 85309B rear panel port Test 2 LO/IF (J11). 2. Connect a test cable between the PNA Port 2 and the 85309B rear panel port Test 2 IF Out (J12). 3. Place a marker on the PNA trace.
3 General Information Service Guide 85309-90082 3-1
General Information General Specifications LO/IF Distribution Unit 85309B General Specifications Environmental: Operating Temperature Range 0 to 50 °C; for indoor use ONLY Storage Temperature Range –40 to 70 °C Operating Altitude 3000 m (9842 ft.) EMC Complies with European EMC Directive 2004/108/EC.
LO/IF Distribution Unit 85309B CAUTION General Information General Specifications Ventilation Requirements: When installing the instrument in a cabinet, the convection into and out of the instrument must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the instrument by 4 C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, forced convection must be used.
General Information 85309B Characteristic Performance LO/IF Distribution Unit 85309B 85309B Characteristic Performance Parameter Value Absolute Maximum Rating LO input port (CW) 20 dBm Reference channel IF input port (CW) 13 dBm Reference channel detector input ±20 VDC Positive Z/blanking input ±10 VDC (Option 40x only) Select HIGH/LOW input ±5.5 VDC Frequency Range Low band1 0.1 GHz to 1.0 GHz High band 0.3 GHz to 18.0 GHz Input Power Range (LO Input Recommended) 0.1 GHz to 1.
LO/IF Distribution Unit 85309B Parameter General Information General Safety Considerations Minimum Maximum IF Channel Small Signal Gain 20 MHz 21 dB 25 dB General Safety Considerations Introduction Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. The documentation contains information and warnings that must be followed by the user to ensure safe operation and to maintain the product in a safe condition.
General Information General Safety Considerations LO/IF Distribution Unit 85309B Statement of Compliance This instrument has been designed and tested in accordance with IEC Publication 61010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the instrument in a safe condition.
LO/IF Distribution Unit 85309B General Information General Safety Considerations Servicing Warnings applicable to this instrument. WARNING For continued protection against fire hazard replace line fuse only with same type and rating: Fuse 5A/250V, Part Number 2110-0680 The use of other fuses or material is prohibited. WARNING Always use the three-prong AC power cord supplied with this instrument with adequate electric rating.
General Information Regulatory Information LO/IF Distribution Unit 85309B Regulatory Information This section contains information that is required by various government regulatory agencies. Instrument Markings Familiarize yourself with these instrument markings and their meanings before operating the instrument. The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation.
LO/IF Distribution Unit 85309B General Information Regulatory Information This symbol on all primary or secondary packaging indicates compliance to China standard GB 18455-2001. IP 20 Degree of protection.
General Information Regulatory Information 3-10 LO/IF Distribution Unit 85309B Service Guide 85309-90082
4 Service Service Guide 85309-90082 4-1
Service Replaceable Parts LO/IF Distribution Unit 85309B Replaceable Parts Introduction This section contains information for ordering replaceable parts for the 85309B LO/IF Distribution Unit. The replaceable parts include major assemblies and cables. Ordering Information To order a part listed in the replaceable parts lists: • include the part number • indicate the quantity required • Contact Keysight Technologies for instructions on where to send the order.
LO/IF Distribution Unit 85309B Service Replaceable Parts This page is intentionally left blank.
Service Replaceable Parts LO/IF Distribution Unit 85309B Replaceable Parts Listings Table 4-1 Major 85309B Assemblies (Standard) Reference Designator Part Number Qty A1 A2 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 A22 A43 AT3 AT4 B1 SW1 85309-60135 85309-60134 0955-0791 0955-2245 5087-7755 5087-7755 85309-80026 85309-80026 5076-6336 5087-7768 85309-61601 5076-6336 0950-5271 85110-80015 33340-67007 33340-67007 54574-68501 85309-61607 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 4-4 Description PC board, Disp
LO/IF Distribution Unit 85309B Figure 4-1 Service Replaceable Parts Major 85309B Assemblies (Standard) Service Guide 85309-90082 4-5
Service Replaceable Parts Table 4-2 LO/IF Distribution Unit 85309B Cables (Standard) Reference Designator Part Number Qty W3 85309-20139 1 A17 OUT - A11 IN W4 85309-20141 1 AT3 OUT - A12 IN W5 85309-20140 1 AT4 OUT - A13 IN W7 85309-20144 1 A16 OUT - RPJ3 W8 85309-60112 1 A16 - RPJ8 W10 85309-20138 1 A17 IN - RPJ1 W11 85309-60109 1 A15 OUT- RPJ9 W12 85309-60109 1 A10 IN- RPJ4 W14 85309-60109 1 A14 OUT- RPJ10 W15 8120-8483 1 A2J1- RPJ5 W16 8120-8483 1 A2J2- R
LO/IF Distribution Unit 85309B Figure 4-2 Service Replaceable Parts Cable Locations (Standard) Service Guide 85309-90082 4-7
Service Replaceable Parts Table 4-3 LO/IF Distribution Unit 85309B Major 85309B Assemblies (Option 001) Reference Designator Part Number Qty A1 A2 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 A22 A24 A25 A26 A43 A44 AT1 AT3 AT4 B1 SW1 85309-60135 85309-60134 0955-0791 0955-2245 5087-7755 5087-7755 85309-80026 85309-80026 5076-6336 5087-7768 85309-61601 5076-6336 0950-5271 5087-7755 5076-6336 85309-80026 85110-80015 85110-80015 33340-67007 33340-67007 33340-67007 54574-68501 85309-61607 1 1 1 1 1 1 1 1 1
LO/IF Distribution Unit 85309B Figure 4-3 Service Replaceable Parts Major 85309B Assemblies (Option 001) Service Guide 85309-90082 4-9
Service Replaceable Parts Table 4-4 LO/IF Distribution Unit 85309B Cables (Option 001) Reference Designator Part Number Qty W3 85309-20139 1 A17 OUT - A11 IN W4 85309-20141 1 AT3 OUT - A12 IN W5 85309-20140 1 AT4 OUT - A13 IN W7 85309-20144 1 A16 OUT - RPJ3 W8 85309-60112 1 A16 - RPJ8 W10 85309-20138 1 A17 IN - RPJ1 W11 85309-60109 1 A15 OUT- RPJ9 W12 85309-60109 1 A10 IN- RPJ4 W14 85309-60109 1 A14 OUT- RPJ10 W15 8120-8483 1 A2J1- RPJ5 W16 8120-8483 1 A2J2-
LO/IF Distribution Unit 85309B Figure 4-4 Service Replaceable Parts Cable Locations (Option 001) Service Guide 85309-90082 4-11
Service Replaceable Parts Table 4-5 LO/IF Distribution Unit 85309B Major 85309B Assemblies (Option 002) Reference Designator Part Number Qty A1 A2 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 A22 A24 A25 A26 A27 A28 A29 A43 A44 A45 AT1 AT2 AT3 AT4 B1 SW1 85309-60135 85309-60134 0955-0791 0955-2245 5087-7755 5087-7755 85309-80026 85309-80026 5076-6336 5087-7768 85309-61601 5076-6336 0950-5271 5087-7755 5076-6336 85309-80026 5087-7755 5076-6336 85309-80026 85110-80015 85110-80015 85110-80015 33340-67007 33
LO/IF Distribution Unit 85309B Figure 4-5 Service Replaceable Parts Major 85309B Assemblies (Option 002) Service Guide 85309-90082 4-13
Service Replaceable Parts Table 4-6 LO/IF Distribution Unit 85309B Cables (Option 002) Reference Designator Part Number Qty W3 85309-20139 1 A17 OUT - A11 IN W4 85309-20141 1 AT3 OUT - A12 IN W5 85309-20140 1 AT4 OUT - A13 IN W7 85309-20144 1 A16 OUT - RPJ3 W8 85309-60112 1 A16 - RPJ8 W10 85309-20138 1 A17 IN - RPJ1 W11 85309-60109 1 A15 OUT- RPJ9 W12 85309-60109 1 A10 IN- RPJ4 W14 85309-60109 1 A14 OUT- RPJ10 W15 8120-8483 1 A2J1- RPJ5 W16 8120-8483 1 A2J2-
LO/IF Distribution Unit 85309B Figure 4-6 Service Replaceable Parts Cable Locations (Option 002) Service Guide 85309-90082 4-15
Service Replaceable Parts Table 4-7 Reference Designator A1 A2 A5 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 A22 A30 A31 A32 A36 A38 A43 AT3 AT4 AT7 AT9 AT10 B1 SW1 SW2 SW3, SW4, SW5 4-16 LO/IF Distribution Unit 85309B Major 85309B Assemblies (Option 400) Part Number Qty Description 85309-60135 85309-60134 85309-60136 0955-0791 0955-2245 5087-7755 5087-7755 85309-80026 85309-80026 5076-6336 5087-7768 85309-61601 5076-6336 0950-5271 85309-80024 85309-80025 85309-80022 0955-2249 0955-2249 85110-80015 3334
LO/IF Distribution Unit 85309B Figure 4-7 Service Replaceable Parts Major 85309B Assemblies (Option 400) Service Guide 85309-90082 4-17
Service Replaceable Parts Table 4-8 LO/IF Distribution Unit 85309B Cables (Option 400) Reference Designator Part Number Qty W1 85309-20151 1 RPJ1 - SW3 (C) W2 85309-20152 1 SW3 (2) - A17 IN W3 85309-20139 1 A17 OUT - A11 IN W4 85309-20141 1 AT3 OUT - A12 IN W5 85309-20140 1 AT4 OUT - A13 IN W6 85309-20156 1 A13 OUT - SW5 (2) W7 85309-20144 1 A16 OUT - RPJ3 W8 85309-60112 1 A16 - RPJ8 W9 85309-20156 1 A12 OUT - SW4 (2) W11 85309-60109 1 A15 OUT- RPJ9 W12 85309-
LO/IF Distribution Unit 85309B Figure 4-8 Service Replaceable Parts Cable Locations (Option 400) Service Guide 85309-90082 4-19
Service Replaceable Parts Table 4-9 Reference Designator A1 A2 A5 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 A22 A24 A25 A26 A30 A31 A32 A36 A37 A38 A43 A44 AT1 AT3 AT4 AT7 AT8 AT9 AT10 B1 SW1 SW2 SW3, SW4, SW5, SW6 4-20 LO/IF Distribution Unit 85309B Major 85309B Assemblies (Option 401) Part Number Qty Description 85309-60135 85309-60134 85309-60136 0955-0791 0955-2245 5087-7755 5087-7755 85309-80026 85309-80026 5076-6336 5087-7768 85309-61601 5076-6336 0950-5271 5087-7755 5076-6336 85309-80026 85309-8
LO/IF Distribution Unit 85309B Figure 4-9 Service Replaceable Parts Major 85309B Assemblies (Option 401) Service Guide 85309-90082 4-21
Service Replaceable Parts Table 4-10 LO/IF Distribution Unit 85309B Cables (Option 401) Reference Designator Part Number Qty W1 85309-20151 1 RPJ1 - SW3 (C) W2 85309-20152 1 SW3 (2) - A17 IN W3 85309-20139 1 A17 OUT - A11 IN W4 85309-20141 1 AT3 OUT - A12 IN W5 85309-20140 1 AT4 OUT - A13 IN W6 85309-20156 1 A13 OUT - SW5 (2) W7 85309-20144 1 A16 OUT - RPJ3 W8 85309-60112 1 A16 - RPJ8 W9 85309-20156 1 A12 OUT - SW4 (2) W11 85309-60109 1 A15 OUT- RPJ9 W12 85309
LO/IF Distribution Unit 85309B Figure 4-10 Service Replaceable Parts Cable Locations (Option 401) Service Guide 85309-90082 4-23
Service Replaceable Parts Table 4-11 Reference Designator A1 A2 A5 A10 A11 A12 A13 A14 A15 A16 A17 A18 A19 A22 A24 A25 A26 A27 A28 A29 A30 A31 A32 A36 A37 A38 A43 A44 A45 A46 AT1 AT2 AT3 AT4 AT7 AT8 AT9 AT10 AT11 B1 SW1 SW2 SW3 - SW7 4-24 LO/IF Distribution Unit 85309B Major 85309B Assemblies (Option 402) Part Number Qty 85309-60135 85309-60134 85309-60136 0955-0791 0955-2245 5087-7755 5087-7755 85309-80026 85309-80026 5076-6336 5087-7768 85309-61601 5076-6336 0950-5271 5087-7755 5076-6336 85309-80026
LO/IF Distribution Unit 85309B Figure 4-11 Service Replaceable Parts Major 85309B Assemblies (Option 402) Service Guide 85309-90082 4-25
Service Replaceable Parts Table 4-12 Reference Designator W1 W2 W3 W4 W5 W6 W7 W8 W9 W11 W12 W14 W15 W16 W23 W25 W30 W31 W32 W33 W34 W35 W36 W37 W38 W39 W40 W41 W42 W43 W44 W45 W47 W48 W49 W50 W51 W54 W58 W59 4-26 LO/IF Distribution Unit 85309B Cables (Option 402) Part Number Qty 85309-20151 85309-20152 85309-20139 85309-20141 85309-20140 85309-20156 85309-20144 85309-60112 85309-20156 85309-60109 85309-60109 85309-60109 8120-8483 8120-8483 85309-20144 8120-5022 85309-20145 85309-20156 85309-20147 853
LO/IF Distribution Unit 85309B Figure 4-12 Service Replaceable Parts Cable Locations (Option 402) Service Guide 85309-90082 4-27
Service Replaceable Parts 4-28 LO/IF Distribution Unit 85309B Service Guide 85309-90082