User Manual

Confidential Page 1 of 3 06/07/2011
F:\agbmi05\Personal\PCBs\HE\Cirrus\CirrusRFIDreaderFCC\CirrusRFIDreaderRegulatoryInetgrationInstr
uctions.doc
a
aa
a
b
bb
b
Medical Imaging
RDE
From:
Jan Vercammen (6233) – MI/TSP-HW/Electronics
Date: July 6, 2011
Concerning: Regulatory and hardware instructions for integration of Cirrus
RFID reader into host applications (concerns FCC ID: HPL5243)
This document describes the regulatory elements and measures for the integration of
a full modular short-range RFID reader into host applications. The internal company
reference of the RFID reader (or reader for short) is A800626.
The reader is a small printed circuit board with local shielding, local power supply regulation and
filtering, here is a short summary of the measures taken to control (spurious) emissions:
The reader transmitter (chip) has a local PCB shield
The reader module power supply is 5V+/-10% and using a local low dropout regulator a
3.3V+/-2% is used on board for all electric circuits, there is also sufficient bulk and high-
frequency decoupling
A common-mode (CM) choke is used on the 5V power supply as a means to limit
unwanted CM conducted- and radiated emissions
The tuning circuit uses low tolerance components (1% or 2%) to limit circuit variations
The full-modular approval allows the integration of the reader into various applications
without the need for radio type testing of the application, however, the user needs to
take a number of measures and comply with regulatory requirements.

Summary of content (3 pages)