Agilent 85309A H20, H21 Distributed Frequency Converter 0.1 to 18 GHz Operating and Service Manual Modification Use this manual modification with instrument serial number 3224A 00499 and above. Use this manual modification with manual part numbers 85310-90001 (printed August 1993) Manual part number: 85309-90056 Printed in USA June 2000 Revision 6.
Notice The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
What You’ll Find in this Manual This Agilent 85309A H20, H21 Operating and Service Manual Modification provides information specific to these options and in contrast to the information provided in the Agilent 85309A Operating and Service Manual (p/n 85309-90001). All other information contained in the Agilent 85309A Operating and Service Manual is still applicable. Contents • Introduction, page 13, provides a description and typical system performance of the Agilent 85309A H20, H21.
Warranty Custom systems are warranted by contractual agreement between Agilent Technologies and the customer. Certification Agilent Technologies, Inc., certifies that this product met its published specifications at the time of shipment from the factory.
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. AGILENT TECHNOLOGIES SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OR MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. EXCLUSIVE REMEDIES. THE REMEDIES PROVIDED HEREIN ARE BUYER’S SOLE AND EXCLUSIVE REMEDIES. AGILENT TECHNOLOGIES SHALL NOT BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY. YEAR 2000.
Service and Support Any adjustment, maintenance, or repair of this product must be performed by qualified personnel. Contact your customer engineer through your local Agilent Technologies Service Center. You can find a list of local service representatives on the Web at: http://www.agilent.com/find/assist If you do not have access to the Internet, one of these centers can direct you to your nearest representative: Contacting Agilent Online assistance: www.agilent.
Safety and Regulatory Information Review this product and related documentation to familiarize yourself with safety markings and instructions before you operate the instrument. This product has been designed and tested in accordance with international standards. WARNING The WARNING notice denotes a hazard. It calls attention to a procedure, practice, or the like, that, if not correctly performed or adhered to, could result in personal injury.
Safety Earth Ground This is a Safety Class I product (provided with a protective earthing terminal). An uninterruptible safety earth ground must be provided from the main power source to the product input wiring terminals, power cord, or supplied power cord set. Whenever it is likely that the protection has been impaired, the product must be made inoperative and secured against any unintended operation.
Typeface Conventions • Used to emphasize important information: Use this software only with the Agilent 85309A H20, H21. • Used for the title of a publication: Refer to the Agilent 85309A H20, H21 Operating and Service Manual Modification. • Used to indicate a variable: Type LOAD BIN filename. Instrument Display • Used to show on-screen prompts and messages that you will see on the display of an instrument: The Agilent 85309A H20, H21 will display the message CAL1 SAVED.
12 Agilent 85309A H20, H21 Operating and Service Manual Modification
Introduction Introduction This operating and service manual modification describes the differences in the Agilent 85309A H20 and H21 options compared to the standard Agilent 85309A LO/IF distribution unit. It also describes the manual changes necessary to document the Agilent 85309A H20 and H21. Description The Agilent 85309A H20 and H21 are broadband, distributed frequency converters that utilize fundamental mixing to provide the performance required for antenna measurement systems.
Introduction • At the recommended +10 dBm setting, spur levels should be no greater than –100 dB. • With a power level of as much as +23 dBm, spurs may be detected as high as –60 dB.
Revised Installation Revised Installation The following modifies the “Installation” section of the Agilent 85309A Operating and Service Manual, pages 2-5 through 2-7. AC Power Connections No line voltage selector setting is required. The ac input power that the option H20 and H21 can accept is 90 to 132 Vac or 198 to 264 Vac at 50-60 Hz.
Revised Operations Revised Operations The following modifies the “Operations” section of the Agilent 85309A Operating and Service Manual, page 3-2 through 3-4. Front and Rear Panel Feature Changes A green LED on the instrument’s front panel (left side) indicates which measurement band has been selected. LED ON indicates HIGH BAND has been selected and LED OFF indicates LOW BAND has been selected. Figures 1 and 2 show typical front and rear panels.
Revised Operations All rear panel functional descriptions are the same for the HIGH or LOW band operations. J15 Rear Panel Port, HIGH/LOW BAND SELECT When the front panel BAND SELECT switch is set to EXT, this port can be used to select the operating band desired by the following TTL levels: HIGH BAND: > + 4 Vdc LOW BAND: < 0.5 Vdc J7 and J8 (using nondiplexed mixers) These two rear panel ports are not available on the Agilent 85309A H20, and H21 options.
Revised General Information Revised General Information The following general information supercedes “General Information, Table 5-3”, in the Agilent 85310A Operating and Service Manual. Typical Agilent 85301B System Performance Data Table 1 Typical Agilent 85301B System Performance Data with Agilent 85309A H20 and H21 Table 5-3a Specification (Typical) GHz *0.8 - 1 **0.
Revised General Information Absolute Maximum Rating Agilent 85309A H20 and H21 Operating Characteristics Parameter Values LO input port (CW) + 23 dBm Reference channel IF input port (CW) + 13 dBm Reference channel detector input ± 20 vdc Pos. Z blanking input ± 10 vdc Select HI/LOW input ± 5.5 vdc The following parameters are unique to Agilent 85309A options H20 and H21. These parameters supercede Tables 5-5, 5-6, and 5-7 in the Agilent 85310A Operating and Service Manual.
Revised General Information Agilent 85381A Cable Length Limits The following figure defines the allowable maximum cable lengths between the LO source and the Agilent 85309A LO/IF distribution unit, and between the Agilent 85309A and the external mixers. The cable lengths are provided when using standard Agilent 85381A RF cable assemblies and with MicroCoax type UFB311A RF low-loss cable assemblies.
Revised General Information Figure 3 External Mixer Configurations (1 of 2) Agilent 85309A H20, H21 Operating and Service Manual Modification 21
Revised General Information Figure 4 External Mixer Configurations (2 of 2) 22 Agilent 85309A H20, H21 Operating and Service Manual Modification
Revised Replaceable Parts Revised Replaceable Parts Agilent 85309A H20 and H21 Major Assemblies The following replaceable parts list supercedes Table 7-19 in the Agilent 85309A Operating and Service Manual (page 7-58). Refer to Figures 3, 4, and 5.
Revised Replaceable Parts Reference Designator Description Agilent Part Number Quantity H20 option Quantity H21 option A38 Filter, 100 MHz high-pass (Test2 Chan) 85309-80012 0 1 A39 Connection adapter, SMA m/f RT Ang (Ref. Chan) 1250-1249 1 1 A40 Connection .
Revised Replaceable Parts Reference Designator Description Agilent Part Number Quantity H20 option Quantity H21 option SW1 Switch, SPDT (part of W24 assy) 3101-0449 1 1 SW2 Switch, SP3T, rotary 3100-3244 1 1 SW3 Switch, RF SPDT, 15 Vdc 18 GHz 8762B #015 1 1 SW4 Switch, RF SPDT, 15 Vdc 18 GHz 8762B #015 1 1 SW5 Switch, RF SPDT, 15 Vdc 18 GHz 8762B #015 1 1 SW6 Switch, RF SPDT, 15 Vdc 18 GHz 8762B #015 0 1 RF Cables The following standard cables list supercedes Table 7-22
Revised Replaceable Parts Non-RF Cables The following standard cables list supercedes Table 7-23 in the Agilent 85309A Operating and Service Manual (page 7-68). Reference Designator Description Agilent Part Number Quantity H20 option Quantity H21 option W8 Coax, flex/ A16 I.F.
Revised Replaceable Parts Reference Designator Description Agilent Part Number Quantity H20 option Quantity H21 option W33 Coax, flex/ A25 I.F. Out to A26 In 8120-5531 0 1 W34 Coax, flex, Test2 IF signal/ A26 Out to RP(J12) 08760-63404 0 1 W40 Wire harness, dc Pwr Intrfc/ A23 to A2(P1) 85309-60052 1 1 Chassis Parts Reference Designator The following chassis parts list supercedes Table 7-25 contained in the Agilent 85309A Operating and Service Manual (page 7-72).
Revised Instrument Diagrams Revised Instrument Diagrams Components Layout Figure 5 Major Components Layout Locations 28 Agilent 85309A H20, H21 Operating and Service Manual Modification
Revised Instrument Diagrams Semirigid RF Cable Locations Figure 6 Semirigid RF Cable Locations Agilent 85309A H20, H21 Operating and Service Manual Modification 29
Revised Instrument Diagrams Agilent 85309A H20 Major Assembly Block Diagram Figure 7 For the Agilent 85309A H20, Figure 3, replaces Figure 7-24 in the Agilent 85309A Operating and Service Manual (page 7-65).
Revised Instrument Diagrams Agilent 85309A H21 Major Assembly Block Diagram Figure 8 For the Agilent 85309A H21, Figure 4, replaces Figure 7-24 in the Agilent 85309A Operating and Service Manual (page 7-65).
Revised Instrument Diagrams 32 Agilent 85309A H20, H21 Operating and Service Manual Modification