Agilent Technologies 85309A H40, H41, H42 Distributed Frequency Converter Operating and Service Manual Supplement
Agilent Technologies 85309A Option H40, H41, H42 Distributed Frequency Converter 0.1 to 18 GHz Operating and Service Manual Modification Use this manual modification with manual part numbers 85310-90001 (printed August 1993) Manual part number: 85309-90079 Printed in USA January 2000 Revision 1.
Notice The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
What You’ll Find in this Manual . . . This Agilent Technologies 85309A H40, H41, H42 Operating and Service Manual Modification provides information specific to these options and in contrast to the information provided in the Agilent Technologies 85309A Operating and Service Manual (85309-90001). All other information contained in the Agilent Technologies 85309A Operating and Service Manual is still applicable.
Warranty Custom systems are warranted by contractual agreement between Agilent Technologies and the customer. Certification Agilent Technologies, Inc., certifies that this product met its published specifications at the time of shipment from the factory.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. EXCLUSIVE REMEDIES. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY. YEAR 2000.
Service and Support Adjustment, maintenance, or repair of this product must be performed by a qualified service technician. Please refer to the table below for your nearest service and support center. Contacting Agilent Technologies Online assistance: www.agilent.
Safety and Regulatory Information Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. This product has been designed and tested in accordance with international standards. WARNING The WARNING notice denotes a hazard. It calls attention to a procedure, practice, or the like, that, if not correctly performed or adhered to, could result in personal injury.
Safety Earth Ground This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.
Typeface Conventions • Used to emphasize important information: Use this software only with the Agilent Technologies . • Used for the title of a publication: Refer to the Agilent Technologies 85309A Option H40, H41, H42 Operating and Service Manual Modification. • Used to indicate a variable: Type LOAD BIN filename. Instrument Display • Used to show on-screen prompts and messages that you will see on the display of an instrument: The Agilent Technologies will display the message CAL1 SAVED.
x Agilent Technologies 85309A Option H40, H41, H42 Operating and Service Manual
Introduction Introduction This operating and service manual modification describes the differences in the Agilent Technologies 85309A H40, H41 and H42 options compared to the standard Agilent Technologies 85309A option 001 LO/IF distribution unit. It also describes the manual changes necessary to document the Agilent Technologies 85309A H40, H41 and H42.
Introduction The spurious signal levels seen on the Agilent Technologies 8510/30 depend on the LO power level setting. • At the recommended +10 dBm setting, spur levels should be no greater than –100 dB. • With a power level of as much as +23 dBm, spurs may be detected as high as –60 dB.
Revised Installation Revised Installation The following modifies the “Installation” section of the Agilent Technologies 85309A Operating and Service Manual, pages 2-5 through 2-7. ac Power Connections No line voltage selector setting is required. The ac input power that the option H40, H41, and H42 can accept is 90 to 132 Vac or 198 to 264 Vac at 50-60 Hz.
Revised Operations Revised Operations The following modifies the “Operations” section of the Agilent Technologies 85309A Operating and Service Manual, page 3-2 through 3-4. Front and Rear Panel Feature Changes A green LED on the instrument’s front panel (left side) indicates which measurement band has been selected. LED ON indicates HIGH BAND has been selected and LED OFF indicates LOW BAND has been selected. Figures 1 and 2 show typical front and rear panels.
Revised Operations All rear panel functional descriptions are the same for the HIGH or LOW band operations. J15 Rear Panel Port, HIGH/LOW BAND SELECT When the front panel BAND SELECT switch is set to EXT, this port can be used to select the operating band desired by the following TTL levels: HIGH BAND: > + 4 Vdc LOW BAND: < 0.5 Vdc J7 and J8 (using nondiplexed mixers) These two rear panel ports are not available on the Agilent Technologies 85309A H40, H41, and H42 options.
Revised General Information Revised General Information The following general information supersedes “General Information, Table 5-3”, in the Agilent Technologies 85310A Operating and Service Manual. Typical Agilent Technologies 85301B System Performance Data Table 1 Typical Agilent Technologies 85301B System Performance Data with Agilent Technologies 85309A H40, H41, H42 Table 5-3a Specification (Typical) a Sensitivity b Compression Level (S/N=1, 0 average) (at 0.1 dB) GHz *0.1 - 0.3 *0.3 - 0.
Revised General Information Absolute Maximum Rating Parameter Agilent Technologies 85309A H40, H41, and H42 Operating Characteristics Values LO input port (CW) + 23 dBm Reference channel IF input port (CW) + 13 dBm Reference channel detector input ± 20 vdc Pos. Z blanking input ± 10 vdc Select HI/LOW input ± 5.5 vdc The following parameters are unique to Agilent Technologies 85309A options H40, H41 and H42.
Revised General Information Agilent Technologies 85381A Cable Length Limits The following figure defines the allowable maximum cable lengths between the LO source and the Agilent Technologies 85309A LO/IF distribution unit, and between the Agilent Technologies 85309A and the external mixers. The cable lengths are provided when using standard Agilent Technologies 85381A RF cable assemblies and with MicroCoax type UFB311A RF low-loss cable assemblies.
Revised General Information Figure 3 Cable Length Configurations (1 of 2) Agilent 85309A H40, H41, H42 Operating and Service Manual Modification 9
Revised General Information Cable Length Configurations (2 of 2) 10 Agilent 85309A H40, H41, H42 Operating and Service Manual Modification
Revised Replaceable Parts Revised Replaceable Parts Agilent Technologies 85309A H40, H41, H42 Major Assemblies The following replaceable parts list supersedes Table 7-19 in the Agilent Technologies 85309A Operating and Service Manual (page 7-58). Refer to Figures 3, 4, and 5.
Revised Replaceable Parts Reference Designator Description Agilent Technologies Part Number Quantity H40 option Quantity H41 option Quantity H42 option A32 Power divider, 1 GHz, 4way 85309-80022 1 1 1 A36 Filter, 100 MHz high-pass (Ref. Chan) 85309-80012 1 1 1 A37 Filter, 100 MHz high-pass (Test1 Chan) 85309-80012 1 1 1 A38 Filter, 100 MHz high-pass (Test2 Chan) 85309-80012 0 1 1 A39 Connection adapter, SMA m/f RT Ang (Ref. Chan) 1250-1249 1 1 1 A40 Connection .
Revised Replaceable Parts Reference Designator Description Agilent Technologies Part Number Quantity H40 option Quantity H41 option Quantity H42 option J3 Coax bulkhead connector, N(f)/SMA(f) 86290-60005 1 1 1 J4 Coax bulkhead connector, N(f)/SMA(f) 86290-60005 1 1 1 J5 Coax bulkhead connector, N(f)/SMA(f) 86290-60005 1 1 1 J6 Coax bulkhead connector, N(f)/SMA(f) 86290-60005 1 1 1 J9 Coax bulkhead connector, N(f)/SMA(f) 86290-60005 1 1 1 J10 Coax bulkhead connector, N(f
Revised Replaceable Parts Reference Designator Description Agilent Technologies Part Number Quantity H40 option Quantity H41 option Quantity H42 option W30 A11 Out (AT1) to A24 In 85309-20108 0 1 1 W31 A24 Out to SW6 (2) 85309-20109 1 0 1 1 1 0 1 1 W32 A25 LO/IF to RP (J11) 85309-20110 W35 A11 Out(AT2) to A27 In 85309-20111 0 0 1 W36 A27 Out to SW7 (2) 85309-20112 1 0 0 1 W37 A28 LO/IF to RP (J13) 85309-20113 1 0 0 1 W41 SW3(1) to A30 In (AT10) 85309-20114 1 1
Revised Replaceable Parts Reference Designator Description Agilent Technologies Part Number Quantity H40 option Quantity H41 option Quantity H42 option W15 Coax, flex, Ref IF signal/ A2(J1) to RP(J5) 8120-6118 1 1 1 W16 Coax, flex, PosZ signal/ A2(J2) to RP(J6) 8120-6118 1 1 1 W17 Ribbon, FP display intrface/ A2(J6) to A1(J1) 85309-60055 1 1 1 W18 Ribbon, Dual RF Amp dc Pwr/ A2(J3) to A12,13 BiasBd.
Revised Replaceable Parts Chassis Parts Reference Designator The following chassis parts list supersedes Table 7-25 contained in the Agilent Technologies 85309A Operating and Service Manual (page 7-72).
Revised Instrument Diagrams Revised Instrument Diagrams Major Components Layout Figure 4 Major Components Layout Locations Agilent 85309A H40, H41, H42 Operating and Service Manual Modification 17
Revised Instrument Diagrams Semirigid RF Cable Locations Figure 5 Semirigid RF Cable Locations 18 Agilent 85309A H40, H41, H42 Operating and Service Manual Modification
Revised Instrument Diagrams Agilent Technologies 85309A H40 Major Assembly Block Diagram For the Agilent Technologies 85309A H40, Figure 3, replaces Figure 7-24 in the Agilent Technologies 85309A Operating and Service Manual (pages 7-65).
Revised Instrument Diagrams Figure 6 Agilent Technologies 85309A H40 Block Diagram 20 Agilent 85309A H40, H41, H42 Operating and Service Manual Modification
Revised Instrument Diagrams Agilent Technologies 85309A H41 Major Assembly Block Diagram For the Agilent Technologies 85309A H41, Figure 4, replaces Figure 7-24 in the Agilent Technologies 85309A Operating and Service Manual (pages 7-65).
Revised Instrument Diagrams Figure 7 Agilent Technologies 85309A H41 Block Diagram 22 Agilent 85309A H40, H41, H42 Operating and Service Manual Modification
Revised Instrument Diagrams Agilent Technologies 85309A H42 Major Assemblies Block Diagram For the Agilent Technologies 85309A H42, Figure 8, replaces Figure 7-24 in the Agilent Technologies 85309A Operating and Service Manual (pages 7-65).
Revised Instrument Diagrams Figure 8 Agilent Technologies 85309A H42 Block Diagram 24 Agilent 85309A H40, H41, H42 Operating and Service Manual Modification