Vacuum Pumps Instrumentation Fittings and Valves LEYBOLD VACUUM GA 01.202/9.02 TRIVAC® B Rotary Vane Vacuum Pump S/D 16, 25 B Cat.-No.
Contents Figures Contents 1 1.1 1.2 1.3 1.4 1.5 1.6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Page Description . . . . . . . . . . . . . . . . . . . . . . . . . .3 Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Supplied Equipment . . . . . . . . . . . . . . . . . . . .5 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . .6 Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . .6 Transportation . . . . . . . . . . . . . . . . . . . . . . . .
Description 1 2 3 Key to Fig. 1 1 Oil filter OF 4-25 2 Exhaust filter AF 16-25 3 Condensate trap AK 16-25 Fig. 1 TRIVAC-B with accessories 1 Description TRIVAC-B pumps are oil-sealed rotary vane pumps. The TRIVAC S16 B and S 25 B are single-stage pumps, and the TRIVAC D16 B and D 25 B are dual-stage pumps. The number in the type designation (16 or 25) indicates the pumping speed in m3 · h-1. TRIVAC-B pumps can pump gases and vapours and evacuate vessels or vacuum systems in the fine vacuum range.
Description Key to Fig. 2 1 Intake port 2 Dirt trap 3 Anti-suckback valve 4 Intake channel 5 Vanes 6 Pump chamber 7 Rotor 8 Cover plate, connection for inert gas ballast 9 Exhaust channel 10 Exhaust valve 11 Internal demister 12 Spring buckles 13 Cover plate, connection for oil filter Fig. 2 Sectional drawing of the TRIVAC-B The gas ballast valve is opened and closed by turning the gas ballast knob (7/5) on the front.
Description Key to Fig. 3 1 Accessories 2 Bearings 3 Non-return valve 4 Pump chamber of the TRIVAC 5 Oil reservoir 6 Oil pump Fig. 3 Schematic of the lubricating system Key to Fig. 4 1 Oil reservoir 2 Spring 3 Control piston 4 Anti-suckback piston 5 Valve seat 6 Valve disk 7 Gas inlet Fig. 4 Hydropneumatic anti-suckback valve 1.2 Supplied Equipment The equipment supplied with the TRIVAC-B pump includes: Pump with motor, including initial filling of N 62 or HE-200 oil, or Artic oil (Cat. Nos.
Description 1.3 Accessories 1.4 Spare Parts Cat. No. / Ref. No.
Description 1.6 Technical Data 50 Hz operation, SI units S 16 B D 16 B S 25 B D 25 B Nominal pumping speed* m3 · h-1 18.9 18.9 29.5 29.5 Pumping speed* m3 · h-1 16.5 16.5 25.77 25.
560 560 482 103 32 102 66 113 27 23 16 507 2752) 103 27 103 28 2752) 507 482 560 560 19 27 26 33 33 103 38 103 37 102 76 - 103 49 103 48 103 36 103 35 102 75 - Cat. No.
Operation Key to Fig. 7 1 Handle 2 Intake port 3 Exhaust port 4 Oil-level glass 5 Gas ballast knob 6 Threaded connection M 16 x 1.5 for inert gas ballast 7 Adapter 8 Cover plate 9 Cover plate; connection for oil filter Abb. 7 Connections and controls 2 Operation below +12 °C (55 °F) (see Section 2.5.3). The max. amount of heat given off approximately corresponds to the rated motor power. 2.
Operation The maximum throughput of the pump is equivalent to the pumping speed of the pump (see Section 1.6). Caution The cross-section of the intake and exhaust lines should be at least the same size as the connection ports of the pump. If the intake line is too narrow, it reduces the pumping speed. If the exhaust line is too narrow, overpressures may occur in the pump; this might damage the shaft seals and cause oil leaks. The maximum pressure in the oil case must not exceed 1.5 bar (absolute).
Operation Star connection Delta connection Fig. 8 Connection diagram for TRIVAC-B with 50 Hz 3-phase motor 2.3.2 Pump with Three-Phase AC Motor TRIVAC-B pumps with a three-phase motor are supplied without accessories for electrical connection. They must be connected via the appropriate cable, and a suitable motor protection switch. Set the switch in accordance with the rating on the motor nameplate. Fig. 8 shows the connection for pumps with 230/400 V, 50 Hz motors.
Operation 2.4.1 Areas of Application Caution The pump is not suitable for pumping of: - ignitable and explosive gases or vapours - oxidants - pyrophorous gases. The pumps are not suitable for pumping of liquids or very dusty media. Suitable protective devices must be installed. Our technical sales department is available for further advice in these matters. 2.4.
Operation 2.5 Operation TRIVAC-B pumps can pump condensable gases and vapours, provided that the gas ballast valve (7/5) is open and the pump has attained its operating temperature. 2.5.1 Pumping of Non-Condensable Gases If the process contains mainly permanent gases, the pump may be operated without gas ballast, provided that the saturation vapour pressure at operating temperature is not exceeded during compression.
Operation 2.6 Switching Off/Shutdown Under normal circumstances, all that you need do is to electrically switch off the TRIVAC-B. 2.6.1 Shut-Down through Monitoring Components Caution No further measures will be required. When pumping condensable media let the pump continue to operate with the gas ballast valve open and the intake line closed before switching off (see Section 2.5.2). When pumping aggressive or corrosive media, let the pump continue to operate even during long non-working intervals (e.g.
Operation 3 Maintenance Caution Disconnect the electrical connections before disassembling the pump. Make absolutely sure that the pump cannot be accidentally started. If the pump has pumped harmful substances, contrary to what has been stated in Section 2.4, ascertain the nature of hazard and take adequate safety measures. Observe all safety regulations. If you send a pump to LEYBOLD for repair please indicate any harmful substances existing in or around the pump.
Maintenance 1 2 3 Key to Fig. 9 1 Oil-fill plug 2 Oil-level mark maximum 3 Oil-level mark minimum 4 Oil-drain plug 4 Abb. 9 Oil change 3.2 Oil Change Screw the oil-drain plug back in (check the flat gasket and reinstall a new one if necessary). For proper operation of the pump, it is essential that the pump has an adequate supply of the correct and clean oil at all times. Remove the oil-fill plug (9/1) and fill in with fresh oil.
Maintenance Key to Fig. 10 1 Oil case 2 Spring buckles 3 Demister 4 Frame for demister 5 Hex. socket screws (5 pcs.) 6 Silencing nozzle 7 Gasket Fig. 10 Removal and fitting of the internal demister 3.3 Cleaning the Dirt Trap The resultant noise at high intake pressures indicates that the internal demister is dirty. A wire-mesh sieve is located in the intake port of the pump to act as a dirt trap for coarse particles. It should be kept clean to avoid a reduction of the pumping speed.
Maintenance Key to Fig. 11 1 Gasket 2 Handle 3 Hex. socket screw 4 Electric motor 5 Intermediate flange 6 Hex. socket screw 7 Coupling 8 Threaded pin Fig. 11 Disassembly and reassembly of the electric motor 3.5 Disassembly and Reassembly of the Electric Motor Caution Before starting work, always disconnect the motor from the mains.Disconnect the wires in the junction box of the motor (threephase models only) or pull the mains plug. Remove the handle (11/2).
Maintenance Key to Fig. 12 1 Coupling element 2 Hexagon socket screw 3 Spring washer 4 Coupling (one half) 5 Key 6 Bushing 7 O-ring 8 Shaft seal 9 Centering disk 10 Hexagon socket screws Position of the shaft seal Fig. 12 Exchanging the outer shaft seal 3.6 Replacing the Outer Shaft Seal Remove the key (12/5). Required tools: If the centering disk is stuck, screw the capscrews (12/10) into the jackscrew holes in the centering disk.
Maintenance If you do not have a shaft seal driver, place the shaft seal on the centering disk and carefully force it in with light blows of the plastic hammer. The shaft seal must not be bent. Carefully push the O-ring (12/7) and the bushing (12/6) onto the shaft. Push the centering disk (12/1) with the shaft seal onto the shaft and up against the end plate; fasten it with the hex. socket screws (12/10). Insert the key (12/5). Mount the pump-half of the coupling (12/4) on the shaft.
Maintenance Key to Fig. 13 1 Hex. nuts 2 Pump module 3 Washer 4 Gasket 5 Coupling element 6 Tie rods Fig. 13 Removing and Remounting the Pump Module 3.8.1 Waste Disposal of Used Pump Materials 3.9 Storing the Pump The corresponding environmental and safety regulations apply. This applies equally to used filters and filter elements (oil filter, exhaust filter and dust filter).
Maintenance 3.10 Maintenance Plan (Recommendation) No. Rotary vane pumps TRIVAC S/D 16 B TRIVAC S/D 25 B Measurement/test quantity Interval Remarks Operating / auxiliary 6m a n-a VE VP t Refer also to the Operating Instructions - Section: indivimaterials dual components. 1 Operate the pump for at least 0.8 hours with gas ballast. 2 Check the oil level, change the oil Oil: N 62 or special and alterif required. native oils, see Section 1.
Maintenance Rotary vane pumps TRIVAC S/D 16 B TRIVAC S/D 25 B No. 7 Change the oil - and Measurement/test quantity Interval Remarks Operating / auxiliary VE VP t 6m a n-a Refer also to the Operating Instructions - Section: individumaterials al components. Oil: N 62 or special and alterx Oil change: native oils. • First oil change after 100 operating hours • Pump switched off and cold. Change the oil when the pump is cold in order to avoid releasing absorbed gases.
Troubleshooting 4 Troubleshooting Fault Possible cause Remedy Repair* Pump does not start. Wiring is malfunctioning. Motor protection switch incorrectly set (3-phase motors only). Operating voltage does not match motor. Motor is malfunctioning. Oil temperature is below 12 °C. Oil is too viscous. Exhaust filter/exhaust line is clogged. Pump is seized up (sign: pump is jammed). Check and repair wiring. Set motor protection switch properly. 2.3 Replace the motor. Replace the motor.
Declaration of Contamination of Vacuum Equipment and Components Copies: Page 1 (white) to manufacturer or representative - Page 2 (yellow) attach to consignment packaging securety - Page 3 (blue) copy for file of sender The repair and/or service of vacuum equipment and components will only be carried out if a correctly completed declaration has been submitted. Non-completion will result in delay. The manufacturer could refuse to accept any equipment without a declaration.
EC Declaration of Conformity We - LEYBOLD Vakuum GmbH - herewith declare that the products defined below meet the basic requirements regarding safety and health of the relevant EEC directives by design, type and versions which are brought into circulation by us. The products conform to the following directives: In case of any product changes made without our approval, this declaration will be void.
GA 01.202/9.
Bonner Straße 498 (Bayenthal) D-50968 Köln Telefon: (0221) 347-0 Telefax: (0221) 347-1250 GA 01.202/9.02 - 03/96 Printed in Germany on chlorine-free bleached paper 03.96 OF/RSP 1.80.6.637.