32” 4K Surgical Display 240-031-050 1
Warning and Cautions Please read this manual and follow its instructions carefully. The words warning, caution, and note carry special meanings and should be carefully reviewed: Warning: Indicates measures to avoid potential serious injury to the user and the patient. Caution: Indicates risks to the equipment. Failure to follow cautions may result in product damage. Note: Provides special information to clarify instructions or present additional useful information.
Cautions 1. Connect the device to an AC adapter connected to a hospital grade power cord ensuring the power cord is plugged into a grounded power outlet to achieve grounding reliability. 2. Do not sterilize the device, as the delicate electronics cannot withstand this procedure. 3. Use only the proprietary surgical display power supply for the display. Completely secure the connection between the DC power cord and the extension cord. 4.
• Connect the device to an outlet on a circuit different from that to which the other device(s) are connected. • Consult the manufacturer or field service technician for help. The warranty is void if any of these warnings or cautions is disregarded. This product contains electrical waste or electronic equipment. It must not be disposed of as unsorted municipal waste and must be collected separately.
About Your Device 32” 4K Surgical Display REF: 240-031-050 32” 4K Surgical Display is 32” LED Display is a wide screen LED surgical display that can support a maximum resolution of UHD (4096 x 2160). The display supports the following video inputs: digital RGB (DVI), HDMI and HDMI (4K). It supports serial communication via the RS232 port and SDC SIDNE port. It also supports USB ports (2) for 5V, 1A power supply for accessories and peripherals.
Package Contents x4 Part Number 240-031-050 0240-031-004 Package Contents 32” 4K Surgical Display (4) M4 x 16mm VESA screws Hospital-grade AC power cord Medical Power Supply Model : BPM150S24F11 (Bridgepower) - HDMI Cable Cable Cover Part Number 0240-030-951 Optional Accessories 15-ft. (5 pin) DC extension cable Model: 1501047***(Bridgepower) ***: blank or 001~999 0240-030-952 75-ft.
Device Features Front panel 1. Auxiliary device status 2. Display Screen 3. Power switch(soft) 4. Power LED* LED shines red to indicate an active connection to a connected device Shows video image Powers the display ON and OFF. Indicates current status 5. PIP Accesses the Picture in Picture adjustment menu. 6. Bright Accesses the Brightness adjustment menu. 7. Specialty Accesses the Specialty selection menu. 8. Input Access the Input selection menu. 9.
This system saves energy by switching your monitor into a low-mode when is has not been used for a certain period of time. Power Management system operates with a VESA DPMS compliant video card installed in your computer. You use a software utility installed on your computer to set up this feature.
Rear Panel 1. Accessory mount Provide an access point for mounting optional accessories. 2. VESA mounting holes Provide access point for mounting the display. (100x100, 200x100) 3. Power switch (hard) Powers the input DC power On and OFF. 4. Handles Aid in display positioning Caution: The handles are not intended to bear the entire weight of the display. 5. Cable cover hinges Attach the bottom of the cable cover to the display 6. Velcro straps Straps aid in cable management. 7.
Cable Cover Installing the Cable Cover 1. Align the left and right hinges of the cable cover onto the bottom rear of the display. 2. Snap on the top section of the cable cover to the aligning clips. Removing the Cable Cover 1. Pinch the left and right clips and pull the cable cover towards you. 2. Remove the cable cover from the left and right hinges.
Display Handles Caution: The handles are intended to aid in positioning the display, not for transporting the display. The handles should not bear the full weight of the display. Removing the Display Handles 1. Using a 3mm hex key, loosen the two M4 x 25mm screws and gently pull the handle away from the display. Installing the Display Handles 1. Align the handle with the screw holes on the rear of the display. 2. With using a 3mm hex key, install the two M4 x 25mm screws to attach the handle.
Setup Stryker Endoscopy considers instructional training, or inservice, an integral part of this device. Your local Stryker Endoscopy sales representative will perform at least one inservice at your convenience to help set up your device and instruct you and your staff on its operation and maintenance. To schedule an inservice, contact your local Stryker Endoscopy representative after your device has arrived.
Connecting the Power Supply 1. Connect the power supply to the 24V input on the display 2. Connect the AC power cord to the power supply*. 3. Connect the AC power, using the supplied hospital-grade power cord. 4. (Optional, not shown) Connect an extension cord between the power supply and display. 5. Install cable cover. * Power supply information: Model Number: BPM150S24F11, Manufacturer: Bridgepower Corp.
Operation Operate the display using the rotary control and the four buttons located on the front panel. A list of the display controls and their functions is provided below. On-Screen Display (OSD) Accessing the On-Screen Display To use the four front-panel buttons: 1. PIP: Press to activate Picture Mode (Picture in Picture, Picture by Picture, Picture on Picture) 2. Bright: Activates the Brightness adjustment menu. 3. Specialty: Activates the Specialty adjustment menu. 4.
Operating On-Screen Display The device OSD helps navigate through various device menus. 1. Press the Rotary Control to activate the OSD menu. 2. Rotate the Rotary Control to move up or down through the menu. The parameter is highlighted when selected. 3. Press the Rotary Control to enter the next level OSD. 4. Rotate the Rotary Control to increase or decrease the value of the selected parameter, or to make a selection on different options. 5.
Troubleshooting Before returning your display for service, consult the troubleshooting list below: 16
Cleaning and Maintenance Warning To avoid electric shock and potentially fatal injury, unplug the display and power supply from the electrical outlet before cleaning. Caution • Do not spray cleaning liquid directly onto the display or the power supply as product damage may result. Spray on the cloth before wiping the unit. • Do not immerse the display or power supply in any liquid as product damage will result.
Technical Specification General Description Classification and Approvals Class 1 Equipment Medical equipment with respect to electric shock, fi re, and mechanical hazards only in accordance with ANSI/AAMI ES60601-1 and CAN/CSA C22.2 No. 60601.1.
Electromagnetic Compatibility Like other electrical medical equipment, 32” 4K Surgical Display requires special precautions to ensure electromagnetic compatibility with other electrical medical devices. To ensure electromagnetic compatibility (EMC), the display must be installed and operated according to the EMC information provided in this manual. The display has been designed and tested to comply with IEC 60601-1-2 requirements for EMC with other devices.
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity 32” 4K Surgical Display is intended for use in the electromagnetic environment specified below. The customer or the user of the display should ensure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic Discharge Floors should be wood, ± 6kV contact ± 6kV contact (ESD) concrete, or ceramic tile.
than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended Separation Distance: d = 1.17√P d = 1.17√P 80 MHz to 800 MHz Conducted RF IEC 61000-4-6 3 Vrms 150 kHz to 80 MHz 3V Radiated RF IEC 61000-4-3 3 V/m 80MHz to 2.5 GHz 3 V/m d = 2.33√P 800 MHz to 2.
0.01 0.12 0.12 0.23 0.1 0.37 0.37 0.74 1 1.17 1.17 2.33 10 3.70 3.70 7.37 100 11.70 11.70 23.30 For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Symbols and Definitions The following symbols appear on the product, its labeling, or the product packaging.