Agilent InfiniiVision 4000 X-Series Oscilloscopes Service Guide s1
Notices © Agilent Technologies, Inc. 2008-2012 Warranty No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions.
In This Service Guide This book provides the service information for the Agilent 4000 X- Series oscilloscopes. This manual is divided into these chapters: 1 Characteristics and Specifications This chapter contains a partial list of characteristics and specifications for the Agilent InfiniiVision 4000 X- Series oscilloscopes. 2 Testing Performance This chapter explains how to verify correct oscilloscope operation and perform tests to ensure that the oscilloscope meets the performance specifications.
7 Safety Notices At the front of the book you will find safety notice descriptions and document warranties. Digital Channels Because all of the oscilloscopes in the Agilent 4000 X-Series have analog channels, the analog channel topics in this book apply to all instruments. Whenever a topic discusses the digital channels, that information applies only to Mixed-Signal Oscilloscope (MSO) models or DSO models that have been upgraded to an MSO.
Contents In This Service Guide 1 Figures 9 Tables 11 3 Characteristics and Specifications Power Requirements 13 14 Measurement Category 15 Measurement Category 15 Measurement Category Definitions 15 Transient Withstand Capability 16 Environmental Conditions Specifications Contact us 2 18 19 Testing Performance Overview 17 21 22 List of Test Equipment Conventions 24 23 To construct the test connector (for use with MSO models only) 25 To test digital channels (MSO models only) 27 To verify
Contents What to Test 28 Verifying Test Results 28 To verify DC vertical gain accuracy To verify dual cursor accuracy To verify bandwidth (-3 dB) 33 39 45 To verify time base accuracy 51 To verify trigger sensitivity 53 Test Internal Trigger Sensitivity 54 Test External Trigger Sensitivity (all models) 57 Agilent 4000 X-Series Oscilloscopes Performance Test Record 3 Calibrating and Adjusting 61 65 User Calibration 67 To perform User Cal 67 User Cal Status 70 4 Troubleshooting 71 Solving Gen
Contents Equipment Required for Troubleshooting Internal Assemblies 82 To prepare for internal assembly troubleshooting 83 Flowchart for Internal Assembly Power Troubleshooting 86 System Board Drawings 87 To check the system board power supply test points 89 To check the line filter board AC output 91 To check the power switch 92 To check the power supply DC output 94 To check the display supplies 95 To check the keyboard supplies 96 To check the fan 97 5 Replacing Assemblies To remove the handle 99 102
Contents To remove the power supply 132 133 6 Replaceable Parts 135 Ordering Replaceable Parts 136 Listed Parts 136 Unlisted Parts 136 Direct Mail Order System 136 Exchange Assemblies 137 Exploded Views 138 Replaceable Parts List 7 Safety Notices Warnings 147 147 To clean the instrument Safety Symbols Index 8 141 148 149 151 4000 X-Series Oscilloscopes Service Guide
Figures Figure 1. Constructing the 8-by-2 Connector 26 Figure 2. Setting Up Equipment for Digital Channel Threshold Accuracy Test 30 Figure 3. Setting up Equipment for DC Vertical Gain Accuracy Test 36 Figure 4. Using a Blocking Capacitor to Reduce Noise 38 Figure 5. Setting up Equipment for Dual Cursor Accuracy Test 42 Figure 6. Using a Blocking Capacitor to Reduce Noise 44 Figure 7. Setting Up Equipment for Bandwidth (-3 dB) Verification Test 47 Figure 8.
Figures Figure 24. Figure 25. Figure 26. Figure 27. Figure 28. Figure 29. Figure 30. Figure 31. Figure 32. Figure 33. Figure 34. Figure 35. Figure 36. Figure 37. Figure 38. Figure 39. Figure 40. Figure 41. Figure 42. Figure 43. Figure 44. Figure 45. Figure 46. Figure 47. Figure 48. Figure 49. Figure 50. Figure 51. Figure 52. Figure 53. Figure 54.
Tables Table 1. Table 2. Table 3. Table 4. List of test equipment 23 Conventions 24 Materials required to construct the test connectors 25 Equipment Required to Test Digital Channel Threshold Accuracy 29 Table 5. Threshold Accuracy Voltage Test Settings 31 Table 6. DC Vertical Gain Accuracy Test Limits 33 Table 7. Equipment Required to Verify DC Vertical Gain Accuracy 34 Table 8. Settings Used to Verify DC Vertical Gain Accuracy 35 Table 9. Equipment Required to Verify Dual Cursor Accuracy 40 Table 10.
Tables 12 4000 X-Series Oscilloscopes Service Guide
Agilent InfiniiVision 4000 X-Series Oscilloscope Service Guide 1 Characteristics and Specifications Power Requirements 14 Measurement Category 15 Environmental Conditions 17 Specifications 18 Contact us 19 This chapter contains a partial list of characteristics and specifications for the Agilent InfiniiVision 4000 X- Series oscilloscopes. For a full list of Agilent InfiniiVision 4000 X- Series oscilloscopes characteristics and specifications see the data sheets. The data sheets are available at www.
1 Characteristics and Specifications Power Requirements Line voltage, frequency, and power: • ~Line 100- 120 Vac, 50/60/400 Hz • 100- 240 Vac, 50/60 Hz • 120 W max 14 4000 X-Series Oscilloscopes Service Guide
Characteristics and Specifications 1 Measurement Category Measurement Category The InfiniiVision 4000 X- Series oscilloscope is intended to be used for measurements in Measurement Category I. WA R N I N G Use this instrument only for measurements within its specified measurement category. Measurement Category Definitions Measurement category I is for measurements performed on circuits not directly connected to MAINS.
1 Characteristics and Specifications Transient Withstand Capability CAUTION Maximum input voltage for analog inputs CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk 50 Ω input: 5 Vrms Input protection is enabled in 50 Ω mode and the 50 Ω load will disconnect if greater than 5 Vrms is detected. However the inputs could still be damaged, depending on the time constant of the signal. The 50 Ω input protection only functions when the oscilloscope is powered on.
Characteristics and Specifications 1 Environmental Conditions Environment Indoor use only. Ambient temperature Operating 0 °C to +55 °C; non-operating –30 °C to +70 °C Humidity Operating: 50% to 95% RH at 40 °C for 5 days. Non-operating: 90% RH at 65 °C for 24 hr. Altitude Maximum operating altitude: 3,000 m (9,842 ft) Overvoltage Category This product is intended to be powered by MAINS that comply to Overvoltage Category II, which is typical of cord-and-plug connected equipment.
1 Characteristics and Specifications Specifications Please see the InfiniiVision 4000 X- Series Oscilloscopes Data Sheet for complete, up- to- date specifications and characteristics. To download a copy of the data sheet please visit: www.agilent.com/find/4000X- Series. Or go to the Agilent home page at www.agilent.com and search for 4000 X-Series oscilloscopes data sheet. To order a data sheet by phone, please contact your local Agilent office. A contact list is provided on the next page.
Characteristics and Specifications 1 Contact us Americas Canada (877) 894-4414 Latin America 305 269 7500 United States (800) 829-4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 81 426 56 7832 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Thailand 1 800 226 008 Europe Austria 0820 87 44 11 Belgium 32 (0) 2 404 93 40 Denmark 45 70 13 15 15 Finland 358 (0) 10 855 2100 France 0825 010 700 Germany 01805 24 63
1 20 Characteristics and Specifications 4000 X-Series Oscilloscopes Service Guide
Agilent InfiniiVision 4000 X-Series Oscilloscope Service Guide 2 Testing Performance Overview 22 List of Test Equipment 23 To construct the test connector (for use with MSO models only) 25 To test digital channels (MSO models only) 27 To verify digital channel threshold accuracy (MSO models only) 28 To verify DC vertical gain accuracy 33 To verify dual cursor accuracy 39 To verify bandwidth (-3 dB) 45 To verify time base accuracy 51 To verify trigger sensitivity 53 Agilent 4000 X-Series Oscilloscopes Perfo
2 Testing Performance Overview To completely test and troubleshoot MSO models, create and use the test connector accessory as described in this chapter. • The test connector is only required for oscilloscopes that have the MSO option licensed (enabled). • The connector is used in the digital channel threshold accuracy test. • The test connector keeps electrical distortion to a minimum and makes it easy for you to connect the oscilloscope probes to function generators and measurement equipment.
Testing Performance 2 List of Test Equipment Below is a list of test equipment and accessories required to perform the performance test verification procedures. Table 1 List of test equipment Equipment Critical Specifications Recommended Model/ Part Number Test connector, 8-by-2* See page 25 for instructions on building test connector. n/a Digital Multimeter 0.1 mV resolution, 0.005% accuracy Agilent 34401A Power Splitter Outputs differ by 0.
2 Testing Performance Table 1 List of test equipment (continued) Recommended Model/ Part Number Equipment Critical Specifications Adapter Type N (m) to BNC (m) Agilent 1250-0082 or Pomona 3288 with Pomona 3533 Blocking capacitor and shorting cap Note: if a BNC blocking capacitor is not available use an SMA blocking capacitor.
Testing Performance 2 To construct the test connector (for use with MSO models only) Agilent 4000 X- Series oscilloscopes that have digital channels enabled require the test connector described below. Follow the steps to build the test connector. Table 3 Materials required to construct the test connectors Description Recommended Part Qty BNC (f) Connector Agilent 1250-1032 or Pomona 4578 1 Berg Strip, 8-by-2 3M .100” x .
2 Testing Performance Jumper (2) Ground Lead (from scope’s MSO cable) Signal Lead (from scope’s MSO cable) 8 x 2 Berg Strip BNC Panel Mount Connector Figure 1 26 Constructing the 8-by-2 Connector 4000 X-Series Oscilloscopes Service Guide
Testing Performance 2 To test digital channels (MSO models only) The acquisition system testing provides confidence that the acquisition system is functioning correctly. It does not, however, check a particular specification. 1 Disconnect all probes from the circuit under test and from any other input source. 2 Using probe leads and grabbers, connect digital channels D0, D1, D2, and D3 to the Probe Comp signal on the center of the front panel. 3 Press the [AutoScale] key.
2 Testing Performance To verify digital channel threshold accuracy (MSO models only) This test verifies the digital channel threshold accuracy specification of the Agilent 4000 X- Series oscilloscopes. Threshold accuracy test limits: ±(100 mV + 3% of threshold setting) When to Test You should perform this test every two years or after 4000 hours of operation, whichever comes first. What to Test Use these instructions to test the threshold settings of digital channels D7- D0.
Testing Performance Table 4 2 Equipment Required to Test Digital Channel Threshold Accuracy Equipment Critical Specifications Recommended Model/Part Digital Multimeter 0.1 mV resolution, 0.005% accuracy Agilent 34401A Oscilloscope Calibrator DC offset voltage 6.
2 Testing Performance 4000 X-Series Oscilloscope Oscilloscope Calibrator Digital Multimeter Probe Cables BNC Tee BNC-Bananna Cable Test Connector Figure 2 Channels 0-7 Channels 8 - 15 Setting Up Equipment for Digital Channel Threshold Accuracy Test 4 Use a BNC- banana cable to connect the multimeter to the other side of the BNC Tee. 5 Connect the BNC Tee to the Channel 1 output of the calibrator as shown in Figure 2.
Testing Performance 2 7 Press the User softkey to the right of the D7 - D0 softkey, then turn the Entry knob ( ) on the front panel of the oscilloscope to set the threshold test settings as shown in Table 5. Table 5 Threshold Accuracy Voltage Test Settings Threshold voltage setting (in oscilloscope User softkey) DC offset voltage setting (on oscilloscope calibrator) +5.00 V +5.250 V ±1 mV dc Lower limit = +4.750 V Upper limit = +5.250 V –5.00 V –4.750 V ±1 mV dc Lower limit = –5.
2 Testing Performance digital high levels. Record the oscilloscope calibrator voltage in the Performance Test Record (see page 61). Before proceeding to the next step, make sure that you have recorded the oscilloscope calibrator voltage levels for each of the threshold settings shown in Table 5. 9 When testing 4000 X- Series MSOs, use the 8- by- 2 test connector to connect digital channels D15- D8 to the output of the oscilloscope calibrator.
Testing Performance 2 To verify DC vertical gain accuracy This test verifies the accuracy of the analog channel DC vertical gain for each channel. In this test, you will measure the dc voltage output of an oscilloscope calibrator using the oscilloscope’s Average - Full Screen voltage measurement and compare the results with the multimeter reading. Table 6 DC Vertical Gain Accuracy Test Limits Models Test Limits Notes 4000 X-Series ±2.
2 Testing Performance Table 7 Equipment Required to Verify DC Vertical Gain Accuracy Recommended Model/Part Equipment Critical Specifications Oscilloscope Calibrator 14 mV to 35 Vdc, 0.1 V resolution Fluke 5820A Digital multimeter Better than 0.
Testing Performance Table 8 2 Settings Used to Verify DC Vertical Gain Accuracy Volts/Div Setting Oscilloscope Calibrator Setting Test Limits 5 V/Div 35 V 34.2 V to 35.8 V 2 V/Div 14 V 13.68 V to 14.32 V 1 V/Div 7V 6.84 V to 7.16 V 500 mV/Div 3.5 V 3.42 V to 3.58 V 200 mV/Div 1.4 V 1.368 V to 1.432 V 100 mV/Div 700 mV 684 mV to 716 mV 50 mV/Div 350 mV 342 mV to 358 mV 20 mV/Div 140 mV 136.8 mV to 143.2 mV 10 mV/Div 70 mV 68.4 mV to 71.
2 Testing Performance c Press Type:; then, turn the Entry knob to select Average - Full Screen, and press Add Measurement. 4 Read the “current” average voltage value as V1. 5 Use the BNC tee and cables to connect the oscilloscope calibrator /power supply to both the oscilloscope and the multimeter (see Figure 3).
Testing Performance 2 6 Adjust the output so that the multimeter reading displays the first Volts/div calibrator setting value in Table 8 (depending on the oscilloscope model). 7 Disconnect the multimeter. 8 Wait until the measurement settles. 9 Read the “current” average voltage value again as V2. 10 Calculate the difference V2 - V1. The difference in average voltage readings should be within the test limits of Table 8 (depending on the oscilloscope model).
2 Testing Performance Use a Blocking Capacitor to Reduce Noise On the more sensitive ranges, such as 1 mV/div, 2 mV/div, and 5 mV/div, noise may be a factor. To eliminate the noise, add a BNC Tee, blocking capacitor, and shorting cap at the oscilloscope channel input to shunt the noise to ground. See Figure 4. If a BNC capacitor is not available, use an SMA blocking capacitor, adapter, and cap. See “Blocking capacitor and shorting cap” in the equipment list on page 24 for details.
Testing Performance 2 To verify dual cursor accuracy This test verifies the dual cursor accuracy for each analog channel. This test is similar to the test for verifying the DC vertical gain, except you will measure the dc voltage output of an oscilloscope calibrator using dual cursors on the oscilloscope and compare the results with the multimeter reading. Dual cursor accuracy test limits: ±[DC vertical gain accuracy + 0.
2 Testing Performance Table 9 Equipment Required to Verify Dual Cursor Accuracy Recommended Model/Part Equipment Critical Specifications Oscilloscope Calibrator 14 mV to 35 Vdc, 0.1 V resolution Fluke 5820A Digital multimeter Better than 0.
Testing Performance Table 10 2 Settings Used to Verify Dual Cursor Accuracy Volts/Div Setting Oscilloscope Calibrator Setting Test Limits 5 V/Div 35 V 34.032 V to 35.968 V 2 V/Div 14 V 13.6128 V to 14.3872 V 1 V/Div 7V 6.8064 V to 7.1936 V 500 mV/Div 3.5 V 3.4032 V to 3.5968 V 200 mV/Div 1.4 V 1.36128 V to 1.43872 V 100 mV/Div 700 mV 680.64 mV to 719.36 mV 50 mV/Div 350 mV 340.32 mV to 359.68 mV 20 mV/Div 140 mV 136.128 mV to 143.87 mV 10 mV/Div 70 mV 68.
2 Testing Performance 4 Use the BNC tee and cables to connect the oscilloscope calibrator /power supply to both the oscilloscope and the multimeter (see Figure 5). Oscilloscope Oscilloscope Calibrator BNC Tee Digital Multimeter BNC (f) to dual bananna adapter Figure 5 Setting up Equipment for Dual Cursor Accuracy Test 5 Adjust the output so that the multimeter reading displays the first Volts/div calibrator setting value in Table 10. 6 Disconnect the multimeter.
Testing Performance 2 8 Press the Y2 softkey, then position the Y2 cursor to the center of the voltage trace using the Entry knob. The ΔY value on the lower line of the display should be within the test limits of Table 10. If a result is not within the test limits, go to the “Troubleshooting” chapter. Then return here. 9 Disconnect the oscilloscope calibrator from the oscilloscope. 10 Repeat this procedure to check the dual cursor accuracy with the remaining Volts/div setting values in Table 10.
2 Testing Performance Use a Blocking Capacitor to Reduce Noise On the more sensitive ranges, such as 1 mV/div, 2 mV/div, and 5 mV/div, noise may be a factor. To eliminate the noise, add a BNC Tee, blocking capacitor, and shorting cap at the oscilloscope channel input to shunt the noise to ground. See Figure 6. If a BNC capacitor is not available, use an SMA blocking capacitor, adapter, and cap. See “Blocking capacitor and shorting cap” in the equipment list on page 24 for details.
Testing Performance 2 To verify bandwidth (-3 dB) This test checks the bandwidth (- 3 dB) of the oscilloscope. In this test you will use a signal generator and a power meter. Table 11 Bandwidth (-3 dB) Test Limits Models Test Limits 1.5 GHz Models All channels (-3 dB), dc to 1.
2 Testing Performance Table 12 Equipment Required to Verify Bandwidth (-3 dB) Equipment Critical Specifications Recommended Model/Part Signal Generator 100 kHz - 1.5 GHz at 200 mVrms Agilent N5181A with 3 GHz option Power Meter 1 MHz - 1.5 GHz ±3% accuracy Agilent N1914A Power Sensor 1 MHz - 1.5 GHz ±3% accuracy Agilent E9304A or N8482A Power Splitter outputs differ by < 0.
Testing Performance 2 Oscilloscope Signal Generator N to BNC Adapter Power Splitter Power Sensor N Cable Power Meter Figure 7 Setting Up Equipment for Bandwidth (-3 dB) Verification Test 2 Set up the power meter. Set the power meter to display measurements in units of watts. 3 Set up the oscilloscope. a Press the [Default Setup] key. b Set channel 1 Coupling to DC. c Set channel 1 Imped to 50 Ohm.
2 Testing Performance d Set the time base to 500 ns/div. e Set the Volts/Div for channel 1 to 200 mV/div. f Press the [Acquire] key, then press the Averaging softkey. g Turn the Entry knob to set # Avgs to 8 averages. 4 Set the signal generator for 1 MHz and six divisions of amplitude. The signal on the oscilloscope screen should be about five cycles at six divisions amplitude. 5 Set up the Amplitude measurement a Press the [Meas] key. b Press the Clear Meas softkey and then the Clear All softkey.
Testing Performance 2 9 Change the signal generator output frequency according to the maximum frequency for the oscilloscope using the following: • 1.5 GHz Models: 1.5 GHz • 1 GHz Models: 1 GHz • 500 MHz Models: 500 MHz • 350 MHz Models: 350 MHz • 200 MHz Models: 200 MHz 10 Referencing the frequency from step 9, set the power meter Cal Factor % to the frequency value on the calibration chart on the power sensor. 11 Set the oscilloscope sweep speed according to the following: • 1.
2 Testing Performance Then after converting the values from the power meter to Vrms: response(dB) = 20 log10 161.6 mV / 185.3 mV = -1.16 dB 210.4 mV/ 211.2 mV 15 The result from step 14 should be within - 3.0 dB. Record the result in the Performance Test Record (see page 61). 16 Move the power splitter from the channel 1 to the channel 2 input. 17 Turn off the current channel and turn on the next channel using the channel keys.
Testing Performance 2 To verify time base accuracy This test verifies the accuracy of the time base. In this test you will measure the absolute error of the time base oscillator and compare the results to the specification. Table 13 Equipment Required to Verify Time Base Accuracy Recommended Model/Part Equipment Critical Specifications Signal Generator 100 kHz - 1.5 GHz, 0.
2 Testing Performance 4 Make the measurement. a Set oscilloscope horizontal sweep speed control to 1 ms/div. b Set horizontal position control to +1 ms (rotate control CCW). c Set the oscilloscope horizontal sweep speed control to 5 ns/div. d Record the number of nanoseconds from where the rising edge crosses the center horizontal grid line to the center vertical grid line. The number of nanoseconds is equivalent to the time base error in ppm.
Testing Performance 2 To verify trigger sensitivity This test verifies the trigger sensitivity. In this test, you will apply a sine wave to the oscilloscope at the upper bandwidth limit. You will then decrease the amplitude of the signal to the specified levels, and check to see if the oscilloscope is still triggered. Table 14 Internal Trigger SensitivityTest Limits BW Models V/div Frequency 200 MHz, 350 MHz, 500 MHz, and 1 GHz < 10 mV/div All greater of 1 div or 5 mVpp >= 10 mV/div All 0.
2 Testing Performance Table 16 Equipment Required to Verify Trigger Sensitivity Equipment Critical Specifications Recommended Model/Part Signal Generator 100 kHz to 1.5 GHz sine waves Agilent N5181A with 3 GHz option Power splitter Outputs differ < 0.15 dB Agilent 11667A Power Meter 1.5 GHz ±3% accuracy Agilent N1914A Power Sensor 1.
Testing Performance 2 Oscilloscope Signal Generator N to BNC Adapter N Cable Figure 8 Setting Up Equipment for Internal Trigger Sensitivity Test b Set channel 1 Imped to 50 Ohm. 4 To verify the trigger sensitivity at the oscilloscope’s maximum bandwidth, set the output frequency of the signal generator to the maximum bandwidth of the oscilloscope: • 1.5 GHz models: 1.5 GHz and 1 GHz. • 1 GHz models: 1 GHz. • 500 MHz models: 500 MHz. • 350 MHz models: 350 MHz. • 200 MHz models: 200 MHz.
2 Testing Performance e Decrease the amplitude from the signal generator until 1 vertical division of the signal (about 5 mVpp) is displayed. The trigger is stable when the displayed waveform is stable. If the trigger is not stable, try adjusting the trigger level. If adjusting the trigger level makes the trigger stable, the test still passes. If adjusting the trigger does not help, see the “Troubleshooting” chapter. Then return here.
Testing Performance 2 Test External Trigger Sensitivity (all models) This test applies to all models. Verify the external trigger sensitivity at these settings: Table 17 External Trigger Sensitivity Test Settings Input Range 100 MHz 200 Mhz 1.6 V 40 mVpp 70 mVpp 8V 200 mVpp 350 mVpp 1 Connect the equipment (see Figure 9). a Use the N cable to connect the signal generator to the power splitter input.
2 Testing Performance Oscilloscope Signal Generator 50 Ohm Feedthrough N to BNC Adapter Power Splitter Power Sensor N Cable Power Meter Figure 9 58 Setting Up Equipment for 4-Channel External Trigger Sensitivity Test 4000 X-Series Oscilloscopes Service Guide
Testing Performance 2 2 Set up the oscilloscope. a Press the [Default Setup] key. b Press the [Mode/Coupling] key; then, press the Mode softkey to select Normal. 3 Change the signal generator output frequency to 100 MHz or 200 MHz. 4 Set the power meter Cal Factor % to the appropriate value (100 MHz or 200 MHz) on the calibration chart on the power sensor. If necessary, do a linear interpolation if a 100 MHz or 200 MHz factor is not included in the power meter’s calibration chart.
2 Testing Performance 8 Record the results as Pass or Fail in the Performance Test Record (see page 61). If the test fails, see the “Troubleshooting” chapter. Then return here.
Testing Performance 2 Agilent 4000 X-Series Oscilloscopes Performance Test Record Serial No. ______________________________________ Test Interval ____________________________________ Recommended Next Testing ________________________ Test by _____________________________ Work Order No. ______________________ Temperature ____________ Threshold Accuracy Test (100 mV + 3% of threshold setting) Limits 4.750 V 5.250 V -5.250 V -4.
2 Testing Performance Dual Cursor Accuracy Range Power Supply Setting 5 V/Div 35 V 2 V/Div 14 V 1 V/Div 7V 500 mV/Div 3.5 V 200 mV/Div 1.4 V 100 mV/Div 700 mV 50 mV/Div 350 mV 20 mV/Div 140 mV 10 mV/Div 70 mV 5 mV/Div 35 mV 2 mV/Div 14 mV 1 mV/Div 7 mV Bandwidth (-3 dB) Model 415x 410x 405x 403x 402x Test Limits 34.0 V to 36.0 V 13.6 V to 14.4 V 6.8 V to 7.2 V 3.4 V to 3.6 V 1.36 V to 1.44 V 680 mV to 720 mV 340 mV to 360 mV 136 mV to 144 mV 68 mV to 72 mV 34 mV to 36 mV 13.2 mV to 14.8 mV 6.2 mV to 7.
Testing Performance External Trigger Sensitivity Input Range: 8 V Generator Setting 200 MHz 100 MHz Input Range: 1.
2 64 Testing Performance 4000 X-Series Oscilloscopes Service Guide
Agilent InfiniiVision 4000 X-Series Oscilloscope Service Guide 3 Calibrating and Adjusting This chapter explains how to adjust the oscilloscope for optimum operating performance.
3 Calibrating and Adjusting CAUTION CAUTION CAUTION 66 REMOVE POWER TO AVOID DAMAGE! Do not disconnect any cables or remove any assemblies with power applied to the oscilloscope. Otherwise, damage to the oscilloscope can occur. USE EXTERNAL FAN TO REDUCE TEMPERATURE! When you must operate the oscilloscope with its cover and main shield removed, use an external fan to provide continuous air flow over the samplers (the ICs with heat sinks on them).
Calibrating and Adjusting 3 User Calibration Perform user- calibration: • Every two years or after 4000 hours of operation. • If the ambient temperature is >10° C from the calibration temperature. • If you want to maximize the measurement accuracy. The amount of use, environmental conditions, and experience with other instruments help determine if you need shorter User Cal intervals. User Cal performs an internal self- alignment routine to optimize the signal path in the oscilloscope.
3 Calibrating and Adjusting 2- channel oscilloscope or four equal- length cables for a 4- channel oscilloscope. Use 50Ω RG58AU or equivalent BNC cables when performing User Cal. a For a 2- channel oscilloscope, connect a BNC tee to the equal length cables. Then connect a BNC(f)- to- BNC(f) (also called a barrel connector) to the tee as shown below.
Calibrating and Adjusting 3 BNC(f)- to- BNC(f) (barrel connector) to the tee as shown below. To Channel 1 To Channel 2 To Channel 3 To Channel 4 Figure 11 Longer cable to TRIG OUT User Calibration cable for 4-channel oscilloscope 4 Connect a BNC cable (40 inches maximum) from the TRIG OUT connector on the rear panel to the BNC barrel connector. 5 Press the [Utility] key; then, press the Service softkey. 6 Begin the Self Cal by pressing the Start User Cal softkey.
3 Calibrating and Adjusting User Cal Status Pressing the User Cal Status softkey displays the following summary results of the previous User Cal, and the status of probe calibrations for probes that can be calibrated. Note that AutoProbes do not need to be calibrated, but InfiniiMax probes can be calibrated.
Agilent InfiniiVision 4000 X-Series Oscilloscope Service Guide 4 Troubleshooting Solving General Problems with the Oscilloscope 72 Verifying Basic Operation 75 Troubleshooting Internal Assemblies 82 Read All Cautions and Warnings Before you begin any troubleshooting, read all Warning and Cautions in the “Troubleshooting” section. This chapter begins with “Solving General Problems with the Oscilloscope”. It tells you what to do in these cases: • If there is no display. • If there is no trace display.
4 Troubleshooting Finally, this chapter describes procedures for “Troubleshooting Internal Assemblies” when performing assembly- level repair: • To prepare for internal assembly troubleshooting. • To check the system board power supply test points. • To check the line filter board AC output. • To check the power switch. • To check the power supply DC output. • To check the display supplies. • To check the keyboard supplies. • To check the fan.
Troubleshooting 4 If there is no trace display ✔ Check that the Intensity (on the front panel) is adjusted correctly. ✔ Recall the default setup by pressing [Default Setup]. This will ensure that the trigger mode is Auto. ✔ Check that the probe clips are securely connected to points in the circuit under test, and that the ground is connected. ✔ Check that the circuit under test is powered on. ✔ Press the [AutoScale] key. ✔ Obtain service from Agilent Technologies, if necessary.
4 Troubleshooting ✔ Check that persistence in the Display menu is turned off, then press the Clear Display softkey. ✔ Press the [Auto Scale] key. If you cannot see a channel ✔ Recall the default setup by pressing [Default Setup]. This will ensure that the trigger mode is Auto. ✔ Check that the oscilloscope probe’s BNC connector is securely attached to the oscilloscope’s input connector. ✔ Check that the probe clips are securely connected to points in the circuit under test.
Troubleshooting 4 Verifying Basic Operation To power-on the oscilloscope 1 Connect the power cord to the rear of the oscilloscope, then to a suitable ac voltage source. The oscilloscope power supply automatically adjusts for input line voltages in the range of 100 to 240 VAC. Ensure that you have the correct line cord (see page 135). The power cord provided is matched to the country of origin. WA R N I N G AVOID INJURY. Always operate the oscilloscope with an approved three conductor power cable.
4 Troubleshooting To perform hardware self test Pressing [Utility] > Service > Hardware Self Test performs a series of internal procedures to verify that the oscilloscope is operating properly. It is recommended you run Hardware Self Test: • After experiencing abnormal operation. • For additional information to better describe an oscilloscope failure. • To verify proper operation after the oscilloscope has been repaired.
Troubleshooting 4 To verify default setup The oscilloscope is designed to turn on with the setup from the last turn on or previous setup. To recall the default setup: 1 Press the [Default Setup] key. This returns the oscilloscope to its default settings and places the oscilloscope in a known operating condition. The major default settings are: • Horizontal: • main mode. • 100 us/div scale. • 0 s delay. • center time reference. • Vertical: • Channel 1 on. • 5 V/div scale. • dc coupling. • 0 V position.
4 Troubleshooting • Other: • Acquire mode normal. • Run/Stop to Run. • cursor measurements off. Figure 12 Default setup screen 2 If your screen looks substantially different, replace the system board. To perform an Auto Scale on the Probe Comp signal 1 Press the [Default Setup] key. The oscilloscope is now configured to its default settings. 2 Connect an oscilloscope probe from channel 1 to the Probe Comp signal terminal on the front panel.
Troubleshooting 4 4 Press [AutoScale]. 5 You should see a waveform on the oscilloscope’s display similar to this: If you see the waveform, but the square wave is not shaped correctly as shown above, perform the procedure “To compensate passive probes” on page 80. If you do not see the waveform, ensure your power source is adequate, the oscilloscope is properly powered- on, and the probe is connected securely to the front- panel analog channel input BNC and to the Demo 2 (Probe Comp) terminal.
4 Troubleshooting To compensate passive probes You should compensate your passive probes to match their characteristics to the oscilloscope’s channels. A poorly compensated probe can introduce measurement errors. 1 Perform the procedure “To perform an Auto Scale on the Probe Comp signal” on page 78 2 Press the channel key to which the probe is connected ([1], [2], etc.). 3 In the Channel Menu, press Probe. 4 In the Channel Probe Menu, press Probe Check; then, follow the instructions on- screen.
Troubleshooting 4 6 Repeat the procedure for each channel. The process of compensating the probes serves as a basic test to verify that the oscilloscope is functional.
4 Troubleshooting Troubleshooting Internal Assemblies The service policy for all bandwidth model oscilloscopes is assembly level replacement. You can use the procedures described in this section to help identify assemblies that need replacement. Generally, you want to make sure cables to the assembly are good and properly seated. Then, you check that the assembly is supplied with the proper power.
Troubleshooting 4 To prepare for internal assembly troubleshooting WA R N I N G WA R N I N G CAUTION CAUTION HAZARDOUS VOLTAGES EXIST — REMOVE POWER FIRST ! The procedures described in this section are performed with power supplied to the oscilloscope and with the protective covers removed. Only trained service personnel who are aware of the hazards involved should perform the procedures. Whenever possible, perform the procedures with the power cord removed from the oscilloscope.
4 Troubleshooting fan power cable connected as shown in the following figure. Figure 14 CAUTION Setup for troubleshooting internal assemblies MAKE SURE EARTH GROUND IS MAINTAINED FOR THE FRONT DECK ! For example, connect a cable with alligator clips between the rear deck chassis and the front deck chassis, or connect a BNC cable from one of the BNCs on the system board to a known grounded BNC on your workbench.
Troubleshooting CAUTION 4 USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS ! When you remove the oscilloscope cover and main shield, use an external fan to provide continuous air flow over the heat sinks. Air flow over the heat sinks is reduced when the cover and main shield are removed, which leads to higher than normal operating temperatures. Have the fan blow air across the system board where the heat sinks are located. Otherwise, damage to the components can occur.
4 Troubleshooting Flowchart for Internal Assembly Power Troubleshooting The following flowchart is a simplified overview of troubleshooting power to the oscilloscope’s internal assemblies.
Troubleshooting 4 System Board Drawings Use these drawings to locate test points on the oscilloscope system board.
4 Troubleshooting Figure 16 88 System Board Test Points/Connectors - Top Side, Right 4000 X-Series Oscilloscopes Service Guide
Troubleshooting 4 To check the system board power supply test points This procedure checks the power supply test points on the system board (see See Figure 16 on page 88). Values outside the expected range help identify bad assemblies. 1 Follow the instructions in “To prepare for internal assembly troubleshooting” on page 83. 2 Connect the negative lead of the multimeter to a ground point on the oscilloscope. 3 Connect the power cord, and turn on the oscilloscope.
4 Troubleshooting 5 Next, check the supplies coming from the system board: Test Point (near J3700) Expected Value Assemblies Supplied VP5V (+5V) +5 V ±3% Keyboard, Indicates supply coming from System Board system board is good. Replace the system board. VP12V (+12V) +12 V ±3% System Board Replace the system board. VP1V8 (+1.8V) +1.8 V ±3% System Board Replace the system board. VP1V4 (+1.4V) +1.4 V ±3% System Board Replace the system board.
Troubleshooting 4 To check the line filter board AC output When the 13 V bulk power is not being properly supplied, this procedure tests the AC input to the power supply to determine whether there is a problem with the line filter board. 1 Follow the instructions in “To prepare for internal assembly troubleshooting” on page 83. 2 Connect the power cord, and turn on the oscilloscope.
4 Troubleshooting • If there is no AC power at J1 of the power supply, there is something wrong with the AC mains section of the line filter assembly, and you need to replace the line filter assembly. • If there is AC power at J1 of the power supply, but the instrument still will not power ON when the power switch (S101) is in the ON position, go to “To check the power switch” on page 92.
Troubleshooting Figure 18 4 Verify power switch operation • With S101 in the OFF position (switch contacts closed) you should measure less than 1 ohm of resistance. • With S101 in the ON position (switch contacts open) you should measure infinite resistance. If you do not measure these two values, there is something wrong with the power switch (S101 on the line filter assembly), and you need to replace the line filter assembly.
4 Troubleshooting To check the power supply DC output This procedure checks the power supply DC output after verifying the AC input is good and checking the power switch operation. 1 Follow the instructions in “To prepare for internal assembly troubleshooting” on page 83. 2 Connect the negative lead of the multimeter to a ground point on the oscilloscope. 3 With the power cord disconnected: a Disconnect the DC supply cable that is plugged into J3700 of the system board.
Troubleshooting 4 To check the display supplies This procedure checks the display supply voltages on the system board test points when the display backlight power cable and the display cable are disconnected. If one or both of these voltages are bad when the cables are connected but good when the cables are disconnected, it indicates problems with the display assembly. 1 Follow the instructions in “To prepare for internal assembly troubleshooting” on page 83.
4 Troubleshooting Test Point (near J2000) Expected Value Assemblies Supplied If Good If Bad VP3V3 (+3.3V) +3.3 V ±3% LCD, System Board If good when the display cable is disconnected from the system board, but bad when the display cable is connected, this indicates a problem with either the display cable or the display assembly, 1 Turn off the oscilloscope, and disconnect the power cord. 2 Replace the display cable. 3 Connect the power cord, and turn on the oscilloscope.
Troubleshooting 4 5 Check the keyboard supplies: Test Point (near J4303) Expected Value Assemblies Supplied If Good VP5V (+5V) +5 V ±3% VP13V_PRB (+13V_PRB) +13 V ±3% VM15V_PRB (-15V_PRB) -13.2 V ±4% VP7V5_PRB (+7.5_PRB) +6.28 V ±4% VM7V5_PRB (-7.5_PRB) -6.
4 Troubleshooting 5 If the fan voltage is approximately +9 Vdc at room temperature, replace the fan. If the fan voltage is not approximately +9 Vdc, replace the system board. The proper voltage range depending on temperature is between +7.9 Vdc to +12.8 Vdc.
Agilent InfiniiVision 4000 X-Series Oscilloscope Service Guide 5 Replacing Assemblies The service policy for 200 MHz and lower bandwidth oscilloscopes is unit replacement, so the instructions in this chapter are primarily for the 350 MHz, 500 MHz, and 1 GHz bandwidth oscilloscopes. This chapter describes how to remove assemblies from an oscilloscope. To install a replacement assembly after you have removed an old one, follow the instructions in reverse order.
5 Replacing Assemblies To remove the line filter board 129 To remove the power supply 132 Tools Used for Disassembly Use these tools to remove and replace the oscilloscope assemblies: • T6, T10, and T20 TORX drivers. • 5/8- inch socket driver or adjustable wrench (for BNC nuts). • Flat head screw driver. See how the Oscilloscope Parts Fit Together An exploded view of the oscilloscope is included in the “Replaceable Parts” chapter.
Replacing Assemblies CAUTION CAUTION 5 REMOVE POWER TO AVOID DAMAGE! Remove power before you begin to remove and replace assemblies. Do not remove or replace assemblies while the oscilloscope is turned on, or damage to the components can occur. AVOID DAMAGE TO ELECTRONIC COMPONENTS! ELECTROSTATIC DISCHARGE (ESD) can damage electronic components. When doing any of the procedures in this chapter, use proper ESD precautions.
5 Replacing Assemblies To remove the handle The strap handle must be removed prior to removing the cabinet. The removal of the Strap handle may also be necessary when mounting oscilloscope in a rack. 1 Using T20 TORX driver, remove the screws holding strap handle caps in place. 2 Lift the strap handle off the cabinet.
Replacing Assemblies 5 To remove the adjustable legs Adjustable legs must be removed prior to removing the cabinet. 1 Using a T20 TORX, remove shoulder screw and washer. 2 Pull adjustable leg assembly from cabinet. Assembly contains latching mechanism and spring.
5 Replacing Assemblies To remove the cabinet Removing the cabinet allows access to the rear deck, fan assembly, power supply cover, power supply assembly, and power switch assembly. 1 Using T20 TORX, remove the three screws securing cabinet to rear deck assembly. 2 Carefully slide cabinet back away from rear deck assembly.
Replacing Assemblies 5 To remove the rear deck assembly Removing the rear deck allows access to the front deck, acquisition board, and inverter board. 1 Remove the BNC securing nuts and washers.
5 Replacing Assemblies 2 Using a T20 TORX, locate and remove all screws securing rear deck to front deck. Figure 24 Removing the rear deck assembly 3 Carefully separate rear deck from front deck. Take care not damage extender switch. WA R N I N G 106 Sheet metal parts may have sharp edges. Handle with care to avoid injury.
Replacing Assemblies 5 4 Disconnect power harness and fan cable from acquisition board.
5 Replacing Assemblies To replace the battery If the battery needs to be replaced, use a CR2032/1HG or CR2032/HGN 3V manganese dioxide lithium battery.
Replacing Assemblies 5 To remove the acquisition board The following illustrates how to remove the Acquisition and Dual inverter printed circuit boards. 1 Using a TORK T6 driver locate and remove the 4 screws on the front of the instrument (4 Channel version). Figure 27 Removing the analog channel BNC securing T6 screws 2 Disconnect Inverter, keyboard and display cables. Note locations for re- connection. It should be noted that cables can be removed from cable clamps at this time as well.
5 Replacing Assemblies 3 Using a TORX T10 driver locate and remove 15 mounting screws.
Replacing Assemblies 5 4 Carefully lift acquisition board off front deck.
5 Replacing Assemblies To remove the touch controller board 1 Using a TORX T10 driver locate and remove the two screws securing the touch controller shield.
Replacing Assemblies Figure 31 4000 X-Series Oscilloscopes Service Guide 5 Lifting the touch controller shield off 113
5 Replacing Assemblies 2 Remove the three screws securing the touch controller board to the front deck. Figure 32 Removing the touch controller board 3 Disconnect all cables, lift board off front deck.
Replacing Assemblies 5 To remove the front panel knobs The front panel knobs must be removed prior to localized front panel overlay installation or front panel disassembly. 1 Gently pull on the front panel knobs to remove them.
5 Replacing Assemblies To remove the front bezel assembly 1 Remove cable shield from front deck. Carefully squeeze the shield so that it clears the sheet metal tabs holding it place then slide it free of the front deck: WA R N I N G Thin sheet metal parts may have sharp edges. Handle with care to avoid injury.
Replacing Assemblies 5 2 Disconnect the end of the keyboard cable that extends thru front deck.
5 Replacing Assemblies 3 Remove the display cable from the cable guide.
Replacing Assemblies 5 4 Remove the bezel from the front deck. The bezel is secured to front deck by molded- in retaining clips located around the perimeter of the bezel. Gently pry these outward (either by hand or using a flat head screwdriver). Working your way around the bezel releasing the clips gently lift the bezel away from the front deck.
5 Replacing Assemblies To remove the display assembly 1 Using a TORX T10 remove the two screws that secure the display assembly to the front deck.
Replacing Assemblies 5 2 Disconnect the backlight power cable. Figure 39 Disconnecting the backlight power cable Take care while threading display cables thru front deck sheet metal openings.
5 Replacing Assemblies To remove the keyboard and keypad 1 Remove the softkey board cable from the keyboard board. Slide the connector clamp toward the softkey board; then, slide the ribbon cable out.
Replacing Assemblies 5 2 To separate the keyboard board from the bezel, carefully pull back locking tabs that secure it.
5 Replacing Assemblies 3 To separate the softkey board from the bezel, carefully pull back locking tabs that secure it.
Replacing Assemblies 5 To remove the fan assembly 1 Ensure that fan power cable has been disconnected from acquisition board. 2 Carefully slide fan assembly (fan and fan mount) to the right, then lift away from rear deck. Note, fan mount is soft and can be damaged by sharp sheet metal edges. Take care that the fan power cable is not damaged when pulling across sheet metal edges.
5 Replacing Assemblies To remove the power supply shield 1 To remove power supply shield, locate and remove using a TORX T20 the four screws securing the power supply shield to the rear deck.
Replacing Assemblies WA R N I N G 5 Thin sheet metal parts may have sharp edges. Handle with care to avoid injury. 2 Once screws have been removed, carefully remove the power supply cover by lifting the cover up and off retaining tabs on rear deck.
5 Replacing Assemblies 3 Carefully remove the air duct by lifting it up and off the rear deck.
Replacing Assemblies 5 To remove the line filter board 1 Disconnect the ground wire from its chassis terminal. Figure 49 Disconnecting the ground wire 2 Disconnect all cables from the line filter board to the power supply and remove the cables from the cable guide.
5 Replacing Assemblies 3 Locate and remove using a TORX T20 driver the single screw securing the assembly to the rear deck. Figure 50 Removing the line filter board 4 Slide assembly to right and lift out of rear deck. 5 Take care that you do not damage the switch extender during removal.
Replacing Assemblies 5 6 To remove the switch extender, gently pry open the extender using a flat head screwdriver.
5 Replacing Assemblies To remove the power supply 1 Disconnect all cables from power supply board. 2 Locate and remove using a TORX T10 driver the four screws securing the power supply assembly to the rear deck.
Replacing Assemblies 4000 X-Series Oscilloscopes Service Guide 5 133
5 134 Replacing Assemblies 4000 X-Series Oscilloscopes Service Guide
Agilent InfiniiVision 4000 X-Series Oscilloscope Service Guide 6 Replaceable Parts This chapter describes how to order replaceable assemblies and parts for the Agilent 4000 X- Series oscilloscopes. Diagrams and parts lists are included for assemblies and hardware that you can order. Before working on the oscilloscope, read the safety summary at the back of this book.
6 Replaceable Parts Ordering Replaceable Parts Listed Parts To order a part in the parts list, quote the Agilent Technologies part number, indicate the quantity desired, and address the order to the nearest Agilent Technologies Sales Office. To find your nearest sales office go to www.agilent.com. Unlisted Parts To order a part not listed in the parts list, include the instrument part number, instrument serial number, a description of the part (including its function), and the number of parts required.
Replaceable Parts 6 In order for Agilent Technologies to provide these advantages, please send a check or money order with each order. Mail order forms and specific ordering information are available through your local Agilent Technologies Sales Office. Addresses and telephone numbers are located in a separate document shipped with the manuals. Exchange Assemblies Some parts used in this instrument have been set up for an exchange program.
6 Replaceable Parts Exploded Views The following exploded views provide a graphical representation of the oscilloscope at the time this manual was released. Not all parts are shown. Your parts may be slightly different than those shown. These views provide reference designator numbers that map to those used in the parts list table in this chapter.
Replaceable Parts 6 MP10 MP1 MP7 MP3 MP4 MP6 MP9 MP2 A7 MP8 A3 MP12 Figure 53 Exploded View 1 of 2 4000 X-Series Oscilloscopes Service Guide 139
6 Replaceable Parts MP12 W1 MP15 H11 A4 MP14 A2 A5 MP28 A6 MP27 MP13 MP29 MP25 A10 A9 A8 MP30 Figure 54 140 MP32 Exploded View 2 of 2 4000 X-Series Oscilloscopes Service Guide
Replaceable Parts 6 Replaceable Parts List The information given for each part consists of the following: • Reference designation. • Agilent Technologies part number. • Total quantity (Qty) in the instrument or on assembly. • Description of the part.
6 Replaceable Parts Table 19 Replaceable Parts (continued) Ref Des Agilent Part Number Qty Description A2 54709-69501-DEF 1 Acquisition board 4000 X-Series 4-Ch - 1.5 GHz (return) A3 54709-66421 1 PCA - 2 Channel Keyboard 4000 X-Series A3 54709-66422 1 PCA - 4 Channel Keyboard 4000 X-Series A4 54709-66425 1 Touch Panel Controller board A5 0950-5419 1 Power Supply Switching AC/DC 150W A6 54709-66424 1 PCA - Line Filter Board A7 54709-44701 1 Display assembly.
Replaceable Parts Table 19 6 Replaceable Parts (continued) Ref Des Agilent Part Number Qty Description H10 54695-26302 2 Screw-Shoulder (not shown) H11 54709-24701 2 Spacer (at bottom of acquisition board) MP1 54702-94301 1 ID Label DSO-X 4022A MP1 54702-94302 1 ID Label MSO-X 4022A MP1 54703-94301 1 ID Label DSO-X 4024A MP1 54703-94302 1 ID Label MSO-X 4024A MP1 54704-94301 1 ID Label DSO-X 4032A MP1 54704-94302 1 ID Label MSO-X 4032A MP1 54705-94301 1 ID Label DS
6 Replaceable Parts Table 19 Replaceable Parts (continued) Ref Des Agilent Part Number Qty Description MP6 54707-94303 1 Front Keyboard Label 2 Channel MP6 54709-94303 1 Front Keyboard Label 4 Channel MP7 54709-60201 1 Panel Bezel Assembly MP8 54709-41901 1 Keyboard Keypad MP9 54709-41903 1 Softkey Keypad MP10 54709-44101 1 Front Cover MP12 54709-00101 1 Front Deck MP13 54709-00102 1 Rear Deck MP14 54709-00602 1 Touch Screen Controller Shield MP15 54709-00603 1
Replaceable Parts Table 19 6 Replaceable Parts (continued) Ref Des Agilent Part Number Qty Description MP30 54709-64401 1 Cabinet Assembly MP31 5042-6805 2 Cap - Strap Handle MP32 54695-44701 2 Foot Pad MP33 54709-94310 * Overlay - Traditional Chinese, 4 channel (not shown) MP33 54709-94309 * Overlay - Traditional Chinese, 2 channel (not shown) MP33 54709-94322 * Overlay - Japanese, 4 channel (not shown) MP33 54709-94321 * Overlay - Japanese, 2 channel (not shown) MP33 5
6 Replaceable Parts Table 19 Replaceable Parts (continued) Agilent Part Number Qty Description 54695-61605 1 Cable - Ground (not shown) 54709-61601 1 Keyboard Cable 54709-61603 1 Cable - DC (not shown) 54709-61605 1 Cable - AC (not shown) 54709-61606 1 Cable - Display (not shown) 54709-61607 1 Cable - Backlight and Touch Screen Controller (not shown) 54709-61608 1 Cable - Line Trigger (not shown) 54709-61609 1 Cable - Remote (not shown) 54709-61610 1 Soft Keyboard Cable (not
Agilent InfiniiVision 4000 X-Series Oscilloscope Service Guide 7 Safety Notices This apparatus has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Measuring Apparatus, and has been supplied in a safe condition. This is a Safety Class I instrument (provided with terminal for protective earthing). Before applying power, verify that the correct safety precautions are taken (see the following warnings).
Safety Notices If you energize this instrument by an auto transformer (for voltage reduction or mains isolation), the common terminal must be connected to the earth terminal of the power source. Whenever it is likely that the ground protection is impaired, you must make the instrument inoperative and secure it against any unintended operation. Service instructions are for trained service personnel. To avoid dangerous electric shock, do not perform any service unless qualified to do so.
Safety Notices Safety Symbols ! Instruction manual symbol: the product is marked with this symbol when it is necessary for you to refer to the instruction manual in order to protect against damage to the product. Hazardous voltage symbol. Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis.
Safety Notices 150 4000 X-Series Oscilloscopes Service Guide
Index A E N accuracy, DC vertical gain, 33 accuracy, dual cursor, 39 Agilent contact information, 19 assemblies exchange, 137 replacing, 99 equipment, test, 23 exchange assemblies, 137 exploded view front panel, 139 instrument, 140 NIST, 67 B fan assembly removal, 125 front panel exploded view, 139 removal, 105 front panel self test, 76 bandwidth verification, 45 F C cabinet, removing, 102, 115 calibration, 67 cautions, 83, 86, 89, 100, 101, 132 certificate of calibration, 67 channels problem solv
Index remove, 129 cabinet, 102, 115 display assembly, 116 fan assembly, 125 front panel, 105 handle, 103 keyboard, 109 power supply, 132 power supply shield, 126 storage lid, 104 replaceable parts list, 141 replacing assemblies, 99 run self-tests, 76 S safety notice, 65, 83 self test, front panel, 76 self test, hardware, 76 self-alignment, user cal, 67 self-tests, 76 specifications, 18 startup sequence, 75 status, User Cal, 70 status, user cal, 67 storage lid, removal, 104 troubleshooting, 71 equipment r