Contents 1 Warnings and Cautions 2 2 4 Warnings Cautions About Your Device 4 5 5 6 7 Intended Use Indications Contraindications Package Content Device Features 10 Setup 10 Connections 12 Basic Video Setup 13 Operation 13 On-Screen Display (OSD) 19 Troubleshooting 21 Periodic Maintenance 21 Cleaning 22 Disposal 23 Technical Specifications 27 Symbols
Warnings and Cautions Please read this manual and follow its instructions carefully. The words warning, caution, and note carry special meanings and should be carefully reviewed: Warning The personal safety of the patient may be involved. Disregarding this information could result in personal injury. Caution Special service procedures or precautions must be followed to avoid damaging the device. Note Special information to make maintenance easier or important information more clear.
Warnings To avoid potential serious injury to the user and the patient and/or damage to this device, please note the following warnings: • Read this manual thoroughly and be familiar with its contents prior to using this device. • Federal law (United States of America) restricts this device to sale by, or on the order of, a physician. • Carefully unpack the device and check if any damage occurred during shipment. • This device is non-sterile and therefore should not be placed in the sterile field.
• To connect to an international power supply, use an attachment plug appropriate for the power outlet, as outlined in the "Technical Specifications" section of this manual. • Unplug the device if it is not to be used for an extended period of time. To disconnect the cord, unscrew the plug first, then pull the cord out by the plug. Never pull the cord itself. • Do not expose the device to moisture or apply liquid cleaners directly to the screen.
About Your Device The WiSe 19" HDTV Surgical Display (display) is a wide screen LCD surgical display that can support a maximum resolution of 1080p. The display supports various video inputs, including digital RGB, analog RGB, serial digital interface (SDI), component video (YPbPr/RGB), S-video, C-video, and wireless RGB. The display features an optional WiSe HDTV Transmitter, which allows it to receive a high-definition video signal over a radio-frequency link (USA and Canada only).
Indications This device is indicated for the following surgical procedures: • • • • • • • • • • • • • General surgery General laparoscopy Nasopharynogoscopy Ear endoscopy Sinusocopy Plastic surgery wherever a laparoscope/endoscope/ arthroscope is indicated for use Laparoscopic cholecystectomy Laparoscopic hernia repair Laparoscopic appendectomy Laparoscopic pelvic lymph node dissection Laparoscopically assisted hysterectomy Laparoscopic & thorascopic anterior spinal fusion Anterior cruciate ligament recons
Package Content 1 2 4 3 5 Package Content Part Number 1. WiSe 19" HDTV Surgical Display 240-030-990 2. (4) M4 × 10 mm VESA screws 3. Hospital-grade AC power cord (USA and Canada only) 4. WiSe 19'' Surgical Power Supply 5. Cable Cover – – 240-030-992 – Optional Accessories Part Number WiSe HDTV Transmitter (USA and Canada only) 240-030-971 15-ft. (5 pin) DC extension cable 240-030-951 75-ft.
Device Features Front panel Operate the display using the rotary control located on the front panel. A list of the display controls and their functions is provided below. 1 2 3 4 5 6 1. Display Screen Shows video image. 2. Power switch (soft) Powers the display ON and OFF. 3. Power LED Indicates current status. Shines green if the display is powered on or is in screen saver mode; blinks red if the display is in standby mode; blinks amber if over voltage. 4.
Rear panel 2 1 3 4 5 8 6 7 1. VESA mounting holes (75 mm) Provide access points for mounting the display. 2. VESA mounting holes (100 mm) Provide access points for mounting the display. 3. Connector labels Indicate the types of video connectors. 4. Power connector Connects to the 13V DC power supply. 5. Power switch (hard) Powers the input DC power ON and OFF. 6. Cable-management Velcro Straps Organize cables (3 straps included). 7. Cable-management cover Covers cables.
Cable Cover Installing the Cable Cover 1. Align the left and right hinges of the cable cover onto the bottom rear of the display. 2. Snap on the top section of the cable cover to the aligning clips. 3. With your fingers, turn the thumbscrews clockwise to tighten and lock the cable cover onto the display. Uninstalling the Cable Cover 1. With your fingers, turn the thumbscrews counter clockwise to loosen. 2.
Setup Stryker Endoscopy considers instructional training, or inservice, an integral part of this device. Your local Stryker Endoscopy sales representative will perform at least one inservice at your convenience to help set up your device and instruct you and your staff on its operation and maintenance. To schedule an inservice, contact your local Stryker Endoscopy representative after your device has arrived.
Connecting the WiSe 19" Surgical Display Power Supply 1 3 2 1. Connect the power supply to the 13V input on the display. 2. Connect the power cord to the power supply. 3. Connect the AC power, using the supplied hospital-grade power cord. 4. (Optional, not shown) Connect an extension cord between the power supply and display. 5. Install cable cover.
Basic Video Setup WiSe 19" HDTV Surgical Display 1 SDC Camera 2 1. Ensure the cable cover has been removed. 2. Route the video output 1 from the camera to the SDC. 3. Route the video output 1 from the SDC to input on the DVI input on the display. 4. Reinstall the cable cover.
Operation Operate the display using the rotary control located on the front panel. A list of the display controls and their functions is provided below. On-Screen Display (OSD) Accessing the On-Screen Display To access the OSD, use the Rotary Control as outlined below: • Turn Right/Left — With the on-screen display menu activated, turning increases/decreases the value of the selected parameter. With the on-screen display deactivated, turning activates the video source selection menu.
OSD Menus Speciality Menu Item Description Color Temperature* Chooses between color temperatures for Standard, Arth, Lap, PACS, or Norm Range — Red Red balance -128 – 127 Green Green balance -128 – 127 Blue Blue balance -128 – 127 Gamma Gamma value 0.1 – 2.
Image Effect Menu Item Description Scale Mode Chooses scale mode between Fill All, V-Fill, H-Fill, One-To-One, or Fill-Aspect Freeze Frame Enables or disables freeze frame PIP Enables PIP (picture in picture) function POP Enables POP (picture on picture) function PBP Enables PBP (picture by picture) function Advanced Menu Item Description Key Lock Sets to key lock mode Auto Source Select Adjusts Auto Source Select between on and off DPMS Chooses DPMS (display power OFF, 30, 60, 90, manage
Information Menu Item Description User Name Entry Enters custom user name display for boot-up display Serial Number Displays device serial number Runtime Displays current device total run time Input Format Displays current input format Note: Actual on-screen display values may vary with updated versions of the firmware and user settings.
Linking Transmitter to Auxiliary Display (for USA and Canada Only) Temporary Linking Caution Equipment that employs RF communications may affect the normal function of the transmitter. When choosing a location for the transmitter, consult the “Electromagnetic Compatibility” section of this manual to ensure proper function. The transmitter functions in the 4.9 – 5.9 GHz spectrum. As necessary, remove or reconfigure other wireless devices from the environment, such as cordless phones and 802.
5. Remove the token from the token slot on the auxiliary display. 6. By repeating steps 4 and 5, as many as two additional auxiliary displays may be linked. All linking must be done within two minutes. 7. Store the token in the transmitter token slot when not in use. Note: If multiple transmitters are within 100 feet of each other, for example, in adjacent operating rooms, link each transmitter to its respective display(s) one set at a time. Wait 15 seconds before linking the next transmitter/display set.
Troubleshooting Before returning your display for service, consult the troubleshooting list below: Display Problem No picture Current Status Remedy Power LED on Using the OSD Menu, adjust the brightness and contrast to maximum, or reset them to their default settings. Power LED off Ensure the power switch at the front and rear of the display are set to ON. Check if the AC power cord is properly connected to the AC adapter and outlet.
Optional Transmitter (for the USA and Canada Only) Problem 20 Current Status Remedy Wireless link not established within 2 minutes (with optional WiSe HDTV transmitter). “Wireless RGB No Signal” Cycle the hard power switch at the rear of the display. Cycle the hard power switch at the front of the transmitter. Wireless link established with some but not all displays (with optional WiSe HDTV transmitter). “Wireless RGB No Signal” Cycle the hard power switch on the affected display only.
Periodic Maintenance Caution Do not expose the display to moisture or apply liquid cleaners directly to the display screen. Spray the cleaning solution into a soft cloth and clean the screen gently. Cleaning 1. Disconnect the display from the power supply before cleaning. 2. Clean the plastic areas of the display with a dry, soft cloth, or a soft cloth lightly moistened with mild detergent solution. Do not use any type of solvent, such as alcohol or benzine, which might damage the finish.
Disposal This product contains electrical waste or electronic equipment. It must not be disposed of as unsorted municipal waste and must be collected separately in accordance with applicable national or institutional related policies relating to obsolete electronic equipment. Dispose of any system accessories according to normal institutional practice relating to potentially contaminated items.
Technical Specifications Display LCD Display Panel 18.95" (481.33 mm) Diagonal (a-Si TFT active matrix LCD) Synchronization 2.5 – 5.0 Vpp separated sync Pixel Pitch 0.2835 (H) × 0.2835 (V) mm Response Time < 5 ms Typ Viewing Angle Right/Left 80 (Typ) Up 75 (Typ), Down 85 (Typ) Display Colors 16.
Transport & Storage Conditions Storage -0.4 – 140°F (-18 – 60°C) Relative Humidity Range 10 – 85% Classification and Approvals Class I Equipment Medical equipment with respect to electric shock, fire, and mechanical hazards only in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1. IPX1 Water Ingress Protection Continuous Operation Compliance Medical Safety Standards IEC 60601-1:1988 + A1:1991 + A2:1995 CAN/CSA C22.2 No 601.1-M90 UL 60601-1:2003 AS/NZS 3200.1.0:1998 CSA 22.2.601.1.
Guidance and Manufacturer’s Declaration: Electromagnetic Emissions The WiSe™ 19" HDTV Surgical Display is intended for use in the electromagnetic environment specified below. The customer or the user of the WiSe™ 19" HDTV Surgical Display should ensure it is used in such an environment.
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity The WiSe™ 19" HDTV Surgical Display is intended for use in the electromagnetic environment specified below. The customer or the user of the WiSe™ 19" HDTV Surgical Display should ensure that it is used in such an environment.
Symbols The following symbols appear on the product, its labeling, or the product packaging.
Stryker Endoscopy 5900 Optical Court San Jose, CA 95138 USA 1-408-754-2000, 1-800-624-4422 www.stryker.com European Representative: Regulatory Manager, Stryker France ZAC Satolas Green Pusignan Av. De Satolas Green 69881 MEYZIEU Cedex, France Stryker Corporation or its divisions or other corporate affiliated entities own, use or have applied for the following trademarks or service marks: WiSe and the Stryker logo. All other trademarks are trademarks of their respective owners or holders.